1.

PURPOSE

This study evaluated the initial stability of different implants placed above the bone level in different types of bone.

MATERIALS AND METHODS

As described by Lekholm and Zarb, cortical layers of bovine bone specimens were trimmed to a thickness of 2 mm, 1 mm or totally removed to reproduce bone types II, III, and IV respectively. Three Implant system (Brånemark System® Mk III TiUnite™, Straumann Standard Implant SLA®, and Astra Tech Microthread™-OsseoSpeed™) were tested. Control group implants were placed in level with the bone, while test group implants were placed 1, 2, 3, and 4 mm above the bone level. Initial stability was evaluated by resonance frequency analysis. Data was statistically analyzed by one-way analysis of variance in confidence level of 95%. The effective implant length and the Implant Stability Quotient (ISQ) were compared using simple linear regression analysis.

RESULTS

In the control group, there was a significant difference in the ISQ values of the 3 implants in bone types III and IV (P<.05). The ISQ values of each implant decreased with increased effective implant length in all types of bone. In type II bone, the decrease in ISQ value per 1-mm increase in effective implant length of the Brånemark and Astra implants was less than that of the Straumann implant. In bone types III and IV, this value in the Astra implant was less than that in the other 2 implants.

CONCLUSION

The initial stability was much affected by the implant design in bone types III, IV and the implant design such as the short pitch interval was beneficial to the initial stability of implants placed above the bone level.  相似文献   

2.

STATEMENT OF PROBLEM

Macroscopic and especially microscopic properties of implant surfaces play a major role in the osseous healing of dental implants. Dental implants with modified surfaces have shown stronger osseointegration than implants which are only turned (machined). Advanced surface modification techniques such as anodic oxidation and Ca-P application have been developed to achieve faster and stronger bonding between the host bone and the implant.

PURPOSE

The purpose of this study was to investigate the effect of surface treatment of titanium dental implant on implant stability after insertion using the rabbit tibia model.

MATERIAL AND METHODS

Three test groups were prepared: sandblasted, large-grit and acid-etched (SLA) implants, anodic oxidized implants, and anodized implants with Ca-P immersion. The turned implants served as control. Twenty rabbits received 80 implants in the tibia. Resonance frequencies were measured at the time of implant insertion, 2 weeks and 4 weeks of healing. Removal torque values (RTV) were measured 2 and 4 weeks after insertion.

RESULTS

The implant stability quotient (ISQ) values of implants for resonance frequency analysis (RFA) increased significantly (P < .05) during 2 weeks of healing period although there were no significant differences among the test and control groups (P > .05). The test and control implants also showed significantly higher ISQ values during 4 weeks of healing period (P < .05). No significant differences, however, were found among all the groups. All the groups showed no significant differences in ISQ values between 2 and 4 weeks after implant insertion (P > .05). The SLA, anodized and Ca-P immersed implants showed higher RTVs at 2 and 4 weeks of healing than the machined one (P < .05). However, there was no significant difference among the experimental groups.

CONCLUSION

The surface-modified implants appear to provide superior implant stability to the turned one. Under the limitation of this study, however, we suggest that neither anodic oxidation nor Ca-P immersion techniques have any advantage over the conventional SLA technique with respect to implant stability.  相似文献   

3.

PURPOSE

Implant stability quotient (ISQ) values have been supposed to predict implant stability. However, the relationship between ISQ values and bone-to-implant contact ratio (BIC%) which is one of the predictors of implant stability is still unclear. The aim of the present study was to evaluate initial ISQ values in relation to BIC% using rabbit model.

MATERIALS AND METHODS

Four New Zealand white rabbits received a total of 16 implants in their tibia. Immediately after implant placement ISQ values were assessed. The measurements were repeated at the time of sacrifice of the rabbits after 4 weeks. Peri-implant bone regeneration was assessed histomorphometrically by measuring BIC% and bone volume to total volume values (bone volume %). The relationships between ISQ values and the histomorphometric output were assessed, and then, the osseointegration prediction model via the initial ISQ values was processed.

