1.

Background

The R3 acetabular system used with its metal liner has higher revision rates when compared to its ceramic and polyethylene liner. In June 2012, the medical and healthcare products regulatory agency issued an alert regarding the metal liner of the R3 acetabular system.

Methods

Six retrieved R3 acetabular systems with metal liners underwent detailed visual analysis using macroscopic and microscopic techniques.

Results

Visual analysis discovered corrosion on the backside of the metal liners. There was a distinct border to the areas of corrosion that conformed to antirotation tab insertions on the inner surface of the acetabular shell, which are for the polyethylene liner. Scanning electron microscopy indicated evidence of crevice corrosion, and energy-dispersive X-ray analysis confirmed corrosion debris rich in titanium.

Conclusion

The high failure rate of the metal liner option of the R3 acetabular system may be attributed to corrosion on the backside of the liner which appear to result from geometry and design characteristics of the acetabular shell.  相似文献   

2.

Background

Cross-linked polyethylene (XLPE) acetabular liners used in cementless total hip arthroplasty (THA) have demonstrated better wear resistance at 10 years compared with conventional polyethylene (CPE) liners. No clinical studies have compared XPLE to CPE liners beyond 10 years.

Methods

We performed a 15-year retrospective cohort study on cementless THA performed in patients with developmental hip dysplasia to measure the differences in polyethylene wear rates and the presence of osteolysis. Twenty-four THAs with XLPE and 17 THAs with CPE were evaluated. The mean age of patients was 55.9 years (41-68) in the XLPE group and 54.4 years (40-67) in the CPE group. The mean follow-up period was 15.1 years (13.9-16.1) in the XLPE group and 15.2 years (14.5-16.0) in the CPE group.

Results

The XLPE group had a significantly lower wear rate at 5 and 10 years compared with the CPE group; however, no significant difference was found at 15 years (XLPE group, 0.040 mm/y; CPE group, 0.034 mm/y). In addition, the incidence of osteolysis did not differ significantly between the groups. However, the incidence of excessive wear between 10 and 15 years after surgery in the XLPE group was significantly higher than that in the CPE group.

Conclusion

XLPE demonstrated no advantage in the wear rate or the incidence of osteolysis at 15 years, despite having superior wear resistance up to 10 years. It is concerning that the incidence of excessive wear was higher in the XLPE group between 10 and 15 years, and this finding should alert the arthroplasty community to this possible problem with the more highly cross-linked polyethylene.  相似文献   

3.

Background

Polyethylene acetabular components are common in hip arthroplasty. Highly cross-linked polyethylene (HXLPE) has lower wear than ultra-high molecular weight polyethylene (UHMWPE). Evidence suggests that wear particles induce inflammation causing periprosthetic osteolysis contributing to implant loosening with wear rates of 0.05 mm/y were considered safe. We aimed to compare incidence and volume of periacetabular osteolysis between HXLPE and UHMWPE using computed tomography.

Methods

Initially, 54 hips in 53 patients were randomized to HXLPE or UHMWPE acetabular liner. At 10 years, 39 hips in 38 patients remained for the radiostereometric analysis' demonstrating significantly lower wear in the HXLPE group. At 12 years, 14 hips in 13 patients were lost to follow-up leaving 25 hips for computed tomography assessment. Images were reconstructed to detect osteolysis and where identified, areas were segmented and volumized.

Results

Osteolysis was observed in 8 patients, 7 from the UHMWPE group and only 1 from the HXLPE group (Fisher exact, P = .042). There was no correlation between the amount of polyethylene wear and osteolysis volume; however, the radiostereometric analysis-measured wear rate in patients with osteolysis from both groups was significantly higher than overall average wear rate.

Conclusion

This data demonstrates lower incidence of periacetabular osteolysis in the HXLPE group of a small cohort. Although numbers are too low to estimate causation, in the context of lower wear in the HXLPE group, this finding supports the hypothesis that HXLPE may not elevate osteolysis risk, and hence does not suggest that HXLPE wear particles are more biologically active than those generated by earlier generations of polyethylene.  相似文献   

4.

