首页 | 本学科首页   官方微博 | 高级检索  
相似文献
 共查询到20条相似文献,搜索用时 15 毫秒
1.
Cavitation memory effects occur when remnants of cavitation bubbles (nuclei) persist in the host medium and act as seeds for subsequent events. In pulsed cavitational ultrasound therapy, or histotripsy, this effect may cause cavitation to repeatedly occur at these seeded locations within a target volume, producing inhomogeneous tissue fractionation or requiring an excess number of pulses to completely homogenize the target volume. We hypothesized that by removing the cavitation memory, i.e., the persistent nuclei, the cavitation bubbles could be induced at random locations in response to each pulse; therefore, complete disruption of a tissue volume may be achieved with fewer pulses. To test the hypothesis, the cavitation memory was passively removed by increasing the intervals between successive pulses, ?t, from 2, 10, 20, 50 and 100, to 200 ms. Histotripsy treatments were performed in red blood cell tissue phantoms and ex vivo livers using 1-MHz ultrasound pulses of 10 cycles at P−/P+ pressure of 21/59 MPa. The phantom study allowed for direct visualization of the cavitation patterns and the lesion development process in real time using high-speed photography; the ex vivo tissue study provided validation of the memory effect in real tissues. Results of the phantom study showed an exponential decrease in the correlation coefficient between cavitation patterns in successive pulses from 0.5 ± 0.1 to 0.1 ± 0.1 as ?t increased from 2–200 ms; correspondingly, the lesion was completely fractionated with significantly fewer pulses for longer ?ts. In the tissue study, given the same number of therapy pulses, complete and homogeneous tissue fractionation with well-defined lesion boundaries was achieved only for ?t ≥ 100 ms. These results indicated that the removal of the cavitation memory resulted in more efficient treatments and homogeneous lesions.  相似文献   

2.
Uterine fibroids (leiomyomas), the most common benign tumors in women of reproductive age, are a frequent cause of abnormal vaginal bleeding and other reproductive complaints among women. This study investigates the feasibility of using histotripsy, a non-invasive, non-thermal focused ultrasound ablation method, to ablate uterine fibroids. Human fibroid samples (n = 16) were harvested after hysterectomy or myomectomy procedures at Carilion Memorial Hospital. Histotripsy was applied to ex vivo fibroids in two sets of experiments using a 700-kHz clinical transducer to apply multicycle histotripsy pulses and a prototype 500-kHz transducer to apply single-cycle histotripsy pulses. Ultrasound imaging was used for real-time treatment monitoring, and post-treatment ablation was quantified histologically using hematoxylin and eosin and Masson trichrome stains. Results revealed that multicycle histotripsy generated diffuse cavitation in targeted fibroids, with minimal cellular ablative changes after treatment with 2000 pulses/point. Single-cycle pulsing generated well-confined bubble clouds with evidence of early coagulative necrosis on histological assessment in samples treated with 2000 pulses/point, near-complete ablation in samples treated with 4000 pulses/point and complete tissue destruction in samples treated with 10,000 pulses/point. This study illustrates that histotripsy is capable of fibroid ablation under certain pulsing parameters and warrants further investigation as an improved non-invasive ablation method for the treatment of leiomyomas.  相似文献   

3.
Histotripsy is a non-invasive ablation method that mechanically fractionates tissue by controlling acoustic cavitation. Previous work has revealed that tissue mechanical properties play a significant role in the histotripsy process, with stiffer tissues being more resistant to histotripsy-induced tissue damage. In this study, we propose a thermal pretreatment strategy to precondition tissues before histotripsy. We hypothesize that a thermal pretreatment can be used to alter tissue stiffness by modulating collagen composition, thus changing tissue susceptibility to histotripsy. More specifically, we hypothesize that tissues will soften and become more susceptible to histotripsy when preheated at ∼60°C because of collagen denaturation, but that tissues will rapidly stiffen and become less susceptible to histotripsy when preheated at ∼90°C because of collagen contraction. To test this hypothesis, a controlled temperature water bath was used to heat various ex vivo bovine tissues (tongue, artery, liver, kidney medulla, tendon and urethra). After heating, the Young's modulus of each tissue sample was measured using a tissue elastometer, and changes in tissue composition (i.e., collagen structure/density) were analyzed histologically. The susceptibility of tissues to histotripsy was investigated by treating the samples using a 750-kHz histotripsy transducer. Results revealed a decrease in stiffness and an increase in susceptibility to histotripsy for tissues (except urethra) preheated to 58°C. In contrast, preheating to 90°C increased tissue stiffness and reduced susceptibility to histotripsy for all tissues except tendon, which was significantly softened due to collagen hydrolysis into gelatin. On the basis of these results, a final set of experiments were conducted to determine the feasibility of using high-intensity focused ultrasound to provide the thermal pretreatment. Overall, the results of this study indicate the initial feasibility of a thermal pretreatment strategy to precondition tissue mechanical properties and alter tissue susceptibility to histotripsy. Future work will aim to optimize this thermal pretreatment strategy to determine if this approach is practical for specific clinical applications in vivo without causing unwanted damage to surrounding or overlying tissue.  相似文献   

