Biomechanical evaluation and comparison of polyetheretherketone rod system to traditional titanium rod fixation
Authors:Ravi K. Ponnappan  Hassan Serhan  Brett Zarda  Ravi Patel  Todd Albert  Alexander R. Vaccaro
Affiliation:1. Department for Trauma and Reconstructive Surgery, University Hospital RWTH Aachen, Pauwelsstraße 30, 52074 Aachen, Germany;2. Institute of General Mechanics, RWTH Aachen, Templergraben 64, 52062 Aachen, Germany;3. Department of Orthopaedic and Trauma Surgery, Medizinisches Zentrum StädteRegion Aachen GmbH, Mauerfeldchen 25, 52146 Aachen, Germany;4. Department of Orthopaedic and Trauma Surgery, University of Cologne, Joseph-Stelzmann-Straße 9, 50924 Cologne, Germany;5. Institute of Molecular and Cellular Anatomy, University Hospital RWTH Aachen, Wendlingweg 2, 52070 Aachen, Germany;6. Department of Trauma, University Hospital Zurich, Raemistr, 100, 8091 Zurich, Switzerland
Abstract:Background contextPolyetheretherketone (PEEK) has been increasingly used as a biomaterial for spinal implants. PEEK lumbar fusion rods have recently become available for use in posterior lumbar fusion procedures.PurposeTo compare Polyetheretherketone Rod System to traditional titanium rod fixation in a cadaveric model and provide mechanical test data for the PEEK system.Study designBiomechanical testing.MethodsCadaveric biomechanical testing was conducted to compare Expedium 5.5 mm PEEK rods to titanium rods of equivalent diameter. Biomaterials testing was performed to determine static and dynamic performance of Expedium 5.5 mm PEEK rods with 6% BaSo4 in compressive bending and torsion.ResultsCadaveric testing demonstrated that PEEK rods can significantly reduce the range of motion of a destabilized segment. The testing showed no significant difference in the stability provided by PEEK and titanium rods in posterolateral fusion (PLF) or posterior lumbar interbody fusion (PLIF) constructs. PEEK static compressive bending tests showed 67 degrees displacement without fracture of the rod. Torsion testing showed 30 degrees of rotation without yield or plastic deformation. Dynamic compression testing revealed two fatigue runouts at 23 degrees.ConclusionsPEEK rods provide comparable stability to titanium rods of equivalent diameter in cadaveric testing. Mechanical testing suggests PEEK rods can withstand far beyond the angular displacements suggested by cadaveric testing and that of normal physiologic range of motion. Potential advantages to PEEK rods include better anterior column load sharing, reduced stress at bone-to-screw interface, and reduced computed tomography and magnetic resonance imaging scatter and artifact.
Keywords:
本文献已被 ScienceDirect 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号