1.
Båth M  Sund P  Månsson LG 《Medical physics》2002,29(10):2286-2297
Two generations of a CCD-based detector system with lens-based optical coupling for digital chest radiography were evaluated in terms of presampling MTF, NPS, NEQ, DQE, linearity in response, and SNR over the detector area. Measurements were performed over a wide exposure range and at several different beam qualities. Neither the presampling MTF nor the DQE showed any general strong beam quality dependence, whereas the NPS and NEQ did when compared at specific entrance air kerma values. The exposure dependency for the DQE was found to be considerable, with the detectors showing low DQE at low exposures, and higher DQE at higher exposures. It was found that the second generation has been substantially improved compared to its predecessor regarding all the relevant parameters. The DQE(0) at an entrance air kerma of 5 microGy increased from 9% to 15%, mainly due to a better system gain (including optical coupling efficiency and matching of the energy of the emitted light photons to the sensitivity of the CCD camera). The first generation of detectors was found to have problems with bad peripheral resolution [MTF(muN/2) <0.1]. This problem was nonexistent for the second generation for which uniform resolution has been obtained [MTF(muN/2)=0.3]. A theoretical calculation of the DQE of two model systems similar to the ones evaluated was also performed, and the results were comparable to the experimentally determined data at high exposures. The model shows that both systems suffer from low optical coupling efficiency due to the large demagnification used. The main conclusion is that although the second generation has been improved, there is still a problem with low system gain leading to relatively modest DQE values, especially at low exposures.  相似文献   

2.
An objective analysis of image quality parameters was performed for a computed radiography (CR) system using both standard single-side and prototype dual-side read plates. The pre-sampled modulation transfer function (MTF), noise power spectrum (NPS), and detective quantum efficiency (DQE) for the systems were determined at three different beam qualities representative of pediatric chest radiography, at an entrance detector air kerma of 5 microGy. The NPS and DQE measurements were realized under clinically relevant x-ray spectra for pediatric radiology, including x-ray scatter radiations. Compared to the standard single-side read system, the MTF for the dual-side read system is reduced, but this is offset by a significant decrease in image noise, resulting in a marked increase in DQE (+40%) in the low spatial frequency range. Thus, for the same image quality, the new technology permits the CR system to be used at a reduced dose level.  相似文献   

3.
An integrated readout computed radiography system (Fuji XU-D1) incorporating dual-side imaging plates (ST-55BD) was analyzed in terms of modulation transfer function (MTF), noise power spectrum (NPS), and detective quantum efficiency (DQE) for standard beam qualities RQA 9 and RQA 5. NPS and DQE were assessed using a detector entrance air kerma consistent with clinical practice for chest radiography. Similar investigation was performed on a standard reader (Fuji FCR 5000) using single-side imaging plates (ST-VI). Negligible differences were found between the MTFs of the two imaging systems for RQA 9, whereas for RQA 5 the single-side system exhibited slightly superior MTF. Regarding noise response, the dual-side system turned out to be better performing for both beam qualities over a wide range of frequencies. For RQA 9, at 8 microGy, the DQE of the dual-side system was moderately higher over the whole frequency range, whereas for RQA 5, at 10 microGy, significant improvement was found at low- and midrange frequencies. As an example, at 1 cycle/mm, the following improvements in the DQE of the dual-side system were observed: +22% (RQA 9, at 8 microGy), +50% (RQA 9, at 30 microGy), and +45% (RQA 5, at 10 microGy).  相似文献   

4.
Influence of cassette type on the DQE of CR systems   总被引:1,自引:0,他引:1  
In our recent paper by Monnin et al. [Med. Phys. 33, 411-420 (2006)], an objective analysis of the relative performance of a computed radiography (CR) system using both standard single-side (ST-VI) and prototype dual-side read (ST-BD) plates was reported. The presampled modulation transfer function (MTF), noise power spectrum (NPS), and detective quantum efficiency (DQE) for the systems were determined at three different beam qualities representative of paediatric chest radiography, at an entrance detector air kerma of 5 microGy. Experiments demonstrated that, compared to the standard single-side read system, the MTF for the dual-side read system was slightly reduced, but a significant decrease in image noise resulted in a marked increase in DQE (+40%) in the low spatial frequency range. However, the DQE improvement for the ST-BD plate decreased with increasing spatial frequency, and, at spatial frequencies above 2.2 mm(-1), the DQE of the dual-side read system was lower than that of the single-side one.  相似文献   

