首页 | 本学科首页   官方微博 | 高级检索  
相似文献
 共查询到20条相似文献,搜索用时 15 毫秒
1.
The incidence of major valve-related complications was evaluated in a series of patients in whom the Hancock pericardial xenograft was used for aortic (AVR; n = 84), mitral (MVR; n = 17) and mitral-aortic (MAVR; n = 13) valve replacement. At 7 years actuarial survival is 66% +/- 8% after AVR, 64% +/- 13% after MVR, and 41% +/- 15% after MAVR, whereas actuarial freedom from valve-related death is 79% +/- 7% after AVR, 78% +/- 13% after MVR, and 81% +/- 12% after MAVR. Actuarial freedom from thromboemboli and anticoagulant-related hemorrhage at 7 years is 93% +/- 4% and 98% +/- 2% after AVR and 83% +/- 10% and 88% +/- 11% after MVR; no such complications occurred after MAVR. Structural valve deterioration determined at reoperation, at autopsy, or by clinical investigation was observed in 34 patients with AVR (10.0 +/- 0.2%/patient-year), in 10 with MVR (10.6 +/- 3.3%/patient-year), and in 9 with MAVR (16.6 +/- 5.5%/patient-year). After AVR, 19 patients underwent reoperation and 2 died before reoperation; 4 patients with MVR underwent reoperation, and 7 patients with MAVR underwent reoperation and 1 died before reoperation. Seventy-eight percent of the current survivors (13 patients with AVR, 7 with MVR, and 1 with MAVR) have clinical evidence of valve failure. At 7 years actuarial freedom from structural deterioration of the Hancock pericardial xenograft is 25% +/- 7% after AVR, 29% +/- 14% after MVR, and 0% after MAVR.(ABSTRACT TRUNCATED AT 250 WORDS)  相似文献   

2.
BACKGROUND: Aspire (Tissuemed) bioprosthesis is a third generation porcine bioprosthesis. 10-year outcome of this bioprosthesis is unknown. METHODS: We report our experience of 139 consecutive prosthesis implanted between 1990-1998. The clinical outcome was reviewed retrospectively. RESULTS: 126 patients (67 males and 59 females), mean age 68.4+/-8.4 years underwent 139 valve replacements. Sites of valve implantation included the aortic in 77 patients (61%); mitral in 35 patients (27%); aortic+mitral in 13 patients (10%) and tricuspid in 1 patient (0.8%). 32/126 patients (25%) also underwent concomitant coronary artery bypass grafting (CABG). 30-day mortality for the whole group was 8.7% (11/126) with no valve related deaths. Follow up was 98.4% complete with a mean follow up of 6.1+/-3.3 years (766 patient-years, range 0-10.2 years).Overall 10-year actuarial survival was 41+/-7% (AVR 49+/-10%, MVR 29+/-11%) and this was influenced by pre-operative poor left ventricular function (EF<30%) (p=0.007) and pre-operative NYHA class III/IV (p=0.001).Overall estimated 10-year actuarial freedom from valve related events (Kaplan-Meier) and valve related events expressed as linearised rates (%/patient-year) were: freedom from structural valve failure 97+/-2% (0.26%/patient-year); non-structural dysfunction 98+/-1% (0.13%/patient-year); freedom from prosthetic valve endocarditis 94+/-3% (0.39%/patient-year); freedom from significant haemorrhagic event 82+/-6% (1.33%/patient-year); freedom from thrombo-embolism 90+/-3% (0.91%/patient-year) and freedom from re-operation was 93+/-3% (0.52%/patient-year). CONCLUSION: In our experience Aspire (Tissuemed) porcine bioprosthesis functions satisfactorily at 10-years with low valve related complications. Further follow-up will determine its long-term durability.  相似文献   

