首页 | 本学科首页   官方微博 | 高级检索  
相似文献
 共查询到20条相似文献,搜索用时 31 毫秒
1.
The volume of current implantable cardioverter defibrillators (ICD) is not convenient for pectoral implantation. One way to reduce the size of the pulse generator is to find a more effective defibrillation pulse waveform generated from smaller volume capacitors. In a prospective randomized crossover study we compared the step-down defibrillation threshold (DFT) of a standard biphasic waveform (STD), delivered by two 250-μF capacitors connected in series with an 80% tilt, to an experimental biphasic waveform delivered by a single 450μF capacitor with a 60% tilt. The experimental waveform delivered the same energy with a lower peak voltage and a longer duration (LVLDj. Intraopera-tively, in 25 patients receiving endocardial (n = 12) or endocardial-subcutaneous array (n = 13) defibrillation leads, the DFT was determined for both waveforms. Energy requirements did not differ at DFT for the STD and LVLD waveforms with the low impedance (32 ± 4Ω) endocardial-subcutaneous array defibrillation lead system (6.4 ± 4.4 J and 5.9 ± 4.2 J, respectively) or increased slightly (P - 0.06) with the higher impedance (42 ± 4 Ω) endocardial lead system (10.4 ± 4.6 J and 12.7 ± 5.7 /. respectively), However, the voltage needed at DFT was one-third lower with the LVLD waveform than with the STD waveform for both lead systems (256 ± 85 V vs 154 ± 53 V and 348 ± 76 V vs 232 ± 54 V, respectively). Thus, a single capacitor with a large capacitance can generate a defibrillation pulse with a substantial lower peak voltage requirement without significantly increasing the energy requirements. The volume reduction in using a single capacitor can decrease ICD device size.  相似文献   

2.
Introduction: Increased defibrillation thresholds (DFTs) with right active pectoral implantable cardioverter defibrillators (ICDs) and/or right proximal coils (SVC) are attributed to poorer vector. However, SVC affects impedance, current flow, and shock waveform phase duration (PD), which exert independent DFT effects.
Objective: Compare DFTs and shock characteristics in SVC On with SVC Off in right ICDs.
Methods and Results : DFT+ testing (n = 42, 62% males, 62 ±15 years, left ventricular ejection fraction (LVEF) 26 ± 11%, ischemic cardiomyopathy 65%, amiodarone 26%) revealed >20% incidence of high DFT (>20J) . Dilated cardiomyopathy and amiodarone increased DFT. Individual impedance variability (25–74 Ω) generated a wide PD range (2.6–8.7 ms). Overall, SVC On reduced impedance by 33% (from 54 ± 10 to 35 ± 5Ω, P< 0.0001), and shortened PD (from 5.45 ± 1.20 to 3.67 ± 0.74 ms, P< 0.01). SVC On affected DFTs in 19/42 (45%) patients. SVC On was beneficial in 12/19. PD shortened but current flow remained unaltered. (In these, SVC Off impedance was >45Ω and PD >5 ms.) SVC On was detrimental in 7/19 despite increasing current flow. In these, PD shortened excessively (median 2.9 ms) because impedance was low (31 ± 4Ω). In 3/6 cases with DFTs >20 J in both SVC On and Off , PD optimization reduced DFT. Overall, selection of best SVC configuration or deliberate PD programming yielded DFTs ≤20 J in >90% patients, reducing need for system modification to <7%.
Conclusions : Right pectoral active ICDs have high DFTs. The SVC coil may be detrimental when pulse waveform excessively shortens. Noninvasive maneuvers, for example, SVC and waveform optimization, may improve DFT.  相似文献   