5.
The purpose of this study was to evaluate and compare the physical characteristics of five clinical systems for digital mammography (GE Senographe 2000D, Lorad Selenia M-IV, Fischer Senoscan, Agfa DM 1000, and IMS Giotto) currently in clinical use. The basic performances of the mammography systems tested were assessed on the basis of response curve, modulation transfer function (MTF), noise power spectrum, noise equivalent quanta (NEQ), and detective quantum efficiency (DQE) in an experimental setting closely resembling the clinical one. As expected, all the full field digital mammography systems show a linear response curve over a dynamic range from 3.5 to 500 microGy (0.998相似文献   

6.
Quantitative image quality results in the form of the modulation transfer function (MTF), normalized noise power spectrum (NNPS) and detective quantum efficiency (DQE) are presented for nine full field digital mammography (FFDM) systems. These parameters are routinely measured as part of the quality assurance (QA) programme for the seven FFDM units covered by our centre. Just one additional image is required compared to the standard FFDM protocol; this is the image of an edge, from which the MTF is calculated. A variance image is formed from one of the flood images used to measure the detector response and this provides useful information on the condition of the detector with respect to artefacts. Finally, the NNPS is calculated from the flood image acquired at a target detector air kerma (DAK) of 100 microGy. DQE is then estimated from these data; however, no correction is currently made for effects of detector cover transmission on DQE. The coefficient of variation (cov) of the 50% point of the MTF for five successive MTF results was 1%, while the cov for the 50% MTF point for an a-Se system over a period of 17 months was approximately 3%. For four a-Se based systems, the cov for the NNPS at 1 mm(-1) for a target DAK of 100 microGy was approximately 4%; the same result was found for four CsI based FFDM units. With regard to the stability of NNPS over time, the cov for four NNPS results acquired over a period of 12 months was also approximately 4%. The effect of acquisition geometry on NNPS was also assessed for a CsI based system. NNPS data acquired with the antiscatter grid in place showed increased noise at low spatial frequency; this effect was more severe as DAK increased. DQE results for the three detector types (a-Se, CsI and CR) are presented as a function of DAK. Some reduction in DQE was found for both the a-Se and CsI based systems at a target DAK of 12.5 microGy when compared to DQE data acquired at 100 microGy. For the CsI based systems, DQE at 1 mm(-1) fell from 0.49 at 100 microGy to 0.38 at 12.5 microGy. For the a-Se units, there was a slightly greater reduction in average DQE at 1 mm(-1), from 0.53 at 100 microGy to 0.31 at 12.5 microGy. Somewhat different behaviour was seen for the CR unit; DQE (at 1 mm(-1)) increased from 0.40 at 100 microGy to 0.49 at 12.5 microGy; however, DQE fell to 0.30 at 420 microGy. DQE stability over time was assessed using the cov of DQE at 1 mm(-1) and a target DAK of 100 microGy; the cov for data acquired over a period of 17 months for an a-Se system was approximately 7%. For comparison with conventional testing methods, the cov was calculated for contrast-detail (cd) data acquired over the same period of time for this unit. The cov for the threshold contrast results (averaged for disc diameters between 0.1 mm and 2 mm) was 6%, indicating similar stability.  相似文献   

