首页 | 本学科首页   官方微博 | 高级检索  
相似文献
 共查询到20条相似文献,搜索用时 31 毫秒
1.
Film dosimetry provides a convenient tool to determine dose distributions, especially for verification of IMRT plans. However, the film response to radiation shows a significant dependence on depth, energy and field size that compromise the accuracy of measurements. Kodak's XV2 film has a low saturation dose (approximately 100 cGy) and, consequently, a relatively short region of linear dose-response. The recently introduced Kodak extended range EDR2 film was reported to have a linear dose-response region extending to 500 cGy. This increased dose range may be particularly useful in the verification of IMRT plans. In this work, the dependence of Kodak EDR2 film's response on the depth, field size and energy was evaluated and compared with Kodak XV2 film. Co-60, 6 MV, 10 MV and 18 MV beams were used. Field sizes were 2 x 2, 6 x 6, 10 x 10, 14 x 14, 18 x 18 and 24 x 24 cm2. Doses for XV2 and EDR2 films were 80 cGy and 300 cGy, respectively. Optical density was converted to dose using depth-corrected sensitometric (Hurter and Driffield, or H&D) curves. For each field size, XV2 and EDR2 depth-dose curves were compared with ion chamber depth-dose curves. Both films demonstrated similar (within 1%) field size dependence. The deviation from the ion chamber for both films was small forthe fields ranging from 2 x 2 to 10 x 10 cm2: < or =2% for 6, 10 and 18 MV beams. No deviation was observed for the Co-60 beam. As the field size increased to 24 x 24 cm2, the deviation became significant for both films: approximately 7.5% for Co-60, approximately 5% for 6 MV and 10 MV, and approximately 6% for 18 MV. During the verification of IMRT plans, EDR2 film showed a better agreement with the calculated dose distributions than the XV2 film.  相似文献   

2.
A new type of radiographic film, Kodak EDR2 film, was evaluated for dose verification of intensity modulated radiation therapy (IMRT) delivered by a static multileaf collimator (SMLC). A sensitometric curve of EDR2 film irradiated by a 6 MV x-ray beam was compared with that of Kodak X-OMAT V (XV) film. The effects of field size, depth and dose rate on the sensitometric curve were also studied. It is found that EDR2 film is much less sensitive than XV film. In high-energy x-ray beams, the double hit process is the dominant mechanism that renders the grains on EDR2 films developable. As a result, in the dose range that is commonly used for film dosimetry for IMRT and conventional external beam therapy, the sensitometric curves of EDR2 films cannot be approximated as a linear function, OD = c * D. Within experimental uncertainty, the film sensitivity does not depend on the dose rate (50 vs 300 MU/min) or dose per pulse (from 1.0 x 10(-4) to 4.21 x 10(-4) Gy/pulse). Field sizes and depths (up to field size of 10 x 10 cm2 and depth = 10 cm) have little effect on the sensitometric curves. Percent depth doses (PDDs) for both 6 and 23 MV x rays were measured with both EDR2 and XV films and compared with ion chamber data. Film data are within 2.5% of the ion chamber results. Dose profiles measured with EDR2 film are consistent with those measured with an ion chamber. Examples of measured IMRT isodose distributions versus calculated isodoses are presented. We have used EDR2 films for verification of all IMRT patients treated by SMLC in our clinic. In most cases, with EDR2 film, actual clinical daily fraction doses can be used for verification of composite isodose distributions of SMLC-based IMRT.  相似文献   

