共查询到20条相似文献,搜索用时 0 毫秒
1.
An experimental comparison of detector performance for direct and indirect digital radiography systems 总被引:11,自引:0,他引:11
Current flat-panel detectors either directly convert x-ray energy to electronic charge or use indirect conversion with an intermediate optical process. The purpose of this work was to compare direct and indirect detectors in terms of their modulation transfer function (MTF), noise power spectrum (NPS), and detective quantum efficiency (DQE). Measurements were made on three flat-panel detectors, Hologic Direct-Ray DR-1000 (DRC), GE Revolution XQ/i (XQ/i), and Philips Digital Diagnost (DiDi) using the IEC-defined RQA5 (approximately 74 kVp, 21 mm Al) and RQA9 (approximately 120 kVp, 40 mm Al) radiographic techniques. The presampled MTFs of the systems were measured using an edge method [Samei et al., Med. Phys. 25, 102 (1998)]. The NPS of the systems were determined for a range of exposure levels by two-dimensional (2D) Fourier analysis of uniformly exposed radiographs [Flynn and Samei, Med. Phys. 26, 1612 (1999)]. The DQEs were assessed from the measured MTF, NPS, exposure, and estimated ideal signal-to-noise ratios. For the direct system, the MTF was found to be significantly higher than that for the indirect systems and very close to an ideal function associated with the detector pixel size. The NPS for the direct system was found to be constant in relation to frequency. For the XQ/i and DRC systems, the DQE results reflected expected differences based on the absorption efficiency of the different detector materials. Using RQA5, the measured DQE values in the diagonal (and axial) direction(s) at spatial frequencies of 0.15 mm(-1) and 2.5 mm(-1) were 64% (64%) and 20% (15%) for the XQ/i system, and 38% (38%) and 20% (20%) for the DRC, respectively. The DQE results of the DiDi system were difficult to interpret due to additional preprocessing steps in that system. 相似文献
2.
Performance evaluation of computed radiography systems 总被引:6,自引:0,他引:6
Recommended methods to test the performance of computed radiography (CR) digital radiographic systems have been recently developed by the AAPM Task Group No. 10. Included are tests for dark noise, uniformity, exposure response, laser beam function, spatial resolution, low-contrast resolution, spatial accuracy, erasure thoroughness, and throughput. The recommendations may be used for acceptance testing of new CR devices as well as routine performance evaluation checks of devices in clinical use. The purpose of this short communication is to provide a tabular summary of the tests recommended by the AAPM Task Group, delineate the technical aspects of the tests, suggest quantitative measures of the performance results, and recommend uniform quantitative criteria for the satisfactory performance of CR devices. The applicability of the acceptance criteria is verified by tests performed on CR systems in clinical use at five different institutions. This paper further clarifies the recommendations with respect to the beam filtration to be used for exposure calibration of the system, and the calibration of automatic exposure control systems. 相似文献
3.
Walter Huda Janice C. Honeyman PhD Meryll M. Frost Edward V. Staab 《Journal of digital imaging》1996,9(1):39-44
A total of 40,000 portable examinations are performed each year at Shands Hospital (Gainesville, FL), a 570-bed teaching hospital. Radiographs are obtained using a screen-film combination with the films digitized for transmission to displays in four intensive care units. A cost-analysis of replacing screen-film with computed radiography (CR) integrated into a filmless picture archiving and communication system (PACS) network was performed. Equipment requirements included two CR units, three high-resolution dual monitor displays, and an archive to store 3 months of image data. The capital costs were amortized over a 5-year period. Capital and operating costs of the proposed expansion to the existing PACS network, together with anticipated cost savings, were determined. The maximum data transfer rate for portable examinations was 150 MByte per hour and approximately 400 GByte of image data are generated each year. These figures were used to determine hhe hardware requirements for handling the acquisition, transfer, and display of the images. Annual costs of the proposed expansion were about $220,000. Cost savings were achieved by elimination of film, including its handling by technologists/library clerks, and amounted to about $200,000 per year. 相似文献
4.