3.
The unipolar system for transvenous defibrillation, consisting of a single right ventricular lead as the cathode and the device shell as anode, has been shown to combine low de- fibrillation thresholds (DFTs) and simple implantation techniques. We compared the defibrillation efficacy of this system with the defibrillation efficacy of a dual lead system with a 12-cm long defibrillation anode placed in the left subclavian vein. The data of 38 consecutive patients were retrospectively analyzed. The implantation of an active can system was attempted in 20 patients (group 1), and of the dual lead system in 18 patients (group 2). Both groups had comparable demographic data, cardiac disease, ventricular function, or clinical arrhythmia. The criterion for successful implantation was a DFT of > 24 J. This criterion was met in all 18 patients of group 2, The active can system could not be inserted in 3 of the 20 group 1 patients because of a DFT > 24 J. In these patients, the implantation of one (n = 2) or two (n = 1) additional transvenous leads was necessary to achieve a DFT ≤ 24). The DFTs of the 17 successfully implanted group 1 patients were not significantly different from the 18 patients in group 2 (12.3 ± 5.7 f vs 10.8 ± 4.8 J). The defibrillation impedance was similar in both groups (50.1 ± 6.1 ± 48.9 ± 5.2 Ω). In group 1, both operation duration (66.8 ± 17 min vs 80.8 ± 11 min; P < 0.05) and fluoroscopy time (3.3 ± 2.1 min vs 5.7 ± 2.9 min; P < 0,05) were significantly shorter. Thus, the active can system allows reliable transvenous defibrillation and a marked reduction of operation duration and fluoroscopy time. The dual lead system, with an increased surface area defibrillation anode, seems to he a promising alternative for active can failures.  相似文献   

4.
The impedance of defibrillation pathways is an important determinant of ventricular defibrillation efficacy. The hypothesis in this study was that the respiration phase (end-inspiration versus end-expiration) mayalter impedance and/or defibrillation efficacy in a "hot can" electrode system. Defibrillation threshold (DFT) parameters were evaluated at end-expiration and at end-inspiration phases in random order by a biphasic waveform in ten anesthetized pigs (body weight: 19.1 ±2.4 kg; heart weight: 97 ± 10g). Pigs were intubated with a cuffed endotracheal tube and ventilated through a Drager SAVrespirator with tidal volume of 400–500 mL. A transvenous defibrillation lead (6 cm long, 6.5 Fr) was inserted into the right ventricular apex. A titanium can electrode (92-cm2 surface area) was placed in the left pectoral area. The right ventricular lead was the anode for the first phase and the cathode for the second phase. The DFT was determined by a "down-up down-up" protocol. Statistical analysis was performed with a Wilcoxon matched pair test. The median impedance at DFT for expiration and inspiration phases were 37.8 ±3.1 Ω and 39.3 ± 3.6 Ω, respectively (P = 0.02). The stored energy at DFT for expiration and inspiration phases were 5.7 ± 1.9 J and 6.0 ± 1.0 J, respectively (P = 0.594). Shocks delivered at end-inspiration exhibited a statistically significant increase in electrode impedance in a "hot can" electrode system. The finding that DFT energy was not significantly different at both respiration phases indicates that respiration phase does not significantly affect defibrillation energy requirements.  相似文献   

5.
Pectorally implanted ICDs that defibrillate with the RV electrode and the ICD housing have gained clinical acceptance. However, it is still debatable whether adding an SVC electrode connected to the housing will further reduce the threshold of defibrillation (DFT). This study utilized eight pigs. DFTs were measured with a 50 V step-down protocol starting at 650 V (20 J). Shock strength for 50% success (E50) was estimated with the average of three reversals. In addition to a dummy device, Lead I (Pacesetter Models 1558 and 1538) or Lead II (Endotak 72) were used. Leads I are active fixation, true bipolar sensing with 5-cm shocking coils. Lead II has an integrated bipolar sensing with a 4.7-cm RV and 6.9-cm SVC shocking coils. A 95 μF defibrillation system was used to deliver a 44% tilt tuned biphasic 1.6/2.5 ms waveform, and to measure lead impedance. The RV electrode was the anode during phase I. With Lead I RV → CAN the DFT was 531 ± 75 V (13.6 ± 3.8 J) and the E50 was 496 ± 89 (12 ± 4.3 J). These were not significantly (NS) different than the DFT for RV → CAN and SVC which was 518 ± 84 V (13 ± 4.2 J) or the E50 which was 476 ± 84 V (11 ± 3.9 J). Similar results were obtained with Lead II. Despite a decrease in lead impedance there was no apparent benefit from the addition of the SVC electrode. Lead I provided equivalent DFT performance to Lead II.  相似文献   

