This report describes transmural penetration of the inferior vena cava (IVC) by a newly introduced IVC filter within 9 days of its placement. A computed tomographic study demonstrated filter penetration with one of the primary struts lodging in the uncinate process of the pancreas. Because of the close resemblance of this new filter to another filter that has not been associated with penetration injuries, the key differences between the two designs were examined and the determination was made that the new filter, unlike the older one, has unprotected primary struts. Only filters with an unprotected primary strut design have been associated with penetration injuries such as the one described in this case.  相似文献   

3.
4.

Purpose

Retrieval of optional caval filters may be impaired by filter tilting, migration, fracture, or embedding in the IVC wall. The goal of this experimental study was to evaluate a new optional filter, convertible by unlocking and removing the filter head.

Methods

Forty-nine Pre-Alp sheep (average weight, 55?kg) were anesthetized. IVC was catheterized via the right femoral vein (n?=?46) or via the internal jugular vein (n?=?3) with a 12.9-F sheath. VenaTech? Convertible? IVC filters were inserted as either permanent filters (n?=?14)?or as filters to be converted. Conversion was immediately after deployment (n?=?19) or delayed after 1, 3, or 6?months (n?=?20). Filter delivery, deployment, and conversion with measurement of migration and tilting were evaluated by cavography. Incorporation of the filter’s stabilizers and arms in the IVC wall was assessed by gross anatomy.

Results

Delivery system insertion, filter release, and immediate conversion were successful in all cases. Delayed conversion was completed in all but one sheep, due to insufficient snare tension. Complimentary balloon-catheter inflation was required in 12 of 20 delayed conversions to achieve filter opening. In all 49 sheep, no thrombosis, migration, or significant tilting occurred. Within 4?weeks of conversion, the filter’s stabilizers and arms were incorporated into the IVC wall. Upon removal, the filter head was free of intimal growth.

Conclusions

The VenaTech? Convertible? optional IVC filter was successfully implanted in all sheep with no migration or tilting. Conversion at various dates by filter head removal was feasible in all but one case.  相似文献   

5.
6.
7.
8.
The purpose of the study was to assess the clinical safety and efficacy of the “RecoveryTM” (Bard) inferior vena cava (IVC) filter. We retrospectively evaluated the clinical and imaging data of patients who had a “RecoveryTM” IVC filter placed between January 2003 and December 2004 in our institution. The clinical presentation, indications, and procedure-related complications during placement and retrieval were evaluated. Follow-up computed tomography (CT) examinations of the abdomen and chest were evaluated for filter-related complications and pulmonary embolism (PE), respectively. “Recovery” filters were placed in 96 patients (72 males and 24 females; age range: 16–87 years; mean: 46 years). Twenty-four patients presented with PE, 13 with deep vein thrombosis (DVT) and 2 with both PE and DVT. The remaining 57 patients had no symptoms of thromboembolism. Indications for filter placement included contraindication to anticoagulation (n = 27), complication of anticoagulation (n = 3), failure of anticoagulation (n = 5), and prophylaxis (n = 61). The device was successfully deployed in the infrarenal (n = 95) or suprarenal (n = 1) IVC through a femoral vein approach. Retrieval was attempted in 11 patients after a mean period of 117 days (range: 24–426). The filter was successfully removed in nine patients (82%). Failure of retrieval was due to technical difficulty (n = 1) and the presence of thrombus in the filter (n = 1). One of the nine patients who had the filter removed developed IVC thrombus after retrieval and another had an intimal tear of the IVC. Follow-up abdominal CT (n = 40) at a mean of 80 days (range: 1–513) showed penetration of the IVC by the filter arms in 11, of which 3 had fracture of filter components. In one patient, a broken arm migrated into the pancreas. Asymmetric deployment of the filter legs was seen in 12 patients and thrombus within the filter in 2 patients. No filter migration or caval occlusion was encountered. Follow-up chest CT (n = 27) at a mean of 63 days (range: 1–386) showed PE in one patient (3%). During clinical follow-up, 12 of 96 patients developed symptoms of PE and only 1 of the 12 had PE on CT. There was no fatal pulmonary embolism in our group of patients following “Recovery” filter placement. However, the current version of the filter is associated with structure weakness, a high incidence of IVC wall penetration, and asymmetric deployment of the filter legs.  相似文献   