3.
Characteristics of sensitometric curves of radiographic films   总被引:4,自引:0,他引:4  
A new type of radiographic film, EDR (extended dose range) film, has been recently become available for film dosimetry. It is particularly attractive for composite isodose verification of intensity modulated radiation therapy because of its low sensitivity relative to the more common Kodak XV film. For XV film, the relationship between optical density and dose, commonly known as the sensitometric curve, depends linearly on the dose at low densities. Unlike XV film, the sensitometric curve of EDR film irradiated by megavoltage x rays is not linearly dependent on the dose at low densities. In this work, to understand the mechanisms governing the shape of the sensitometric curves, EDR film was studied with kilovoltage x rays, 60Co gamma rays, megavoltage x rays, and electron beams. As a comparison, XV film was also studied with the same beams mentioned above. The model originally developed by Silberstein [J. Opt. Soc. Am. 35, 93-107, 1945)] is used to fit experimental data. It is found that the single hit model can be used to predict the sensitometric curve for XV films irradiated by all beams used in this work and for EDR films exposed to kilovoltage x rays. For EDR film irradiated by 60Co gamma rays, megavoltage x rays, and electron beams, the double hit model is used to fit the sensitometric curves. For doses less than 100 cGy, a systematic difference between measured densities and that predicted by the double hit model is observed. Possible causes of the observed differences are discussed. The results of this work provide a theoretical explanation of the sensitometric behavior of EDR film.  相似文献   

4.
We have investigated the dependence of the measured optical density on the incident beam energy, field size and depth for a new type of film, Kodak extended dose range (Kodak EDR). Film measurements have been conducted over a range of field sizes (3 x 3 cm2 to 25 x 25 cm2) and depths (d(max) to 15 cm), for 6 MV and 15 MV photons within a solid water phantom, and the variation in sensitometric response (net optical density versus dose) has been reported. Kodak EDR film is found to have a linear response with dose, from 0 to 350 cGy, which is much higher than that typically seen for Kodak XV film (0-50 cGy). The variation in sensitometric response for Kodak EDR film as a function of field size and depth is observed to be similar to that of Kodak XV film; the optical density varied in the order of 2-3% for field sizes of 3 x 3 cm2 and 10 x 10 cm2 at depths of d(max), 5 cm and 15 cm in the phantom. Measurements for a 25 x 25 cm2 field size showed consistently higher optical densities at depths of d(max), 5 cm and 15 cm, relative to a 10 x 10 cm2 field size at 5 cm depth, with 4-5% differences noted at a depth of 15 cm. Fractional depth dose and profiles conducted with Kodak EDR film showed good agreement (2%/2 mm) with ion chamber measurements for all field sizes except for the 25 x 25 cm2 at depths greater than 15 cm, where differences in the order of 3-5% were observed. In addition, Kodak EDR film measurements were found to be consistent with those of Kodak XV film for all fractional depth doses and profiles. The results of this study indicate that Kodak EDR film may be a useful tool for relative dosimetry at higher dose ranges.  相似文献   

5.
For intensity modulated radiation therapy (IMRT) dose distribution verification, multidimensional measurements are required to quantify the steep dose-gradient regions. High resolution, two-dimensional dose distributions can be measured using radiographic film. However, the photon energy response of film is known to be a function of depth, field size, and photon beam energy, potentially reducing the accuracy of dose distribution measurements. The dosimetric properties of the recently developed Kodak EDR2 film were investigated and compared to those of Kodak XV film. The dose responses of both film types to 6 MV and 18 MV photon beams were investigated for depths of 5 cm, 10 cm, and 15 cm and field sizes of 4x4 cm2 and 15x15 cm2. This analysis involved the determination of sensitometric curves for XV and EDR2 films, the determination of dose profiles from exposed XV and EDR2 films, and comparison of the film-generated dose profiles to ionization chamber measurements. For the combinations of photon beam energy, depth, and field size investigated here, our results indicate that the sensitometric curves are nearly independent of field size and depth of calibration. For a field size of 4x4 cm2, a single sensitometric curve for either EDR2 and XV film can be used for the determination of relative dose profiles. For the larger field size, the sensitometric curve for EDR2 film is superior to XV film in regions where the dose falls below 20% of the central axis dose, due to the effects that the increased low energy scattered photon contributions have on film response. The limited field size and depth dependence of sensitometric data measured using EDR2 film, along with the inherently wide linear dose-response range of EDR2 film, makes it better suited to the verification of IMRT dose distributions.  相似文献   

