首页 | 本学科首页   官方微博 | 高级检索  
     

二尖瓣置换术后两种类型机械瓣对血流动力学的影响
引用本文:黄建新,曹红格,冯晓华,王岩,刘爱琴,杜巧,封淑文. 二尖瓣置换术后两种类型机械瓣对血流动力学的影响[J]. 武警医学, 2010, 21(7): 560-562
作者姓名:黄建新  曹红格  冯晓华  王岩  刘爱琴  杜巧  封淑文
作者单位:武警北京总队第二医院特检科,100037;河北省邢台市医学专科学校第二附属医院功能科,054000;北京军区装备部门诊部,100144
摘    要: 目的 评价不同类型机械瓣对二尖瓣置换术后血流动力学的影响.方法 本院行二尖瓣置换术96例,分为双叶瓣及倾斜碟瓣两组,比较手术前后及手术1年后两组间血流动力学的超声参数.结果 两组术前二维超声参数:各房室内径及主、肺动脉内径、二尖瓣口面积差异均无统计学意义,血流动力学参数中E峰、A峰、平均跨瓣压差(mean pressure gradient,MPG)、肺动脉收缩压差异均无统计学意义;术后10 d内的二维超声参数比较:双叶瓣及倾斜碟瓣两组的E峰速分别为(1.43±0.70)m/s和(1.86±0.65)m/s (P<0.05),有效瓣口面积(effective orifice area,EOA)分别为(2.61±0.61)cm2、(2.31±0.63)cm2(P<0.05);肺动脉收缩压(pulmonary artery systolic pressure,PASP)分别为(33.9±14.2)mmHg 和(32.8±9.4)mmHg(1 mmHg=0.133 kPa,P>0.05);MPG分别为(4.76±2.8)mmHg和(5.2±3.7)mmHg(P>0.05).手术1年后复查,双叶瓣及倾斜碟瓣两组的E峰速分别为(1.48±0.9)m/s和(1.92±0.4)m/s(P<0.05),EOA为(2.52±0.32)cm2和(2.01±0.2)cm2 (P<0.05),PASP为(33.4±22.5)mmHg 和(36.4±24.7)mmHg(P>0.05).结论 双叶机械瓣对风湿性心脏病二尖瓣置换术后血流动力学的影响强于倾斜碟瓣.

关 键 词:风湿性心脏病  机械瓣  血流动力学  
收稿时间:2010-02-08

Hemodynamic effect of different mechanical valves in patients of rheumatic heart disease with a replaced mitral valve
HUANG Jianxin,CAO Hongge,FENG Xiaohua,WANG Yan,LIU Aiqin,DU Qiao,FENG Shuwen. Hemodynamic effect of different mechanical valves in patients of rheumatic heart disease with a replaced mitral valve[J]. Medical Journal of the Chinese People's Armed Police Forces, 2010, 21(7): 560-562
Authors:HUANG Jianxin  CAO Hongge  FENG Xiaohua  WANG Yan  LIU Aiqin  DU Qiao  FENG Shuwen
Affiliation:1. Department of Special Examination, Beijing Municipal Corps Second Hospital, Chinese Peopleg Armed Police Forces, Beijing 100037, China. 2. Department of Pharmacy,the Second Affiliated Hospital of Xingtai Medical School, Xingtai 054000, China. 3. The Clinic of the Military Armament Department, Beijing Military Command, Beijing 100144, China)
Abstract:Objective To assess the hemodynamic effect of different types of mechanical valves on patients with rheumatic heart disease (RHD) who received replacement of a mitral valve. Methods Ninety - six patients receiving replacement of the mitral valve were retrospectively analysed. They were divided into bileaflet valve group ( group 1 ) and tilting disk valve group ( group 2). He- modynamic changes in the two groups were compared. Results The hemodynamic indices were not significantly different in the two groups before operation. Ten days after operation, the peak Evelocity in group 1 was significantly lower than that in group 2, ( 1.43 ± 0.70) m/s vs(1.86 ±0.65)m/s, (P〈0.05) ; the effective orifice area(EOA) was (2.61±0.61) cm^2 in group 1 vs (2.31 ±0. 63 ) cm^2 in group 2 ( P 〈 0.05 ). One year after operation, the peak E velocity in group 1 was significantly lower than that in group 2, [ ( 1.48 ± 0.9) m/s vs ( 1.92 ± 0.4) mm, P 〈 0.05 ] , and the effective orifice area (EOA) was (2.52 ± 0.32) cm^2 in group 1 vs ( 2.01 ± 0.2 ) cm^2 in group 2 ( P 〈 0.05 ) Conclusions The bileaflet valve has a better dynamic influence than the tilting disk valve in patients of RHD with a replaced mitral valve.
Keywords:rheumatic heart disease  prosthetic heart valve  hemodynamics