4.
Extensive mechanical tissue fractionation can be achieved using successive high intensity ultrasound pulses (“histotripsy”). Histotripsy has many potential medical applications where noninvasive tissue removal is desired (e.g., tumor ablation). There is a concern that debris generated by histotripsy-induced tissue fractionation might be an embolization hazard. The aim of this study is to measure the size distribution of these tissue debris particles. Histotripsy pulses were produced by a 513-element 1 MHz array transducer, an 18-element 750 kHz array transducer and a 788 kHz single element transducer. Peak negative pressures of 11 to 25 MPa, pulse durations of 3 to 50 cycles, pulse repetition frequencies of 100 Hz to 2 kHz were used. Tissue debris particles created by histotripsy were collected and measured with a particle sizing system. In the resulting samples, debris <6 μm in diameter constituted >99% of the total number of tissue particles. The largest particle generated by one of the parameter sets tested was 54 μm in diameter, which is smaller than the clinic filter size (100 μm) used to prevent embolization. The largest particles generated using other parameter sets were larger than 60 μm but the value could not be specified using our current setup. Exposures with shorter pulses produced lower percentages of large tissue debris (>30 μm) in comparison to longer pulses. These results suggest that the tissue debris particle size distribution is adjustable by altering acoustic parameters if necessary. (E-mail: zhenx@umich.edu)  相似文献   

5.
Histotripsy is a focused ultrasound therapy for non-invasive tissue ablation. Unlike thermally ablative forms of therapeutic ultrasound, histotripsy relies on the mechanical action of bubble clouds for tissue destruction. Although acoustic bubble activity is often characterized as chaotic, the short-duration histotripsy pulses produce a unique and consistent type of cavitation for tissue destruction. In this review, the action of histotripsy-induced bubbles is discussed. Sources of bubble nuclei are reviewed, and bubble activity over the course of single and multiple pulses is outlined. Recent innovations in terms of novel acoustic excitations, exogenous nuclei for targeted ablation and histotripsy-enhanced drug delivery and image guidance metrics are discussed. Finally, gaps in knowledge of the histotripsy process are highlighted, along with suggested means to expedite widespread clinical utilization of histotripsy.  相似文献   

6.
Elastic, anisotropic tissue such as tendon has proven resistant to mechanical fractionation by histotripsy, a subset of focused ultrasound that uses the creation, oscillation and collapse of cavitation bubbles to fractionate tissue. Our objective was to fabricate an optically transparent hydrogel that mimics tendon for evaluation of histotripsy bubble dynamics. Ex vivo bovine deep digital flexor tendons were obtained (n = 4), and varying formulations of polyacrylamide (PA), collagen and fibrin hydrogels (n = 3 each) were fabricated. Axial sound speeds were measured at 1 MHz for calculation of anisotropy. All samples were treated with a 1.5-MHz focused ultrasound transducer with 10-ms pulses repeated at 1 Hz (p+ = 127 MPa, p- = 35 MPa); treatments were monitored with passive cavitation imaging and high-speed photography. Dehydrated fibrin gels were found to be the most similar to tendon in cavitation emission energy (fibrin = 0.69 ± 0.24, tendon = 0.64 ± 0.19 [× 1010 V2]) and anisotropy (fibrin = 3.16 ± 1.12, tendon = 19.4). Bubble cloud area in dehydrated fibrin (0.79 ± 0.14 mm2) was significantly smaller than most other tested hydrogels. Finally, anisotropy was found to have moderately strong linear relationships with cavitation energy and bubble cloud size (r = –0.65 and –0.80, respectively). Dehydrated fibrin shows potential as a repeatable, transparent, tissue-mimicking hydrogel for evaluation of histotripsy bubble dynamics in elastic, anisotropic tissues.  相似文献   