7.
This paper presents pre-sampling modulation transfer function (MTF), normalized noise power spectrum (NNPS) and detective quantum efficiency (DQE) results for an amorphous selenium (a-Se) full field digital mammography system. MTF was calculated from the image of an angled 0.5 mm thick Cu edge, acquired without additional beam filtration. NNPS data were acquired at detector air-kerma levels ranging from 9.1 microGy to 331 microGy, using a standard mammography x-ray spectrum of 28 kV, Mo/Mo target/filter combination and 4 cm of PMMA additional filtration. Prior to NNPS estimation, the image statistics were assessed using a variance image. This method was able to easily identify a detector artefact and should prove useful in routine quality assurance (QA) measurements. Detector DQE, calculated from the NNPS and MTF data, dropped to 0.3 for low detector air-kerma settings but reached an approximately constant value of 0.6 above 50 microGy at the detector. Subjective image quality data were also obtained at these detector air-kerma settings using the CDMAM contrast-detail (c-d) test object. The c-d data reflected the trend seen in DQE, with threshold contrast increasing at low detector air-kerma values. The c-d data were then compared against predictions made using two established models, the Rose model and a standard signal detection theory model. Using DQE(0), the Rose model gave results within approximately 15% on average for all the detector air-kerma values studied and for detail diameters down to 0.2 mm. Similar agreement was also found between the measured c-d data and the signal detection theory results, which were calculated using an ideal human visual response function and a system magnification of unity. The use of full spatial frequency DQE improved the agreement between the calculated and observer results for detail sizes below 0.13 mm.  相似文献   

8.
This work compares the detector performance and image quality of the new Kodak Min-R EV mammography screen-film system with the Fuji CR Profect detector and with other current mammography screen-film systems from Agfa, Fuji and Kodak. Basic image quality parameters (MTF, NPS, NEQ and DQE) were evaluated for a 28 kV Mo/Mo (HVL = 0.646 mm Al) beam using different mAs exposure settings. Compared with other screen-film systems, the new Kodak Min-R EV detector has the highest contrast and a low intrinsic noise level, giving better NEQ and DQE results, especially at high optical density. Thus, the properties of the new mammography film approach those of a fine mammography detector, especially at low frequency range. Screen-film systems provide the best resolution. The presampling MTF of the digital detector has a value of 15% at the Nyquist frequency and, due to the spread size of the laser beam, the use of a smaller pixel size would not permit a significant improvement of the detector resolution. The dual collection reading technology increases significantly the low frequency DQE of the Fuji CR system that can at present compete with the most efficient mammography screen-film systems.  相似文献   

9.
Mackenzie A  Honey ID 《Medical physics》2007,34(8):3345-3357
The performances of two generations of computed radiography (CR) were tested and compared in terms of resolution and noise characteristics. The main aim was to characterize and quantify the noise sources in the images. The systems tested were (1) Agfa CR 25.0, a flying spot reader with powder phosphor image plates (MD 40.0); and (2) the Agfa DX-S, a line-scanning CR reader with needle crystal phosphor image plates (HD 5.0). For both systems, the standard metrics of presampled modulation transfer function (MTF), normalized noise power spectra (NNPS) and detective quantum efficiency (DQE) were measured using standard radiation quality RQA5 as defined by the International Electrotechnical Commission. The various noise sources contributing to the NNPS were separated by using knowledge of their relationship with air kerma, MTF, absorption efficiency and antialiasing filters. The DX-S MTF was superior compared with the CR 25.0. The maximum difference in MTF between the DX-S scan and CR 25.0 subscan directions was 0.13 at 1.3 mm(-1). For a nominal detector air kerma of 4 microGy, the peak DQE of the DX-S was 43 (+/-3)%, which was over double that of the CR 25.0 of 18 (+/-2)%. The additive electronic noise was negligible on the CR 25.0 but calculated to be constant 3.4 x 10(-7) (+/-0.4 x 10(-7)) mm2 at 3.9 microGy on the DX-S. The DX-S has improved image quality compared with a traditional flying spot reader. The separation of the noise sources indicates that the improvements in DQE of the DX-S are due not only to the higher quantum, efficiency and MTF, but also the lower structure, secondary quantum, and excess noise.  相似文献   