3.
The Carpentier-Edwards porcine valve prosthesis has afforded our patients a satisfactory quality of life and a low incidence of valve-related complications at follow-up periods of up to five years. From December, 1975 to March, 1980, 768 prostheses were implanted in 700 patients (aortic valve replacement [AVR], 334; mitral valve replacement [MVR], 292; tricuspid valve replacement [TVR], 6; and multiple valve replacement, 68). One hundred and thirty-seven patients (19.6%) had had previous cardiac operations. Concomitant aortocoronary bypass was performed in 127 patients (18.1%). There were 52 hospital deaths, for a mortality of 7.4% (AVR, 4.8%; MVR, 9.2%; multiple valve replacement, 11.8%). Total follow-up was 1,047 patient-years (range, 6 to 60 months, mean, 19.4 months). There were 33 late deaths (AVR, 1.7% per patient-year; MVR, 4.0% per patient-year; multiple valve replacement, 8.1% per patient-year). Eight percent of AVR patients and 47% of MVR and multiple valve replacement patients were taking anticoagulants. The valve-related complications (expressed as events per 100 patient-years) were as follows: (1) thromboembolism (AVR, 0.94; MVR, 1.42; multiple valve replacement, 4.62); (2) infective endocarditis (AVR, 0.94; MVR, 0.24; multiple valve replacement, 2.31); (3) periprosthetic leak (AVR, 0.94; MVR, 0.71; multiple valve replacement, 3.46); and (4) valve dysfunction (MVR, 0.24). The only case of valve dysfunction was a calcified mitral prosthesis in a 13-year-old girl. Actuarial survival, including operative deaths, was as follows: AVR, 90.5% at 36 months; MVR, 84% at 36 months; and multiple valve replacement, 74% at 24 months. Of surviving patients, 93.6% were in New York Heart Association Class I or II at follow-up evaluation.  相似文献   

4.
Between 1983 and 1987 the Mitroflow pericardial prosthesis was implanted in 354 patients, ranging in age from 14 to 94 years (mean 60.1 years). The early mortality was 5.7% and the late mortality 2.9% per patient-year. The total cumulative follow-up was 853 years (mean 2.4 years). Patient survival at 4 years for aortic valve replacement (AVR) was 81.5% and for mitral valve replacement (MVR) 74.8%. The overall rate of valve-related complications was 4.8% per patient-year (41 complications): thromboembolism, 15; hemorrhage related to antithromboembolic therapy, 1; prosthetic valve endocarditis, 15; nonstructural dysfunction, 3; and structural valve deterioration, 7. At 4 years the freedom from thromboembolism was 91.5% +/- 2.7% for AVR and 91.1% +/- 4.0% for MVR, from prosthetic valve endocarditis 93.5% +/- 2.3% for AVR and 94.0% +/- 2.9% for MVR, from structural valve deterioration 97.3% +/- 2.1% for AVR and 92.6% +/- 3.2% for MVR, from valve-related mortality 96.9% +/- 1.4% for AVR and 97.5% +/- 1.8% for MVR, and from reoperation 93.5% +/- 2.8% for AVR and 83.1% +/- 5.1% for MVR. The freedom from the composite of all valve-related complications at 4 years was 81.1% +/- 4.2% for AVR and 75.3% +/- 2.8% for MVR. The Mitroflow valve has provided satisfactory clinical performance at the 4-year evaluation. Structural valve deterioration is greater in the mitral position than in the aortic position. Long-term evaluation of the Mitroflow valve is necessary to determine the impact of structural valve deterioration on its clinical performance.  相似文献   

5.
We examined hematological complications in 415 patients having valve replacement with the St. Jude mechanical prosthesis (212, aortic valve replacement [AVR]; 159, mitral valve replacement [MVR]; and 44, AVR + MVR). There were 164 men and 251 women with a mean age of 59 years (range, 20 to 88 years). Preoperatively 386 patients were in New York Heart Association functional classes III and IV. There were 154 associated procedures (37%), the most common being myocardial revascularization. Overall hospital mortality was 7.5% (31/415), 7% after AVR, 8% after MVR, and 7% after AVR + MVR. All operative survivors were anticoagulated with Coumadin (crystalline warfarin sodium) to maintain the prothrombin time at 1.5 times control. During a mean follow-up of 21 months (range, 6 to 60 months), there were 29 late deaths (7.6%) and 5 patients (1.3%) lost to follow-up. No patient experienced structural valve degeneration. At 48 months, actuarial freedom from thromboembolism was 87% +/- 3% after AVR and 91% +/- 9% after MVR; from anticoagulation-related hemorrhage, 97% +/- 3% after AVR and 91% +/- 3% after MVR; and from hemolysis, 100% after AVR and 98% +/- 2% after MVR. Freedom from all valve-related morbidity at 4 years was 82% +/- 5% after AVR and 75% +/- 10% after MVR. Actuarial survival at 48 months was 80% +/- 4% after AVR and 65% +/- 7% after MVR.  相似文献   