6.
SANDSTEDT, B., et al. : Bidirectional Defibrillation Using Implantable Defibrillators: A Prospective Randomized Comparison Between Pectoral and Abdominal Active Generators. The objective of this study was to compare the effects of active abdominal and pectoral generator positions on DFTs in a bidirectional tripolar ICD system. Twenty-five consecutive patients had ICD systems implanted under general anesthesia. A transvenous single lead bipolar defibrillation system and an active 57-cc test emulator in the abdominal and pectoral positions were used in the same patient. A randomized, alternating stepdown protocol was used starting at 15 J with 3-J decrements until failure. The mean implantation time was  114 ± 23 minutes  , the mean arrhythmia duration was  14.5 ± 1.5 seconds  , and the mean recovery time was  5.4 ± 1.1 minutes  . The mean DFTs in the abdominal and pectoral positions were  10.9 ± 5.1  and  9.7 ± 5.2 J  , respectively (NS), the mean intraindividual DFT difference (abdominal minus pectoral) was  −0.89 ± 4.15 J  (  range −9.5 to + 5.8 J  ). The 95% confidence interval showed a  −2.60 to + 0.82 J  mean difference (NS). The DFT was < 15 J in 72% and 88% of the patients and the defibrillation impedance was  41 ± 3  and  44 ± 3 Ω  , abdominal versus pectoral positions. There was no difference in DFT between active abdominal and pectoral generator bidirectional tripolar defibrillation. The pectoral position may be considered the primary option, but in cases of high DFTs the abdominal site should be considered an alternative to adding a subcutaneous patch. In some patients, the anatomy may favor an abdominal position. Possible differences in the long-term functionality on the leads are not yet well known and need to be further evaluated.  相似文献   

7.
Introduction: A superior performance of a tuned waveform based on duration using an assumed cardiac membrane time constant of 3.5 ms and of a 50/50% tilt waveform over a standard 65/65% tilt waveform has been documented before. However, there has been no direct comparison of the tuned versus the 50/50% tilt waveforms.
Methods: In 34 patients, defibrillation thresholds (DFTs) for tuned versus 50/50% tilt waveforms in a random order were measured by using the optimized binary search method. High voltage lead impedance was measured and used to select the pulse widths for tuned and 50/50% tilt defibrillation waveforms.
Results: Delivered energy (7.3 ± 4.6 J vs 8.7 ± 5.3 J, P = 0.01), stored energy (8.2 ± 5.1 J vs 9.7 ± 5.6 J, P = 0.01), and delivered voltage (405.9 ± 121.7 V vs 445.0 ± 122.6 V, P = 0.008) were significantly lower for the tuned than for the 50/50% tilt waveform. In four patients with DFT ≥15 J, the tuned waveform lowered the mean energy DFT by 2.8 J and mean voltage DFT by 45 V. For all patients, the mean peak delivered energy DFT was reduced from 29 J to 22 J (24% decrease). Multiple regression analysis showed that a left ventricular ejection fraction <20% is a significant predictor of this advantage.
Conclusion: Energy and voltage DFTs are lowered with an implantable cardioverter defibrillator that uses a tuned waveform compared to a standard 50% tilt biphasic waveform.  相似文献   

8.
The effect of initial phase polarity on the DFT of two pectorally implanted biphasic ICDs was tested in a randomized, prospective manner at the time of implantation. Twenty-two consecutive patients with VT or VF who received either the Medtronic PCD 7219C fewel device (10 patients) or PCD 7219D fewel device (12 patients) were studied. DFT testing was performed in a standard step-down manner. Both initial phase polarities—initial defibrillation current flowing from active can/SVC coil (± subcutaneous patch) to the RV coil (RV-) or from RV coil to active can/SVC coil (RV+)—were tested in random order. The mean DFT achieved with RV+ compared with RV- was lower for the 7219C patient group (6.6 ±3.1 vs 10.8 ± 5.5 J; P = 0.007). A similar trend was observed forthe 7219D group, though the difference did not reach statistical significance (12.0 ± 4.0 vs 16.3 ± 7.3 J; P = 0.07). Seven of the 10 patients in the 7219C group had a lower DFT with RV+, while the initial phase polarity made no difference in 3. In the 7219D group, 7 patients had a lower DFT using RV+, 2 patients had a lower DFT using RV-, and the initial phase polarity made no difference in 3. In conclusion, this study demonstrates that changing the polarity of the initial phase of a biphasic shock wave form can have a significant impact on the DFT achieved at the time of ICD implantation.  相似文献   