6.
Film dosimetry is an important tool for the verification of irradiation techniques. The shape of the sensitometric curve depends on the type of film as well as on the irradiation and processing conditions. Existing data concerning the influence of irradiation geometry on the sensitometric curve are conflicting. In particular the variation of optical density, OD, with field size and depth in a phantom shows large differences in magnitude between various authors. This variation, as well as the effect of beam energy and film plane orientation on OD, was therefore investigated for two types of film, Kodak X-Omat V and Agfa Structurix D2. Films were positioned in a solid phantom, either perpendicular or (almost) parallel to the beam axis, and irradiated to different dose levels using various photon beams (Co-60, 6 MV, 15 MV, 18 MV, 45 MV). It was found that the sensitometric curves of the Kodak film derived at different depths are almost identical for the four x-ray beams. For the Kodak film the differences in OD with depth are less than 2%, except for the Co-60 beam, where the difference is about 4% at 10 cm depth for a 15 cm x 15 cm field. The slope of the sensitometric curve of the Agfa film is somewhat more dependent on photon beam energy, depth and field size. The sensitometric curves of both types of film are almost independent of the film plane orientation, except for shallow depths. For Co-60 and for the same dose, the Kodak and Agfa films gave at dose maximum an OD lower by 4% and 6%, respectively, for the parallel compared to the perpendicular geometry. Good dosimetric results can be obtained if films from the same batch are irradiated with small to moderate field sizes (up to about 15 cm x 15 cm), at moderate depths (up to about 15 cm), using a single calibration curve, e.g., for a 10 cm x 10 cm field.  相似文献   

7.
Matchline dosimetry in step and shoot IMRT fields: a film study   总被引:2,自引:0,他引:2  
The Varian millennium 120 multileaf collimator has curved leaf ends. Transmission through the leaf ends generates a small asymmetric penumbral dose effect. This design can lead to hot spots between neighbouring beam segments during step and shoot IMRT dose delivery. We have observed some matchlines with film for clinical beams optimized using the pinnacle radiotherapy treatment planning system; hence we sought to verify the optimum leaf offset required to minimize the matchline effect. An in-house program was created to control the MLC leaf banks in 2 cm steps with a 2 cm gap. The gap was varied by the following offset values from 0.0 to 0.1 cm. Two types of radiographic films (Kodak EDR and XV films) and a radiochromic film (Gafchromic MD-55-2) were used to measure the optical density maps. The films were positioned in a solid water phantom perpendicular to the beam axis and irradiated at d(max) using a 6 MV photon beam. An ion chamber (IC4) was used to measure point doses for normalization in a beam umbral minima position. The relative mean peak to valley dose ratios measured with no leaf offset were 1.31, 1.30 and 1.31 for the XV, EDR2 and Gafchromic films, respectively. For a 0.07 cm gap per leaf and a performance of end leaf repeatability of 0.01 cm, the central matchline was reduced to about 1.0 for all dosimeters, with two mini-peaks measured as 1.05, 1.05 and 1.08 each side of the matchline, for XV, EDR2 and Gafchromic, respectively. The average relative dose across the umbra for this offset was XO-mat V = 1.01, EDR = 1.01 and radiochromic film = 1.02, respectively. While we expected the beam penumbral tails from segment neighbours to cause overprediction of the dose in the central valley regions due to the energy response of radiographic films, by normalizing all dosimeters to an ion chamber reading in the minimum we could not observe any major shape distortion between the radiographic film and radiochromic film results. In conclusion, relative doses measured by radiographic and radiochromic films agree well with IC4 within +/-2%.  相似文献   

