1.

Background

Total hip arthroplasty in the lateral position involves particularly large variance in the sagittal tilt of pelvis fixation, which affects the imprecision of the cup anteversion leading to poor outcomes. We have added an additional compression pad to an existing device, also to be used in the lateral position, but theoretically enabling fixation on the anatomical pelvic plane (APP) serving as the reference plane. The present study aims to evaluate the usefulness of this device in comparison with the conventional device.

Methods

We have studied 141 patients who underwent total hip arthroplasty at our hospital. Two frontal plain x-rays of the pelvis were obtained preoperatively for each patient after pelvis fixation; one with the conventional lateral fixation device and the other with an APP lateral fixation device. The sagittal tilt of the pelvis in each position was measured with 3D templating software, and variance in the sagittal tilt was compared between the 2 devices.

Results

The mean bias in sagittal tilt relative to the functional pelvic plane (FPP) in the conventional device was ?5.0° ± 4.8° (minus mean backward tilt) and was within 5° relative to the functional pelvic plane in 43%. The mean bias in the sagittal tilt relative to the APP in the APP lateral position device was 1.7° ± 3.1° (forward tilt) and was within 5° relative to the APP in 89%. The APP lateral device significantly reduced the variance in the sagittal tilt.

Conclusion

This device holds promise as a means of reducing the sagittal tilt in a simple, minimally invasive, and highly cost-effective manner.  相似文献   

2.

Background

Accurate cup positioning is one of the most challenging aspects of total hip arthroplasty (THA). Undetected movement of the patient during THA surgery can lead to inaccuracies in cup anteversion and inclination, increasing the potential for dislocation and revision surgery. Investigations into the magnitude of patient motion during THA are not well represented in the literature.

Methods

We analyzed intraoperative pelvic motion using a novel navigation device used to assist surgeons with cup position, leg length, and offset during THA. This device uses an integrated accelerometer to measure motion in 2 orthogonal degrees of freedom. We reviewed the data from 99 cases completed between February and September 2016.

Results

The mean amount of pitch recorded per patient was 2.7° (standard deviation, 2.2; range, 0.1°-9.9°), whereas mean roll per patient was 7.3° (standard deviation, 5.5; range, 0.3°-31.3°). Twenty-one percent (21 of 99) of patients demonstrated pitch of >4°. Sixty-nine percent (68 of 99) of patients demonstrated >4° of roll, and 25% (25 of 99) of patients demonstrated roll of ≥10°.

Conclusion

Our findings indicate that while the majority of intraoperative motion is <4°, many patients experience significant roll, with a large proportion rolling >10°. This degree of movement has implications for acetabular cup position, as failure to compensate for this motion can result in placement of the cup outside the planned safe zone, thus, increasing the potential for dislocation. Further study is warranted to determine the effect of this motion on cup position, leg length, and offset.  相似文献   

3.
BackgroundAnterior and posterior pelvic tilt appears to play a role in total hip arthroplasty (THA) stability. When changing from the standing to the sitting position, the pelvis typically rotates posteriorly while the hips flex and this affects the femoro-acetabular positions. This case-control study compares changes in 3-D acetabular cup orientation during functional pelvic tilt between posterior THA dislocations vs stable THAs.MethodsStanding and sitting 3-D cup orientation was compared between fifteen posterior dislocations vs 233 prospectively followed stable THAs. 3-D cup orientation was calculated using previously validated trigonometric algorithms on biplanar radiographs. Those algorithms combine the angles in the three anatomical planes (coronal inclination, transverse version, and sagittal ante-inclination) in the standing position with the change in sagittal pelvic tilt from standing to sitting to calculate the 3-D orientation in the sitting position.ResultsThe standing cup orientation of the dislocated THAs was only characterized by a lower coronal inclination (P = .039). Compared with the controls, from standing to sitting, they showed less posterior pelvic tilt (P < .001). This led to a significant lower coronal inclination (P < .001) and sagittal ante-inclination (P < .001) in the sitting position but similar transverse version (P = .366).ConclusionsComparing posterior THA dislocations to stable THAs, there is a lower increase of all three orientation angles from standing to sitting. This leads to a decreased sitting coronal inclination and sagittal ante-inclination which may lead to an increased risk of impingement ensued by THA instability. By contrast, the transverse version was not significantly different in both positions. This confirms the importance of biplanar data on functional cup orientation.Level of EvidenceDiagnostic, Level III.  相似文献   

4.

