首页 | 本学科首页   官方微博 | 高级检索  
相似文献
 共查询到20条相似文献,搜索用时 93 毫秒
1.
Background and purpose — Trabecular metal (TM) cups have demonstrated favorable results in acetabular revision and their use in primary total hip arthroplasty (THA) is increasing. Some evidence show that TM cups might decrease periprosthetic infection (PPI) incidence. We compared the survivorship of TM cups with that of other uncemented cups in primary THA, and evaluated whether the use of TM cups is associated with a lower risk of PPI.

Patients and methods — 10,113 primary THAs with TM cup and 85,596 THAs with other uncemented cups from 2 high-quality national arthroplasty registries were included. The mean follow-up times were 3.0 years for the TM cups and 3.8 years for the other uncemented cups.

Results — The overall survivorship up to 8 years for TM cups and other uncemented cups was 94.4% and 96.2%, respectively (p = < 0.001). Adjusting for relevant covariates in a Cox regression model the TM cups had a persistently higher revision risk than other uncemented cups (HR =1.5, 95% CI 1.4–1.7, p = < 0.001). There was a slightly higher, though not statistically significant, revision rate for PPI in the TM group (1.2, 95% CI 1.0–1.6, p = 0.09).

Interpretation — Risk of revision for any reason was higher for the TM cup than for other uncemented cups in primary THA. In contrast to our hypothesis, there was no evidence that the revision rate for PPI was lower in the TM cup patients. Regardless of the promising early and mid-term results for TM cups in hip revision arthroplasty, we would like to sound a note of caution on the increasing use of the TM design, especially in uncomplicated primary THAs, where uncemented titanium cups are considered to provide a reliable outcome.  相似文献   


2.
Background and purpose — Revision total hip arthroplasty (THA) is associated with higher dislocation rates than primary THA. We compared the risk of dislocation within 6 months and all-cause re-revision during the whole study period using either the dual-mobility cup or the unipolar cup.

Methods — We used a prospective hospital registry-based cohort including all total and cup-only revision THAs performed between 2003 and 2013. The cups used were either dual-mobility or unipolar; the choice was made according to the preference of the surgeon. 316 revision THAs were included. The mean age of the cohort was 69 (25–98) years and 160 THAs (51%) were performed in women. The dual-mobility group (group 1) included 150 THAs (48%) and the mean length of follow-up was 31 (0–128) months. The unipolar group (group 2) included 166 THAs (53%) and the mean length of follow-up was 52 (0–136) months.

Results — The incidence of dislocation within 6 months was significantly lower with the dual-mobility cup than with the unipolar cup (2.7% vs. 7.8%). The unadjusted risk ratio (RR) was 0.34 (95% CI: 0.11–1.02) and the adjusted RR was 0.28 (95% CI: 0.09–0.87). The number of patients needed to treat with a dual-mobility cup in order to prevent 1 case of dislocation was 19. The unadjusted incidence rate ratio for all-cause re-revision in the dual-mobility group compared to the unipolar group was 0.6 (95% CI: 0.3–1.4).

Interpretation — Use of a dual-mobility rather than a unipolar cup in revision THA reduced the risk of dislocation within 6 months.  相似文献   


3.
Background and purpose — The use of a cemented cup together with an uncemented stem in total hip arthroplasty (THA) has become popular in Norway and Sweden during the last decade. The results of this prosthetic concept, reverse hybrid THA, have been sparsely described. The Nordic Arthroplasty Register Association (NARA) has already published 2 papers describing results of reverse hybrid THAs in different age groups. Based on data collected over 2 additional years, we wanted to perform in depth analyses of not only the reverse hybrid concept but also of the different cup/stem combinations used.

Patients and methods — From the NARA, we extracted data on reverse hybrid THAs from January 1, 2000 until December 31, 2013. 38,415 such hips were studied and compared with cemented THAs. The Kaplan-Meier method and Cox regression analyses were used to estimate the prosthesis survival and the relative risk of revision. The main endpoint was revision for any reason. We also performed specific analyses regarding the different reasons for revision and analyses regarding the cup/stem combinations used in more than 500 cases.