6.
T E David  S Armstrong  Z Sun 《The Annals of thoracic surgery》1992,54(4):661-7; discussion 667-8
The Hancock II bioprosthesis was used for heart valve replacement in 614 patients from 1982 to 1990. Aortic valve replacement (AVR) was performed in 376 patients, mitral valve replacement (MVR) in 195, and aortic and mitral valve replacement (DVR) in 43. The mean age was 62.7 years, and 78% of all patients were in New York Heart Association functional class III or IV before operation. Coronary artery bypass graft was necessary in 232 patients and replacement of ascending aorta in 55. There were 31 operative deaths (AVR, 4%; MVR, 6%; DVR, 9%). Follow-up was complete in 98.5% of the patients and extended from 12 to 103 months, with a mean of 49 months. At the last follow-up, 85% of the patients were in New York Heart Association class I or II. The actuarial survival at 8 years was 79% +/- 3% for AVR, 68% +/- 4% for MVR, and 65% +/- 10% for DVR. The freedom from stroke at 8 years was 93% +/- 2% for AVR, 83% +/- 5% for MVR, and 90% +/- 5% for DVR. At the end of 8 years 96% +/- 1% of all patients were free from endocarditis, 92% +/- 1% were free from primary tissue failure, and 89% +/- 3% were free from reoperation. The actuarial freedom from valve-related death at 8 years was 98% +/- 1% for AVR, 86% +/- 5% for MVR, and 91% +/- 6% for DVR. Hemodynamic assessment was obtained by Doppler echocardiography in all operative survivors and demonstrated satisfactorily effective valve orifices and transvalvular gradients. The clinical results obtained with the Hancock II bioprosthesis have been gratifying, particularly in the aortic position. This bioprosthesis is our biological valve of choice.  相似文献   

7.
Omniscience valves were implanted in sixty-two patients. Twenty-eight of these patients underwent aortic valve replacement (AVR), 15 had mitral valve replacement (MVR) and 8 had aortic and mitral valve replacement (DVR). Post-operative events occurred in nine (5.9%/patient year) of the AVR group, in three (1.7%/patient year) of the MVR group and in three (5.4%/patient year) of the DVR group. The actuarial freedom from all events at five years in the AVR, MVR and DVR was 74 +/- 8%, 88 +/- 6%, 67 +/- 16%, respectively. Cardiac death occurred in four (2.5%/patient year) of the AVR, one (0.6%/patient year) of the MVR and two (3.6%/patient year) of the DVR. The freedom at five years in the AVR, MVR and DVR was 88 +/- 6%, 96 +/- 4%, and 77 +/- 14%, respectively. Valve-related complications were noted in four patients. Post-operative cerebral hemorrhage was seen in three of the AVR. Maximum opening angle of the Omniscience valve was 39.1 +/- 4.5 degrees at the aortic position and 44.6 +/- 9.7 degrees at the mitral position. Omnicarbon valves implanted in ninety-five patients, fifty-eight of these patients underwent AVR, 24 had MVR and 13 had DVR. Events occurred post-operatively in four (2.6%/patient year) of the AVR group, in three (12.2%/patient year) of the MVR group, but in none of the DVR group. The freedom at five years was 89 +/- 6% in the AVR and 84 +/- 8% at three years in the MVR. Post-operative cardiac death occurred in one (0.7%/patient year) of the AVR and in two (8.1%/patient year) of the MVR.(ABSTRACT TRUNCATED AT 250 WORDS)  相似文献   