9.
Recent technological advances have resulted in high success rates for implantation of nonthoracotomy defibrillation lead systems. Further decreases in defibrillator size, facilitating pectoral placement, will depend in part on lowering defibrillation energy requirements. The purpose of this study was to determine if endocardial defibrillation energy requirements are influenced by electrode size. Thirteen adult mongrel dogs were studied under general anesthesia. A9 Fr integrated bipolar pace/sense/defibrillation electrode (cathode) was positioned transvenously at the RV apex. The second defibrillation electrode (anode) was positioned at the junction of the RA and SVC. Two diameters of the proximal electrode, 7 Fr and 11 Fr, were sequentially tested in random order in each animal. The DFT for each electrode was determined using a 50-V up-down method. Energy, leading edge voltage, and current, current distribution, and total resistance were measured. The mean defibrillation voltage threshold with the 11 Fr proximal electrode was significantly less than with the 7 Fr proximal electrode (551.1 ± 76.5 V vs 588.5 ± 54.6 V, P < 0.01). Similarly, the mean DFT with the 11 Fr electrode was less than with the 7 Fr electrode (20.7 ±5.7J vs 23.3 ± 4.4 J.P < 0.01). Lower DFTs were found using the larger electrode in 11 of the 13 animals studied. However, there was no difference in defibrillation lead impedance between the two electrode systems. Endocardial defibrillation energy requirements may be lowered with a larger diameter proximal electrode. The mechanism by which this occurs may be due to a more even distribation of current gradients with the larger electrode. Determination of the optimal electrode size requires evaluation in humans, as this may allow further reduction in defibrillation energy requirements and defibrillator size.  相似文献   

10.
Nonthoracotomy lead systems for implantable cardioverter defibrillators (ICDs) have reduced operative mortality and morbidity as compared to epicardial lead systems but are usually associated with higher defibrillation thresholds (DFTs). The purpose of this prospective randomized trial was to investigate if the second defibrillation electrode in the left subclavian vein can increase defibrillation efficacy and decrease DFT as compared to the superior vena cava (SVC) position in nonthoracotomy lead systems for ICDs. Seventeen patients (mean age; 49.9 ± 11.3 years, mean ejection fraction; 46.1%± 15.8%) were implanted with an investigational unipolar electrode (Medtronic 13001) used as the defibrillation anode. DFT testing was started in the SVC (n = 10, group A) or the left subclavian vein (n = 7, group B), and repeated in the alternative position starting at the DFT of the initial position. Fifteen patients were eligible for analysis (group A: n = 9, group B: n = 6). With the electrode in the SVC, ventricular fibrillation could be successfully terminated in 9 out of 15 patients (60%). In the left subclavian vein the success rate was 100% (P < 0.01). Mean DFT in the SVC was 13.0 ± 5.2 J and in the left subclavian vein 10.2 ± 4.9 J. DFTs in the left subclavian vein were either lower (group A: n = 5/9, group B: n = 5/6) or equal to the results in the SVC position (P < 0.001). Thus, the left subclavian vein appears to be a superior alternative for positioning of the defibrillation anode as compared to the SVC for nonthoracotomy lead systems using two separate leads.  相似文献   

11.
Antiarrhythmic agents with a Class III action are known to increase defibrillation efficacy. We investigated whether a Class III drug simply shifts the dose-response curve for defibrillation or more extensively alters the curve. Forty-five dogs were divided into four groups according to the shock waveform and the presence or absence of treatment with a novel Class III drug, MS-551 (2 mg/kg bolus + 0.02 mg/kg per min). In addition to the conventional transcardiac DFT, dose-response curves were obtained by fitting the results of 40 fibrillation-defibrillation sequences at five shock strengths to a logistic model. MS-551 significantly decreased DFT regardless of the shock waveform (control vs MS-551 = 306 ± 79 V vs 229 ± 72 V [monophasic shock, P < 0.05], or 227 ± 42 V vs 176 ± 26 V [biphasic shock, P < 0.005]). The dose-response curves in dogs treated with MS-551 had a gentler slope than those without treatment, and the ratio of the voltages corresponding to 50% and 90% defibrillation success (V90/V50) was significantly greater in the MS-551 group (monophasic: 1.21 ± 0.06 vs 1.62 ± 0.42 [P < 0.005], biphasic: 1.20 ± 0.05 vs 1.37 ± 0.18 [P < 0.01]). The V90/DFT ratio was also significantly larger in the MS-551 group (monophasic: 1.22 ± 0.12 vs 1.66 ± 0.37 (P < 0.001); biphasic: 1.19 ± 0.11 vs 1.44 ± 0.79 [P < 0.005]). Thus, this Class III drug decreased the shock strength corresponding to relatively higher success rate (˜90%) less markedly than that for moderate success rate (˜50%). These results suggest that a Class III drug does not simply shift the dose response curve in proportion to the change in DFT, but more extensively alters its configuration.  相似文献   

