Screw angulation affects bone-screw stresses and bone graft load sharing in anterior cervical corpectomy fusion with a rigid screw-plate construct: a finite element model study
Authors:Mozammil Hussain  Raghu N Natarajan  Amir H Fayyazi  Brian R Braaksma  Gunnar BJ Andersson  Howard S An
Institution:1. Institute for Biomechanics, ETH Zurich, Vladimir-Prelog-Weg 3, 8093 Zurich, Switzerland;2. Biomechanics Section, KU Leuven – University of Leuven, Celestijnenlaan 300, 3001 Leuven, Belgium;1. Key Laboratory for Biomechanics and Mechanobiology of Ministry of Education, Beijing Advanced Innovation Center for Biomedical Engineering, School of Biological Science and Medical Engineering, Beihang University, Beijing 100083, China;2. Department of Orthopedic Surgery, Peking University Third Hospital, Beijing 100191, China;1. University of Zielona Góra, Zielona Góra, Poland;2. St. Raphael Hospital in Krakow, Krakow, Poland
Abstract:Background contextAnterior corpectomy and reconstruction with bone graft and a rigid screw-plate construct is an established procedure for treatment of cervical neural compression. Despite its reliability in relieving symptoms, there is a high rate of construct failure, especially in multilevel cases.PurposeThere has been no study evaluating the biomechanical effects of screw angulation on construct stability; this study investigates the C4–C7 construct stability and load-sharing properties among varying screw angulations in a rigid plate-screw construct.Study designA finite element model of a two-level cervical corpectomy with static anterior cervical plate.MethodsA three-dimensional finite element (FE) model of an intact C3–T1 segment was developed and validated. From this intact model, a fusion model (two-level C5, C6] anterior corpectomy) was developed and validated. After corpectomy, allograft interbody fusion with a rigid anterior screw-plate construct was created from C4 to C7. Five additional FE models were developed from the fusion model corresponding to five different combinations of screw angulations within the vertebral bodies (C4, C7): (0°, 0°), (5°, 5°), (10°, 10°), (15°, 15°), and (15°, 0°). The fifth fusion model was termed as a hybrid fusion model.ResultsThe stability of a two-level corpectomy reconstruction is not dependent on the position of the screws. Despite the locked screw-plate interface, some degree of load sharing is transmitted to the graft. The load seen by the graft and the shear stress at the bone-screw junction is dependent on the angle of the screws with respect to the end plate. Higher stresses are seen at more divergent angles, particularly at the lower level of the construct.ConclusionThis study suggests that screw divergence from the end plates not only increases load transmission to the graft but also predisposes the screws to higher shear forces after corpectomy reconstruction. In particular, the inferior screw demonstrated larger stress than the upper-level screws. In the proposed hybrid fusion model, lower stresses on the bone graft, end plates, and bone-screw interface were recorded, inferring lower construct failure (end-plate fractures and screw pullout) potential at the inferior construct end.
Keywords:
本文献已被 ScienceDirect 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号