8.
BACKGROUND: All patients undergoing St. Jude Medical valve replacement at the Medical University of South Carolina since January 1979 have been followed prospectively at 12-month intervals. METHODS: This report describes long-term experience in 710 adult patients undergoing isolated aortic (AVR) (418) or mitral valve replacements (MVR) (292) with this prosthesis from January 1979 to December 1996. RESULTS: Ages ranged from 19 to 84 years (54.8 +/- 15.1 AVR, 51.8 +/- 12.9 MVR; mean +/- SD). Male gender predominated in the AVR group (70%) and female gender in the MVR group (62%). One hundred and fifty-seven patients (22%) had associated coronary artery bypass grafting (AVR 27%, MVR 15%). Thirty-day operative mortality was 5.3% (22/418) in the AVR group and 5.1% (15/292) in the MVR group. Follow-up is 96.9% complete and ranges from 1 month to 16.9 years (AVR, 2,376 patient-years, mean 5.7 +/- 4.5 years; MVR, 1,868 patient-years, mean 6.4 +/- 4.8 years). In the AVR group, 120 late deaths have occurred and actuarial survival was 78.0 +/- 2.3%, 58.0 +/- 3.2%, and 36.8 +/- 4.8%; at 5, 10, and 15 years, respectively. Forty-six patients have sustained 55 thromboembolic (TE) events (2.3%/patient-year). Fifty-one patients had anticoagulant-related bleeding complications (2.7%/patient-year). The mean improvement in New York Heart Association (NYHA) functional class from preoperative to postoperative was 3.0 +/- 0.8 to 1.7 +/- 0.1 (p < 0.05). In the MVR group, there have been 84 late deaths, and the actuarial survival was 79.3 +/- 2.5%, 60.1 +/- 3.5%, and 49.3 +/- 4.1% at 5, 10, and 15 years, respectively. Fifty-two patients have had 64 TE events (3.5%/patient-year). Twenty-three patients had anticoagulant-related bleeding complications (1.6%/patient-year). The mean improvement in NYHA functional class was from 3.3 +/- 0.6 to 1.8 +/- 0.1. There were no mechanical failures in either group. CONCLUSIONS: With a follow-up now extending to 17 years, the St. Jude Medical valve continues to be a reliable mechanical prosthesis with low and stable rates of valve-related complications.  相似文献   

9.
We analyzed midterm results using the Bicarbon valve in a single center. Forty-four patients had aortic valve replacement (AVR), 48 had mitral valve replacement (MVR), and 13 had both aortic and mitral valve replacement (DVR). The mean age of the 105 patients was 61.2 +/- 11.3 years. The mean follow-up was 1.8 +/- 1.1 years with a cumulative follow-up of 188 patient-years. There were 5 early deaths (4.7%: 4 in the AVR group and 1 in the MVR group) and 5 late deaths (2.7% per patient-year: 3 malignancy, 1 cerebral hemorrhage, 1 myocardial infarction). Survival at 3 years was 91 +/- 4% in the AVR group, 92 +/- 5% in the MVR group, and 66 +/- 23% in the DVR group. The linearized incidence of thromboembolic complications, hemorrhagic complications, and paravalvular leaks in all patients was 1.06 +/- 2.34%, 1.60 +/- 2.53%, and 0.53 +/- 2.22% per patient-year, respectively. No other complications were observed. In conclusion, the Bicarbon prosthetic heart valve has shown excellent clinical results associated with a low incidence of valve-related complications.  相似文献   

10.
From April 1, 1979 to August 31, 1983, 228 patients underwent isolated aortic (AVR) (118) or mitral (MVR) (90) valve replacements with a new tilting disc valve prosthesis, the St. Jude prosthesis, at the Medical University of South Carolina. Age ranged from 6 to 84 years (mean 49.1 +/- 19.2 AVR, 44.5 +/- 16.5 MVR). Male sex predominated in the AVR group (68%) and female sex in the MVR group (68%). Thirty-five patients (16.8%) had associated coronary bypass surgery (AVR 23.7%, MVR 7.8%). There were seven deaths (3.4%) occurring during the same hospitalization (AVR: 3/118, 2.5%; MVR: 4/90, 4.4%). Follow-up is 97.6% complete and ranges from 1 to 54 months (mean 19.6 +/- 12.4). In the AVR group, nine late deaths have occurred and actuarial survival at 42 months is 86.7 +/- 3.8%. Three patients have sustained thromboembolic episodes for a linearized rate of 1.6% patient-year, and the probability of remaining free of thromboembolism at 42 months is 96.9 +/- 1.8%. The mean improvement in functional class from preoperative to postoperative is 3.1 +/- 0.7 to 1.2 +/- 0.4 (p less than 0.001). In the MVR group, there have been four late deaths, and the actuarial survival at 42 months is 89.3 +/- 3.8%. Two patients have sustained thromboembolic complications for a linearized rate of 1.2%/patient-year, and the probability of remaining free of thromboembolism at 42 months is 97.2 +/- 2%. The mean improvement in functional class from before to after surgery is 3.2 +/- 0.7 to 1.3 +/- 0.5 (p less than 0.001). There have been no thromboses of the St. Jude valve in the mitral or aortic position, no mechanical failures, and no patient has had significant valve-related hemolysis. Because of this experience, the St. Jude Medical heart valve prosthesis is our prosthesis of choice for any patient undergoing valve replacement with a mechanical prosthesis.  相似文献   