Background

This study's purpose was to determine the impact of THA implantation on pelvic motion, and to assess motion in patients with a history of lumbar fusion or prosthetic dislocation.

Methods

This was an IRB-approved, prospective investigation of 3 cohorts: (1) patients without a history of lumbar surgery undergoing THA (group A), (2) patients with a lumbar fusion (group B), and (3) patients with a THA prosthetic dislocation (group C). All patients received both standing and sitting lateral pelvis images to measure sacral slope and pelvic tilt in the sagittal plane.

Results

Fifty-eight patients were enrolled (24 group A, 27 group B, and 7 group C), with no differences in age, gender, or body mass index (P = .1-.7). In group A, the mean change in sacral slope from standing to sitting was 22.1° ± 15.2° preoperatively and 19.5° ± 14.8° postoperatively. However, in 13 patients, the difference in pelvic motion from the standing to seated position, from preoperatively to postoperatively, was >5° and in 10 patients, this difference was >10°.The change in standing to sitting sacral slope was significantly less in patients with a lumbar fusion (9.8° ± 8.2°) and history of prosthetic dislocation (12.5° ± 4.7°) vs group A (P < .001 and P = .008).

Conclusion

Implantation of a THA can increase or decrease sagittal plane pelvic motion from the standing to seated position with a high degree of variability. Thus, the ability to predict ideal component positioning from preoperative images may be challenging.  相似文献   

5.

Background

Developmental hip dysplasia is the most common cause of secondary hip osteoarthritis. Due to severe acetabular bone deficiency, cup positioning in total hip arthroplasty (THA) of dysplastic hips remains a surgical challenge. The aim was to analyze the functional outcome of far proximal cup positions in primary THA.

Methods

Fifty patients (61 hips) with THA for severe dysplastic osteoarthritis and a far proximal cup position were included. Patients were divided according to the heights of the implanted cups with increasing vertical distance from the interteardrop line (group A: 55-65 mm, group B: 65-75 mm, group C: >75 mm). Functional outcome was assessed at latest follow-up (38 ± 16 months) by Lower Extremity Functional Score, Tegner Activity Score, and Harris Hip Score (HHS). Patients answered a Patient Satisfaction Questionnaire. Leg length discrepancy was estimated radiographically.

Results

The Lower Extremity Functional Score significantly decreased in C (45.3 ± 25) compared to A (66.7 ± 15.3) and B (67.9 ± 9.9). The Tegner Activity Score significantly increased in all subgroups from preoperative to postoperative (2.2 ± 1.3 to 4.1 ± 1.4; P < .05). The mean overall HHS was 89.3 ± 14.7 (A: 89.5 ± 14.3, B: 94.3 ± 6.5, C: 78.3 ± 22.1). The HHS domains of activity of daily life and gait were significantly reduced in C (P < .05). Patients described a high satisfaction level with the surgery. No significant differences were found with regard to preoperative and postoperative leg lengthening (P = .881). Neither dislocations, impingement problems nor neurologic complications were observed.

Conclusion

Primary THA without any concomitant surgical interventions with a far proximal cup position offers a safe and effective treatment option in severe dysplastic hip osteoarthritis.  相似文献   

6.

Background

Movement of the pelvis during implantation of total hip arthroplasty (THA) has a major influence on the positioning of the acetabular cup. Strong traction caused by retractors leads to iatrogenic pelvic lift and can thus be partly responsible for cup malpositioning. The objective of this study was to investigate such factors that influence pelvic lift.

Methods

The dynamic movement of the pelvis was measured during implantation of THA in 67 patients. This was done by measuring the acceleration using the SensorLog app on a smartphone.

Results

At its maximum, the pelvis was lifted by an average of 6.7°. When impacting the press-fit cup, the surgical side was raised by 4.4° compared with the time of skin incision. This lift at the time of cup implantation correlates significantly with the body mass index and the patient's abdominal and pelvic circumference.