Background

We evaluated the clinical and radiographic outcomes, including femoral head penetration, of total hip arthroplasty performed using a specific polyethylene (PE) liner in small Asian patients at 10 years after the index surgery. In addition, we investigated whether femoral head penetration was affected by patient-related, implant-related, and surgical factors.

Methods

Between August 2002 and June 2005, for cementless primary total hip arthroplasty, we used acetabular PE liners that were manufactured from GUR 1050 resin, machined from isostatic compression-molded bar stock, and sterilized with a gamma ray irradiation in argon gas. We assessed 82 hips in 78 patients who received these liners.

Results

The mean Harris hip score improved from 41.0 preoperatively to 84.5 at 10 years postoperatively. Periprosthetic osteolysis was observed in 7 hips (9.8%). No acetabular component migration was detected, and no revision surgery was performed 10 years postoperatively. The mean steady-state wear rate was 0.031 mm/y, which was lower than the wear rate for other conventional PE liners of the previous studies. Among the patient-related, implant-related, and surgical factors, sex was significantly associated with the mean steady-state wear rate, with a higher rate in male patients than in female patients.

Conclusion

PE acetabular liners used in small Asian patients show similar clinical outcomes and reduced wear compared with those of other liners. In addition, sex is significantly associated with the mean steady-state wear rate, and the steady-state wear rate is higher in male patients than in female patients.  相似文献   

5.

Background

The aim of our study was to evaluate the clinical and radiographic outcomes of malseating of the acetabular liner in ceramic-on-ceramic total hip arthroplasty (THA).

Methods

Outcomes for 160 ceramic-on-ceramic THAs, contributed by 116 women and 39 men, were evaluated. Clinical and radiographic measurements were obtained over a 5- to 15-year follow-up for analysis.

Results

Liner malseating was identified in 20% of cases. Outcomes for 32 cases with liner malseating (group A) were compared to outcomes for 128 joints with correct liner seating (group B). The Harris hip score at the last follow-up was 90.1 for group A and 89.6 for group B. Osteolysis was identified in 5 cases in group A (15.6%), compared to 3 cases in group B (P < .001). No significant between-group differences were identified with regard to ceramic fracture, audible squeaking, loosening of components, and revision THA. The mean annual liner wear rate was comparable between groups, 0.0045 mm/y for group A and 0.0039 mm/y for group B. The 10-year Kaplan-Meier survivorship, based on an end point of revision THA, was 100% for group A and 99.0% for group B.

Conclusion

Over a moderate-length follow-up of 5-15 years, malseating of the acetabular liner was not associated with negative clinical outcomes or THA survivorship. Malseating did increase the incidence of osteolysis, a risk factor for adverse effects. Long-term follow-up studies are needed to fully quantify the effects of malseating of the acetabular liner.  相似文献   

6.

Background

Malseating of ceramic liner appeared as a matter of concern in multibearing metal, although multibearing cup can be coupled with hard liners as well as polyethylene liner. In this metal shell, the inner taper angle should be 10° for the modularity, while standard metal shells for ceramic liner have an inner taper angle of 18°. However, there has been no study in the effect of taper angle to the risk of malseating. We evaluated whether the taper angle of metal shell might affect the malseating of ceramic liner, and dissociation force of ceramic liner from metal shell.

Methods

Three surgeons manually inserted ceramic liners into two designs of metal shell with different tapers angles (10° and 18°). We compared malseating rates of ceramic liners and push-out strengths, which means dissociation force of the ceramic liner from the metal shell, between these two metal shell designs.

Results

The malseating rates in 10° metal shell were higher than those in 18° metal shell (23.3% vs 0%, P < .05). The mean dissociation force (1148.8 ± 46.7 N) in 10° taper cup was higher than that (389.7 ± 108.3 N) in 18° taper cup (P < .01).

Conclusion

Our results suggest that surgeon should be cautious about malseating of ceramic liner when using multi-bearing metal shell with inner taper angle of 10°.