RESULTS

Initial ISQ values showed significant correlation with the BIC%. The bone volume % did not show any significant association with the ISQ values.

CONCLUSION

In the limitation of this study, resonance frequency analysis is a useful clinical method to predict the BIC% values and examine the implant stability.  相似文献   

4.

PURPOSE

The purpose of this study was to assess the difference in efficacy between calcium metaphosphate (CMP)-coated implant fixtures and conventional resorbable blasted media (RBM) processed implant fixtures.

MATERIALS AND METHODS

This study targeted 50 implants from 44 patients who visited Dankook University Dental Hospital. Implantations were done separately for RBM treated and CMP-coated implants, although their design was the same. Calcium metaphosphate has a quicker biodegradation process through hydrolysis compared to other phosphate calcium groups. For the first year of the implantation, the resorption volume of marginal bone analyzed via radiography and perio-test value were measured, under the check plan. Their analyses were composed of a non-inferiority trials test. A 95% level of reliability was used.

RESULTS

In the comparative analysis of the resorption volume of marginal bone and the perio-test value, no statistically significant difference was found between the CMP-coated implants and RBM implants.

CONCLUSION

One year after the implant placement, CMP-coated implants were found not to be inferior to the conventional RBM implants.  相似文献   

5.

PURPOSE

The aim of this study was to investigate a comparison of implant bone bed preparation with Er,Cr:YSGG laser and conventional drills on the relationship between implant stability quotient (ISQ) values and implant insertion variables.

MATERIALS AND METHODS

Forty implants were inserted into two different types of pig rib bone. One group was prepared with conventional drills and a total of 20 implants were inserted into type I and type II bone. The other group was prepared with a Er,Cr:YSGG laser and a total of 20 implants were inserted into type I and type II bone. ISQ, maximum insertion torque, angular momentum, and insertion torque energy values were measured.

RESULTS

The mean values for variables were significantly higher in type I bone than in type II bone (P < .01). In type I bone, the ISQ values in the drill group were significantly higher than in the laser group (P < .05). In type II bone, the ISQ values in the laser group were significantly higher than in the drill group (P < .01). In both type I and type II bone, the maximum insertion torque, total energy, and total angular momentum values between the drill and laser groups did not differ significantly (P ≥ .05). The ISQ values were correlated with maximum insertion torque (P < .01, r = .731), total energy (P < .01, r = .696), and angular momentum (P < .01, r = .696).

CONCLUSION

Within the limitations of this study, the effects of bone bed preparation with Er,Cr:YSGG laser on the relationship between implant stability quotient (ISQ) values and implant insertion variables were comparable to those of drilling.  相似文献   

6.

Objective

The objective of this research was to evaluate implant stability following sinus lift with two grafting materials, and to compare it with the results obtained for the implants placed in a pristine posterior maxilla.

Materials and methods

The study included 44 healthy patients with an existing indication for sinus lift procedure (test group). 46 implants were placed following sinus lift with a pure-phase beta-tricalcium phosphate, while 39 implants were placed following augmentation with 60% hydroxyapatite with 40% beta-tricalcium phosphate material. The control group consisted of 48 healthy patients who were treated with 85 implants but without bone augmentation in posterior maxilla. Astra Tech OsseoSpeed implants were placed in all subjects. Resonance frequency analysis was used in both groups for determining implant stability 4 months after insertion. A mean implant stability quotient (ISQ) was calculated on the basis of 3 measurements.

Results

No statistical difference was observed in ISQ values of implants placed with and without augmentation procedure (p=0,789). Statistically significant difference was not found when ISQ values of implants placed following particular grafting material were compared with ISQ values of corresponding implants in a pristine bone (p=0.697 and p=0.402).

Conclusions

This study demonstrated that the implant stability is comparable among implants placed in the posterior maxilla regardless of sinus lift and grafting procedure. Implants placed in the grafted posterior maxilla can be predictably loaded as the implants placed in a non-grafted, pristine maxilla.Key words: Dental Implants, Sinus Floor Augmentation, Resonance Frequency Analysis, Beta-tricalcium Phosphate, Hydroxyapatite  相似文献   

7.