8.
Dosimetric accuracy of Kodak EDR2 film for IMRT verifications   总被引:4,自引:0,他引:4  
Patient-specific intensity-modulated radiotherapy (IMRT) verifications require an accurate two-dimensional dosimeter that is not labor-intensive. We assessed the precision and reproducibility of film calibrations over time, measured the elemental composition of the film, measured the intermittency effect, and measured the dosimetric accuracy and reproducibility of calibrated Kodak EDR2 film for single-beam verifications in a solid water phantom and for full-plan verifications in a Rexolite phantom. Repeated measurements of the film sensitometric curve in a single experiment yielded overall uncertainties in dose of 2.1% local and 0.8% relative to 300 cGy. 547 film calibrations over an 18-month period, exposed to a range of doses from 0 to a maximum of 240 MU or 360 MU and using 6 MV or 18 MV energies, had optical density (OD) standard deviations that were 7%-15% of their average values. This indicates that daily film calibrations are essential when EDR2 film is used to obtain absolute dose results. An elemental analysis of EDR2 film revealed that it contains 60% as much silver and 20% as much bromine as Kodak XV2 film. EDR2 film also has an unusual 1.69:1 silver:halide molar ratio, compared with the XV2 film's 1.02:1 ratio, which may affect its chemical reactions. To test EDR2's intermittency effect, the OD generated by a single 300 MU exposure was compared to the ODs generated by exposing the film 1 MU, 2 MU, and 4 MU at a time to a total of 300 MU. An ion chamber recorded the relative dose of all intermittency measurements to account for machine output variations. Using small MU bursts to expose the film resulted in delivery times of 4 to 14 minutes and lowered the film's OD by approximately 2% for both 6 and 18 MV beams. This effect may result in EDR2 film underestimating absolute doses for patient verifications that require long delivery times. After using a calibration to convert EDR2 film's OD to dose values, film measurements agreed within 2% relative difference and 2 mm criteria to ion chamber measurements for both sliding window and step-and-shoot fluence map verifications. Calibrated film results agreed with ion chamber measurements to within 5 % /2 mm criteria for transverse-plane full-plan verifications, but were consistently low. When properly calibrated, EDR2 film can be an adequate two-dimensional dosimeter for IMRT verifications, although it may underestimate doses in regions with long exposure times.  相似文献   

9.
Recently the EDR2 (extended dose range) film has been introduced commercially for applications in radiation therapy dosimetry. In addition to characterizing the wide dynamic range, several authors have reported a reduced energy dependence of this film compared to that of X-Omatic Verification (XV) films for megavoltage photon beams. However, those investigations were performed under limited geometrical conditions. We have investigated the dosimetric performance of EDR2 film for the verification of IMRT fields at more clinically relevant conditions by comparing the film doses with the doses measured with an ion chamber and XV films. The effects of using a low energy scattered photon filter on EDR2 film dosimetry was also studied. In contrast to previous reports our results show that EDR2 film still exhibits considerable energy dependence (a maximum discrepancy of 9%, compared with an ion chamber) at clinically relevant conditions (10 cm depth for IMRT fields). However, by using the low-energy filters the discrepancy is reduced to within 3%. Therefore, EDR2 film, in combination with the filters, is found to be a promising two-dimensional dosimeter for verification of IMRT treatment fields.  相似文献   

10.
Since its introduction the PinPoint (PTW-Freiburg) micro-ionization chamber has been proposed for relative dosimetry (output factors, depth dose curves, and beam profiles) as well as for determination of absolute dose of small high-energy photon beams. This paper investigates the dosimetric performance of a new design (type 31014) of the PinPoint ion chamber with a central aluminum electrode. The study included characterization of inherent and radiation-induced leakage, ion collection efficiency and polarity effect, relative response of the chamber, measurement of beam profiles, and depth dose curves. The 6 and 15 MV photon beams of a Varian 2100 C/D were considered. At the nominal operating voltage of 400 V the PinPoint type 31014 chamber was found to present a strong field size dependence of the polarity correction factor and an excess of the collected charge, which can lead to an underestimation of the collection efficiency if determined with the conventional "two-voltage" method. In comparison to the original PinPoint design (type 31006) the authors found for type 31014 chamber no overresponse to large-area fields if polarity correction is applied. If no correction is taken into consideration, the authors found the chamber's output to be inaccurate for large-area fields (0.5% accuracy limited up to the 12 x 12 and 20 x 20 cm2 field for the 6 and 15 MV beams, respectively), which is a direct consequence of the stem and polarity effects due to the chamber's very small sensitive volume (0.015 cc) and cable irradiation. Beam profiles and depth dose curves measured with type 31014 PinPoint chamber for small and medium size fields were compared to data measured with a 0.125 cc ion chamber and with high-resolution Kodak EDR2 films. Analysis of the penumbra (80%-20% distance) showed that the spatial resolution of type 31014 PinPoint ion chamber approaches (penumbra broadening < or = 0.6 mm) EDR2 film results.  相似文献   