7.
Our recent studies have shown that high-intensity pulsed ultrasound can achieve mechanical tissue fragmentation, a process we call histotripsy. Histotripsy has many medical applications where noninvasive tissue removal or significant tissue disruption is needed (e.g., cancer therapy). The primary aim of this study is to investigate tissue regions treated by histotripsy and to characterize the boundary between the treated and untreated zones using transmission electron microscopy (TEM). The nature of the tissue disruption suggests many clinical applications and provides insights on the physical mechanism of histotripsy. Fresh ex vivo porcine kidney tissues were treated using histotripsy. A 1 MHz 100 mm diameter focused transducer was used to deliver 15 cycle histotripsy pulses at a peak negative pressure of 17 MPa and a pulse repetition frequency (PRF) of 100 Hz. Each lesion was produced by a 3 × 3 (lateral) × 4 (axial) grid with 2 mm between adjacent lateral and 3 mm between axial exposure points using mechanical scanning. Two thousand pulses were applied to each exposure point to achieve tissue fragmentation. After treatment, the tissue was processed and examined using TEM. Extensive fragmentation of the tissues treated with histotripsy was achieved. TEM micrographs of the tissue treated by histotripsy, showing no recognizable cellular features and little recognizable subcellular structures, demonstrates the efficacy of this technique in ablating the targeted tissue regions. A boundary, or transition zone, of a few microns separated the affected and unaffected areas, demonstrating the precision of histotripsy tissue targeting. TEM micrographs of the tissue treated by histotripsy showed no discernable cellular structure within the treated region. Histotripsy can minimize fragmentation of the adjoining nontargeted tissues because, as a nonlinear threshold phenomenon, damage can be highly localized. The potential for high lesion precision is evident in the TEM micrographs. (E-mail: fwinterr@umich.edu)  相似文献   

8.
Boiling histotripsy (BH) is a focused ultrasound technology that uses millisecond-long pulses with shock fronts to induce mechanical tissue ablation. The pulsing scheme and mechanisms of BH differ from those of cavitation cloud histotripsy, which was previously developed for benign prostatic hyperplasia. The goal of the work described here was to evaluate the feasibility of using BH to ablate fresh ex vivo human prostate tissue as a proof of principle for developing BH for prostate applications. Fresh human prostate samples (N = 24) were obtained via rapid autopsy (<24 h after death, institutional review board exempt). Samples were analyzed using shear wave elastography to ensure that mechanical properties of autopsy tissue were clinically representative. Samples were exposed to BH using 10- or 1-ms pulses with 1% duty cycle under real-time B-mode and Doppler imaging. Volumetric lesions were created by sonicating 1–4 rectangular planes spaced 1 mm apart, containing a grid of foci spaced 1–2 mm apart. Tissue then was evaluated grossly and histologically, and the lesion content was analyzed using transmission electron microscopy and scanning electron microscopy. Observed shear wave elastography characterization of ex vivo prostate tissue (37.9 ± 22.2 kPa) was within the typical range observed clinically. During BH, hyperechoic regions were visualized at the focus on B-mode, and BH-induced bubbles were also detected using power Doppler. As treatment progressed, hypoechoic regions of tissue appeared, suggesting successful tissue fractionation. BH treatment was twofold faster using shorter pulses (1 ms vs. 10 ms). Histological analysis revealed lesions containing completely homogenized cell debris, consistent with histotripsy-induced mechanical ablation. It was therefore determined that BH is feasible in fresh ex vivo human prostate tissue producing desired mechanical ablation. The study supports further work aimed at translating BH technology as a clinical option for prostate ablation.  相似文献   