10.
Our work is to investigate and understand the factors affecting the imaging performance of amorphous selenium (a-Se) flat-panel detectors for digital mammography. Both theoretical and experimental methods were developed to investigate the spatial frequency dependent detective quantum efficiency [DQE(f)] of a-Se flat-panel detectors for digital mammography. Since the K edge of a-Se is 12.66 keV and within the energy range of a mammographic spectrum, a theoretical model was developed based on cascaded linear system analysis with parallel processes to take into account the effect of K fluorescence on the modulation transfer function (MTF), noise power spectrum (NPS), and DQE(f) of the detector. This model was used to understand the performance of a small-area prototype detector with 85 microm pixel size. The presampling MTF, NPS, and DQE(f) of the prototype were measured, and compared to the theoretical calculation of the model. The calculation showed that K fluorescence accounted for a 15% reduction in the MTF at the Nyquist frequency (fNy) of the prototype detector, and the NPS at fNy was reduced to 89% of that at zero spatial frequency. The measurement of presampling MTF of the prototype detector revealed an additional source of blurring, which was attributed to charge trapping in the blocking layer at the interface between a-Se and the active matrix. This introduced a drop in both presampling MTF and NPS at high spatial frequency, and reduced aliasing in the NPS. As a result, the DQE(f) of the prototype detector at fNy approached 40% of that at zero spatial frequency. The measured and calculated DQE(f) using the linear system model have reasonable agreement, indicating that the factors controlling image quality in a-Se based mammographic detectors are fully understood, and the model can be used to further optimize detector imaging performance.  相似文献   

11.
The image quality of a dedicated mammography computed radiography (CR) system was characterized. A unique feature of this system is that it collects image signals from both sides of the storage phosphor. Measurements of the modulation transfer function (MTF), noise power spectrum (NPS), and detective quantum efficiency (DQE) were made. This work included improvements in our measurement methods to specifically account for the detrimental effects of system glare on the MTF and to accurately characterize the low-frequency NPS components. Image quality measurements were performed using a 25 kVp beam filtered with 2 mm Al and an exposure range of 1 to 100 mR (87 to 870 microGy). The DQE was found to decrease with increasing exposure due to an increased contribution of storage phosphor structure noise. The DQE of this system was compared to similar measurements made using a standard CR system. The dual-side read system demonstrated superior DQE compared to the standard system. The decrease in DQE with increasing exposure was more severe for the standard system than the dual-side read system. This finding suggests that the CR system noise was reduced for the dual-side read system compared to the standard system.  相似文献   

12.
The aim of this study was to characterize the effect of an image processing algorithm (FineView) on both quantitative image quality parameters and the threshold contrast detail response of the GE Senographe DS full-field digital mammography system. The system was characterized using signal transfer property, pre-sampling modulation transfer function (MTF), normalized noise power spectrum (NNPS) and detective quantum efficiency (DQE) of the system. An algorithmic modulation transfer function (MTF(a)) was calculated from images acquired at a reduced detector air kerma (DAK) and with the FineView algorithm enabled. Two sets of beam conditions were used: Mo/Mo/28 kV and Rh/Rh/29 kV, both with 2 mm added Al filtration at the x-ray tube. Images were acquired with and without FineView at four DAK levels from 14 to 378 μGy. The threshold contrast detail response was assessed using the CDMAM contrast-detail test object which was imaged under standard clinical conditions with and without FineView at three DAK levels from 24 to 243 μGy. The images were scored by both human observers and by automated scoring software. Results indicated an improvement of up to 125% at 5 mm?1 in MTF(a) when FineView was activated, particularly at high DAK levels. A corresponding increase of up to 425% at 5 mm?1 was also seen in the NNPS, again with the same DAK dependence. FineView did not influence DQE, an indication that the signal to noise ratio transfer of the system remained unchanged. FineView did not affect the threshold contrast detectability of the system, a result that is consistent with the DQE results.  相似文献   