Conclusion

Every surgeon performing THA must be aware of the pelvic lift during an operation. Especially in patients with a high body mass index, a large abdominal circumference, or a large pelvic circumference, there is an increased risk of malpositioning of the acetabular cup. When impacting the cup, we recommend releasing the traction of the retractor, so that the pelvis can tilt back into its natural position, and thus, the anticipated cup positioning can be implemented as exactly as possible.  相似文献   

7.
Total hip arthroplasty (THA) is regarded as one of the most successful surgical procedures of modern times yet continues to be associated with a small but significant complication rate. Many early failures may be associated with poor component positioning with, in particular, acetabular component orientation dependent on the subjective judgement of the surgeon. In this paper, we compare the manufacturers' instructions on acetabular cup orientation with the literature-based recommended safety zones and surgical technique, by transforming them onto a single, clinically-relevant framework in which the different reference systems, safety guidelines and current instrumentation surgical techniques can be evaluated. The observed limited consensus between results reflects ongoing uncertainty regarding the optimum acetabular component positioning. As malpositioning of the acetabular cup increases the risk of revision surgery, any ambiguity over the correct position can have a causal effect. Our analysis highlights the need for a surgical reference system which can be used to describe the position of the acetabular cup intra-operatively.  相似文献   

8.
《The Journal of arthroplasty》2020,35(12):3621-3626
BackgroundThis study is prospectively conducted to evaluate surgical complications of monolithic dual mobility cup total hip arthroplasty (THA) in elderly patients with fractured neck of the femur.MethodsNinety-seven patients (97 hips) with displaced femoral neck fracture who gave informed consent for participation were prospectively enrolled. Their mean age was 76.6 years (range, 60-95 years), and the mean bone mineral density T-score of neck of the femur was −2.8 (range, −1.2 to −5.5). All patients underwent THA with monolithic dual mobility cup, and computed tomography scans were obtained to evaluate radiographic parameters including anteversion, inclination, and loosening of acetabular cups, and periprosthetic acetabular fractures.ResultsWith regard to cup orientation, mean inclination angle was 40.2° (range, 23.5°-63°) and mean anteversion was 32.6° (range, 7°-66.2°). The proportion of surgical outliers was 10.3% (10/97) in inclination and 35.1% (34/97) in anteversion. Early cup loosening within 2 weeks was detected in 2 hips. Periprosthetic acetabular fractures were identified in 6 hips (6/97, 6.2%). Of the 6 fractures, 5 nondisplaced fractures were healed with conservative management, but 1 fracture with displacement eventually led to cup loosening and the patient underwent revision surgery. Reoperation rate of the monolithic dual mobility cup was 4.1% (4/97).ConclusionThe use of the monolithic dual mobility was associated with improper cup fixation and periprosthetic acetabular fractures in the elderly with poor bone stock, although the dual mobility cup lowered the risk of early dislocation after THA.  相似文献   

9.
Diagnosis of a pelvic discontinuity before revision total hip arthroplasty is critical for adequate preoperative planning. The lateral view of the pelvis or high-angle oblique views can aid in visualizing the posterior column when hip hardware obscures the view on standard anteroposterior and Judet views of the pelvis. These views are easy to obtain and can provide valuable information when planning revision total hip arthroplasty.  相似文献   

10.
BackgroundAlthough pelvic osteotomy (PO) is an important surgical procedure that can alleviate symptoms and potentially slow progression of osteoarthritis in patients with development dysplasia of the hip, some patients eventually require conversion to total hip arthroplasty (THA). This study aimed to determine the outcome of conversion THA in patients with prior PO.MethodsForty nine patients with a history of prior PO who underwent conversion THA at a single institution were matched at a 1:3 ratio based on the date of surgery, age, gender, and body mass index with 147 developmental dysplasia of the hip patients who underwent primary THA without prior PO. A retrospective chart review was performed to compare outcomes at a minimum follow-up of 2 years.ResultsPatients with prior PO required more supplemental screw fixation for the acetabular component (59.2% vs 38.1%, P = .016), more autologous bone grafting (24.5% vs 11.6%, P = .048), had a longer mean operative time (106.0 vs 79.8 minutes, P < .001), and greater estimated blood loss (350.0 vs 206.8 mL, P = .015). Patients with prior PO had smaller cup version angle (26.0° vs 29.0°, P = .012) and greater discrepancy in the limb length (10.3 vs 7.26 mm, P = .041). Eight hips (16.3%) with prior PO and 6 (4.1%) without osteotomy required reoperation (P = .008). There was no difference in outcome scores at the latest follow-up.Conclusion: THA after prior PO is technically demanding, leading to longer operative times, greater blood loss, and variation in implant placement. Although functional outcomes are similar, THA after a prior PO is more likely to require reoperation.ConclusionTHA after prior PO is technically demanding, leading to longer operative times, greater blood loss, and variation in implant placement. Although functional outcomes are similar, THA after a prior PO is more likely to require reoperation.  相似文献   