11.
Childress NL  Dong L  Rosen II 《Medical physics》2002,29(10):2384-2390
A method for measuring a film sensitometric curve using a single sheet of film exposed with a two field step-and-shoot MLC treatment was developed and tested with Kodak XV2 and EDR2 films. With this technique a film sensitometric curve can be completed in only 10 minutes, making it practical to generate new film calibrations daily. This method is applicable to film calibrations for all purposes, but is particularly useful in IMRT treatment verification due to the method's use of small fields. This method agrees with the traditional large-field multifilm calibration within 0.5% and will produce sensitometric curves with errors less than 1% throughout the dose range, including uncertainties in dose delivery, film response, and optical density measurements. OD values for XV2 and EDR2 films were consistent in the middle of exposure areas at high depths, but the XV2 film penumbra regions showed large amounts of over-response as the calibration depth increased. If XV2 film is used for IMRT treatment verification, it is necessary to reduce the fluence of low energy photons in areas around the film by using thin lead shields. EDR2 film was shown to have minimal energy dependence, as it accurately represented penumbra areas and yielded identical sensitometric curves generated with 6 and 18 MV photons. However, its darker tint may make it more sensitive to scanning laser film digitizers' horizontal nonuniformities. This single film method proved to be superior to the traditional calibration method and allows fast daily calibrations of films for highly accurate IMRT delivery verifications.  相似文献   

12.

Purpose:

To investigate the optimal sensitometric curves of extended dose range (EDR2) radiographic film in terms of depth, field size, dose range and processing conditions for dynamic intensity modulated radiation therapy (IMRT) dosimetry verification with 6 MV X-ray beams.

Materials and methods:

A Varian Clinac 23 EX linear accelerator with 6 MV X-ray beam was used to study the response of Kodak EDR2 film. Measurements were performed at depths of 5, 10 and 15 cm in MedTec virtual water phantom and with field sizes of 2x2, 3x3, 10x10 and 15x15 cm2. Doses ranging from 20 to 450 cGy were used. The film was developed with the Kodak RP X-OMAT Model M6B automatic film processor. Film response was measured with the Vidar model VXR-16 scanner. Sensitometric curves were applied to the dose profiles measured with film at 5 cm in the virtual water phantom with field sizes of 2x2 and 10x10 cm2 and compared with ion chamber data. Scanditronix/Wellhofer OmniProTM IMRT software was used for the evaluation of the IMRT plan calculated by Eclipse treatment planning.

Results:

Investigation of the reproducibility and accuracy of the film responses, which depend mainly on the film processor, was carried out by irradiating one film nine times with doses of 20 to 450 cGy. A maximum standard deviation of 4.9% was found which decreased to 1.9% for doses between 20 and 200 cGy. The sensitometric curves for various field sizes at fixed depth showed a maximum difference of 4.2% between 2x2 and 15x15 cm2 at 5 cm depth with a dose of 450 cGy. The shallow depth tended to show a greater effect of field size responses than the deeper depths. The sensitometric curves for various depths at fixed field size showed slightly different film responses; the difference due to depth was within 1.8% for all field sizes studied. Both field size and depth effect were reduced when the doses were lower than 450 cGy. The difference was within 2.5% in the dose range from 20 to 300 cGy for all field sizes and depths studied. Dose profiles measured with EDR2 film were consistent with those measured with an ion chamber. The optimal sensitometric curve was acquired by irradiating film at a depth of 5 cm with doses ranging from 20 to 450 cGy with a 3×3 cm2 multileaf collimator. The optimal sensitometric curve allowed accurate determination of the absolute dose distribution. In almost 200 cases of dynamic IMRT plan verification with EDR2 film, the difference between measured and calculated dose was generally less than 3% and with 3 mm distance to agreement when using gamma value verification.