Comparison of imaging properties of a computed radiography system and screen-film systems 总被引:1,自引:0,他引:1
To compare the diagnostic quality of images obtained with a computed radiography (CR) system based on storage phosphor technology with that obtained with conventional screen-film systems, a dual-image recording technique was devised. With this technique, a CR imaging plate is placed behind a screen-film system in a conventional cassette. This makes it possible to obtain two images simultaneously, one from each system, in a clinical examination with the same patient positioning, the same degree of patient motion, the same geometric unsharpness, and no additional exposure. The modulation transfer functions (MTFs) of the CR system with and without the dual-image recording technique were greater at low frequencies, but lower at high frequencies, that the MTFs of the screen-film systems used. The noise Wiener spectra of the CR images at the plane of the imaging plate were greater than those of the screen-film systems, but were comparable to those of the screen-film systems at the plane of the printed film due to the reduction in image size. Clinical chest images obtained with the dual-image recording technique appeared comparable, probably because of the image size reduction and the use of mild unsharp mask processing. 相似文献
5.
Osteoporosis is a disease that results in an increased risk of bone fracture due to a loss of bone mass and deterioration of bone structure. Bone mineral density (BMD) provides a measure of bone mass and is frequently measured by bone densitometry systems to diagnose osteoporosis. In addition, computerized radiographic texture analysis (RTA) is currently being investigated as a measure of bone structure and as an additional diagnostic predictor of osteoporosis. In this study, we assessed the ability of a peripheral bone densitometry (PD) system to yield images useful for RTA. The benefit of such a system is that it measures BMD by dual-energy x-ray absorptiometry and therefore provides high- and low-energy digital radiographic images. The bone densitometry system investigated was the GE/Lunar PIXI, which provides 512 x 512 digital images of the heel or forearm (0.2 mm pixels). We compared texture features of heel images obtained with this PD system to those obtained on a Fuji computed radiography (CR) system (0.1 mm pixels). Fourier and fractal-based texture features of images from 24 subjects who had both CR and BMD exams were calculated, and correlation between the two systems was analyzed. Fourier-based texture features characterize the magnitude, frequency content, and orientation of the trabecular bone pattern. Good correlation was found between the two modalities for the first moment (FMP) with r=0.71 (p value<0.0001) and for minimum FMP with r=0.52 (p value=0.008). Root-mean-square (RMS) did not correlate with r=0.31 (p value>0.05), while the standard deviation of the RMS did correlate with r=0.79 (p value<0.0001). Good correlation was also found between the two modalities for the fractal-based texture features with r=0.79 (p value<0.0001) for the global Minkowski dimension and r=0.63 (p value=0.0007) for the fractal dimension from a box counting method. The PD system therefore may have the potential for yielding heel images suitable for RTA. 相似文献
6.
In a companion paper, we reported on the construction and testing of a xenon gas ionization detector for use in line scanned projection radiography. Experimental results indicate that the detector has sufficient resolution for chest radiography, but higher resolution is required for mammography. Theoretical analysis suggested that a detector pressurized with Freon 13B1 would have better resolution and higher x-ray energy efficiency than a xenon-filled detector for energies below 60 keV. In this paper we compared, theoretically and experimentally, Freon 13B1 to xenon as the detector gas. For a 120-kVp x-ray spectrum, the detector filled with 2.0 MPa of xenon had less channel-to-channel crosstalk, a higher quantum efficiency (QE), and twice the output signal than the detector filled with 1.4 MPa of Freon (highest possible pressure at room temperature), while for a 60-kVp spectrum, crosstalk is the same, but the detector has slightly higher QE and 1.4 times the energy efficiency when filled with Freon instead of xenon. We conclude that xenon is better for high-kVp imaging, while Freon at a lower pressure is slightly better for low-kVp imaging. 相似文献
7.
The physics of computed radiography 总被引:3,自引:0,他引:3
Rowlands JA 《Physics in medicine and biology》2002,47(23):R123-R166
Cassette-based computed radiography (CR) systems have continued to evolve in parallel with integrated, instant readout digital radiography (DR) systems. The image quality of present day CR systems is approaching its theoretical limits but is significantly inferior to DR. Further improvements in CR image quality require improved concepts. The aim of this review is to identify the fundamental limitations in CR performance. This will provide a background for the development of new approaches to improve photostimulable phosphor CR systems. It will also guide research in designing better CR systems to possibly compete with DR systems in terms of image quality parameters such as detective quantum efficiency and yet maintain CR convenience in being portable and more economical. 相似文献
8.
Charles E Willis 《Journal of digital imaging》2007,20(1):1-3; author reply 4-5
9.