Results — We found a higher rate of revision for reverse hybrids than for cemented THAs, with an adjusted relative risk of revision (RR) of 1.4 (95% CI: 1.3–1.5). At 10 years, the survival rate was 94% (CI: 94–95) for cemented THAs and 92% (95% CI: 92–93) for reverse hybrids. The results for the reverse hybrid THAs were inferior to those for cemented THAs in patients aged 55 years or more (RR =1.1, CI: 1.0–1.3; p < 0.05). We found a higher rate of early revision due to periprosthetic femoral fracture for reverse hybrids than for cemented THAs in patients aged 55 years or more (RR =3.1, CI: 2.2–4.5; p < 0.001).

Interpretation — Reverse hybrid THAs had a slightly higher rate of revision than cemented THAs in patients aged 55 or more. The difference in survival was mainly caused by a higher incidence of early revision due to periprosthetic femoral fracture in the reversed hybrid THAs.  相似文献   


4.
Background and purpose — Most registry studies regarding highly crosslinked polyethylene (XLPE) have focused on the overall revision risk. We compared the risk of cup and/or liner revision for specific cup and liner designs made of either XLPE or conventional polyethylene (CPE), regarding revision for any reason and revision due to aseptic loosening and/or osteolysis.

Patients and methods — Using the Nordic Arthroplasty Register Association (NARA) database, we identified cup and liner designs where either XLPE or CPE had been used in more than 500 THAs performed for primary hip osteoarthritis. We assessed risk of revision for any reason and for aseptic loosening using Cox regression adjusted for age, sex, femoral head material and size, surgical approach, stem fixation, and presence of hydroxyapatite coating (uncemented cups).

Results — The CPE version of the ZCA cup had a risk of revision for any reason similar to that of the XLPE version (p = 0.09), but showed a 6-fold higher risk of revision for aseptic loosening (p < 0.001). The CPE version of the Reflection All Poly cup had an 8-fold elevated risk of revision for any reason (p < 0.001) and a 5-fold increased risk of revision for aseptic loosening (p < 0.001). The Charnley Elite Ogee/Marathon cup and the Trilogy cup did not show such differences.

Interpretation — Whether XLPE has any advantage over CPE regarding revision risk may depend on the properties of the polyethylene materials being compared, as well as the respective cup designs, fixation type, and follow-up times. Further research is needed to elucidate how cup design factors interact with polyethylene type to affect the risk of revision.  相似文献   


5.
Background and purpose — Recurrent dislocation is the commonest cause of early revision of a total hip arthropasty (THA). We examined the effect of femoral head size and surgical approach on revision rate for dislocation, and for other reasons, after total hip arthroplasty (THA).

Patients and methods — We analyzed data on 166,231 primary THAs and 3,754 subsequent revision THAs performed between 2007 and 2015, registered in the Dutch Arthroplasty Register (LROI). Revision rate for dislocation, and for all other causes, were calculated by competing-risk analysis at 6-year follow-up. Multivariable Cox proportional hazard regression ratios (HRs) were used for comparisons.

Results — Posterolateral approach was associated with higher dislocation revision risk (HR =1) than straight lateral, anterolateral, and anterior approaches (HR =0.5–0.6). However, the risk of revision for all other reasons (especially stem loosening) was higher with anterior and anterolateral approaches (HR =1.2) and lowest with posterolateral approach (HR =1). For all approaches, 32-mm heads reduced the risk of revision for dislocation compared to 22- to 28-mm heads (HR =1 and 1.6, respectively), while the risk of revision for other causes remained unchanged. 36-mm heads increasingly reduced the risk of revision for dislocation but only with the posterolateral approach (HR =0.6), while the risk of revision for other reasons was unchanged. With the anterior approach, 36-mm heads increased the risk of revision for other reasons (HR =1.5).

Interpretation — Compared to the posterolateral approach, direct anterior and anterolateral approaches reduce the risk of revision for dislocation, but at the cost of more stem revisions and other revisions. For all approaches, there is benefit in using 32-mm heads instead of 22- to 28-mm heads. For the posterolateral approach, 36-mm heads can safely further reduce the risk of revision for dislocation.  相似文献   


6.
Background and purpose — Given similar functional outcomes with mobile and fixed bearings, a difference in survivorship may favor either. This study investigated the risk of aseptic loosening for the most used subtypes of mobile-bearing rotating-platform knees, in Norway and Australia.