9.
10.
11.
A 29-year-old woman with acute iliofemorocaval thrombosis was discovered to have suprarenal caval agenesis with azygous continuation, hepatic congestion, and fibrosis as a result of chronic Budd–Chiari syndrome. Three staged procedures were performed: pharmacomechanical thrombolysis of acute thromboses, transfemoral liver biopsy and hemodynamic assessment, and percutaneous endovascular creation of a “neocava” lined with endografts. Symptomatic improvement and patency were maintained at 12-week follow-up.  相似文献   

12.

Purpose

To prospectively assess the Sentry bioconvertible inferior vena cava (IVC) filter in patients requiring temporary protection against pulmonary embolism (PE).

Materials and Methods

At 23 sites, 129 patients with documented deep vein thrombosis (DVT) or PE, or at temporary risk of developing DVT or PE, unable to use anticoagulation were enrolled. The primary end point was clinical success, including successful filter deployment, freedom from new symptomatic PE through 60 days before filter bioconversion, and 6-month freedom from filter-related complications. Patients were monitored by means of radiography, computerized tomography (CT), and CT venography to assess filtering configuration through 60 days, filter bioconversion, and incidence of PE and filter-related complications through 12 months.

Results

Clinical success was achieved in 111 of 114 evaluable patients (97.4%, 95% confidence interval [CI] 92.5%–99.1%). The rate of freedom from new symptomatic PE through 60 days was 100% (n = 129, 95% CI 97.1%–100.0%), and there were no cases of PE through 12 months for either therapeutic or prophylactic indications. Two patients (1.6%) developed symptomatic caval thrombosis during the first month; neither experienced recurrence after successful interventions. There was no filter tilting, migration, embolization, fracture, or caval perforation by the filter, and no filter-related death through 12 months. Filter bioconversion was successful for 95.7% (110/115) at 6 months and for 96.4% (106/110) at 12 months.

Conclusions

The Sentry IVC filter provided safe and effective protection against PE, with a high rate of intended bioconversion and a low rate of device-related complications, through 12 months of imaging-intense follow-up.  相似文献   

13.
The Tempofilter II is a widely used temporary vena cava filter. Its unique design, which includes a long tethering catheter with a subcutaneous anchor, facilitates the deployment and retrieval of the device. Despite this, the Tempofilter II has been used only in the inferior vena cava of patients with lower extremity deep venous thrombosis. In this article, we present a case of superior vena cava filtering using the Tempofilter II in patients with upper extremity deep venous thrombosis.  相似文献   

14.

Purpose

To identify factors associated with advanced inferior vena cava filter (IVCF) retrieval to raise awareness on technical considerations, retrieval efficiency, and patient safety.

Materials and Methods

A single-center retrospective review was performed of 203 consecutive retrievable IVC filters placed between 2007 and 2014. Attempted retrievals were classified as advanced if the routine “snare and sheath” technique was initially unsuccessful after multiple attempts, or an alternate endovascular maneuver or access site was utilized. Patient and filter characteristics were recorded.

Results

203 attempted retrievals were reviewed (48.7 % male, 51.2 % female, mean age 52.7 years, mean dwell time 109 days). Advanced retrievals were observed in 20 patients (9.8 %) (15 females, 5 males). Fluoroscopy time (p ≤ 0.01, 34.3 ± 21.1 and 5.3 ± 4.5 min for advanced retrievals and routine retrievals respectively, same below), gender (p = 0.031), and retrieval tilt angle (p ≤ 0.01, 5.7 ± 5.10° vs. 11.9 ± 11.03°) were associated with advanced retrievals. Females were 3.16 times more likely to have an advanced retrieval performed than males with a significantly higher tilt angle in those with advanced retrieval. History of cancer (p = 0.502), dwell time (p = 0.916), retrieval caval diameter (p = 0.053), placement caval diameter (p = 0.365), filter type (p = 0.710), strut perforation (p = 0.506), placement tilt angle (p = 0.311), and age (p = 0.558) were not found significantly associated with advanced retrievals.