PURPOSE

This study investigated the influence of bone quality and surgical technique on the implant stability quotient (ISQ) value. In addition, the influence of interfacial bone quality, directly surrounding the implant fixture, on the resonance frequency of the structure was also evaluated by the finite element analysis.

MATERIALS AND METHODS

Two different types of bone (type 1 and type 2) were extracted and trimmed from pig rib bone. In each type of bone, the same implants were installed in three different ways: (1) Compaction, (2) Self-tapping, and (3) Tapping. The ISQ value was measured and analyzed to evaluate the influence of bone quality and surgical technique on the implant primary stability. For finite element analysis, a three dimensional implant fixture-bone structure was designed and the fundamental resonance frequency of the structure was measured with three different density of interfacial bone surrounding the implant fixture.

RESULTS

In each group, the ISQ values were higher in type 1 bone than those in type 2 bone. Among three different insertion methods, the Tapping group showed the lowest ISQ value in both type 1 and type 2 bones. In both bone types, the Compaction groups showed slightly higher mean ISQ values than the Self-tapping groups, but the differences were not statistically significant. Increased interfacial bone density raised the resonance frequency value in the finite element analysis.

CONCLUSION

Both bone quality and surgical technique have influence on the implant primary stability, and resonance frequency has a positive relation with the density of implant fixture-surrounding bone.  相似文献   

8.

PURPOSE

The purpose of this study was to examine the abutment screw stability of screw- and cement-retained implant-supported dental prosthesis (SCP) after simulated cement washout as well as the stability of SCP cements after complete loosening of abutment screws.

MATERIALS AND METHODS

Thirty-six titanium CAD/CAM-made implant prostheses were fabricated on two implants placed in the resin models. Each prosthesis is a two-unit SCP: one screw-retained and the other cemented. After evaluating the passive fit of each prosthesis, all implant prostheses were randomly divided into 3 groups: screwed and cemented SCP (Control), screwed and noncemented SCP (Group 1), unscrewed and cemented SCP (Group 2). Each prosthesis in Control and Group 1 was screwed and/or cemented, and the preloading reverse torque value (RTV) was evaluated. SCP in Group 2 was screwed and cemented, and then unscrewed (RTV=0) after the cement was set. After cyclic loading was applied, the postloading RTV was measured. RTV loss and decementation ratios were calculated for statistical analysis.

RESULTS

There was no significant difference in RTV loss ratio between Control and Group 1 (P=.16). No decemented prosthesis was found among Control and Group 2.

CONCLUSION

Within the limits of this in vitro study, the stabilities of SCP abutment screws and cement were not significantly changed after simulated cement washout or screw loosening.  相似文献   

9.

Purpose

This retrospective chart review study assessed patient records to determine implant insertion torque (IT) and implant stability quotient (ISQ) values during implant placement to evaluate the correlation with cortical bone anchorage (mono- or bicortical).

Methods

Primary stability data (IT during implant placement surgery and ISQ values immediately after implant placement) and cone beam computed tomography of 33 patients (165 implants) were assessed. Patients were divided into the following groups: G1, implants with apical cortical bone contact; G2, implants with bicortical bone contact (apical and cervical regions); and G3, implants with cervical cortical bone contact.

Results

Sixty-eight implants were excluded due to cortical bone contact on regions other than implant apical or cervical. Ninety-seven implants were therefore assessed for this study. No implant failure was found after a mean 70.42-month follow-up time. Implants with bicortical anchorage (G2) showed higher IT (64.1 Ncm) during implant placement and higher ISQ values (76) (p?<?0.05). Monocortical implants (G1, apical, and G3, cervical) showed similar IT (G1 52.3 and G3 54.3) and ISQ values (G1 71.9 and G3 73) (p?>?0.05). No correlation (Pearson correlation coefficient) was found between the two stability measurement devices for the different cortical bone anchorages that were analyzed (G1 0.190, G2 0.039, and G3 ??0.027) (p?>?0.05).