Patients and methods — Primary TKRs reported to the Norwegian and Australian joint registries, between 2003 and 2014, were analyzed with aseptic loosening as primary end-point and all revisions as secondary end-point. We hypothesized that no difference would be found in the rate of revision between rotating-platform and the most used fixed-bearing TKRs, or between keeled and non-keeled tibia. Kaplan–Meier estimates and curves, and Cox regression relative risk estimates adjusted for age, sex, and diagnosis were used for comparison.

Results — The rotating-platform TKRs had an increased risk of revision for aseptic loosening compared with the most used fixed-bearing knees, in Norway (RR =6, 95% CI 4–8) and Australia (RR =2.1, 95% CI 1.8–2.5). The risk of aseptic loosening as a reason for revision was highest in Norway compared with Australia (RR =1.7, 95% CI 1.4–2.0). The keeled tibial component had the same risk of aseptic loosening as the non-keeled tibia (Australia). Fixation method and subtypes of the tibial components had no impact on the risk of aseptic loosening in these mobile-bearing knees.

Interpretation — The rotating-platform TKRs in this study appeared to have a higher risk of revision for aseptic loosening than the most used fixed-bearing TKRs.  相似文献   


7.
Background and purpose — In a previous registry report, short-term implant survival of hip resurfacing arthroplasty (HRA) in Finland was found to be comparable to that of total hip arthroplasty (THA). Since then, it has become evident that adverse reactions to metal debris (ARMDs) may also be associated with HRA, not only with large-diameter head metal-on-metal THA. The aim of the study was to assess medium- to long-term survivorship of HRA based on the Finnish Arthroplasty Register (FAR).

Patients and methods — 5,068 HRAs performed during the period 2001–2013 in Finland were included. Kaplan-Meier survival analysis was used to calculate survival probabilities and their 95% confidence intervals (CIs). Cox multiple regression, with adjustment for age, sex, diagnosis, femoral head size, and hospital volume was used to analyze implant survival of HRA devices with revision for any reason as endpoint. The reference group consisted of 6,485 uncemented Vision/Bimetric and ABG II THAs performed in Finland over the same time period.

Results — The 8-year survival, with any revision as an endpoint, was 93% (CI: 92–94) for Birmingham Hip Resurfacing (BHR), 86% (CI: 78–94) for Corin, 91% (CI: 89–94) for ReCap, 92% (CI: 89–96) for Durom, and was 72% (CI: 69–76) for the Articular Surface Replacement (ASR). The 10-year survival, with any revision as an endpoint, for reference THAs was 92% (CI: 91–92) and for all HRAs it was 86% (CI: 84–87%). Female HRA patients had about twice the revision risk of male patients. ASR had an inferior outcome: the revision risk was 4-fold higher than for BHR, the reference implant.

Interpretation — The 10-year implant survival of HRAs is 86% in Finland. According to new recommendations from NICE (The National Institute for Health and Care Excellence), an HRA/THA should have a revision rate of 5% or less at 10 years. None of the HRAs studied achieved this goal.  相似文献   


8.
Background and purpose — The original Müller acetabular reinforcement ring (ARR) was developed to be used for acetabular revisions with small cavitary and/or segmental defects or poor acetabular bone quality. Long-term data for this device are scarce. We therefore investigated long-term survival and radiographic outcome for revision total hip arthroplasty using the ARR.

Patients and methods — Between October 1984 and December 2005, 259 primary acetabular revisions using an ARR were performed in 245 patients (259 hips). The mean follow-up time was 10 (0–27) years; 8 hips were lost to follow-up. The cumulative incidence for revision was calculated using a competing risk model. Radiographic assessment was performed for 90 hips with minimum 10 years’ follow-up. It included evaluation of osteolysis, migration and loosening.

Results — 16 ARRs were re-revised: 8 for aseptic loosening, 6 for infection, 1 for suspected infection, and 1 due to malpositioning of the cup. The cumulative re-revision rate for aseptic loosening of the ARR at 20 years was 3.7% (95% CI 1.7–6.8%). Assuming all patients lost to follow-up were revised for aseptic loosening, the re-revision rate at 20 years was 6.9% (95% CI 4.1–11%). The overall re-revision rate of the ARR for any reason at 20 years was 7.0% (95% CI 4.1–11%). 21 (23%) of the 90 radiographically examined ARR had radiographic changes: 12 showed isolated signs of osteolysis but were not loose; 9 were determined loose on follow-up, of which 5 were revised.