11.
Long-term evaluation of patients undergoing combined mitral and aortic valve replacement (MVR + AVR) with a porcine bioprosthesis provides the opportunity for a direct comparison of the durability of the mitral versus the aortic porcine bioprosthesis in the same patient. From 1970 to 1983, 71 patients underwent MVR + AVR with Hancock porcine bioprostheses. There were 46 men an 25 women ranging in age from 21 to 64 years (mean, 47.5 +/- 5 years). Sixteen patients (22.5%) died at operation. The survivors were followed from 0.2 to 11.5 years (mean, 5.7 +/- 3 years). Duration of follow-up was 313 patient-years and was 100% complete. Overall late mortality was 6.7 +/- 1.4% per patient-year (linearized incidence), and actuarial survival was 54.2 +/- 8% at 11 years. Endocarditis occurred in 4 patients (linearized incidence of 1.3 +/- 0.6% per patient-year); thromboembolic events were sustained by 4 patients (linearized incidence of 1.3 +/- 0.6% per patient-year); the event was fatal in 1 patient. Actuarial freedom from thromboembolism was 90 +/- 4.8% at 11 years. Reoperation for primary tissue failure was performed in 11 patients (linearized incidence of 3.5 +/- 1% per patient-year) with no deaths; in 7 patients both bioprostheses were explanted, and in 4, only the mitral bioprosthesis was replaced. The durability of explanted aortic and mitral porcine bioprostheses was not significantly different, and the evaluation of seven pairs of explanted aortic and mitral bioprostheses showed similar amounts of calcification. Actuarial freedom from reoperation because of primary tissue failure was 44.6 +/- 13.7% at 11 years.(ABSTRACT TRUNCATED AT 250 WORDS)  相似文献   

12.
AIM: The goal of aortic valve replacement (AVR) surgery in the elderly (= or >75 years) is to extend survival and minimize valve-related morbidity, mortality and reoperation. As the elderly population lives longer, those with implanted valves are at risk of suffering valve related complications. We hypothesize that bioprostheses are appropriate for the elderly. METHODS: The follow-up evaluation of 966 patients with valves (AVR, 666; mitral valve replacements [MVR], 226; multiple valve replacements [MR], 74) implanted between 1975 and 1999 was examined. There were 879 bioprotheses (BP) and 87 mechanical prostheses (MP). The mean age was 78.9+/-3.3 years (range 75-94.6 years). Concomitant coronary artery bypass was performed in AVR in 51.7%, MVR in 50.4% and MR in 28.4%. Valve type, valve lesion, coronary artery bypass (previous/concomitant), age and gender were considered as independent predictors of composites and survival. The total follow-up was 3905 patient-years. RESULTS: Early mortality was for AVR 9.6% (64), MVR 15.0% (34) and MR 25.7% (19). The late mortality was for AVR 8.8%, MVR 10.4% and MR 8.8%/patient-year. The only independent predictor of survival and valve-related mortality, morbidity and reoperation was age for survival in those with AVR, hazard ratio 1.15 [CL 1.03-1.27] p=0.0094). The BP reoperative rate was 0.5%/patient-year (reoperation was fatal in 6/15) of total, MP reoperative rate was 0% [reasons for reoperation structural valve deterioration (4), non-structural dysfunction (6), prosthetic valve endocarditis (5), reoperation fatality due to non-structural dysfunction (2), prosthetic valve endocarditis (4)]. Overall patient survival at 10 and 15 years, respectively, was 30.5+/-2.4% and 3.6+/-2.2% irrespective of valve position and type. Overall actual and actuarial freedom from valve-related morbidity at 15 years was 96.8+/-0.9% and 93.7+/-2.3%, respectively. Actual and actuarial overall freedom from valve-related mortality at 15 years was 84.3+/-2.4% and 58.4+/-0.9%, respectively. Overall actual and actuarial freedom from valve related reoperation at 15 years was 95.8+/-1.6% and 74.8+/-16.9%, respectively. CONCLUSIONS: BP valves are further confirmed to be a good option for AVR in patients = or >75 years of age.  相似文献   