9.
Although primarily known as an ablative modality, histotripsy can increase the efficacy of lytic therapy in a retracted venous clot model. Bubble cloud oscillations are the primary mechanism of action for histotripsy, and the type of bubble activity is dependent on the pulse duration. A retracted human venous clot model was perfused with and without the thrombolytic recombinant tissue plasminogen activator (rt-PA). The clot was exposed to histotripsy pulses of single- or five-cycle duration and peak negative pressures of 0–30 MPa. Bubble activity within the clot was monitored via passive cavitation imaging. The combination of histotripsy and rt-PA was more efficacious than rt-PA alone for single- and five-cycle pulses with peak negative pressures of 25 and 20 MPa, respectively. For both excitation schemes, the detected acoustic emissions correlated with the degree of thrombolytic efficacy. These results indicate that rt-PA and single- or multicycle histotripsy pulses enhance thrombolytic therapy.  相似文献   

10.
Histotripsy has been previously applied to target various cranial locations in vitro through an excised human skull. Recently, a transcranial magnetic resonance (MR)-guided histotripsy (tcMRgHt) system was developed, enabling pre-clinical investigations of tcMRgHt for brain surgery. To determine the feasibility of in vivo transcranial histotripsy, tcMRgHt treatment was delivered to eight pigs using a 700-kHz, 128-element, MR-compatible phased-array transducer inside a 3-T magnetic resonance imaging (MRI) scanner. After craniotomy to open an acoustic window to the brain, histotripsy was applied through an excised human calvarium to target the inside of the pig brain based on pre-treatment MRI and fiducial markers. MR images were acquired pre-treatment, immediately post-treatment and 2–4 h post-treatment to evaluate the acute treatment outcome. Successful histotripsy ablation was observed in all pigs. The MR-evident lesions were well confined within the targeted volume, without evidence of excessive brain edema or hemorrhage outside of the target zone. Histology revealed tissue homogenization in the ablation zones with a sharp demarcation between destroyed and unaffected tissue, which correlated well with the radiographic treatment zones on MRI. These results are the first to support the in vivo feasibility of tcMRgHt in the pig brain, enabling further investigation of the use of tcMRgHt for brain surgery.  相似文献   

11.
Hepatocellular carcinoma (HCC), or liver cancer, is one of the fastest growing cancers in the United States. Current liver ablation methods are thermal based and share limitations resulting from the heat sink effect of blood flow through the highly vascular liver. In this study, we explore the feasibility of using histotripsy for non-invasive liver ablation in the treatment of liver cancer. Histotripsy is a non-thermal ablation method that fractionates soft tissue through the control of acoustic cavitation. Twelve histotripsy lesions ∼1 cm3 were created in the livers of six pigs through an intact abdomen and chest in vivo. Histotripsy pulses of 10 cycles, 500-Hz pulse repetition frequency (PRF), and 14- to 17-MPa estimated in situ peak negative pressure were applied to the liver using a 1-MHz therapy transducer. Treatments were performed through 4–6 cm of overlying tissue, with 30%–50% of the ultrasound pathway covered by the rib cage. Complete fractionation of liver parenchyma was observed, with sharp boundaries after 16.7-min treatments. In addition, two larger volumes of 18 and 60 cm3 were generated within 60 min in two additional pigs. As major vessels and gallbladder have higher mechanical strength and are more resistant to histotripsy, these remained intact while the liver surrounding these structures was completely fractionated. This work shows that histotripsy is capable of non-invasively fractionating liver tissue while preserving critical anatomic structures within the liver. Results suggest histotripsy has potential for the non-invasive ablation of liver tumors.  相似文献   

12.
Histotripsy is a non-invasive, non-thermal ablation technique that uses high-amplitude, focused ultrasound pulses to fractionate tissue via acoustic cavitation. The goal of this study was to illustrate the potential of histotripsy with electronic focal steering to achieve rapid ablation of a tissue volume at a rate matching or exceeding those of current clinical techniques (~1–2 mL/min). Treatment parameters were established in tissue-mimicking phantoms and applied to ex vivo tissue. Six-microsecond pulses were delivered by a 250-kHz array. The focus was electrically steered to 1000 locations at a pulse repetition frequency of 200 Hz (0.12% duty cycle). Magnetic resonance imaging and histology of the treated tissue revealed a distinct region of necrosis in all samples. Mean lesion volume was 35.6 ± 4.3 mL, generated at 0.9–3.3 mL/min, a speed faster than that of any current ablation method for a large volume. These results suggest that histotripsy has the potential to achieve non-invasive, rapid, homogeneous ablation of a tissue volume.  相似文献   