Interpretation — Our data suggest that the long-term survival and radiographic results of the ARR in primary acetabular revision are excellent.  相似文献   


9.
Background and purpose — Clindamycin has not been compared with other antibiotics for prophylaxis in arthroplasty. Since 2009, the Swedish Knee Arthroplasty Register (SKAR) has been collecting information on the prophylactic antibiotic regime used at every individual operation. In Sweden, when there is allergy to penicillin, clindamycin has been the recommended alternative. We examined whether there were differences in the rate of revision due to infection depending on which antibiotic was used as systemic prophylaxis.

Patients and methods — Patients who had a total knee arthroplasty (TKA) performed due to osteoarthritis (OA) during the years 2009–2015 were included in the study. Information on which antibiotic was used was available for 80,018 operations (55,530 patients). Survival statistics were used to calculate the rate of revision due to infection until the end of 2015, comparing the group of patients who received cloxacillin with those who received clindamycin as systemic prophylaxis.

Results — Cloxacillin was used in 90% of the cases, clindamycin in 7%, and cephalosporins in 2%. The risk of being revised due to infection was higher when clindamycin was used than when cloxacillin was used (RR =1.5, 95% CI: 1.2–2.0; p = 0.001). There was no significant difference in the revision rate for other causes (p = 0.2).

Interpretation — We advise that patients reporting allergic reaction to penicillin should have their allergic history explored. In the absence of a clear history of type-I allergic reaction (e.g. urticaria, anaphylaxis, or bronchospasm), we suggest the use of a third-generation cephalosporin instead of clindamycin as perioperative prophylaxis when undergoing a TKR. No recommendation can be given regarding patients with type-1 allergy.  相似文献   


10.
Background and purpose — The use of uncemented revision stems is an established option in 2-stage procedures in patients with periprosthetic joint infection (PJI) after total hip arthroplasty (THA). However, in 1-stage procedures, they are still rarely used. There are still no detailed data on radiological outcome after uncemented 1-stage revisions. We assessed (1) the clinical outcome, including reoperation due to persistent infection and any other reoperation, and (2) the radiological outcome after 1- and 2-stage revision, using an uncemented stem.

Patients and methods — Between January 1993 and December 2012, an uncemented revision stem was used in 81 THAs revised for PJI. Patients were treated with 1- or 2-stage procedures according to a well-defined algorithm (1-stage: n = 28; 2-stage: n = 53). All hips had a clinical and radiological follow-up. Outcome parameters were eradication of infection, re-revision of the stem, and radiological changes. Survival was calculated using Kaplan-Meier analysis. Radiographs were analyzed for bone restoration and signs of loosening. The mean clinical follow-up time was 7 (2–15) years.

Results — The 7-year infection-free survival was 96% (95% CI: 92–100), 100% for 1-stage revision and 94% for 2-stage revision (95% CI: 87–100) (p = 0.2). The 7-year survival for aseptic loosening of the stem was 97% (95% CI: 93–100), 97% for 1-stage revision (95% CI: 90–100) and 97% for 2-stage revision (95% CI: 92–100) (p = 0.3). No further infection or aseptic loosening occurred later than 7 years postoperatively. The radiographic results were similar for 1- and 2-stage procedures.

Interpretation — Surgical management of PJI with stratification to 1- or 2-stage exchange according to a well-defined algorithm combined with antibiotic treatment allows the safe use of uncemented revision stems. Eradication of infection can be achieved in most cases, and medium- and long-term results appear to be comparable to those for revisions for aseptic loosening.  相似文献   


11.
Background and purpose — To further improve the success of joint replacement surgery, attention needs to be paid to variations associated with improved or worsened outcomes. We investigated the association between the type of bone cement used and the risk of revision surgery after primary total hip replacement.

Methods — We conducted a prospective study of data from the National Joint Registry for England and Wales between April 1, 2003 and December 31, 2013. 199,205 primary total hip replacements performed for osteoarthritis where bone cement was used were included. A multilevel over-dispersed piecewise Poisson model was used to estimate differences in the rate of revision by bone cement type adjusted for implant type, head size, age, sex, ASA grade, and surgical approach.

Results — The rate of revision was higher in DePuy CMW3 medium viscosity with gentamicin (IRR 2.0, 95% CI 1.5–2.7) and DePuy SmartSet high viscosity plain (IRR 2.7, 95% CI 1.1–5.5), and lower in DePuy CMW1 high viscosity plain (IRR 0.44, 95% CI 0.19–0.89) bone cements compared with Heraeus Palacos high viscosity with gentamicin. Revision rates were similar between plain and antibiotic-loaded bone cement.