Clinical relevance

When surgeon use multi-bearing metal shell with inner taper angle of 10°, our results suggest that surgeon should be cautious about malseating of ceramic liner.  相似文献   

7.

Background

The use of porous tantalum for the acetabular component in primary total hip arthroplasty (THA) has demonstrated excellent short-term and midterm results. However, long-term data are scarce. The purpose of this prospective study is to report the long-term clinical and radiologic outcome following use of an uncemented porous tantalum acetabular component in primary THA with a minimum follow-up of 17.5 years, in a previously studied cohort of patients.

Methods

We prospectively followed 128 consecutive primary THAs in 140 patients, between November 1997 and June 1999. A press-fit porous tantalum monoblock acetabular component was used in all cases. All patients were followed clinically and radiographically for a mean of 18.1 years (range, 17.5-19 years).

Results

Mean age of patients at the time of operation was 60.4 years. Harris hip score, Oxford hip score, and range of motion were dramatically improved in all cases (P < .001). At last follow-up, all cups were radiographically stable with no evidence of migration, gross polyethylene wear, progressive radiolucencies, osteolytic lesions, or acetabular fractures. The survivorship with reoperation for any reason as end point was 92.8% and the survivorship for aseptic loosening as an end point was 100%.

Conclusion

The porous tantalum monoblock cup in primary THA demonstrated excellent clinical and radiographic outcomes with no failures because of aseptic loosening at a mean follow-up of 18.1 years.  相似文献   

8.

Background

There are relatively few 20-year results of uncemented acetabular components, and most of these are modular designs. This study reports the 20-year results of a monoblock press-fit acetabular component.

Methods

A total of 122 total hip arthroplasties (111 patients) using the Morscher cup were reviewed at a mean of 19.7 years. The average age at implantation was 57.3 years (range, 36-74 years), and 81 (66%) were men.

Results

Twenty-two patients (25 hips) had died. Seven hips were revised, including 5 acetabular revisions. Six patients (6 hips) declined to participate but were known not to have been revised. The mean Oxford hip score was 41.1 (range, 22-48), and the mean reduced Western Ontario and McMaster Universities Osteoarthritis Index score was 5.7/48 (range, 0-24). Eccentric wear was seen in 13 (15.7%) and major osteolysis in 14 (17%) of 82 surviving hips with radiographs. The all-cause revision rate was 0.32 per 100 observed component years (95% confidence interval [CI], 0.13-0.66). The 20-year Kaplan-Meier survival was 93.4% (CI, 86.6-96.8) for all-cause revisions, 95.5% (CI, 89.4-98.1) for any acetabular revision, and 97.1% (CI, 91.2-99.1) for acetabular aseptic loosening, wear, or osteolysis.

Conclusion

The Morscher acetabular component has continued to perform well at 20 years despite using conventional polyethylene with results that match or surpass other cementless acetabulae.  相似文献   

9.

Background

Cross-linked polyethylene (XLPE) has demonstrated significantly reduced wear and osteolysis into the second decade for total hip arthroplasty. There is a relative paucity of data with ≥36-mm bearings. Issues include potential effects of reduced liner thickness and component position on wear, osteolysis, and mechanical failure of the bearing.

Methods

Radiographs of 48 primary total hip arthroplasties with ≥36-mm modular XLPE bearings were analyzed at a minimum 5 years postoperative on serial radiographs using a validated, edge-detection-based algorithm. Subgroups were examined to assess the effect of bearing diameter, liner thickness, acetabular abduction angle, and acetabular anteversion on XLPE wear.

Results

There was no significant difference in volumetric wear when subgroups were stratified by component factors: liner thickness (<6.5 mm vs ≥6.5 mm) 40.69 mm3/y vs 24.47 mm3/y, respectively (P = .315); acetabular component abduction angle (<45° vs ≥45°): 38.68 mm3/y vs 27.8 mm3/y, respectively (P = .522); acetabular anteversion (<20° vs ≥20°): 41.32 mm3/y vs 31.79 mm3/y, respectively (P = .521). There were no dislocations, mechanical failures, or revisions. There were 7 hips with volumetric wear rates ≥80 mm3/y; 1 had possible osteolysis.