12.
Nonthoracotomy defibrillator systems can be implanted with a lower morbidity and mortality, compared to epicardial systems. However, implantation may be unsuccessful in up to 15% of patients, using a monophasic waveform. It was the purpose of this study to prospectively examine the efficacy of a second chest patch electrode in a nonthoracotomy defibrillator system. Fourteen patients (mean age 62 ± 11 years, ejection fraction = 0.29 ± 0.12) with elevated defibrillation thresholds, defined as ≥ 24 J, were studied. The initial lead system consisted of a right ventricular electrode (cathode), a left innominate vein, and subscapular chest patch electrode (anodes). If the initial defibrillation threshold was ≥ 24 J, a second chest patch electrode was added. This was placed subcutaneously in the anterior chest (8 cases), or submuscularly in the subscapular space (6 cases). This resulted in a decrease in the system impedance at the defibrillation threshold, from 72.3 ± 13.3 Ω to 52.2 ± 8.6 Ω. Additionally, the defibrillation threshold decreased from ≥ 24 J, with a single patch, to 16.6 ± 2.8 J with two patches. These changes were associated with successful implantation of a nonthoracotomy defibrillator system in all cases. In conclusion, the addition of a second chest patch electrode (using a subscapular approach) will result in lower defibrillation thresholds in patients with high defibrillation thresholds, and will subsequently increase implantation rates for nonthoracotomy defibrillators.  相似文献   

13.
A new, thinner (10 Fr) and more flexible, single-pass transvenous endocardial ICD lead, Endotak DSP, was compared with a conventional lead, Endotak C, as a control in a prospective randomized multicenter study in combination with a nonactive can ICD. A total of 123 patients were enrolled, 55 of whom received a down-sized DSP lead. Lead-alone configuration was successfully implanted in 95% of the DSP patients vs 88% in the control group. The mean defibrillation threshold (DFT) was determined by means of a step-down protocol, and was identical in the two groups, 10.5 ± 4.8 J in the DSP group versus 10.5 ± 4.8 J in the control group. At implantation, the DSP mean pacing threshold was lower, 0.51 ± 0.18 V versus 0.62 ± 0.35 V (p < 0.05) in the control group, and the mean pacing impedance higher, 594 ± 110 Ω vs 523 ± 135 Ω (p < 0.05). During the follow-up period, the statistically significant difference in thresholds disappeared, while the difference in impedance remained. Tachyarrhythmia treatment by shock or antitachycardia pacing (ATP) was delivered in 53% and 41%, respectively, of the patients with a 100% success rate. In the DSP group, all 28 episodes of polymorphic ventricular tachycardia or ventricular fibrillation were converted by the first shock as compared to 57 of 69 episodes (83%) in the control group (p < 0.05). Monomorphic ventricular tachycardias were terminated by ATP alone in 96% versus 94%. Lead related problems were minor and observed in 5% and 7%, respectively. In summary, both leads were safe and efficacious in the detection and treatment of ventricular tachyarrhythmias. There were no differences between the DSP and control groups regarding short- or long-term lead related complications.  相似文献   

14.
The stability of the defibrillation threshold (DFT) early after implantation of an implantable cardioverter defibrillator was evaluated in 15 patients. All but one patient had a three lead nonthoracotomy system using a subcutaneous patch, a right ventricular endocardial lead, and a lead in coronary sinus (n = 5) or superior vena cava (n = 9). Shocks were delivered using simultaneous in nine, sequential in three, and single pathway (coronary sinus not used) in one patient. DFTs were measured at implant (n = 15), 2–8 days postoperation (postop, n = 15), and 4–6 weeks later (n = 8). The DFT was defined as the lowest energy shock that resulted in successful defibrillation. The DFT was assessed with output beginning at 18 joules or 2–4 joules above the implant DFT. All shocks were delivered in 2- to 4-joule increments or decrements. DFTs were significantly higher postoperatively than DFTs at implant (22.7 ± 7.0 J vs 16.9 ± 3.9 J; P < 0.05), Eight of 15 patients had DFT determined at all three study periods. In these patients, DFT increased at postop (22.8 ± 8.3 J vs 16.4 ± 3.9 J at implant: P < 0.05) and returned to baseline at 4–6 weeks (16 ± 7.1) vs 16.4 ± 3.9 J at implant; P = N.S.). Thus, in patients with a multilead nonthoracotomy system, a DFT rise was observed early after implant. The DFT appears to return to baseline in 4–6 weeks. These results have important implications for programming energy output after implantable cardioverter defibrillator implantation.  相似文献   