Conclusion:

EDR2 film can be used for accurate verification of composite isodose distributions of dynamic IMRT when the optimal sensitometric curve has been established.  相似文献   

13.
INTRODUCTION AND PURPOSE: Conventional x-ray films and radiochromic films have inherent challenges for high precision radiotherapy dosimetry. Here we have investigated basic characteristics of optically stimulated luminescence (OSL) of irradiated films containing carbon-doped aluminum oxide (Al2O3:C) for dosimetry in therapeutic photon and electron beams. MATERIALS AND METHODS: The OSL films consist of a polystyrene sheet, with a top layer of a mixture of single crystals of Al2O3:C, ground into a powder, and a polyester base. The total thickness of the films is 0.3 mm. Measurements have been performed in a water equivalent phantom, using 4, 6, 10, and 18 MV photon beams, and 6-22 MeV electron beams. The studies include assessment of the film response (acquired OSL signal/delivered dose) on delivered dose (linearity), dose rate (1-6 Gy/min), beam quality, field size and depth (6 MV, ranges 4 x 4-30 x 30 cm2, dmax-35 cm). Doses have been derived from ionization chamber measurements. OSL films have also been compared with conventional x-ray and GafChromic films for dosimetry outside the high dose area, with a high proportion of low dose scattered photons. In total, 787 OSL films have been irradiated. RESULTS: Overall, the OSL response for electron beams was 3.6% lower than for photon beams. Differences between the various electron beam energies were not significant. The 6 and 18 MV photon beams differed in response by 4%. No response dependencies on dose rate were observed. For the 6 MV beam, the field size and depth dependencies of the OSL response were within +/-2.5%. The observed inter-film response variation for films irradiated with the same dose varied from 1% to 3.2% (1 SD), depending on the measurement day. At a depth of 20 cm, 5 cm outside the 20 x 20 cm2 6 and 18 MV beams, an over response of 17% was observed. In contrast to GafChromic and conventional x-ray films, the response of the Al2O3:C films is linear in the clinically relevant dose range 0-200 cGy. CONCLUSIONS: Measurement of the OSL signal of irradiated films containing Al2O3:C is a promising technique for film dosimetry in radiotherapy with no or small response variations with dose rate, beam quality, field size and depth, and a linear response from 0 to 200 cGy.  相似文献   

14.
The higher sensitivity to low-energy scattered photons of radiographic film compared to water can lead to significant dosimetric error when the beam quality varies significantly within a field. Correcting for this artifact will provide greater accuracy for intensity modulated radiation therapy (IMRT) verification dosimetry. A procedure is developed for correction of the film energy-dependent response by creating a pencil beam kernel within our treatment planning system to model the film response specifically. Film kernels are obtained from EGSnrc Monte Carlo simulations of the dose distribution from a 1 mm diameter narrow beam in a model of the film placed at six depths from 1.5 to 40 cm in polystyrene and solid water phantoms. Kernels for different area phantoms (50 x 50 cm2 and 25 x 25 cm2 polystyrene and 30 x 30 cm2 solid water) are produced. The Monte Carlo calculated kernel is experimentally verified with film, ion chamber and thermoluminescent dosimetry (TLD) measurements in polystyrene irradiated by a narrow beam. The kernel is then used in convolution calculations to, predict the film response in open and IMRT fields. A 6 MV photon beam and Kodak XV2 film in a polystyrene phantom are selected to test the method as they are often used in practice and can result in large energy-dependent artifacts. The difference in dose distributions calculated with the film kernel and the water kernel is subtracted from film measurements to obtain a practically film artifact free IMRT dose distribution for the Kodak XV2 film. For the points with dose exceeding 5 cGy (11% of the peak dose) in a large modulated field and a film measurement inside a large polystyrene phantom at depth of 10 cm, the correction reduces the fraction of pixels for which the film dose deviates from dose to water by more than 5% of the mean film dose from 44% to 6%.  相似文献   