13.
We are developing a modular detector for applications in full field digital mammography and for diagnostic breast imaging. The detector is based on a design that has been refined over the past decade for applications in x-ray crystallography [Kalata et al., Proc. SPIE 1345, 270-279 (1990); Phillips et al. ibid. 2009, 133-138 (1993), Phillips et al., Nucl. Instrum. Methods Phys. Rev. A 334, 621-630 (1993)]. The full field mammographic detector, currently undergoing clinical evaluation, is formed from a 19 cm x 28 cm phosphor screen, read out by a 2 x 3 array of butted charge-coupled device (CCD) modules. Each 2k x 2k CCD is optically coupled to the phosphor via a fiber optic taper with dimensions of 9.4 cm x 9.4cm at the phosphor. This paper describes the imaging performance of a two-module prototype, built using a similar design. In this paper we use cascaded linear systems analysis to develop a model for calculating the spatial frequency dependent noise power spectrum (NPS) and detective quantum efficiency (DQE) of the detector using the measured modulation transfer function (MTF). We compare results of the calculation with the measured NPS and DQE of the prototype. Calculated and measured DQEs are compared over a range of clinically relevant x-ray exposures and kVps. We find that for x-ray photon energies between 10 and 28 keV, the detector gain ranges between 2.5 and 3.7 CCD electrons per incident x-ray, or approximately 5-8 electrons per absorbed x ray. Using a Mo/Mo beam and acrylic phantom, over a detector entrance exposure range of approximately 10 to 80 mR, the volume under the measured 2-d NPS of the prototype detector is proportional to the x-ray exposure, indicating quantum limited performance. Substantial agreement between the calculated and measured values was obtained for the frequency and exposure dependent NPS and DQE over a range of tube voltage from 25 to 30 kVp.  相似文献   

14.
The purpose of this study was to measure experimentally the physical performance of a prototype mammographic imager based on a direct detection, flat-panel array design employing an amorphous selenium converter with 70 microm pixels. The system was characterized for two different anode types, a molybdenum target with molybdenum filtration (Mo/Mo) and a tungsten target with rhodium filtration (W/Rh), at two different energies, 28 and 35 kVp, with approximately 2 mm added aluminum filtration. To measure the resolution, the presampled modulation transfer function (MTF) was measured using an edge method. The normalized noise power spectrum (NNPS) was measured by two-dimensional Fourier analysis of uniformly exposed mammograms. The detective quantum efficiencies (DQEs) were computed from the MTFs, the NNPSs, and theoretical ideal signal to noise ratios. The MTF was found to be close to its ideal limit and reached 0.2 at 11.8 mm(-1) and 0.1 at 14.1 mm(-1) for images acquired at an RQA-M2 technique (Mo/Mo anode, 28 kVp, 2 mm Al). Using a tungsten technique (MW2; W/Rh anode, 28 kVp, 2 mm Al), the MTF went to 0.2 at 11.2 mm(-1) and to 0.1 at 13.3 mm(-1). The DQE reached a maximum value of 54% at 1.35 mm(-1) for the RQA-M2 technique at 1.6 microC/kg and achieved a peak value of 64% at 1.75 mm(-1) for the tungsten technique (MW2) at 1.9 microC/kg. Nevertheless, the DQE showed strong exposure and frequency dependencies. The results indicated that the detector offered high MTFs and DQEs, but structured noise effects may require improved calibration before clinical implementation.  相似文献   

15.
The purpose of this study was to investigate physical characteristics of two full field digital mammography (FFDM) systems (GE Senographe Essential and DS). Both are indirect conversion (x ray to light) alpha-Si flat panels coupled with a CsI(Tl) scintillator. The examined systems have the same pixel size (100 microm) but a different field of view: a conventional size 23 x 19.2 cm2 and a large field 24 X 30.7 cm2, specifically designed to image large breasts. In the GE Senographe Essential model relevant improvements in flat panel design were implemented and new deposition tools for metal, alpha-Si, and CsI(Tl) were introduced by GE. These changes in detector design are expected to be beneficial for advanced applications such as breast tomosynthesis. The presampling modulation transfer function (MTF), normalized noise power spectrum (NNPS), and detective quantum efficiency (DQE) were measured for a wide range of exposure (25-240 microGy) with a RQA-M2 technique (28 kVp with a Mo/Mo target/filter combination and 2 mm of additional aluminum filtration). At 1, 2, and at 4 lp/mm MTF is equal to 0.9, 0.76, and 0.46 for the conventional field detector and to 0.85, 0.59, and 0.24 for the large field detector. The latter detector exhibits an improved NNPS due to a lower electronic noise and a better DQE that reaches 60%. In addition a contrast-detail analysis was performed with CDMAM 3.4 phantom and CDCOM software: GE Senographe DS showed statistically significant poorer detection ability in comparison with the GE Senographe Essential. These results could have been expected, at least qualitatively, considering the relative DQE of the two systems.  相似文献   