Conclusions

Insertion torque values and implant stability quotients were influenced by cortical bone contact. No significant correlation was found between IT and ISQ values—higher insertion torque values do not necessarily lead to higher implant stability quotients.
  相似文献   

10.

Aim

The aim of this study was to evaluate the implant stability of Bredent Sky Blue implants of different diameters following one stage sinus lift procedure.

Material and methods

This study included 9 male patients with an existing indication for unilateral or bilateral sinus lift procedure. As grafting materials, combination of allograft material MinerOss® cortical & cancellous and Ossceram nano were used.

Results

All implants were considered successful and ISQ levels were measured by Osstell ISQ device. The ISQ values were from 68 to 84. The mean values of stability of Bredent Sky Blue implants of different diameters following one stage sinus lift procedure were 77.73 ± 2.93 (MD) and 77.98 ± 2.72 (VO).Key words: dental implants, implant stability, sinus floor augmentation  相似文献   

11.

PURPOSE

To determine the change in stability of single-stage, three different design of implant systems in humans utilizing resonance frequency analysis for early healing period (24 weeks), without loading.

MATERIAL AND METHODS

Twenty-five patients were included into this study. A total of 45 implants, three different design of implant systems (group A,C,R) were placed in the posterior maxilla or mandible. The specific transducer for each implant system was used. ISQ (implant stability quotient) reading were obtained for each implant at the time of surgery, 3, 6, 8, 10, 12, 24 weeks postoperatively. Data were analyzed for different implant type, bone type, healing time, anatomical locations.

RESULTS

For each implant system, a two-factor mixed-model ANOVA demonstrated that a significant effect on ISQ values (group A = 0.0022, C = 0.017, R = 0.0018). For each implant system, in a two-factor mixed model ANOVA, and two-sample t-test, the main effect of jaw position (P > .005) on ISQ values were not significant.

CONCLUSIONS

All the implant groups A, C and R, the change patterns of ISQ over time differed by bone type. Implant stability increased greatly between week 0 and week six and showed slow increase between week six and six months (plateau effect).  相似文献   

12.

STATEMENT OF PROBLEM

How the ISQ values measured by Osstell™ and Osstell™ Mentor are related, and whether the ISQ values acquired from the two machines changes in accordance with changes in implant stability are not yet fully understood.

PURPOSE

The aim of this study was to find out correlation between the ISQ values acquired from Osstell™ and Osstell™ Mentor, and to evaluate the clinical effectiveness and accuracy of two devices.

MATERIAL AND METHODS

Sixty two implants were inserted into 47 patients, and their ISQ values were measured using Osstell™ and Osstell™ Mentor. In the first stage surgery, the ISQ values of forty four implants inserted into thirty five patients were measured. In the second stage surgery, the values of fifty implants inserted into thirty seven patients were measured. The values were analyzed to determine the difference between the mean ISQ values of Osstell™ and Osstell™ Mentor. In addition, the correlation between implants used in the first and second stage of surgery with regard to their types and areas of insertion were analyzed. The difference between the ISQ values of 32 implants in each patient during the first and second stage was analyzed. The statistical assessment was carried out using SPSS V. 12.0 for Win. (SPSS Inc., Chicago, USA). The Pearson correlation coefficient was used to examine the correlation between Osstell™ and Osstell™ Mentor in the first and second stages of surgery, whereas the difference between their ISQ values was evaluated using a paired t-test.

RESULTS

In the first stage, the mean ISQ value for Osstell™ and Osstell™ Mentor was 70.84 and 75.09, respectively, showing a significant difference (P < .01). In the second stage, the mean ISQ value of Osstell™ and Osstell™ Mentor was 71.76 and 75.94, respectively, also showing a significant difference (P < .01). The difference between the ISQ values in patients in the first and the second stages was significant with both instruments.

CONCLUSION

The significant difference in the values obtained using the Osstell™ and Osstell™ Mentor between the first and second stages of implant surgery indicates that these values can be a convenient and precise way for evaluating the implant stability in clinical practice.  相似文献   

13.