13.
Between November, 1978, and December, 1983, 736 patients had valve replacement with the St. Jude Medical valve prosthesis. There were 478 patients with aortic valve replacement (AVR), 188 with mitral valve replacement (MVR), 63 with double valve replacement, and 7 with tricuspid valve replacement (they were not included in this study). The mean age at the time of operation was 46.7 years for patients having AVR and 48.6 years for those having MVR and AVR + MVR. Follow-up totaled 1,116 patient-years (range, 4 to 82 months). Early (30-day) mortality was lowest for isolated MVR (2.3%) and AVR (3.7%), and increased with reoperation or when associated procedures were combined with valve replacement. Patients undergoing reoperation or having associated procedures made up 49% of the AVR and 54% of the MVR groups. All patients were advised of the need for long-term anticoagulation with warfarin sodium. Nine patients (7 with AVR, 1 with MVR, 1 with AVR + MVR) had suspected or confirmed episodes of systemic thromboembolism, a linearized incidence of 0.99% per patient-year for AVR, 0.36% per patient-year for MVR, and 0.98% per patient-year for AVR + MVR. Eight patients with AVR underwent reoperation for prosthetic valve endocarditis (5 of the 8 patients had endocarditis prior to initial valve replacement). There were no instances of structural valve failure. There were 37 late deaths. Actuarial survival at 5 years (excluding early mortality, 95% confidence limits) was 89.8% for AVR, 84.8% for MVR, and 95.2% for AVR + MVR.(ABSTRACT TRUNCATED AT 250 WORDS)  相似文献   

14.
All patients undergoing a heart valve replacement with a glutaraldehyde-preserved Hancock xenograft from June, 1974, through June, 1977, were reviewed. This analysis included 403 patients: 202 having mitral valve replacement (MVR), 131 having aortic valve replacement (AVR), and 70 having mitral and aortic valve replacement (M/AVR). The hospital mortality was 9.9% (40/403): 10.8% (22/202) for MVR, 4.5% (6/131) for AVR, and 17.1% (12/70) for M/AVR. Of 363 patients discharged from the hospital, six were lost to follow-up at 24 and 37 months and they are excluded from the figures. There were 41 late deaths, which represents an incidence of 1.47% per patient-year for MVR (16/177), 1.95% per patient-year for AVR (14/122), and 3.35% per patient-year for the M/AVR (11/58). There were 37 thromboembolic events in 32 patients. This represents a linearized incidence of 1.93%, 0.97%, and 2.74% per patient-year for MVR, AVR, and M/AVR, respectively. None of the accidents was fatal, and four of the 32 patients were on a regimen of controlled anticoagulation at the time of embolism. Primary tissue failure was observed in 34 patients (231 MVR, eight AVR, and five MAVR). The probability of being free from primary tissue failure at 8 years is 85.3% +/- 3.7% for MVR, 85% +/- 11.6% for AVR, and 81% +/- 8.9% for M/AVR. These current results indicate that the Hancock xenograft valve can be considered as a valid alternative in heart valve replacement.  相似文献   

15.
The Duromedics (Baxter Healthcare Corp., Edwards CVS Div., Irvine, Calif.) mechanical cardiac valvular prosthesis was implanted in 480 patients between 1984 and 1987 at the Montreal Heart Institute, the Hospital Clinic of Barcelona and the teaching hospitals of the University of British Columbia. The mean age of the patients was 52 years. The early mortality was 7.9% and the late mortality was 4.1% per patient-year. The overall survival at 4 years for aortic valve replacement (AVR) was 87.0% +/- 3.7% and for mitral valve replacement (MVR) was 81.9% +/- 2.9%. There were 16 valve-related reoperations in 14 patients--for prosthetic valve endocarditis in 9 patients, for thromboembolism in 1 patient and for nonstructural dysfunction in 4 patients. The freedom from thromboembolism at 3 and 4 years was 94.3% +/- 3.1% for AVR and 95.1% +/- 1.8% for MVR. The freedom from prosthetic valve endocarditis at 3 and 4 years was 95.3% +/- 2.2% for AVR and 96.2% +/- 1.6% for MVR. The freedom from structural valve deterioration for all positions was 100%. The freedom from reoperation at 4 years was 95.3% +/- 2.1% for AVR and 92.3% +/- 4.3% for MVR and from valve-related death was 98.7% +/- 1.3% for AVR and 96.2% +/- 1.6% for MVR. The freedom from all valve-related complications at 4 years was 87.7% +/- 3.8% for AVR and 85.7% +/- 2.9% for MVR. Long-term evaluation of the Duromedics prosthesis is required to determine the influence of documented structural valve deterioration.  相似文献   

