Cardiac functional analysis with multi-detector row CT and segmental reconstruction algorithm: comparison with echocardiography, SPECT, and MR imaging
Authors:Yamamuro Masaki  Tadamura Eiji  Kubo Shigeto  Toyoda Hiroshi  Nishina Takeshi  Ohba Muneo  Hosokawa Ryohei  Kimura Takeshi  Tamaki Nagara  Komeda Masashi  Kita Toru  Konishi Junji
Affiliation:Department of Nuclear Medicine and Diagnostic Imaging, Kyoto University Graduate School of Medicine, 54 Shogoinkawahara, Sakyo-ku, Kyoto 606-8507, Japan. myamamu@kuhp.kyoto-u.ac.jp
Abstract:PURPOSE: To evaluate accuracy of cardiac functional analysis with multi-detector row computed tomography (CT) and segmental reconstruction algorithm over a range of heart rates. MATERIALS AND METHODS: Institutional review board approval was obtained. Informed consent was not required. Multi-detector row CT (500-msec rotation time, 8 x 1-mm detector collimation) and magnetic resonance (MR) imaging were performed in 50 patients (28 men, 22 women; age range, 46-84 years; mean age, 67 years). Two-dimensional echocardiography was performed in 41 patients, and electrocardiographically (ECG)-gated single photon emission computed tomography (SPECT) was performed in 27. End-diastolic volume (EDV), end-systolic volume (ESV), ejection fraction (EF), and left ventricular (LV) mass were estimated with multi-detector row CT and compared with values estimated with MR imaging, which served as the reference standard. Additionally, EF values estimated with multi-detector row CT, echocardiography, and SPECT were compared with those estimated with MR imaging. Systemic error and degree of agreement of global functional parameters measured with MR imaging and other modalities were assessed. In a second analysis, linear regression analysis was added. RESULTS: EF estimated with multi-detector row CT agreed and correlated well with EF estimated with MR imaging (bias +/- standard deviation, -1.2% +/- 4.6; r = 0.96). Agreement and correlation were similar for EDV (-0.35 mL +/- 15.2; r = 0.97), ESV (1.1 mL +/- 8.6; r = 0.99), and LV mass (2.5 mL +/- 15.0; r = 0.96). Standard deviation of EF difference between multi-detector row CT and MR imaging was significantly less than that between echocardiography and MR imaging (P < .001) or that between SPECT and MR imaging (P < .001). CONCLUSION: Various LV functional parameters were measured with multi-detector row CT with a segmental approach, and measurements correlated and agreed with those obtained with MR imaging. Moreover, functional analysis with multi-detector row CT was more accurate than that with two-dimensional echocardiography or ECG-gated SPECT.
Keywords:
本文献已被 PubMed 等数据库收录!
点击此处可从《Radiology》浏览原始摘要信息
点击此处可从《Radiology》下载全文
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号