11.
BackgroundPatients undergoing primary total hip arthroplasty (THA) following pelvic radiation have historically had poor survivorship free of aseptic acetabular component loosening. However, several series have reported improved results with tantalum acetabular components. The purpose of this study is to assess implant survivorship, radiographic results, and clinical outcomes of contemporary, porous titanium acetabular components in the setting of prior pelvic radiation.MethodsWe retrospectively reviewed 33 patients (38 hips) with prior pelvic radiation between 2006 and 2016 who underwent primary THA. The mean overall pelvic radiation dose was 6300 cGy with a mean latency period to THA of 5 years. The most common acetabular component was Pinnacle (DePuy Synthes) in 76%. Eight-seven percent of cups were fixed with screws. The mean age at primary THA was 74 years, 76% were male, and the mean body mass index was 30 kg/m2. Mean follow-up was 5 years.ResultsThe 10-year survivorship free of revision for aseptic loosening, free of any revision, and free of any reoperation were 100%, 89%, and 89%, respectively, when accounting for death as a competing risk. There were 3 revisions; one each for taper corrosion, recurrent dislocation, and infection. Radiographically, all cups had evidence of osteointegration and none had radiographic evidence of loosening. The mean Harris Hip Score improved from 50 to 84 postoperatively (P < .0001).ConclusionContemporary porous titanium acetabular components with supplemental screws provided excellent implant fixation in patients with prior therapeutic pelvic radiation. At 10 years, 100% of these components were free of revision for aseptic loosening and 100% were radiographically well-fixed.Level of EvidenceLevel IV, Therapeutic.  相似文献   

12.
13.
We evaluated the clinical and radiographic results of primary total hip arthroplasty for 81 dysplastic hips (71 patients) using cementless Spongiosa Metal II cups (ESKA Implants, Lübeck, Germany). The mean follow-up period was 6.4 years (minimum 5 years), and the preoperative mean Japanese Orthopaedic Association hip score had improved from 45.2 to 87.4 points at the latest follow-up. The radiographic outcome was no aseptic loosening in all 81 hips. The hip center was located significantly more superior than in the contralateral normal hip in 45 patients, but the difference was less than 10 mm; however, there was no significant difference in the lateral position of the hip center. The use of a Spongiosa Metal II cup for dysplastic hips provided satisfactory 5- to 10-year clinical and radiographic results.  相似文献   

14.
目的探讨人工全髋关节置换术中髋臼内陷的处理方法及早期疗效。方法 2006年7月至2012年12月,收治继发性髋臼内陷症21例28髋,男12例17髋,女9例11髋;年龄26~68岁,平均56.3岁;原发疾病:类风湿关节炎9例14髋,强直性脊柱炎6例8髋,感染4例4髋(结核和化脓性关节炎各2例2髋),髋臼骨折畸形愈合2例2髋;轻度髋臼内陷6髋、中度髋臼内陷12髋、重度髋臼内陷10髋。患者均行人工全髋关节置换,术中单纯植入生物型臼杯6髋,打压植骨后直接植入生物型臼杯18髋,直接植入骨水泥杯2髋,打压植骨后骨水泥杯固定2髋。结果术后21例28髋切口均一期愈合,无感染及下肢深静脉血栓形成等并发症发生。患者均获随访,随访时间12~78个月,平均42个月。末次随访时,髋关节Harris评分为(87±4.1)分,术前Harris髋关节评分平均(45±3.3)分,较术前有显著差异,优22髋,良6髋,优良率100%。臼杯平均外翻角(42±3.5)°。植骨均骨性融合,无臼杯再次内陷及松动。髋关节活动度增加:屈曲增加(64.3±4.5)°,外展增加(20.5±3.5)°,内旋增加(16.5±2.5)°,外旋增加(19±2.5)°。末次随访时X线片显示所有假体均获得骨性稳定。术前股骨头内陷3~30 mm,平均13 mm;术后股骨头内陷0~5 mm,平均2 mm。人工股骨头旋转中心外移3~28 mm,平均10.5 mm。假体位置良好,无松动、下沉,植骨与髋臼融合,无髋臼再次内陷。结论人工全髋关节置换治疗髋臼内陷时,恢复患髋股骨偏心距及髋臼旋转中心,可获满意早期疗效。  相似文献   

15.

Background

Total hip arthroplasty (THA) provides a successful salvage option for failed acetabular fractures. The complexity of arthroplasty for a failed acetabular fracture will depend on the fracture pattern and the initial management of the fracture. Our objective was to compare the midterm outcome of THA between patients who presented with failed acetabular fractures following initial surgical or nonsurgical treatment.