Interpretation — The majority of bone cements performed similarly well, excluding DePuy SmartSet high viscosity and CMW3 high viscosity with gentamicin, which both had higher revision rates. We found no clear differences by viscosity or antibiotic content.  相似文献   


12.
Objective: To identify the availability and unmet need of home adaptations (HAs) among the Swiss population with spinal cord injury (SCI).

Design: Cross-sectional study.

Setting: Swiss Spinal Cord Injury Community Survey 2012.

Participants: Individuals aged 16 or older with chronic SCI living in Switzerland.

Interventions: Not applicable.

Outcome measures: The availability of ten HAs (self-report) was analyzed by sex, age, living situation, indoor mobility, SCI severity, SCI etiology and time since SCI. The unmet need (self-report of not having a HA but needing it) of HAs was analyzed by financial hardship.

Results: Among the 482 study participants (mean age 55.2 years, standard deviation 15.0 years, 71.6% males), 85.1% had at least one HA. The most frequent HA was a wheelchair accessible shower (62.7%). Availability of HAs markedly varied with indoor mobility (e.g. 38.4% of participants using a wheelchair had a stair lift compared to 17.4% of those walking) and with SCI severity (e.g. 54.8% of those with complete paraplegia had a wheelchair accessible kitchen worktop compared to 26.0% of those with incomplete paraplegia). Unmet need was highest for adjustable kitchen worktops (78.7% of those with a need) and adjustable kitchen cabinets (75.7%) and lowest for wheelchair accessible showers (9.4%) and grab bars next to the toilet (8.5%). No significant differences in unmet need were found when stratifying for financial hardship.

Conclusion: Availability of HAs is dependent on indoor mobility and SCI severity. There is a considerable degree of unmet need for selected HAs, which couldn't be explained by financial hardship.  相似文献   


13.
Background and purpose — Patient-reported outcome (PRO) is recognized as an important tool for evaluating the outcome and satisfaction after total hip arthroplasty (THA). We wanted to compare patient-reported outcome measure (PROM) scores from patients with ceramic-on-ceramic (CoC) THAs and those with metal-on-metal (MoM) THAs to scores from patients with metal-on-polyethylene (MoP) THAs, and to determine the influence of THA-related noise on PROM scores.

Patients and methods — We conducted a nationwide cross-sectional questionnaire survey in a cohort of patients identified from the Danish Hip Arthroplasty Registry. The PROMs included were: hip dysfunction and osteoarthritis and outcome score (HOOS), EQ-5D-3L, EQ VAS, UCLA activity score, and questions about noise from the THA. The response rate was 85% and the number of responders was 3,089. Of these, 45% had CoC THAs, 17% had MoM THAs, and 38% had MoP THAs, with a mean length of follow-up of 7, 5, and 7 years, respectively.

Results — Compared to MoP THAs, the mean PROM scores for CoC and MoM THAs were similar, except that CoC THAs had a lower mean score for HOOS Symptoms than did MoP THA. 27% of patients with CoC THAs, 29% with MoM THAs, and 12% with MoP THAs reported noise from their hip. For the 3 types of bearings, PROM scores from patients with a noisy THA were statistically significantly worse than those from patients with a silent MoP THA. The exception was noisy CoC and MoM THAs, which had the same mean UCLA activity score as silent MoP THAs.

Interpretation — A high proportion of patients reported noise from the THA, and these patients had worse PROM scores than patients with silent MoP THAs.  相似文献   


14.
Background and purpose — Hip dislocation is one of the most common complications following total hip arthroplasty (THA). Several factors that affect dislocation have been identified, including acetabular cup positioning. Optimal values for cup inclination and anteversion are debatable. We performed a systematic review to describe the different methods for measuring cup placement, target zones for cup positioning, and the association between cup positioning and dislocation following primary THA.

Methods — A systematic search of literature in the PubMed database was performed (January and February 2016) to identify articles that compared acetabular cup positioning and the risk of dislocation. Surgical approach and methods for measurement of cup angles were also considered.