The properties of the human visual system (HVS) relevant to the diagnostic process are described after a brief introduction on the general problems and advantages of using soft copy for primary radiology interpretations. At various spatial and temporal frequencies the contrast sensitivity defines the spatial resolution of the eye-brain system and the sensitivity to flicker. The adaptation to the displayed radiological scene and the ambient illumination determine the dynamic range for the operation of the HVS. Although image display devices are determined mainly by state-of-the-art technology, analysis of the HVS may suggest technical characteristics for electronic displays that will help to optimize the display to the operation of the HVS. These include display size, spatial resolution, contrast resolution, luminance range, and noise, from which further consequences for the technical components of a monitor follow. It is emphasized that routine monitor quality control must be available in clinical practice. These image quality measures must be simple enough to be applied as part of the daily routine. These test instructions might also serve as elements of technical acceptance and constancy tests. 相似文献
10.
A computed tomography (CT) imaging technique called variable resolution x-ray (VRX) detection provides variable image resolution ranging from that of clinical body scanning (1 cy/mm) to that of microscopy (100 cy/mm). In this paper, an experimental VRX CT scanner based on a rotating subject table and an angulated storage phosphor screen detector is described and tested. The measured projection resolution of the scanner is > or = 20 lp/mm. Using this scanner, 4.8-s CT scans are made of specimens of human extremities and of in vivo hamsters. In addition, the system's projected spatial resolution is calculated to exceed 100 cy/mm for a future on-line CT scanner incorporating smaller focal spots (0.1 mm) than those currently used and a 1008-channel VRX detector with 0.6-mm cell spacing. 相似文献
11.
Characterization of noise sources for two generations of computed radiography systems using powder and crystalline photostimulable phosphors 总被引:1,自引:0,他引:1
The performances of two generations of computed radiography (CR) were tested and compared in terms of resolution and noise characteristics. The main aim was to characterize and quantify the noise sources in the images. The systems tested were (1) Agfa CR 25.0, a flying spot reader with powder phosphor image plates (MD 40.0); and (2) the Agfa DX-S, a line-scanning CR reader with needle crystal phosphor image plates (HD 5.0). For both systems, the standard metrics of presampled modulation transfer function (MTF), normalized noise power spectra (NNPS) and detective quantum efficiency (DQE) were measured using standard radiation quality RQA5 as defined by the International Electrotechnical Commission. The various noise sources contributing to the NNPS were separated by using knowledge of their relationship with air kerma, MTF, absorption efficiency and antialiasing filters. The DX-S MTF was superior compared with the CR 25.0. The maximum difference in MTF between the DX-S scan and CR 25.0 subscan directions was 0.13 at 1.3 mm(-1). For a nominal detector air kerma of 4 microGy, the peak DQE of the DX-S was 43 (+/-3)%, which was over double that of the CR 25.0 of 18 (+/-2)%. The additive electronic noise was negligible on the CR 25.0 but calculated to be constant 3.4 x 10(-7) (+/-0.4 x 10(-7)) mm2 at 3.9 microGy on the DX-S. The DX-S has improved image quality compared with a traditional flying spot reader. The separation of the noise sources indicates that the improvements in DQE of the DX-S are due not only to the higher quantum, efficiency and MTF, but also the lower structure, secondary quantum, and excess noise. 相似文献
12.
A computed tomography imaging technique called variable resolution x-ray (VRX) detection provides detector resolution ranging from that of clinical body scanning to that of microscopy (1 cy/mm to 100 cy/mm). The VRX detection technique is based on a new principle denoted as "projective compression" that allows the detector resolution element to scale proportionally to the image field size. Two classes of VRX detector geometry are considered. Theoretical aspects related to x-ray physics and data sampling are presented. Measured resolution parameters (line-spread function and modulation-transfer function) are presented and discussed. A VRX image that resolves a pair of 50 micron tungsten hairs spaced 30 microns apart is shown. 相似文献
13.
An energy discriminating x-ray detector has been developed for dual-energy, scan projection digital radiography. The detector is comprised of a pair of x-ray intensifying screen/linear photodiode arrays, aligned one behind the other. Energy discrimination is achieved by employing a low atomic number phosphor in the front screen and a high atomic number phosphor in the back screen. The x-ray response, modulation transfer function, and defective quantum efficiency of the detector are reported along with the experimental methodology utilized for the measurements. Also presented is an analysis which indicates that in a typical patient's lung field, the detector can resolve the projected density (g/cm2) of a 3-mm-thick, 1-cm2 area of bone to better than 1.5%. 相似文献
14.