Background

Implant restorations have become a high predictable treatment option. Several caracteristics such as surgical technique and implant design can influence the treatment outcomes. The aim of the present study was to evaluate the influence of implant macro-design on primary stability measured with resonance frequency analysis (RFA) and insertion torque (IT).

Material and Methods

A total of 47 implants divided in two groups: Test group (TI): 22 Tapered MIS® Seven implants; Control group (CI): 25 cylindrical Astra® Osseospeed implants. All implants were inserted following the manufacturers’ standard protocols. Implant primary stability was measured at the moment of implant placement by registering insertion torque values (ITv) and ISQ values by means of Osstell™ Mentor (ISQv) (Integration Diagnostic Ltd., Goteborg, Sweden).

Results

In the mandible, mean ISQv for tapered implants (TI) was 71.67±5.16 and for cylindrical implants (CI) 57.15±4.83 (p=0.01). Mean insertion torque was 46.67±6.85 Ncm for TI and 35.77±6.72 Ncm for CI (p=0.01). In the maxilla, mean ISQ was 67.2±4.42 for tapered implants and 49.17±15.30 for cylindrical implants (p=0.01). Mean insertion torque for TI was 41.5±6.26 Ncm and for CI 39.17±6.34 Ncm (p>0.05). For tapered implants, no correlation could be found between implant diameter and primary stability. But for cylindrical implants there was a statistically significant correlation between implant diameter and primary stability: ITv (p=0.03); ISQv (p=0.04).

Conclusions

Within the limits of the present study, tapered shaped implants achieve higher primary stability measured through ISQ and insertion torque values. Moreover, for cylindrical implants positive correlation has been established between implant diameter and primary stability. Key words:Primary stability, tapered, cylindrical, conical, implant macro-design.  相似文献   

14.

STATEMENT OF PROBLEM

A few of studies which compared and continuously measured the stability of various surface treated implants in the same individual had been performed.

PURPOSE

We aim to find the clinical significance of surface treatments by observing the differences in the stabilization stages of implant stability.

MATERIAL AND METHODS

Eight different surface topographies of dental implants were especially designed for the present study. Machined surface implants were used as a control group. 4 nano-treated surface implants (20 nm TiO2 coating surface, heat-treated 80 nm TiO2 coating surface, CaP coating surface, heat treated CaP coating surface) and 3 micro-treated surface implants [resorbable blast media (RBM) surface, sandblast and acid-etched (SAE) surface, anodized RBM surface] were used as experiment groups. All 24 implants were placed in 3 adult dogs. Periotest® & ISQ values measured for 8 weeks and all animals were sacrificed at 8 weeks after surgery. Then the histological analyses were done.

RESULTS

In PTV, all implants were stabilized except 1 failed implants. In ISQ values, The lowest stability was observed at different times for each individual. The ISQ values were showed increased tendency after 5 weeks in every groups. After 4 to 5 weeks, the values were stabilized. There was no statistical correlation between the ISQ values and PTV. In the histological findings, the bone formation was observed to be adequate in general and no differences among the 8 surface treated implants.

CONCLUSIONS

In this study, the difference in the stability of the implants was determined not by the differences in the surface treatment but by the individual specificity.  相似文献   

15.
Primary stability in low-density bone is crucial for the long-term success of implants. Tapered implants have shown particularly favourable properties under such conditions. The aim of this study was to compare the primary stability of tapered titanium and novel cylindrical zirconia dental implant systems in low-density bone. Fifty implants (25 tapered, 25 cylindrical) were placed in the anterior maxillary bone of cadavers meeting the criteria of low-density bone. The maximum insertion (ITV) and removal (RTV) torque values were recorded, and the implant stability quotients (ISQ) determined. To establish the isolated influence of cancellous bone on primary stability, the implantation procedure was performed in standardized low-density polyurethane foam bone blocks (cancellous bone model) using the same procedure. The primary stability parameters of both implant types showed significant positive correlations with bone density (Hounsfield units) and cortical thickness. In the cadaver, the cylindrical zirconia implants showed a significantly higher mean ISQ when compared to the tapered titanium implants (50.58 vs 37.26; P < 0.001). Pearson analysis showed significant positive correlations between ITV and ISQ (P = 0.016) and between RTV and ISQ (P = 0.035) for the cylindrical zirconia implants; no such correlations were observed for the tapered titanium implants. Within the limitations of this study, the results indicate that cylindrical zirconia implants represent a comparable viable treatment option to tapered titanium implants in terms of primary implant stability in low-density human bone.  相似文献   

16.