16.
Clinical results with porcine bioprostheses were reviewed for 990 patients who underwent heart valve replacement from January, 1974, to December, 1980. Eight hundred and seventy-four Hancock, 283 Carpentier-Edwards, and 10 Liotta bioprostheses were used. In 23 patients, 26 mechanical prostheses were implanted as well. Overall operative mortality was 60 out of 990 (6.06%): 30 out of 506 (5.9%) for mitral valve replacement (MVR), 13 out of 287 (4.5%) for aortic valve replacement (AVR), 1 out of 4 (25%) for tricuspid valve replacement, 0 out of 2 for pulmonary valve replacement, and 16 out of 191 (8.4%) for multiple valve replacement. Cumulative follow-up covered 1,793 patient-years. (Actuarial survival at 7 years was 76.6 +/- 3% for MVR. At 6 years, it was 83.2 +/- 2.8% for AVR and 55 +/- 13.5% for multiple valve replacement.) Prosthesis-related survival at 7 years was 91.7 +/- 1.9% for MVR, and at 6 years, it was 96.6 +/- 1.5% for AVR and 95.1 +/- 2.2% for multiple valve replacement. Bioprosthesis survival, considering deaths or complications that led to reoperation as final events, was 84.2 +/- 3.7% at 7 years for mitral valves and 87.7 +/- 3.8% at 6 years for aortic valves. Emboli per 100 patient-years numbered 3.2 for MVR, 0.5 for AVR, and 1.6 for multiple valve replacement. Twenty-seven patients underwent reoperation, 12 for perivalvular leak, 5 for endocarditis, 6 for valve thrombosis, and 4 for primary tissue failure (linearized rates of 0.7, 0.3, 0.3, and 0.2% per patient-year, respectively).(ABSTRACT TRUNCATED AT 250 WORDS)  相似文献   

17.
Two hundred and two patients (97 female and 105 male; mean age: 45. 5+/-9 years) received CarboMedics bileaflet valves during a period of eight years. Ninety-one patients received mitral, 72 aortic and 39 aortic+mitral valve prosthesis. Tricuspid plasty and coronary artery bypass surgery were the concomitant operations in 17 and 12 patients, respectively. The mean follow-up period was 24.7 months and the ratio was 91%. Overall operative mortality was 3.96% (8 patients); 2.78% for aortic valve replacement (AVR), 3.29% for mitral valve replacement (MVR) and 7.7% for double valve replacement (DVR). The late mortality rate was 2.89% for AVR, 2.2% for MVR and 8. 3% for DVR. The main cause of mortality was low cardiac output. The overall survival rate was 91.5% in 2 years. The actuarial freedom from thromboembolism in 2 years was 97% for AVR, 95% for MVR and 84% for DVR. No mortality due to heamorrhagic events was observed. CarboMedics prosthetic heart valves may be used satisfactorily with a low incidence of valve-related morbidity and mortality.  相似文献   

18.
All patients who had a mitral or aortic Hancock valve replacement between June, 1974, and June, 1979, were reviewed. A total of 734 bioprostheses were implanted in 632 patients: 291 had mitral (MVR), 239 had aortic (AVR), and 102 had both mitral and aortic valve replacement (MVR + AVR). In 228 patients, an associated surgical procedure was necessary. It involved conservative valve operation in 205 of them. The hospital mortality was 9.6% (28) for MVR (11.5% with associated operation), 4.6% (11) for AVR (8.7% with associated operation), and 13.7% (14) for MVR + AVR (13.0% with associated operation).The follow-up period was between 1 and 6 years with a total follow-up of 934.6, 714.6, and 288.3 patient-years for MVR, AVR, and MVR + AVR, respectively. The late mortality was 0.96% (9), 1.53% (11), and 2.08% (6) per patient-year for MVR, AVR, and MVR + AVR, respectively. The thromboembolic rate was 1.49%, 0.14%, and 2.08% per patient-year for MVR, AVR, and MVR + AVR, respectively. There were twelve valve failures (six were due to rupture; four, thrombosis; one, insufficiency because of intrinsic failure; and one, stenosis without evident cause at reoperation). This represents a failure rate of 0.53%, 0.13%, and 2.08% per patient-year for MVR, AVR, and MVR + AVR, respectively. These results encourage us to continue our routine use of the glutaraldehyde xenografts as the safest valve substitute at present.  相似文献   