13.
The left ventricular diastolic time constant (Tau) cannot be practically measured non-invasively. Thus, the aim of this study was to investigate a new method for the evaluation of Tau using continuous-wave (CW) Doppler in dogs with mitral regurgitation. Guided by ultrasound, we created 12 beagle models of mitral regurgitation and acute ischemic left ventricular diastolic dysfunction. Raw audio signals of the CW Doppler spectra were collected, and new mitral regurgitation Doppler spectra were observed after computer re-processing. The new Doppler spectra contour line was constructed using MATLAB (Version R2009), and two time intervals, t1–t2 and t1–t3, were measured on the descending branch of the mitral regurgitation Doppler spectrum and were substituted into Bai's equation group. The Doppler-derived Tau (Tau-d) was resolved and compared with the simultaneous catheter-derived Tau (Tau-c). No significant difference (p?>?0.05) between Tau-d (49.33?±?18.79?ms) and Tau-c (48.76?±?17.60?ms) was found. A correlation analysis between Tau-d and Tau-c suggested a strong positive relationship (r?=?0.85, p?=?0.000). Bland-Altman plots of Tau-d and Tau-c revealed fair agreement. Compared with previous non-invasive approaches, this method is simpler and more accurate. There is a strong positive relationship and fair agreement between Tau-d and Tau-c.  相似文献   

14.
Large-volume soft tissue hematomas are a serious clinical problem, which, if untreated, can have severe consequences. Current treatments are associated with significant pain and discomfort. It has been reported that in an in vitro bovine hematoma model, pulsed high-intensity focused ultrasound (HIFU) ablation, termed histotripsy, can be used to rapidly and non-invasively liquefy the hematoma through localized bubble activity, enabling fine-needle aspiration. The goals of this study were to evaluate the efficiency and speed of volumetric histotripsy liquefaction using a large in vitro human hematoma model. Large human hematoma phantoms (85 cc) were formed by recalcifying blood anticoagulated with citrate phosphate dextrose/saline–adenine–glucose–mannitol solution. Typical boiling histotripsy pulses (10 or 2 ms) or hybrid histotripsy pulses using higher-amplitude and shorter pulses (0.4 ms) were delivered at 1% duty cycle while continuously translating the HIFU focus location. Histotripsy exposures were performed under ultrasound guidance with a 1.5-MHz transducer (8-cm aperture, F# = 0.75). The volume of liquefied lesions was determined by ultrasound imaging and gross inspection. Untreated hematoma samples and samples of the liquefied lesions aspirated using a fine needle were analyzed cytologically and ultrastructurally with scanning electron microscopy. All exposures resulted in uniform liquid-filled voids with sharp edges; liquefaction speed was higher for exposures with shorter pulses and higher shock amplitudes at the focus (up to 0.32, 0.68 and 2.62 mL/min for 10-, 2- and 0.4-ms pulses, respectively). Cytological and ultrastructural observations revealed completely homogenized blood cells and fibrin fragments in the lysate. Most of the fibrin fragments were less than 20 μm in length, but a number of fragments were up to 150 μm. The lysate with residual debris of that size would potentially be amenable to fine-needle aspiration without risk for needle clogging in clinical implementation.  相似文献   

15.
Clinically available thrombolysis techniques are limited by either slow reperfusion (drugs) or invasiveness (catheters) and carry significant risks of bleeding. In this study, the feasibility of using histotripsy as an efficient and noninvasive thrombolysis technique was investigated. Histotripsy fractionates soft tissue through controlled cavitation using focused, short, high-intensity ultrasound pulses. In vitro blood clots formed from fresh canine blood were treated by histotripsy. The treatment was applied using a focused 1-MHz transducer, with five-cycle pulses at a pulse repetition rate of 1 kHz. Acoustic pressures varying from 2 to 12 MPa peak negative pressure were tested. Our results show that histotripsy can perform effective thrombolysis with ultrasound energy alone. Histotripsy thrombolysis only occurred at peak negative pressure ≥6 MPa when initiation of a cavitating bubble cloud was detected using acoustic backscatter monitoring. Blood clots weighing 330 mg were completely broken down by histotripsy in 1.5 to 5 min. The clot was fractionated to debris with >96% weight smaller than 5 μm diameter. Histotripsy thrombolysis treatment remained effective under a fast, pulsating flow (a circulatory model) as well as in static saline. Additionally, we observed that fluid flow generated by a cavitation cloud can attract, trap and further break down clot fragments. This phenomenon may provide a noninvasive method to filter and eliminate hazardous emboli during thrombolysis. (E-mail: adamdm@umich.edu)  相似文献   