Methods

Forty-seven patients underwent cementless THA ± acetabular reconstruction following failed treatment of acetabular fractures. Twenty-seven were initially treated by surgery (group A) and 20 had nonsurgical treatment (group B). Intraoperative measures, preoperative and follow-up clinical, radiological, and functional outcomes were compared between the 2 groups.

Results

The mean surgical time, blood loss, and need for blood transfusion were significantly less in group A (P < .05). Acetabular reconstruction to address cavitary or segmental defects was needed in a significantly higher number of patients in group B (P = .006). Significant improvement in modified Merle d'Aubigne and Oxford scores was seen postsurgery in both groups. Acetabular component survival with aseptic loosening as end point was 98%. Overall survival rate with infection, revision, or loosening as end point was 93% at a mean follow-up of 7 years ± 17 months.

Conclusion

THA for a failed acetabular fracture is greatly facilitated by initial surgical treatment. Although functional results and survivorship were similar in both groups, failed nonsurgical treatment in complex fractures is associated with migrated femoral head and extensive acetabular defects requiring complex acetabular reconstruction.  相似文献   

16.

Background

Sitting radiographs have been used as a pre-operative tool to plan patient-specific total hip arthroplasty (THA) component position that would improve hip stability. Previous work has demonstrated that spinal mobility may impact functional acetabular position when seated. We sought to determine whether patients who dislocate following THA have different sitting spinopelvic alignment or acetabular component orientation compared to patients who did not dislocate.

Methods

A consecutive series of 1000 patients underwent post-operative low-dose biplanar spine-to-ankle lateral radiographs in standing and sitting positions 1 year following THA. Twelve patients (1% of all patients) experienced hip dislocation. Patients were categorized as having normal lumbar spines (without radiographic arthrosis) or as having lumbar multi-level degenerative disc disease. Measurements of spinopelvic alignment parameters (including sacral slope, lumbar lordosis, and proximal femur angles) and acetabular component orientation in sitting position (functional inclination and functional anteversion) were performed.

Results

Patients who dislocated had significantly less spine flexion, less change in pelvic tilt, and more hip flexion from standing to sitting positions compared to patients with normal spines. In sitting position, dislocators had acetabular components with less functional inclination and less functional anteversion.

Conclusion

This study demonstrates that patients with fixed spinopelvic alignment from standing to sitting position are at higher risk of hip dislocation. Imaging patients from standing to sitting position using this technique can provide valuable information on whether a patient has fixed spinopelvic alignment with postural changes and is therefore at higher risk of dislocation.  相似文献   

17.
BackgroundMalposition of the acetabular component during total hip arthroplasty (THA) is associated with increased risk of dislocation, reduced range of motion, and accelerated wear. The purpose of this study is to compare cup positioning with a portable, accelerometer-based hip navigation system and conventional surgical technique.MethodsIn a prospective, randomized, clinical study, cups were implanted with a portable, accelerometer-based hip navigation system (navigation group; n = 55) or conventional technique (conventional group; n = 55). THA was conducted in the lateral position and through posterior approach. The cup position was determined postoperatively on pelvic radiograph and computed tomography scans.ResultsAn average cup abduction of 39.2° ± 4.6° (range, 27° to 50°) and an average cup anteversion of 14.6° ± 6.1° (range, 1° to 27.5°) were found in the navigation group, and an average cup abduction of 42.9° ± 8.0° (range, 23° to 73°) and an average cup anteversion of 11.6° ± 7.7° (range, −12.1° to 25°) in the conventional group. A smaller variation in the navigation group was indicated for cup abduction (P = .001). The deviations from the target cup position were significantly lower in the navigation group (P = .001, .016). While only 37 of 55 cups in the conventional group were inside the Lewinnek safe zone, 51 of 55 cups in the navigation group were placed inside this safe zone (P = .006). The navigation procedure took a mean of 10 minutes longer than the conventional technique.ConclusionUse of the portable, accelerometer-based hip navigation system can improve cup positioning in THA.  相似文献   

18.

Background

Cup anteversion and inclination are important for avoiding implant impingement and dislocation in total hip arthroplasty. However, functional cup anteversion and cup inclination also change as the pelvic sagittal inclination (PSI) changes. Therefore, PSI in both supine and standing positions was measured in a large cohort in this study.