Conclusion

Larger-diameter XLPE wear was not measurably affected by liner thickness, acetabular abduction angle, or acetabular anteversion. However, there is a trend for increasing volumetric wear with increasing bearing size. Wear outliers do occur, and continued follow-up of larger-diameter XLPE bearings is warranted.  相似文献   

10.

Background

In total hip arthroplasty (THA) of hips with coxa vara, the femoral stems might be inserted in a varus alignment. To avoid varus insertion, we designed a technique, which we termed “trochantoplasty.” In this procedure, the medial half of the greater trochanter was removed during THA.

Methods

We evaluated 30 patients (31 hips) who had coxa vara deformity and underwent THA using trochantoplasty at the mean follow-up of 5 years (range, 3-9 years).

Results

All stems were inserted in the neutral position. One Vancouver type 1 periprosthetic femoral fracture occurred after a fall at postoperative 2 months. At the latest follow-up, the mean power of abductor was 4.3 (range, 3-5). Four patients had moderate limp whereas 26 patients had slight limp. The abduction at 90° flexion ranged from 15° to 45° (mean, 35°). There was no revision. All prostheses had bone-ingrown stability without any detectable wear or osteolysis. The mean Harris hip score was improved from 66.9 to 89.4 points.

Conclusion

Trochantoplasty can be used to avoid varus insertion of the femoral stem while performing THA in patients with coxa vara deformity without compromising the abductor mechanism.  相似文献   

11.

Background

Liner exchange for articular component wear in total hip joint replacements (THJRs) is a common procedure, often thought to be benign with reliable outcomes. Recent studies, however, suggest high failure rates of liner exchange revisions with significant complications. The primary aim of this study was, therefore, to analyze the survivorship of isolated liner exchange for articular component wear, and secondarily to assess the influence of patient demographics (gender, age, and American Society of Anaesthesiologists [ASA] ratings) on rerevisions following isolated liner exchange for wear.

Methods

A retrospective review of the 15-year New Zealand Joint Registry (1999-2014) was performed, analyzing the outcomes of isolated liner exchange for articular component wear. The survivorship as defined as rerevision with component exchange was determined and 10-year Kaplan-Meier survivorship curves were constructed. These revision rates were compared to age, gender, and ASA rating groups using a log-rank test.

Results

The 10-year survivorship of THJR following liner exchange revision for liner wear was 75.3%. If a rerevision was required, the median time to rerevision was 1.33 years with a rerevision rate of 3.33 per 100 component years (95% confidence interval 2.68-4.08/100 component years). The principle reasons for rerevision were dislocation (48.4%) and acetabular component loosening (20.9%). There was no statistically significant difference in rerevision rates based on gender, age categories, or ASA scores.

Conclusion

THJR isolated liner exchange for liner wear is not a benign procedure with a survivorship of 75.3% at 10 years. Surgeons contemplating liner exchange revisions should be cognisant of this risk and should adequately assess component position and stability preoperatively.  相似文献   

12.

Background

Changes in spinal alignment and pelvic tilt alter acetabular orientation in predictable ways, which may have implications on stability of total hip arthroplasty (THA). Patients with sagittal spinal deformity represent a subset of patients who may be at particularly high risk of THA instability because of postural compensation for abnormal spinal alignment.

Methods

Using standing stereoradiography, we evaluated the spinopelvic parameters, acetabular cup anteversion, and inclination of 139 THAs in 107 patients with sagittal spinal deformity. Standing images were compared with supine pelvic radiographs to evaluate dynamic changes in acetabular cup position. Dislocation and revision rates were procured through retrospective chart review. The spinal parameters and acetabular cup positions among dislocators were compared with those who did not dislocate.