PURPOSE

The purpose of this study was to compare the coronal bone level and patient satisfaction in 1-implant and 2-implant assisted mandibular overdentures.

MATERIALS AND METHODS

Twenty patients who had maladaptive mandibular dentures were treated in this study. Patients were randomly divided into two groups. The first group received 1 implant (Simple line II, Implantium, South Korea) in their mandibular midline and the second group received 2 implants in their B and D regions (according to Misch''s category). If the primary stability of each implant was at least 60 ISQ, ball attachment was placed and denture relined with soft liner. After 6 weeks, retentive cap incorporated with hard acrylic resin. In the 6 and 12 months recalls, periapical digital radiograph were made and visual analogue scale questionnaires were used to record patient satisfaction. The Friedman test was done for comparing the presurgical and postsurgical parameters in each group and the U-Mann Whitney test (P<.05) was done for comparison of post-treatment results between the two groups.

RESULTS

All implants achieved sufficient primary stability to be immediately loaded. Patient satisfaction was high, and there were no significant differences between two groups (P>.05). In addition, mean marginal bone loss was 0.6 ± 0.67 mm in the first group and 0.6 ± 0.51 mm in the second group, after 12 month. Mean marginal bone loss showed no significant differences between two groups.

CONCLUSION

This preliminary one-year result indicated that mandibular overdentures anchored to a single implant can be a safe and cost-effective method as a starting step for implant-overdenture treatment.  相似文献   

17.

Objectives

The objectives of this in vitro study were to investigate the sensitivity and reliability of the Osstell™ systems (Resonance Frequency Analysis – RFA) compared to the Periotest® system in implant bone simulated conditions. Three conditions were simulated: (1) the direct fixture-bone contact and fibrous tissue fixture contact, (2) The different levels of horizontal bone loss, and (3) The hardening implant–bone interface.

Materials and methods

Forty-nine dental implant fixtures were placed in the center of acrylic cubes. In Part I seven fixtures were placed in direct contact with acrylic and another seven were placed in contact with polyvinyl siloxane impression material as soft interface. Part II: four sets of 0, 2, 4 and 6 mm horizontally exposed fixture samples were made (seven fixtures in each set). Part III: seven fixtures were placed in contact with a thin mix of autopolymerizing resin. The stability of these fixtures was measured using Osstell™ and Periotest® systems.

Results

The mean Periotest® value(PTV) and Osstell™ measurements showed a significant difference between the direct contact and soft interface (P < 0.001). These values also showed statistically significant difference between the different levels of horizontally exposed fixture groups (P < 0.001). The level of horizontal fixture exposure was strongly correlated with the PTVS (r = 0.967) and strongly negatively correlated with Implant Stability Quotient (r = −0.946). A strong correlation was found between the Osstell™ readings and the change in the stiffness of the autopolymerizing resin fixture interface group (r = 0.986).

Conclusions

Both Osstell™ and Periotest® systems proved to be sensitive in measuring dental implant stability in hard and in soft interfaces. Osstell™ also proved to be sensitive in detecting changes in the fixture interface stiffness. Osstell™ system proved to be more reliable compared to Periotest® system in measuring dental implant stability in hard and in soft interfaces.  相似文献   

18.

PURPOSE

This study aimed to evaluate the effect of implant thread depth on primary stability in low density bone.

MATERIALS AND METHODS

The insertion torque was measured by inserting Ti implants with different thread depths into solid rigid polyurethane blocks (Sawbones) with three different bone densities (0.16 g/cm3, 0.24 g/cm3, and 0.32 g/cm3). The insertion torque value was evaluated with a surgical engine. The static compressive strength was measured with a universal testing machine (UTM) and the Ti implants were aligned at 30° against the loading direction of the UTM. After the static compressive strength test, the Ti implants were analyzed with a Measurescope.