Results — 28 articles were determined to be relevant to our research question. Some articles demonstrated that cup positioning influenced postoperative dislocation whereas others did not. The majority of articles could not identify a statistically significant difference between dislocating and non-dislocating THA with regard to mean angles of cup anteversion and inclination. Most of the articles that assessed cup placement within the Lewinnek safe zone did not show a statistically significant reduction in dislocation rate. Alternative target ranges have been proposed by several authors.

Interpretation — The Lewinnek safe zone could not be justified. It is difficult to draw broad conclusions regarding a definitive target zone for cup positioning in THA, due to variability between studies and the likely multifactorial nature of THA dislocation. Future studies comparing cup positioning and dislocation rate should investigate surgical approach separately. Standardized tools for measurement of cup positioning should be implemented to allow comparison between studies.  相似文献   


15.
Background and purpose — Fast-track protocols have been successfully implemented in many hospitals as they have been shown to result in shorter length of stay (LOS) without compromising results. We evaluated the effect of fast-track implementation on the use of institutional care and results after total hip replacement (THR).

Patients and methods — 3,193 THRs performed in 4 hospitals between 2009–2010 and 2012–2013 were identified from the Finnish Hospital Discharge Register and the Finnish Arthroplasty Register. Hospitals were classified as fast-track (Hospital A) and non-fast-track (Hospitals B, C, and D). We analyzed LOS, length of uninterrupted institutional care (LUIC, including LOS), discharge destination, readmission, revision rate, and mortality in each hospital. We compared these outcomes for THRs performed in Hospital A before and after fast-track implementation and we also compared outcomes, excluding readmission rates, with the corresponding outcomes for the other hospitals.

Results — After fast-track implementation, median LOS in Hospital A diminished from 5 to 2 days (p < 0.001) and (median) LUIC from 6 to 3 (p = 0.001) days. No statistically significant changes occurred in discharge destination. However, the reduction in LOS was combined with an increase in the 42-day readmission rate (3.1% to 8.3%) (p < 0.001). A higher proportion of patients were at home 1 week after THR (p < 0.001) in Hospital A after fast-tracking than before.

Interpretation — The fast-track protocol reduces LUIC but needs careful implementation to maintain good quality of care throughout the treatment process.  相似文献   


16.
Background and purpose — A large number of fixation methods of hamstring tendon autograft (HT) are available for anterior cruciate ligament reconstruction (ACLR). Some studies report an association between fixation method and the risk of revision ACLR. We compared the risk of revision of various femoral and tibial fixation methods used for HT in Scandinavia 2004–2011.

Materials and methods — A register-based study of 38,666 patients undergoing primary ACLRs with HT, with 1,042 revision ACLRs. The overall median follow-up time was 2.8 (0–8) years. Fixation devices used in a small number of patients were grouped according to design and the point of fixation.

Results — The most common fixation methods were Endobutton (36%) and Rigidfix (31%) in the femur; and interference screw (48%) and Intrafix (34%) in the tibia. In a multivariable Cox regression model, the transfemoral fixations Rigidfix and Transfix had a lower risk of revision (HR 0.7 [95% CI 0.6–0.8] and 0.7 [CI 0.6–0.9] respectively) compared with Endobutton. In the tibia the retro interference screw had a higher risk of revision (HR 1.9 [CI 1.3–2.9]) compared with an interference screw.

Interpretation — The choice of graft fixation influences the risk of revision after primary ACLR with hamstring tendon autograft.  相似文献   


17.
Background and purpose — Porous tantalum cups have been introduced as an alternative to various reinforcement rings in revision hip surgery. We hypothesized that porous tantalum cups would be superior to Müller acetabular roof reinforcement rings (MARRs) in revision hip surgery with re-revision for aseptic loosening as the primary outcome measure.

Patients and methods — 207 hips operated with either a porous tantalum cup (TM cup, n = 111) or a MARR (n = 96) at index procedure were identified in our local arthroplasty register. Acetabular defects were classified according to Paprosky. There were 96 men and 111 women with a median age of 71 (35–95) years, presenting acetabular defect size type I in 39 cases, IIA in 22, IIB in 27, IIC in 43, IIIA in 32, and IIIB in 37 cases. Analysis of medical records identified all patients with subsequent re-revision and reasons for re-revisions. Kaplan-Meier survival functions were used to estimate implant survival.