Conclusions

Women are at increased risk for advanced filter retrieval secondary to a significant change in filter tilt over time compared to men, independent of filter type or competing demographic or filter risks, likely placing them at increased risk for higher procedural fluoroscopy times.
  相似文献   

15.

Purpose

To evaluate efficacy and safety of a novel device that combines an inferior vena cava (IVC) filter and central venous catheter (CVC) for prevention of pulmonary embolism (PE) in critically ill patients.

Materials and Methods

In a multicenter, prospective, single-arm clinical trial, the device was inserted at the bedside without fluoroscopy and subsequently retrieved before transfer from the intensive care unit (ICU). The primary efficacy endpoint was freedom from clinically significant PE or fatal PE 72 hours after device removal or discharge, whichever occurred first. Secondary endpoints were incidence of acute proximal deep venous thrombosis (DVT), catheter-related thrombosis, catheter-related bloodstream infections, major bleeding events, and clinically significant thrombus (occupying > 25% of volume of filter) detected by cavography before retrieval.

Results

The device was placed in 163 critically ill patients with contraindications to anticoagulation; 151 (93%) were critically ill trauma patients, 129 (85%) had head or spine trauma, and 102 (79%) had intracranial bleeding. The primary efficacy endpoint was achieved for all 163 (100%) patients (95% confidence interval [CI], 97.8%–100%, P < .01). Diagnosis of new or worsening acute proximal DVT was time dependent with 11 (7%) occurring during the first 7 days. There were no (0%) catheter-related bloodstream infections. There were 5 (3.1%) major bleeding events. Significant thrombus in the IVC filter occurred in 14 (8.6%) patients. Prophylactic anticoagulation was not initiated for a mean of 5.5 days ± 4.3 after ICU admission.

Conclusions

This novel device prevented clinically significant and fatal PE among critically ill trauma patients with low risk of complications.  相似文献   

16.

Purpose

To report technical success, outcomes, and patency of iliocaval stent reconstruction for inferior vena cava (IVC) filter–bearing iliocaval thrombosis.

Materials and Methods

A total of 120 patients with 123 IVC filters and symptomatic iliocaval thrombosis underwent stent reconstruction. Mean patient age was 55 years (range, 19-88 y). Filters included 70 (57%) retrievable and 53 (43%) permanent filters. Symptoms included lower extremity swelling or pain (n = 93), ulcers (n = 8), phlegmasia (n = 7), back pain (n = 5), shortness of breath (n = 4), worsening renal function (n = 2), and stenosis identified during translumbar catheter placement (n = 1). Clinical success was defined as decrease in clinical, etiology, anatomy, and pathophysiology (CEAP) score of at least 1; resolution of presenting symptoms; or normalization of renal function in patients with juxtarenal or suprarenal thrombosis on presentation. Technical aspects of reconstruction, technical success, complications, 6-month clinical response, and 6-, 12-, and 24-month primary, primary-assisted, and secondary stent patency rates were recorded.

Results

Stent reconstruction was technically successful in all 120 patients, 63 of whom (53%) underwent thrombolysis. Thirty filters (24%) were retrieved, and 93 (76%) were excluded with stent placement across the indwelling filter. Six minor and 2 major complications occurred. Clinical success was achieved in 115 patients (96%) at 6 months. Six-, 12-, and 24-month primary iliocaval stent patency rates were 96.4%, 94.8%, and 87.2%, respectively. Twenty-four month primary-assisted and secondary patency rates were 90.3% and 94.2%, respectively.

Conclusions

Iliocaval stent reconstruction is an effective treatment for filter-associated thrombosis with 100% technical success and 96% clinical success at 6 months. Technical and clinical outcomes in patients who underwent filter retrieval versus filter exclusion were similar.  相似文献   