16.
The physical characteristics of a clinical charge coupled device (CCD)-based imager (Senovision, GE Medical Systems, Milwaukee, WI) for small-field digital mammography have been investigated. The imager employs a MinR 2000 (Eastman Kodak Company, Rochester, NY) scintillator coupled by a 1:1 optical fiber to a front-illuminated 61 x 61 mm CCD operating at a pixel pitch of 30 microns. Objective criteria such as modulation transfer function (MTF), noise power spectrum (NPS), detective quantum efficiency (DQE), and noise equivalent quanta (NEQ) were employed for this evaluation. The results demonstrated a limiting spatial resolution (10% MTF) of 10 cy/mm. The measured DQE of the current prototype utilizing a 28 kVp, Mo-Mo spectrum beam hardened with 4.5 cm Lucite is approximately 40% at close to zero spatial frequency at an exposure of 8.2 mR, and decreases to approximately 28% at a low exposure of 1.1 mR. Detector element nonuniformity and electronic gain variations were not significant after appropriate calibration and software corrections. The response of the imager was linear and did not exhibit signal saturation under tested exposure conditions.  相似文献   

17.
Samei E  Flynn MJ 《Medical physics》2003,30(4):608-622
Current flat-panel detectors either directly convert x-ray energy to electronic charge or use indirect conversion with an intermediate optical process. The purpose of this work was to compare direct and indirect detectors in terms of their modulation transfer function (MTF), noise power spectrum (NPS), and detective quantum efficiency (DQE). Measurements were made on three flat-panel detectors, Hologic Direct-Ray DR-1000 (DRC), GE Revolution XQ/i (XQ/i), and Philips Digital Diagnost (DiDi) using the IEC-defined RQA5 (approximately 74 kVp, 21 mm Al) and RQA9 (approximately 120 kVp, 40 mm Al) radiographic techniques. The presampled MTFs of the systems were measured using an edge method [Samei et al., Med. Phys. 25, 102 (1998)]. The NPS of the systems were determined for a range of exposure levels by two-dimensional (2D) Fourier analysis of uniformly exposed radiographs [Flynn and Samei, Med. Phys. 26, 1612 (1999)]. The DQEs were assessed from the measured MTF, NPS, exposure, and estimated ideal signal-to-noise ratios. For the direct system, the MTF was found to be significantly higher than that for the indirect systems and very close to an ideal function associated with the detector pixel size. The NPS for the direct system was found to be constant in relation to frequency. For the XQ/i and DRC systems, the DQE results reflected expected differences based on the absorption efficiency of the different detector materials. Using RQA5, the measured DQE values in the diagonal (and axial) direction(s) at spatial frequencies of 0.15 mm(-1) and 2.5 mm(-1) were 64% (64%) and 20% (15%) for the XQ/i system, and 38% (38%) and 20% (20%) for the DRC, respectively. The DQE results of the DiDi system were difficult to interpret due to additional preprocessing steps in that system.  相似文献   