15.
Kodak EDR2 film is a widely used two-dimensional dosimeter for intensity modulated radiotherapy (IMRT) measurements. Our clinical use of EDR2 film for IMRT verifications revealed variations and uncertainties in dose response that were larger than expected, given that we perform film calibrations for every experimental measurement. We found that the length of time between film exposure and processing can affect the absolute dose response of EDR2 film by as much as 4%-6%. EDR2 films were exposed to 300 cGy using 6 and 18 MV 10 x 10 cm2 fields and then processed after time delays ranging from 2 min to 24 h. An ion chamber measured the relative dose for these film exposures. The ratio of optical density (OD) to dose stabilized after 3 h. Compared to its stable value, the film response was 4%-6% lower at 2 min and 1% lower at 1 h. The results of the 4 min and 1 h processing time delays were verified with a total of four different EDR2 film batches. The OD/dose response for XV2 films was consistent for time periods of 4 min and 1 h between exposure and processing. To investigate possible interactions of the processing time delay effect with dose, single EDR2 films were irradiated to eight different dose levels between 45 and 330 cGy using smaller 3 x 3 cm2 areas. These films were processed after time delays of 1, 3, and 6 h, using 6 and 18 MV photon qualities. The results at all dose levels were consistent, indicating that there is no change in the processing time delay effect for different doses. The difference in the time delay effect between the 6 and 18 MV measurements was negligible for all experiments. To rule out bias in selecting film regions for OD measurement, we compared the use of a specialized algorithm that systematically determines regions of interest inside the 10 x 10 cm2 exposure areas to manually selected regions of interest. There was a maximum difference of only 0.07% between the manually and automatically selected regions, indicating that the use of a systematic algorithm to determine regions of interest in large and fairly uniform areas is not necessary. Based on these results, we recommend a minimum time of 1 h between exposure and processing for all EDR2 film measurements.  相似文献   

16.
Olch AJ 《Medical physics》2002,29(9):2159-2168
Film-based quality assurance (QA) is an important element of any intensity modulated radiation therapy (IMRT) program. XV2 film is often used for IMRT QA, however, it has saturation and energy response limitations which hinder accurate film dosimetry. A new commercially released ready-pack film has been introduced that has an extended dose range (EDR2), reportedly allowing measured doses above 600 cGy without saturation. Also, this film may have less energy dependence due to its composition. The purpose of this paper is to study and compare the two types of film with respect to absolute dose accuracy for IMRT plans, percent depth dose accuracy for square fields between 2 and 20 cm, ability to measure composite plan isodoses and single beam fluence maps for IMRT cases, and sensitivity to processor variations over time. In 19 IMRT patient QA tests, the EDR2 film was able to achieve an absolute dose accuracy of better than 2% vs over 4% for XV2 film. The EDR2 film was able to reproduce ionization chamber and diode-measured percent depth doses to 20 cm depth generally to within 1% over the range of field sizes tested compared to about 10% for the XV2 film. When compared to calculations, EDR2 film agreed better than XV2 film for both composite plan isodoses and single beam fluence intensity maps. The EDR2 film was somewhat more resistant to processor changes over time than the XV2 film, with a standard deviation of dose reproducibility of less than 2% compared to 6%, respectively.  相似文献   

17.
Radiographic or silver halide film is a well-established 2D dosimeter with an unquestioned spatial resolution. But its higher sensitivity to low-energy photons has to be taken into consideration. Metal compensators or physical modulators to deliver intensity modulated radiation therapy (IMRT) are known to change the beam energy spectrum and to produce scattered photons and contaminating electrons. Therefore the reliability of film dosimetry in compensator-based IMRT might be questioned. Conflicting data have been reported in the literature. This uncertainty about the validity of film dosimetry in compensator-based IMRT triggered us to conduct this study. First, the effect of MCP-96 compensators of varying thickness on the depth dose characteristics was investigated using a diamond detector which has a uniform energy response. A beam hardening effect was observed at 6 MV that resulted in a depth dose increase that remained below 2% at 20 cm depth. At 25 MV, in contrast, beam softening produced a dose decrease of up to 5% at the same depth. Second, dose was measured at depth using EDR2 film in perpendicular orientation to both 6 MV and 25 MV beams for different compensator thicknesses. A film dose underresponse of 1.1% was found for a 30 mm thick block in a 25 MV beam, which realized a transmission factor of 0.243. The effect induced by the compensators is higher than the experimental error but still within the accepted overall uncertainty of film dosimetry in clinical IMRT QA. With radiographic film as an affordable QA tool, the physical compensator remains a low threshold technique to deliver IMRT.  相似文献   