Results

The rate of THA dislocation in this cohort was 8.0%, with a revision rate of 5.8% for instability. Patients who sustained dislocations had significantly higher spinopelvic tilt, T1-pelvic angle, and mismatch of lumbar lordosis and pelvic incidence. Among all patients, 78% had safe anteversion while supine, which decreased significantly to 58% when standing due to increases in spinopelvic tilt. Among dislocating THA, 80% had safe anteversion, 80% had safe inclination, and 60% had both parameters within the safe zone.

Conclusion

In this cohort, patients with THA and concomitant spinal deformity have a particularly high rate of THA instability despite having an acetabular cup position traditionally thought of as within acceptable alignment. This dislocation risk may be driven by the degree of spinal deformity and by spinopelvic compensation. Surgeons should anticipate potential instability after hip arthroplasty and adjust their surgical plan accordingly.  相似文献   

13.

Background

There are limited studies to evaluate long-term clinical and radiographic outcomes of alumina delta ceramic-on-ceramic bearings in cementless total hip arthroplasty (THA). The purpose of this study was to evaluate the clinical and radiographic results, prevalence of osteolysis, squeaking, and fracture of ceramic material associated with the use of the alumina delta ceramic-on-alumina delta ceramic bearing in cementless THA in patients aged <50 years.

Methods

We reviewed the cases of 277 patients (334 hips) who underwent a cementless THA using alumina delta ceramic-on-alumina delta ceramic when they were 50 years or younger at the time of surgery. Demographic data; Harris Hip Score; Western Ontario McMaster Universities Osteoarthritis Index; and University of California, Los Angeles activity score were recorded. Radiographic and computerized tomographic evaluations were used to evaluate implant fixation and osteolysis. Squeaking sound and ceramic fracture were documented. The mean follow-up was 7.8 years (range, 6-9).

Results

The mean postoperative Harris Hip Score, Western Ontario and McMaster Universities Osteoarthritis Index score, University of California, Los Angeles activity score were 93 points, 15 points, and 8.6 points, respectively. Two patients had thigh pain (grade 7 points). All acetabular components and all but 2 femoral components were well fixed. Thirty-three hips (10%) exhibited clicking sound, and 2 hips (0.6%) exhibited squeaking sound. No hip had osteolysis or ceramic head or liner fracture.

Conclusion

Our minimum 6-year follow-up results with the use of alumina delta ceramic-on-alumina delta ceramic bearings in patients aged <50 years suggest that cementless THA provides a high rate of survivorship without evidence of osteolysis or fracture of ceramic material.  相似文献   

14.

Background

Dual mobility total hip arthroplasty (THA) components improve stability, yet use of a modular cobalt alloy acetabular liner may be associated with metal ion release. This study's purpose was to measure blood metal ion levels in young, active patients receiving a dual mobility THA prosthesis.

Methods

This is a prospective study of young, active patients undergoing primary THA. Twenty-six patients received a 22-mm cobalt alloy (n = 10) or a 28-mm ceramic (n = 16) femoral head, a modular cobalt chrome acetabular liner, with a highly cross-linked polyethylene insert (dual mobility). Seventeen control patients received a 32-mm cobalt alloy (n = 6), oxidized zirconium (n = 5), or ceramic (n = 6) femoral head and polyethylene acetabular liner (conventional). All patients received a cementless, titanium femoral stem. Blood metal ion levels (μg/L) were measured preoperatively and at 1 year postoperatively.

Results

No difference was present for age or body mass index (P = .5 and .9). At 1 year postoperatively, mean cobalt levels were greater in the dual mobility cohort (0.23 ± 0.39 vs 0.15 ± 0.07, P < .001). Four patients in the dual mobility cohort had a cobalt level outside the reference range (0.03-0.29), with values from 0.34 to 1.81 μg/L. One patient in the conventional cohort had a cobalt level outside the reference range with a value of 0.39 μg/L.

Conclusion

The presence and clinical significance of increased cobalt levels in 4 patients with the use of a modular dual mobility prosthesis demonstrates the necessity of continued surveillance.  相似文献   

15.