18.
The use of cadmium tungstate (CdWO4) and cesium iodide [CsI(Tl)] scintillation detectors is studied in megavoltage computed tomography (MVCT). A model describing the signal acquired from a scintillation detector has been developed which contains two steps: (1) the calculation of the energy deposited in the crystal due to MeV photons using the EGSnrc Monte Carlo code; and (2) the transport of the optical photons generated in the crystal voxels to photodiodes using the optical Monte Carlo code DETECT2000. The measured detector signals in single CdWO4 and CsI(Tl) scintillation crystals of base 0.275 x 0.8 cm2 and heights 0.4, 1, 1.2, 1.6 and 2 cm were, generally, in good agreement with the signals calculated with the model. A prototype detector array which contains 8 CdWO4 crystals, each 0.275 x 0.8 x 1 cm3, in contact with a 16-element array of photodiodes was built. The measured attenuation of a Cobalt-60 beam as a function of solid water thickness behaves linearly. The frequency dependent modulation transfer function [MTF(f)], noise power spectrum [NPS(f)], and detective quantum efficiency [DQE(f)] were measured for 1.25 MeV photons (in a Cobalt-60 beam). For 6 MV photons, only the MTF(f) was measured from a linear accelerator, where large pulse-to-pulse fluctuations in the output of the linear accelerator did not allow the measurement of the NPS(f). A two-step Monte Carlo simulation was used to model the detector's MTF(f), NPS(f) and DQE(f). The DQE(0) of the detector array was found to be 26% and 19% for 1.25 MeV and 6 MV photons, respectively. For 1.25 MeV photons, the maximum discrepancies between the measured and modeled MTF(f), relative NPS(f) and the DQE(f) were found to be 1.5%, 1.2%, and 1.9%, respectively. For the 6 MV beam, the maximum discrepancy between the modeled and the measured MTF(f) was found to be 2.5%. The modeling is sufficiently accurate for designing appropriate detectors for MVCT.  相似文献   

19.
The imaging performance of an amorphous selenium (a-Se) flat-panel detector for digital fluoroscopy was experimentally evaluated using the spatial frequency dependent modulation transfer function (MTF), noise power spectrum (NPS), and detective quantum efficiency (DQE). These parameters were investigated at beam qualities and exposures within the range typical of gastrointestinal fluoroscopic imaging (approximately 0.1 - 10 microR, 75 kV). The investigation does not take into consideration the detector cover, which in clinical use will lower the DQE measured here by its percent attenuation. The MTF was found to be less than the expected aperture response and the NPS was not white which together indicate presampling blurring. The cause of this blurring was attributed to charge trapping at the interface between two different layers of the a-Se. The effect on the DQE was also consistent with presampling blur, which reduces the aliasing in the NPS and thereby reduces the spatial frequency dependence of the DQE. (The DQE was independent of spatial frequency from 0.12 to 0.73 mm(-1) due to antialiasing of the NPS.) Moreover, the first zero of the measured MTF and the aperture response appeared at the same spatial frequency (6.66 mm(-1) for a pixel of 150 microm). Hence, the geometric fill factor (77%) was increased to an effective fill factor of 99 +/- 1%. A large scale ( approximately 32 pixels) correlation in the noise due to the configuration of the readout electronics caused increased noise power in the gate line NPS at low spatial frequency (< 0.1 mm(-1)). The DQE (f = 0) was exposure independent over a large range of exposures but became exposure dependent at low exposures due to the electronic noise.  相似文献   

20.
Image quality evaluation of a desktop computed radiography system   总被引:2,自引:0,他引:2  
The modulation transfer function (MTF), noise power spectrum (NPS), and detective quantum efficiency (DQE) of the Lumisys ACR-2000 desktop computed radiography (CR) reader were measured and compared to equivalent measurements acquired from a Fuji AC-3 CR system. The one-dimensional (1D) MTF was measured from an image of a sharp edge and the 1D NPS was derived from a 2D NPS measured from a uniform field exposure. The energy dependent ideal input signal to noise ratio of the incident x-ray beams was estimated using published x-ray spectra and attenuation coefficients. Measurements were acquired using Agfa, Fuji, and Kodak storage phosphor plates and it was concluded that use of the Fuji plates resulted in the highest system DQE for the ACR-2000. The DQE was measured using exposures of 0.10, 1.0, and 10.0 mR from 70 and 120 kVp x-ray beams filtered with aluminum. The DQE of the Lumisys ACR-2000 was lower than that of the Fuji AC-3.  相似文献   