15.
Preliminary Clinical Results of a Biphasic Waveform and an RV Lead System   总被引:1,自引:0,他引:1  
Biphasic defibrillation waveforms have provided a reduction in defibrillation thresholds in transvenous ICD systems. Although a variety of biphasic waveforms have been tested, the optimal pulse durations and tilts have yet to be identified. A multicenter clinical study was conducted to evaluate the performance of a new ICD biphasic waveform and new RV active fixation steroid eluting lead system. Fifty-three patients were entered into the study. Mean age was 63 years with a mean ejection fraction of 36.8%. Primary indication for implantation was monomorphic ventricular tachycardia alone (54.7%). Forty-eight patients (90.6%) were implanted with an RV shocking lead and active can alone as the anodal contact. The ICD can was the cathode. In four cases (7.5%), an additional SVC or CS had was used due to a high DFT with the RV lead alone. In an additional case, a chronic SVC lead was used although the RV-Can DFT was acceptable. DFT for all cases at implant was 9.8 ± 3.7 J. Repeat testing at 3 months for a subset of patients showed a reduction in DFT (7.4 ± 3.0 J), P value = 0.03. Sensing and pacing characteristics of the RV lead system remained excellent during the study period (acute 0.047 ± 0.005 ms at 5.4 V and 9.9 ± 6.2 mV R wave; chronic 0.067 ± 0.11 ms at 5.4 V and 9.3 ± 5.4 mV R wave). It is concluded that this lead system provides good acute and chronic sensing and pacing characteristics with good DFT values in combination with this waveform.  相似文献   

16.
We hypothesized that a long thin right ventricular (RV) electrode would have equivalent defibrillation threshold (DFT) performance to a short thick electrode with approximately the same surface area. This could lead to thinner transvenous lead systems, which would be easier to implant. A thin (5.1 French) lead was compared to a standard control (10.7 French). The thin lead had an 8-cm RV electrode length with a surface area of 4.26 cm2. The standard lead had a RV electrode length of 3.7 cm and a surface area of 4.12 cm2. A 140-μFrench capacitor 65%/65% tilt biphasic defibrillation shock was delivered between the RV electrode and a 14-cm2 subcutaneous patch. DFTs were determined following 10 seconds of fibrillation in 11 dogs by a triple determination averaging technique. The thin lead had a lower resistance (77.1 ± 27.4 Ω vs 88.9 ± 30.3 Ω, P < 0.001) than did the thick lead. There was no significant difference in stored energy DFTs (9.9 ± 2.5 vs 10.3 ± 2.7, P = 0.098 2-sided, P = 0.049 1-sided). This was in spite of the fact that the long thin lead had a portion of its RV coil extending above the tricuspid valve and, thus, not contributing efficiently to the ventricular gradients in the small dog heart. We conclude that a long thin right ventricular electrode and a standard short thick electrode had equivalent defibrillation performance. This preliminary result should be confirmed in clinical studies as it could lead to significantly thinner transvenous lead systems.  相似文献   

17.
A potential benefit of two-lead transvenous defibrillation systems is the ability to independently position the defibrillation electrodes, changing the vector field and possibly decreasing the DFT, Using the new two-lead transvenous TVL lead system, we studied whether DFT is influenced by SVC lead position and whether there is an optimal position. TVL leads and Cadence pulse generators were implanted in 24 patients. No intraoperative or perioperative complications were observed. In each patient, the DFTs were determined for three SVC electrode positions, which were tested in random order: the brachiocephalic vein, the mid-RA, and the RA-SVC junction. The mean DFTs in the three positions were not statistically different, nor was any single lead position consistently associated with lower DFTs. However, an optimal electrode position was identified in 83% of patients, and the DFT from the best lead position for each patient was significantly lower than for any one of the electrode positions (P < 0.01). The mean safety margin for the best SVC lead position was approximately 27 J. These results demonstrate the advantage of a two-lead system, as well as the importance of testing multiple SVC lead positions when the patient's condition permits. Both of these factors can decrease the DFT and maximize the defibrillation safety margin. This will become increasingly important as pulse generator capacitors become smaller (as part of the effort to decrease generator size) and the energy output of the generators consequently decreases.  相似文献   

