New generation 3-dimensional echocardiography for left ventricular volumetric and functional measurements: comparison with cardiac magnetic resonance.
Authors:Nikolay P Nikitin  Cristina Constantin  Poay Huan Loh  Justin Ghosh  Elena I Lukaschuk  Anna Bennett  Sarah Hurren  Farqad Alamgir  Andrew L Clark  John G F Cleland
Institution:Department of Cardiology, Academic Unit, The University of Hull, Castle Hill Hospital, Castle Road, Cottingham, Kingston-upon-Hull HU16 5JQ, UK. n.p.nikitin@hull.ac.uk
Abstract:AIMS: Non-invasive assessment of left ventricular (LV) structure and function is important in the evaluation of cardiac patients. This study was designed to test the accuracy and reproducibility of new generation 3-dimensional echocardiography (3DE) in measuring volumetric and functional LV indices as compared with current "gold standard" of non-invasive cardiac imaging, cardiac magnetic resonance (CMR). METHODS AND RESULTS: Sixty-four subjects with good acoustic windows, including 40 cardiac patients with LV ejection fraction (EF)<45%, 14 patients with EF>45% and 10 normal volunteers underwent 3DE using a commercially available Philips Sonos 7500 scanner equipped with a matrix phase-array x4 xMATRIX transducer, and CMR on a 1.5 T Signa CV/i scanner (GE Medical Systems). Volumetric assessment was performed with analytical 4D-LV-Analysis software (TomTec) for 3DE and MRI-Mass software (Medis) for CMR. We found no significant differences in LV end-diastolic volume (EDV), end-systolic volume (ESV) and EF with excellent correlations between the indices measured using 3DE and CMR (r=0.97, r=0.98, and r=0.94, respectively). Bland-Altman analysis showed bias of 7 ml for EDV, 3 ml for ESV and -1% for EF with 3DE with corresponding limits of agreement (2SD) of 28 ml, 22 ml and 10%, respectively. Intraobserver and interobserver variabilities were for EDV: 3% and 4% (3DE) vs 2% and 2% (CMR), for ESV: 3% and 6% (3DE) vs 2% and 3% (CMR), and for EF: 4% and 4% (3DE) vs 2% and 4% (CMR), respectively. CONCLUSION: New generation 3DE provides accurate and reproducible quantification of LV volumetric and functional data in subjects with good acoustic windows as compared with CMR.
Keywords:Imaging  Echocardiography  Left ventricular function  Magnetic resonance imaging
本文献已被 Oxford 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号