18.
Zeidan OA  Li JG  Low DA  Dempsey JF 《Medical physics》2004,31(10):2730-2737
In this study, we compared the dosimetric properties of four of the most commonly used films for megavoltage photon-beam dosimetry when irradiated under identical conditions by small multileaf-collimator (MLC) defined beamlets. Two silver-halide films (SHFs), Kodak XV2 and EDR2, and two radiochromic films (RCFs), Gafchromic HS and MD55-2, were irradiated by MLC-defined 1 x 1 cm2 beamlets from a Varian 2100 C/D linac equipped with a 120-leaf MLC. The beamlets were delivered with the accelerator gantry set laterally (90 degrees rotation) upon a solid-water compression film phantom at 100 cm source-to-surface distance which was positioned with the films parallel to the beam axis. Beamlets were delivered at central axis, 5.0 cm, and 10.5 cm off-axis for both leaf-end and leaf-side defined beamlets. The film dosimetry was performed using a quantitative optical density (OD) imaging system that was validated in a previous study. No significant differences between SHF and RCF measurements were observed in percentage depth doses, horizontal depth profiles, or two-dimension spatial isodose distributions in both the central axis and off-axis measurements. We found that regardless of the type of film used, RCF or SHF, a consistent data set for small beam dose modeling was generated. Previous validation studies based on the use of RCF and OD imaging system would indicate that all film produce an accurate result for small beam characterization.  相似文献   

19.
In the current work, EPR (electron paramagnetic resonance) dosimetry using alanine films (134 microm thick) was utilized for dose measurements in inhomogeneous phantoms irradiated with radiotherapy photon beams. The main phantom material was PMMA, while either Styrofoam or aluminium was introduced as an inhomogeneity. The phantoms were irradiated to a maximum dose of about 30 Gy with 6 or 15 MV photons. The performance of the alanine film dosimeters was investigated and compared to results from ion chamber dosimetry, Monte Carlo simulations and radiotherapy treatment planning calculations. It was found that the alanine film dosimeters had a linear dose response above approximately 5 Gy, while a background signal obscured the response at lower dose levels. For doses between 5 and 60 Gy, the standard deviation of single alanine film dose estimates was about 2%. The alanine film dose estimates yielded results comparable to those from the Monte Carlo simulations and the ion chamber measurements, with absolute differences between estimates in the order of 1-15%. The treatment planning calculations exhibited limited applicability. The current work shows that alanine film dosimetry is a method suitable for estimating radiotherapeutical doses and for dose measurements in inhomogeneous media.  相似文献   

20.
The accuracy of high-energy x-ray dosimetry can be improved by taking account of differences between the compositions of the chamber wall and the buildup cap or dosimetry phantom. The fraction of the ionization due to secondary electrons arising in the chamber wall has been determined as a function of wall thickness for 60Co gamma rays and x rays in the range of 2-25 MV for Farmer-type chambers. Secondary electrons arising in the accelerator head were removed from the x-ray beams by a magnetic field placed just in front of the ionization chamber. For 60Co gamma rays, the fraction increases from 40% to 100% as the wall thickness increases from 0.05 to 0.55 g cm-2. For a 0.05 g cm-2 wall, fraction decreases from 60% to 10% as the x-ray energy is increased from 2 to 25 MV. Limited data obtained with different chambers suggest that the fraction is independent of chamber wall composition when the thickness is expressed in g cm-2.  相似文献   

设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号