18.
Approximately 30% of implantable cardioverter defibrillator (ICD) patients still die of sudden death. A major cause of these sudden deaths is the failure to defibrillate because of failure to achieve a low defibrillation threshold (DFT). Anti‐arrhythmic drugs can have a profound positive or negative effect on the DFT. Unfortunately, present clinical practice continues to feature many procedures and tactics that have minimal to negative DFT benefit. In addition, many demonstrated helpful tactics are not understood or followed. This review covers the optimal RV (right ventricular) coil position and polarity, superior vena cava (SVC) coil positioning and usage, pulse width settings, and azygous vein coil implants. Specifically, the RV coil should be set to an anodal polarity and never ‘reversed’. The optimal RV coil position appears to be along the mid‐septum. The SVC coil should be kept out of the right atrium and placed in the innominate vein junction. The SVC coil should be always on for high impedance patients. For low impedance patients, the SVC coil should be set on or off depending on which setting gives the lowest DFT. Pulse widths should be set to correspond to optimally charging and discharging a cardiac membrane time constant of between 3.5 and 4.5 ms. For the highest DFT patients, a separate coil should be placed in the azygous vein and connected to the ICD ‘SVC’ port. Anachronistic approaches such as the use of polarity reversal, apical RV coil tip forcing, and subcutaneous arrays are also discussed.  相似文献   

19.
The purpose of the current study was to characterize the effects of transvenous ICD shocks on myocardial impedance. Rather than recording impedance during shocks, it was measured during continuous pacing in order to minimize confounding effects such as electrode polarization. Pacing impedance (reflecting the combined impedances of the electrode-tissue interface, myocardium, and blood pool) was measured every 5 seconds before and after 58 single shocks in 22 patients undergoing ICD implantation with a Transvene (n = 14) or Endotak (n = 8) lead. There was a progressive and long-lasting decrease in impedance after shocks. The magnitude of this change was similar for 0.6-J test shocks and shocks ≥ 5 J (28 ± 32 Ω vs 23 ± 16 Ω P = 0.8). However, the drop in impedance was more abrupt after high energy shocks. Because impedance continued to decline throughout the 5-minute interval between shocks, successive shocks had a cumulative effect, with a decrease of 46 ± 42 Ω after four discharges. In conclusion, a progressive decline in pacing impedance is a characteristic response to transvenous ICD discharges.  相似文献   

20.
Conventional transvenous defibrillation is performed with an ICD using a dual current pathway. The defibrillation energy is delivered from the RV electrode to the superior vena cava (SVC) electrode and the metallic case (CAN) of the ICD. Biventricular defibrillation uses an additional electrode placed in the LV free wall with sequential shocks to create an additional current vector. Clinical studies of biventricular defibrillation have reported a 45% reduction in mean defibrillation threshold (DFT) energy. The aim of the study was to use computational methods to examine the biventricular defibrillation fields together with their corresponding DFTs in a variety of patient derived models and to compare them to simulations of conventional defibrillation. A library of thoracic models derived from nine patients was used to solve for electric field distributions. The defibrillation waveform consisted of a LV --> SVC + CAN monophasic shock followed by a biphasic shock delivered via the RV --> SVC + CAN electrodes. When the initial voltage of the two shocks is the same, the simulations show that the biventricular configuration reduces the mean DFT by 46% (3.5 +/- 1.3 vs 5.5 +/- 2.7 J, P = 0.005). When the leading edge of the biphasic shock is equal to the trailing edge of the monophasic shock, there is no statistically significant difference in the mean DFT (4.9 +/- 1.9 vs 5.5 +/- 2.7 J, P > 0.05) with the DFT decreasing in some patients and increasing in others. These results suggest that patient-specific computational models may be able to identify those patients who would most benefit from a biventricular configuration.  相似文献   

设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号