Background

Coronal plane alignment is one of the contributing factors to polyethylene wear in total knee arthroplasty.

Methods

Based on 95 retrieved polyethylene inserts, wear and damage patterns were analyzed in relationship to the overall mechanical alignment and to the position of the tibial component.

Results

A progression of wear was observed with progressively mechanical varus alignment. However, there was significantly more damage in the lateral compartment in the mild and moderate varus group compared to the valgus group. No difference in damage was seen between all groups for tibial component positioning in valgus or varus.

Conclusion

Progressive wear was observed with progressively varus alignment with more damage at the lateral side. This observation is unique and might be explained by lateral condylar lift-off inducing impact and shear loading in the varus group.  相似文献   

16.

Background

A number of studies on total hip arthroplasty have compared highly cross-linked polyethylene (HXLPE) with conventional polyethylene (CPE) liners beyond 10 years. However, the impact of HXLPE on the wear-related reoperation rate is unclear. The purpose of this study was to evaluate the clinical advantage of using a single manufacturer's HXLPE in terms of reducing the reoperation rate.

Methods

The study was a follow-up retrospective cohort study over a mean of 12 years that examined patients aged 45-70 years with cementless total hip arthroplasty using a 26-mm-diameter cobalt-chromium head. Sixty-seven patients (79 hips; HXLPE group = 41 hips, CPE group = 38 hips) were evaluated for a minimum 10-year follow-up. Kaplan-Meier survival analysis was performed, with wear-related reoperations and radiographic osteolysis serving as the end points. The polyethylene wear rate was also assessed.

Results

The mean 12-year follow-up rates of survivorship that were evaluated using wear-related reoperations as the end point were 100% and 91.4% in the HXLPE and CPE groups, respectively (P = .007), and the mean 12-year follow-up rates of survivorship with osteolysis as the end point were 100% and 36.2%, respectively (P < .001). Compared with the CPE group, the HXLPE group presented a significantly reduced wear rate (HXLPE group, 0.035 mm/y; CPE group, 0.118 mm/y).

Conclusion

A unique strength of this study is that we assessed a single manufacturer's HXLPE while keeping most other implant parameters uniform. This study reveals the clinical advantage of using a single manufacturer's HXLPE in terms of a reduced wear-related reoperation rate at a mean 12-year follow-up.  相似文献   

17.

Background

Our aim was to examine how academic adult reconstructive surgeons have interpreted evidence on femoral head material in total hip arthroplasty (THA).

Methods

A 16-question survey to evaluate attitudes toward ceramic and cobalt-chrome head use was emailed to 274 faculty at 42 US adult reconstruction fellowship programs.

Results

With 116 respondents, the response rate was 42.2%. Faculty use ceramic heads 72.9% of the time. The most common reason why respondents do not use ceramic heads is cost (44.8%). Ninety-four percent of faculty have observed head-neck taper corrosion in cobalt-chrome on polyethylene THA, while 9.5% of faculty have observed head-neck taper corrosion in ceramic on polyethylene THA. Only 6.0% of surgeons have seen Biolox Delta ceramic fracture.

Conclusion

Adult reconstruction thought leaders are guided by evidence suggesting that with ceramic heads, taper corrosion and fracture are rare. Cost and personal experience also strongly influence their implant selection. Efforts to equalize cost of ceramic and cobalt-chrome heads may free surgeons to practice in a purely evidence-based fashion.  相似文献   

18.

Background

Patients with destructive acetabular lesions are at high risk for morbidity. Harrington described reinforcing acetabular diseased bone with methylmethacrylate, supplemented by metal fixation, to allow total hip reconstruction. However, all published studies using this technique report dislocations. We believe that initial use of constrained liners would help alleviate dislocation risk, without increasing component loosening or polyethylene wear.

Methods

Sixty-eight patients who underwent Harrington hip arthroplasties for metastatic cancer from August 2005 to March 2015 were identified. All patients had a constrained acetabular liner implanted during their index procedure. Electronic medical records and radiographs were reviewed. A literature review of studies with similar technique was used as a historical control.