15.
16.
Hiroshi Kondoh MD Junpei Ikezoe Kiyonari Inamura Chikazumi Kuroda Takahiro Kozuka 《Journal of digital imaging》1994,7(4):193-195
This study examined whether hard-copy radiographs produced from computed radiography (CR) images show the subtle interstitial pulmonary disease equally well to conventional screen-film radiographs, because a digital radiography should be chosen for introduction of the digital picture archiving and communication system (PACS) for the new Osaka University Hospital.1,2 Eleven radiologists examined 20 abnormal and 20 control chest radiographs presented in each of three groups: conventional screen-film radiographs and two sizes of hard-copy radiographs made from CR images. This study of digital image quality of chest examinations found that some findings on conventional screen-film radiography images are not reproduced by current CR (2,000×2,000×10 bits in matrix), especially when the experienced radiologists were observed. This finding suggested improvements are needed in CR before CR of chest should fully replace conventional screen-film radiography. 相似文献
17.
Pongnapang N 《Biomedical imaging and intervention journal》2005,1(2):e12-Dec;1(2):e12
Computed Radiography (CR) has become a major digital imaging modality in a modern radiological department. CR system changes workflow from the conventional way of using film/screen by employing photostimulable phosphor plate technology. This results in the changing perspectives of technical, artefacts and quality control issues in radiology departments. Guidelines for better image quality in digital medical enterprise include professional guidelines for users and the quality control programme specifically designed to serve the best quality of clinical images. Radiographers who understand technological shift of the CR from conventional method can employ optimization of CR images. Proper anatomic collimation and exposure techniques for each radiographic projection are crucial steps in producing quality digital images. Matching image processing with specific anatomy is also important factor that radiographers should realise. Successful shift from conventional to fully digitised radiology department requires skilful radiographers who utilise the technology and a successful quality control program from teamwork in the department. 相似文献
18.
Experiments with an ionisation detector were performed in order to determine whether it was possible to obtain high energy photon beam images for radiotherapy treatment verification. A small prototype detector with a field of view of 78 mm X 78 mm and constructed from printed circuit boards was used. The imaging area was a matrix ionisation chamber, filled with air or liquid (2,2,4-trimethylpentane). A minicomputer was used to control the data acquisition electronics and to reconstruct and restore the images. The images were displayed on a viewing console for computed tomography images. The liquid filled detector with a front-rear board separation of 1.0 mm gave the best results. The spatial resolution was about 3.8 mm with a density resolution of 0.5% for a data acquisition time of 120 s. Comparison of the liquid detector images with corresponding metal screen-film detector images showed that the image qualities were the same. An important advantage of the ionisation detector image is that grey scale modification, sharpening and smoothing by digital processing can easily be performed. 相似文献
19.
20.
Olch AJ 《Medical physics》2005,32(9):2987-2999
Computed radiography (CR) systems have been gaining adoption as digital replacements for film for diagnostic and therapy imaging. As a result, film processors are being removed from service, leaving a void for the medical physicists who use film and processors for two-dimensional mega-voltage beam dosimetry. This is the first report to evaluate the ability of a commercial CR reader and storage phosphor plate system to accurately quantitate absolute dose and dose distributions from a 6 MV photon beam. There are potential advantages and disadvantages of current CR systems compared to film systems. CR systems inherently produce a linear dose-response over several logs of dose. However, the barium in the storage phosphor has a higher atomic number than the silver in film, resulting in significant energy sensitivity. The purpose of this work is to fully characterize the impact of these and other features of this CR system relevant to dosimetry. The tests performed and reported on in this study include uniformity of readout across a uniform field, geometrical accuracy, intra- and interday reproducibility, signal decay with time and with light exposure, dose-to-signal calibration, high dose effects, obliquity effects, perpendicular and parallel calibration results, field size and depth of measurement effects and the use of lead filters to minimize them, and intensity modulated radiation therapy quality assurance test results compared to that for film. Practical techniques are provided to optimize the accuracy of the system as a dosimetric replacement for film. 相似文献