Methods

A total of 422 patients (median age, 61; range, 15-87) who underwent total hip arthroplasty were the subjects of this study. There were 83 patients with primary osteoarthritis (OA), 274 patients with developmental dysplasia–derived secondary OA, 48 patients with osteonecrosis, and 17 patients with rapidly destructive coxopathy (RDC). Preoperative PSI in supine and standing positions was measured by automated computed topography segmentation and landmark localization of the pelvis followed by intensity-based 2D-3D registration, and the number of cases in which PSI changed more than 10° posteriorly was calculated. Hip disease, sex, and age were analyzed if they were related to a PSI change of more than 10°.

Results

The median PSI was 5.1° (interquartile range, 0.4°-9.4°) in supine and ?1.3° (interquartile range, ?6.5° to 4.2°) in standing position. There were 79 cases (19%) in which the PSI changed more than 10° posteriorly from supine to standing. Elder age and patients with primary OA and RDC were revealed to be the related factors.

Conclusion

PSI changed more than 10° posteriorly from supine to standing in 19% of cases. Age and diagnosis of primary OA and RDC were the significant factors for the posterior rotation.  相似文献   

19.
《The Journal of arthroplasty》2020,35(12):3758-3764
BackgroundSpinopelvic pathology has been identified as a potential risk factor for instability after total hip arthroplasty. Spinopelvic radiographic parameters used to diagnose spinopelvic disease may also impact optimal cup placement. The purpose of this study was to assess the arthroplasty surgeon accuracy in making spinopelvic measurements.MethodsFive fellowship-trained or in-training arthroplasty surgeons reviewed 44 lateral lumbar radiographs in two sessions. All evaluators were instructed how to perform measurements but had little experience in doing so. Traditional measurements included the pelvic tilt (PT), pelvic incidence (PI), and sacral slope (SS), and novel measurements included the acetabular ante-inclination (AI) and pelvic femoral angle. Surgeon measurements were compared with those made by experienced engineers. intraclass coefficients (ICCs) were calculated and interpreted.ResultsThe interobserver reliability for 4 of 5 of the spinopelvic parameters was good to excellent (ICC >0.75) with the highest reliability seen for PI measurement (ICC = 0.939). Only moderate interobserver reliability was observed for AI measurement (ICC = 0.559). Intraobserver reliability was the highest for the PI and SS, ranging from moderate to good (ICC, 0.718 to 0.896). The lowest intraobserver reliability was seen for the AI (ICC range, 0.026 to 0.545) and pelvic femoral angle (ICC range, 0.035 to 0.828). Surgeon measurements of PT and SS were compared with engineer measurements with extremely poor correlation observed (ICC <0 for all).ConclusionSurgeon intraobserver and interobserver reliability in making novel radiographic measurements was low compared with traditional radiographic measurements. Surgeon reliability in making traditional measurements of the PT and SS was very poor compared with experienced assessors using software-based measurements.  相似文献   

20.

Background

The purpose of the study was to evaluate the implant positions and clinical results of revision total hip arthroplasty (THA) using an imageless navigation with the concept of combined anteversion.

Methods

A total of 40 cementless revision THAs (24 men and 16 women) using an imageless navigation with the concept of combined anteversion were retrospectively evaluated. The concept of combined anteversion was applied in cup positioning based on Widmer's equation (cup anteversion + 0.7 × stem anteversion). The mean follow-up period was 80.7 months. Postoperatively, the inclination of the cup was evaluated on standard anteroposterior view of the radiograph, and the anteversion of the cup and femoral stem was evaluated using computed tomography scan. A cup inclination of 40° ± 10° and combined anteversion of the cup and femoral stem of 37° ± 10° based on Widmer's equation were regarded as the “safe zone.”

Results

The average anteversion of the revised femoral stems was 15.3° ± 2.9° (range, 9.5°-21.5°), whereas that of the remained femoral stems was 17.4° ± 9.7° (range, 4.2°-29.8°). The inclination, anteversion of the cup, and combined anteversion after revision THA were 42.3° ± 3.1° (range, 32.1°-48.2°), 25.0° ± 2.9° (range, 16.9°-29.5°), and 36.1° ± 3.4° (range, 27.2°-42.9°), respectively. Therefore, the position of the implants, relative to the safe zone, showed no outliers after the revision surgery. Neither dislocation nor osteolysis was observed after the surgery.

Conclusion

Favorable results of this study indicate that imageless navigation helps the surgeon in placing the components of revision THA in the safe zone. This study also shows that when this safe zone is consistently obtained, then no postoperative dislocations were observed in these patients over the 6-year follow-up period.  相似文献   