16.
Infected abscesses are walled-off collections of pus and bacteria. They are a common sequela of complications in the setting of surgery, trauma, systemic infections and other disease states. Current treatment is typically limited to antibiotics with long-term catheter drainage, or surgical washout when inaccessible to percutaneous drainage or unresponsive to initial care efforts. Antibiotic resistance is also a growing concern. Although bacteria can develop drug resistance, they remain susceptible to thermal and mechanical damage. In particular, short pulses of focused ultrasound (i.e., histotripsy) generate mechanical damage through localized cavitation, representing a potential new paradigm for treating abscesses non-invasively, without the need for long-term catheterization and antibiotics. In this pilot study, boiling and cavitation histotripsy treatments were applied to subcutaneous and intramuscular abscesses developed in a novel porcine model. Ultrasound imaging was used to evaluate abscess maturity for treatment monitoring and assessment of post-treatment outcomes. Disinfection was quantified by counting bacteria colonies from samples aspirated before and after treatment. Histopathological evaluation of the abscesses was performed to identify changes resulting from histotripsy treatment and potential collateral damage. Cavitation histotripsy was more successful in reducing the bacterial load while having a smaller treatment volume compared with boiling histotripsy. The results of this pilot study suggest focused ultrasound may lead to a technology for in situ treatment of acoustically accessible abscesses.  相似文献   

17.
A novel histotripsy approach based on fundamental and second harmonic superposition and incorporating hundred-microsecond-long pulses and two-stage pulse protocol is proposed in this study to rapidly generate mechanically homogenized lesions. Two pulse stages were applied: stage 1, pulses with a pulse duration of 500–600 μs and pulse repetition frequency of 100 Hz, and stage 2, multiple periods, each composed of multiple pulses with the same pulse duration and pulse repetition frequency as those in stage 1, but with an off-time of 600 ms between periods. A custom-designed 1.1/2.2-MHz two-element confocal-annular array, with an f-number of 0.69, and lateral and axial full width at half-maximum pressure dimensions of approximately 1.0 and 6.0 mm, was used. The peak positive/negative pressures at the focus were +22/–7 MPa for 1.1 MHz and +56/–14 MPa with shock wave for 2.2 MHz. To investigate the feasibility of this approach, experiments were designed and performed in tissue-mimicking polyacrylamide gel phantoms with bovine serum albumin and in ex vivo porcine tissues. Cavitation and boiling activities were observed through high-speed photography, and the corresponding acoustic emissions were recorded through passive cavitation detection. Ex vivo experimental results revealed that complete tissue homogeneous regions with regular long tear shape and typical dimensions of 5.80 ± 0.19 mm in axial and 2.20 ± 0.26 mm in lateral were successfully generated in porcine kidney samples. The hematoxylin and eosin staining evidenced that the lesions were thoroughly homogenized and sharply demarcated from untreated regions. These results indicated that the histotripsy approach using fundamental and second harmonic superposition combined with hundred-microsecond pulses and two-stage pulse protocol can efficiently obtain a mechanical disruption of soft tissues with spatial precision, and this approach may have the potential to be developed as a useful tool for precise tumor treatment.  相似文献   