Results

Forty-seven patients (69.2%) died 22.7 ± 22.7 months after surgery, at age 62.4 ± 17.2 years. At an average clinical follow-up of 9.9 ± 14.2 months (range 2-50 months), we found no incidents of dislocation. No patients had component failure requiring revision or evidence of loosening on x-ray. Thirteen papers, comprising 185 patients, demonstrated 39 patients (21.1%) with dislocations and 13 patients (7.0%) requiring revision surgery for aseptic loosening.

Conclusion

Modified Harrington technique has demonstrated relative durability when reconstructing metastatic acetabular disease, although dislocation occurs in approximately 20% of patients. This study demonstrates that initial use of a constrained liner can alleviate postoperative dislocations without increasing risk of perioperative complications. Even modest reduction in complications during the limited lifespan of oncologic patients can have dramatic positive effects on quality of life.  相似文献   

19.

Background

The advent of highly cross-linked polyethylene (HCLPE) has significantly improved total hip arthroplasty survivorship. HCLPE has been shown to improve wear properties in midterm outcomes when compared to traditional polyethylene liners; however, there is a paucity of studies evaluating long-term outcomes. In addition, there is concern that wear rates may accelerate as the implant ages. Thus, the aims of this study are to report on the longest-to-date follow-up of a specific first-generation HCLPE liner and to determine whether there is a change in the annual wear rate over time.

Methods

Forty hips in 38 patients which were previously reported on in a midterm study were included in this long-term follow-up study. Patients in this cohort all received total hip arthroplasty between March 1999 and August 2004 using the Crossfire HCLPE liner. Annual wear rates (mm/y) were calculated for this cohort. Patients were contacted and asked about complications or revision procedures they may have had since the index procedure.

Results

Clinical follow-up averaged 12.9 years with a range of 7-18 years. The average follow-up duration was 12.5 years with a range of 10-17 years. Linear wear was found to be 0.056 ± 0.036 mm/y. Osteolysis was not observed in any of the patients with greater than 10-year radiographic follow-up. Furthermore, only 1 patient required revision surgery following a mechanical fall.

Conclusion

Our study demonstrates the long-term wear rates associated with HCLPE liners continue to match rates published in midterm studies. Previously, we have reported that this cohort had an average annual wear rate of 0.05 mm/y over 10 years. This most recent report demonstrates a similar wear rate with up to 18-year follow-up.  相似文献   

20.

Background

The incidence of hip fractures is growing with the increasing elderly population. Typically, hip fractures are treated with open reduction internal fixation, hemiarthroplasty, or total hip arthroplasty (THA). Failed hip fracture fixation is often salvaged by conversion THA. The total number of conversion THA procedures is also supplemented by its use in treating different failed surgical hip treatments such as acetabular fracture fixation, Perthes disease, slipped capital femoral epiphysis, and developmental dysplasia of the hip. As the incidence of conversion THA rises, it is important to understand the perioperative characteristics of conversion THA. Some studies have demonstrated higher complication rates in conversion THAs than primary THAs, but research distinguishing the 2 groups is still limited.

Methods

Perioperative data for 119 conversion THAs and 251 primary THAs were collected at 2 centers. Multivariable linear regression was performed for continuous variables, multivariable logistic regression for dichotomous variables, and chi-square test for categorical variables.

Results

Outcomes for conversion THAs were significantly different (P < .05) compared to primary THA and had longer hospital length of stay (average 3.8 days for conversion THA, average 2.8 days for primary THA), longer operative time (168 minutes conversion THA, 129 minutes primary THA), greater likelihood of requiring metaphysis/diaphysis fixation, and greater likelihood of requiring revision type implant components.

Conclusion

Our findings suggest that conversion THAs require more resources than primary THAs, as well as advanced revision type components. Based on these findings, conversion THAs should be reclassified to reflect the greater burden borne by treatment centers.  相似文献   