19.
BACKGROUND: From February 1985 to December 1994, 781 Omnicarbon valve prostheses were implanted in 647 patients. These were 357 male and 290 female patients with a mean age of 53.5+/-10.5 years (range, 4 to 78 years). Before operation, 81% of the patients were in New York Heart Association class III or IV, 16% were in class II, and only 3% were in class I. METHODS: There were 227 aortic valve replacements (AVR) (35%), 286 mitral valve replacements (MVR) (44%), and 134 double-valve replacements (DVR) (21%) (AVR + MVR). Follow-up was 96.3% complete and consisted of 2,746 patient-years (mean follow-up, 4.6 years, and maximum follow-up, 10.7 years). RESULTS: Hospital mortality rates were 7.0% for AVR, 8.0% for MVR, and 8.2% for DVR. The annualized rate of anticoagulant-related hemorrhage was 0.8% per patient-year, and thromboembolism occurred at a rate of 0.7% per patient-year. No structural failure was observed during 10-year follow-up. Twenty-one instances of nonstructural dysfunction (two, pannus growth, and 19, dehiscence) of the Omnicarbon valve occurred in 20 patients, an incidence of 0.8% per patient-year. Hemolytic anemia was observed only in the presence of valvular dehiscence (6 of 19). Eight patients (0.3% per patient-year) had development of prosthetic valve endocarditis (4, AVR; 2, MVR; and 2 DVR). At the end of 10 years of follow-up, 91% of the survivors were in New York Heart Association class I or II. The overall survival rate at 10 years was 82.5%+/-2.6% (85.0%+/-3.9%, AVR; 81.0%+/-4.1%, MVR; and 82.5%+/-2.6%, DVR). Considering only valve-related deaths, the survival rate at 10 years was 91.9%+/-2.4% (90.0%+/-2.7%, AVR; 93.1%+/-3.8%, MVR; and 90.0%+/-1.8%, DVR). CONCLUSIONS: Clinical results over a 10-year follow-up are excellent with the Omnicarbon prosthesis.  相似文献   

20.
BACKGROUND: In this study, we present a single center experience with Bicarbon bileaflet valve in 307 patients with rheumatic heart disease. METHODS: Between August 1998 and September 2000, 307 patients underwent heart valve replacement using the Bicarbon bileaflet valve (Sorin Biomedica, Saluggia, Italy) with an average age of 47.19+/-13.21 years (range 14 to 80 years), consisting of 147 males and 160 females at Alkan Hospital, Cardiovascular Surgery Department. Aortic valve replacement (AVR) was performed in 77 patients, mitral valve replacement (MVR) in 156 patients and double valve replacement (DVR) in 74 patients. RESULTS: The early mortality rate was 3.3% (10/307) and there was no late mortality. The actuarial survival rate, including hospital mortality, was 96.74+/-1.01% for the whole group, 96.5+/-1.5% for the MVR group, 97.4+/-1.8% for the AVR group and 97.3+/-1.9% for the DVR group at 35 months. One patient had obstructive valve thrombosis with MVR. The 35 months actuarial freedom from valve thrombosis was 99.58+/-0.4% for the whole group. Four patients were reoperated and the 35 months actuarial freedom from reoperation was 98.53+/-0.7% for the whole group, 98.65+/-0.9% for the MVR group, 96.73+/-02% for the DVR group and 100% for the AVR group. No instances of perivalvular leak, hemolysis, endocarditis or embolism were observed during the entire follow-up period. Mean follow-up duration was 16.5+/-7.9 months (ranged 4 to 35 months). CONCLUSION: We have presented our mid-term results with the Sorin Bicarbon bileaflet valve in patients with rheumatic heart disease, which provided good clinical performance combined with meticulous patient care and advanced surgical techniques.  相似文献   