Estimation of adult patient doses for common diagnostic X-ray examinations in Wad-madani,Sudan: derivation of local diagnostic reference levels
Authors:I I Suliman  T S Mohammedzein
Institution:1. Department of Radiology and Molecular Imaging, Medical Physics Section, College of Medicine and Health Sciences, Sultan Qaboos University, P.O. Box 35, Postal Code 123,, Al-Khod, Oman
2. Sudan Atomic Energy Commission, Radiation Safety Institute, Jamma Str., P.O. Box 3001, Khartoum, Sudan
Abstract:The objectives of this study were to estimate patient dose in some common diagnostic X-ray examinations. Radiation doses were estimated for 307 patients in six public hospitals comprising 7 X-ray units in Wad-madani, Sudan. Entrance surface air kerma (ESAK) was estimated in a three step protocol: First, X-ray unit output Y(d) was measured at a distance, d for different peak tube voltages and tube loadings (mAs). Next, incident air kerma (Ki) was calculated from Y(d) using inverse square law combined with patient exposure factors. ESAK was calculated from Ki using backscatter factor, B. Mean ESAK values are comparable to those reported in other countries and are below reference dose levels. The estimated mean ESAK values are: 0.3, 2.2, 2.2, 2.9, 2.8, 3.1, and 7.5 mGy for chest PA, Skull AP/PA, Skull LAT, Abdomen, Pelvis AP, Lumbar Spine AP and Lumbar Spine LAT examinations, respectively. The results are used for dose optimization, and to propose local diagnostic reference levels.