RESULTS

The Ti implants with deeper thread depth showed statistically higher mean insertion torque values (P<.001). Groups A and group B had similar maximum static compressive strengths, as did groups C and D (P>.05). After the static compressive strength, the thread shape of the Ti implants with deeper thread depth did not show any breakage but did show deformation of the implant body and abutment.

CONCLUSION

The implants with deeper thread depth had higher mean insertion torque values but not lower compressive strength. The deep threads had a mechanical stability. Implants with deeper thread depth may increase the primary stability in areas of poor quality bone without decreasing mechanical strength.  相似文献   

19.

STATEMENT OF PROBLEM

Despite an improved bone reactions of Mg-incorporated implants in the animals, little yet has been carried out by the experimental investigations in functional loading conditions.

PURPOSE

This study investigated the clinical and histologic parameters of osseointegrated Mg-incorporated implants in early loading conditions.

MATERIAL AND METHODS

A total of 36 solid screw implants (diameter 3.75 mm, length 10 mm) were placed in the mandibles of 6 beagle dogs. Test groups included 18 Mg-incorporated implants. Turned titanium implants served as control. Gold crowns were inserted 4 weeks after implant placement and the dogs were immediately put on a food diet. Implants were observed for 10 weeks after loading. Radiographic assessments and stability tests were performed at the time of fixture installation, 2nd stage surgery, 4 weeks after loading, and 10 weeks after loading. Histological observations and morphometrical measurements were also performed.

RESULTS

Of 36 implants, 33 displayed no discernible mobility, corresponding to successful clinical function. There was no statistically significant difference between test implants and controls in marginal bone levels (P = .46) and RFA values. The mean BIC% in the Mg-implants was 54.5 ± 8.4%. The mean BIC% in the turned implant was 45.3 ± 12.2%. These differences between the Mg-implant and control implant were statistically significant (P = .005).

CONCLUSIONS

The anodized, Mg-incorporated implant demonstrated significantly more bone-to-implant contact (BIC) in early loading conditions.

CLINICAL IMPLICATIONS

The results of this study in beagle dogs suggest the possibility of achieving predictable stability of early loaded free-standing dental implants with Mg-incorporated surface.  相似文献   

20.

Statement of problem

The success of implants for bone-anchored hearing aids (BAHA) relies on proper osseointegration at the bone-implant interface. Clinical evaluation of implant stability is important in prescribing loading, identifying the risk of failure, and monitoring the long-term health of the implant.

Purpose

The purpose of this in vitro study was to evaluate 2 measurement systems for BAHA implant stability: the most commonly used, Osstell implant stability quotient (ISQ), and a newly developed advance system for implant stability testing (ASIST).

Material and methods

BAHA implants (Oticon Medical Ponto and Cochlear BAHA Connect systems) were installed in plastic materials with adhesive to simulate implants integrated in bone with varying levels of interface stiffness. Different lengths of BAHA abutments were used with each implant specimen, and stability measurements were obtained with both the Osstell ISQ and the ASIST systems. The measurement systems were evaluated in terms of sensitivity to differences in interface stiffness and the effect of abutment length on the stability measurement. Repeated measures ANOVA followed by post hoc t tests were used for the comparisons with a Bonferroni adjusted alpha value of .05/15 = .003 to control for potential type 1 errors.

Results

Changing the abutment length of a single implant installation had minimal effect on the ASIST stability coefficient, whereas large variations were observed in the Osstell implant stability quotient (ISQ). The Osstell showed a clear relationship of decreasing ISQ with increasing abutment length for both the Oticon Medical and the Cochlear implant systems. Both the ASIST and the Osstell were found to be sensitive to changes in interface properties, with the ASIST being more sensitive to these changes.

Conclusions

The ASIST system is more sensitive to changes in interface properties and shows smaller variation because of changes in abutment length than the Osstell ISQ system.  相似文献   