Results — With re-revision for aseptic loosening as the endpoint, the 6-year unadjusted cumulative survival was 97% (95% CI: 94–100) for TM cups and 96% (CI: 92–100) for MARR (p = 0.6). Using re-revision for any reason as the endpoint, 6-year survival was 87% (CI: 81–94) for TM cups and 95% (CI: 90–99) for MARR (p = 0.06). The main reason for re-revision in the TM group was dislocation (n = 10), followed by loosening (n = 3), whereas the main reason for re-revision in the MARR group was aseptic loosening (n = 8). Duration of the index procedure and perioperative blood loss were lower in the TM group.

Interpretation — Both TM and MARR lead to good 6-year results in acetabular revision surgery. The methods differ in their respective failure mechanisms. We conclude that TM cups are a valuable treatment option in acetabular revision surgery, but the reasons underlying dislocations after the use of TM cups must be analyzed further.  相似文献   


18.
Background and purpose — Psychosocial factors are important risk factors for poor outcomes in the first year after total knee replacement (TKR), however their impact on long-term outcomes is unclear. We aimed to identify preoperative psychosocial risk factors for poor outcomes at 1 year and 5 years after TKR.

Patients and methods — 266 patients were recruited prior to TKR surgery. Knee pain and function were assessed preoperatively and at 1 and 5 years postoperative using the WOMAC Pain score, WOMAC Function score and American Knee Society Score (AKSS) Knee score. Preoperative depression, anxiety, catastrophizing, pain self-efficacy and social support were assessed. Statistical analyses involved multiple linear regression and mixed effect linear regression.

Results — Higher anxiety was a risk factor for worse pain at 1 year postoperative. No psychosocial factors were associated with any outcomes at 5 years postoperative. Analysis of change over time found that patients with higher pain self-efficacy had lower preoperative pain and experienced less improvement in pain up to 1 year postoperative. Higher pain self-efficacy was associated with less improvement in the AKSS up to 1 year postoperative but more improvement between 1 and 5 years postoperative.

Interpretation — Preoperative anxiety was found to influence pain at 1 year after TKR. However, none of the psychosocial variables were risk factors for a poor outcome at 5 years post­operative, suggesting that the negative effects of anxiety on outcome do not persist in the longer-term.  相似文献   


19.
Background: Upper cervical spine fractures are traumatic injuries typically associated with high-energy trauma and have a high morbidity and mortality rate. We describe a case of upper cervical spine fracture occurring due to high-energy trauma that resulted in significant myositis ossificans (MO) of the longus coli muscle.

Study Design: Case Report and literature review.

Methods: Retrospective review of medical records.

Results: The patient was treated non-operatively for the neck rotation and MO of his longus coli muscle and had gradual improvement of symptoms.

Conclusions: To our knowledge, this is the first report of symptomatic MO of the longus coli following traumatic cervical spine fracture in an adult. Treatment with supportive measures appears to achieve satisfactory outcome.  相似文献   


20.
Background and purpose — There is debate as to the relative merits of unicompartmental and total knee arthroplasty (UKA, TKA). Although the designer surgeons have achieved good results with the Oxford UKA there is concern over the reproducibility of these outcomes. Therefore, we evaluated published long-term outcomes of the Oxford Phase 3 UKA.

Patients and methods — We searched databases to identify studies reporting ≥10 year outcomes of the medial Oxford Phase 3 UKA. Revision, non-revision, and re-operation rates were calculated per 100 component years (% pa).

Results — 15 studies with 8,658 knees were included. The annual revision rate was 0.74% pa (95% CI 0.67–0.81, n = 8,406) corresponding to a 10-year survival of 93% and 15-year survival of 89%. The non-revision re-operation rate was 0.19% pa (95% CI 0.13–0.25, n = 3,482). The re-operation rate was 0.89% pa (95% CI 0.77–1.02, n = 3,482). The most common causes of revision were lateral disease progression (1.42%), aseptic loosening (1.25%), bearing dislocation (0.58%), and pain (0.57%) (n = 8,658). Average OKS scores were 40 at 10 years (n = 3,417). The incidence of medical complications was 0.83% (n = 1,443).

Interpretation — Very good outcomes were achieved by both designer and non-designer surgeons. The PROMs, medical complication rate, and non-revision re-operation rate were better than those found in meta-analyses and publications for TKA but the revision rate was higher. However, if failure is considered to be all re-operations and not just revisions, then the failure rate of UKA was less than that of TKA.  相似文献   