18.
Boiling histotripsy (BH) is a high-intensity focused ultrasound (HIFU)–based method of mechanical tissue fractionation that utilizes millisecond-long bursts of HIFU shock waves to cause boiling at the focus in milliseconds. The subsequent interaction of the incoming shocks with the vapor bubble mechanically lyses surrounding tissue and cells. The acoustic parameter space for BH has been investigated previously and an inverse dependence between the HIFU frequency and the dimensions of a BH lesion has been observed. The primary goal of the present study was to investigate in more detail the ablation rate and reliability of BH in the frequency range relevant to treatment of deep abdominal tissue targets (1–2 MHz). The second goal was to investigate the effect of focal peak pressure levels and shock amplitude on BH lesion formation, given a constant duty factor, a constant ratio of the pulse duration to the time to reach boiling and a constant number of BH pulses. A custom-built 12-element sector array HIFU transducer with F-number = 1.05 was used in all experiments. BH pulses at 5 different frequencies (1, 1.2, 1.5, 1.7 and 1.9 MHz) were delivered to optically transparent polyacrylamide gel phantoms and ex vivo bovine liver and myocardium tissue to observe cavitation and boiling bubble activity with high-speed photography and B-mode ultrasound imaging, correspondingly. In gel phantoms, a cavitation bubble cloud was shown to form prefocally and to shield the focus in all exposures at 1 and 1.2 MHz and in the highest amplitude exposures at 1.5–1.7 MHz; shielding was not observed at 1.9 MHz. In ex vivo tissue, this shielding effect was observed in 25% of exposures when peak negative in situ pressure exceeded 10.2 MPa at 1 MHz and 14.5 MPa at 1.5 MHz. When shielding occurred, the exposures resulted in mild tissue disruption in the prefocal region, but not liquefaction. The dimensions of liquefied lesions followed the inverse proportionality trend with frequency; consequently, the frequency range of 1.2–1.5 MHz appeared to be preferable for BH exposures in terms of the compromise between the ablation rate and reliability. The lesion size was independent of the duration of the BH pulses (or the total “HIFU on” time), provided that the number of pulses was constant and boiling was induced within each pulse. Thus, the use of shorter (1 ms vs. 10 ms), higher amplitude BH pulses allowed up to 10-fold reduction in treatment time for a given duty factor.  相似文献   

19.
Histotripsy is a focused ultrasound ablation therapy being developed for the treatment of liver tumors. A recent study investigating the feasibility of using histotripsy for the ablation of cholangiocarcinoma (CC), bile duct cancer that is difficult to treat with current therapies because of its location near critical structures and fibrous tissue, reported the feasibility of treating CC in an acute mouse model. Here, we investigate histotripsy for the in vivo ablation of CC in a chronic study using a 1-MHz transducer at an applied dose of 500 pulses/point. A pilot study determined that treating the CC tumors plus a 1- to 2-mm margin induced significant injuries to intestinal tissues, thus precluding the use of this strategy. Next, histotripsy was applied to CCs (n = 6) with the treatment contained to the tumor. Post-treatment, the ablation was visualized using ultrasound, and subjects were monitored over time. Histotripsy achieved an average of 73% reduction of tumor diameter 26 d after treatment, with no significant adverse events. Notably, three of six treated tumors were undetectable after 2.5 wk. The treated animals were found to have significantly increased tumor progression-free and overall survival. Overall, results indicate that histotripsy can be used as a safe and effective method for treating CC.  相似文献   

20.
Recent studies have shown that shockwave heating and millisecond boiling in high-intensity focused ultrasound fields can result in mechanical fractionation or emulsification of tissue, termed boiling histotripsy. Visual observations of the change in color and contents indicated that the degree of thermal damage in the emulsified lesions can be controlled by varying the parameters of the exposure. The goal of this work was to examine thermal and mechanical effects in boiling histotripsy lesions using histologic and biochemical analysis. The lesions were induced in ex vivo bovine heart and liver using a 2-MHz single-element transducer operating at duty factors of 0.005–0.01, pulse durations of 5–500 ms and in situ shock amplitude of 73 MPa. Mechanical and thermal damage to tissue was evaluated histologically using conventional staining techniques (hematoxylin and eosin, and nicotinamide adenine dinucleotide-diaphorase). Thermal effects were quantified by measuring denaturation of salt soluble proteins in the treated region. According to histologic analysis, the lesions that visually appeared as a liquid contained no cellular structures larger than a cell nucleus and had a sharp border of one to two cells. Both histologic and protein analysis showed that lesions obtained with short pulses (<10 ms) did not contain any thermal damage. Increasing the pulse duration resulted in an increase in thermal damage. However, both protein analysis and nicotinamide adenine dinucleotide-diaphorase staining showed less denaturation than visually observed as whitening of tissue. The number of high-intensity focused ultrasound pulses delivered per exposure did not change the lesion shape or the degree of thermal denaturation, whereas the size of the lesion showed a saturating behavior suggesting optimal exposure duration. This study confirmed that boiling histotripsy offers an effective, predictable way to non-invasively fractionate tissue into sub-cellular fragments with or without inducing thermal damage.  相似文献   