首页 | 本学科首页   官方微博 | 高级检索  
相似文献
 共查询到20条相似文献,搜索用时 15 毫秒
1.
INTRODUCTION: Atrial defibrillation can be achieved with standard implantable cardioverter defibrillator leads, which has led to the development of combined atrial and ventricular devices. For ventricular defibrillation, use of an active pectoral electrode (active can) in the shocking pathway markedly reduces defibrillation thresholds (DFTs). However, the effect of an active pectoral can on atrial defibrillation is unknown. METHODS AND RESULTS: This study was a prospective, randomized, paired comparison of two shock configurations on atrial DFTs in 33 patients. The lead system evaluated was a dual-coil transvenous defibrillation lead with a left pectoral pulse generator emulator. Shocks were delivered either between the right ventricular coil and proximal atrial coil (lead) or between the right ventricular coil and an active can in common with the atrial coil (active can). Delivered energy at DFT was 4.2 +/- 4.1 J in the lead configuration and 5.0 +/- 3.7 J in the active can configuration (P = NS). Peak current was 32% higher with an active can (P < 0.01), whereas shock impedance was 18% lower (P < 0.001). Moreover, a low threshold (< or = 3 J) was observed in 61% of subjects in the lead configuration but in only 36% in the active can configuration (P < 0.05). There were no clinical predictors of the atrial DFT. CONCLUSION: These results indicate that low atrial DFTs can be achieved using a transvenous ventricular defibrillation lead. Because no benefit was observed with the use of an active pectoral electrode for atrial defibrillation, programmable shock vectors may be useful for dual-chamber implantable cardioverter defibrillators.  相似文献   

2.
INTRODUCTION: Atrial defibrillation can be achieved with standard implantable cardioverter defibrillator (ICD) leads, but the optimal shocking configuration is unknown. The objective of this prospective study was to compare atrial defibrillation thresholds (DFTs) with three shocking configurations that are available with standard ICD leads. METHODS AND RESULTS: This study was a prospective, randomized, paired comparison of shocking configurations on atrial DFTs in 58 patients. The lead system evaluated was a transvenous defibrillation lead with coils in the superior vena cava (SVC) and right ventricular apex (RV) and a left pectoral pulse generator emulator (Can). In the first 33 patients, atrial DFT was measured with the ventricular triad (RV --> SVC + Can) and unipolar (RV --> Can) shocking pathways. In the next 25 patients, atrial DFT was measured with the ventricular triad and the proximal triad (SVC --> RV + Can) configurations. Delivered energy at DFT was significantly lower with the ventricular triad compared to the unipolar configuration (4.7 +/- 3.7 J vs 10.1 +/- 9.5 J, P < 0.001). Peak voltage and shock impedance also were significantly reduced (P < 0.001). There was no significant difference in DFT energy when the ventricular triad and proximal triad shocking configurations were compared (3.6 +/- 3.0 J vs 3.4 +/- 2.9 J for ventricular and proximal triad, respectively, P = NS). Although shock impedance was reduced by 13% with the proximal triad (P < 0.001), this effect was offset by an increased current requirement (10%). CONCLUSION: The ventricular triad is equivalent or superior to other possible shocking pathways for atrial defibrillation afforded by a dual-coil, active pectoral lead system. Because the ventricular triad is also the most efficacious shocking pathway for ventricular defibrillation, this pathway should be preferred for combined atrial and ventricular defibrillators.  相似文献   

3.
INTRODUCTION: Many patients with implantable cardioverter defibrillators (ICDs) have older lead systems, which are usually not replaced at the time of pulse generator replacement unless a malfunction is noted. Therefore, optimization of defibrillation with these lead systems is clinically important. The objective of this prospective study was to determine if an active abdominal pulse generator (Can) affects chronic defibrillation thresholds (DFTs) with a dual-coil, transvenous ICD lead system. METHODS AND RESULTS: The study population consisted of 39 patients who presented for routine abdominal pulse generator replacement. Each patient underwent two assessments of DFT using a step-down protocol, with the order of testing randomized. The distal right ventricular (RV) coil was the anode for the first phase of the biphasic shocks. The proximal superior vena cava (SVC) coil was the cathode for the Lead Alone configuration (RV --> SVC). For the Active Can configuration, the SVC coil and Can were connected electrically as the cathode (RV --> SVC + Can). The Active Can configuration was associated with a significant decrease in shock impedance (39.5 +/- 5.8 Omega vs. 50.0 +/- 7.6 Omega, P < 0.01) and a significant increase in peak current (8.3 +/- 2.6 A vs. 7.2 +/- 2.4 A, P < 0.01). There was no significant difference in DFT energy (9.0 +/- 4.6 J vs. 9.8 +/- 5.2 J) or leading edge voltage (319 +/- 86 V vs. 315 +/- 83 V). An adequate safety margin for defibrillation (> or =10 J) was present in all patients with both shocking configurations. CONCLUSION: DFTs are similar with the Active Can and Lead Alone configurations when a dual-coil, transvenous lead is used with a left abdominal pulse generator. Since most commercially available ICDs are only available with an active can, our data support the use of an active can device with this lead system for patients who present for routine pulse generator replacement.  相似文献   

4.
The majority of cardioverter/defibrillator (ICD) implantations are currently performed with a non-thoracotomy approach. From November 1993 to January 1995, 46 patients underwent implantation of a PCD 7219C with an "active-can" lead configuration at our institution. While the chronic stability of the defibrillation threshold (DFT) for an epicardial lead system is well established, the results are still inconsistent for non-thoracotomy lead systems. Accordingly, the aim of the present study was to compare the acute and chronic defibrillation thresholds of the ICDs implanted with an "active-can" lead system in order to assess the chronic stability of these systems. The defibrillation energy requirements were measured at implant, prior to hospital discharge, three, six and twelve months after implantation of the defibrillator. The patient group consisted of 8 females and 38 males with a mean age of 57.2 years. The mean left ventricular ejection fraction was 43.8%. The most frequent underlying heart disease was coronary artery disease in 31 of 46 patients. Eight patients had idiopathic dilated cardiomyopathy. In 39 of 46 patients, the defibrillation threshold could be successfully determined at all 4 time points after implantation. The mean defibrillation energy requirement at the time of implantation was 9.2 +/- 5.9 Joules (J). The subsequent mean energy requirements were 7.6 +/- 4.8 J at pre-hospital discharge, 8.6 +/- 5.7 J at the 3 month, 8.1 +/- 6.0 J at the 6 month and 8.6 +/- 5.8 J at the 12 month follow-up visits. The mean defibrillation threshold was lowest at the time of prehospital discharge, significantly lower than at the time of initial implantation (p = 0.021). However, at all later time points up to one year, there was no significant difference in the DFT as compared with the time of initial implantation. Comparing the DFT at the time of implantation and the DFT at all other time points, there were no significant differences (9.23 vs. 8.56 J, p = 0.291). Although there was an initial decrease in the DFT at seven to ten days, the long-term stability of the DFT up to one year remained stable in the devices with the "active-can" lead system.  相似文献   

5.
OBJECTIVES: The purpose of this study was to identify clinical predictors of atrial defibrillation thresholds (DFTs) with standard implantable cardioverter-defibrillator (ICD) leads. BACKGROUND: Atrial defibrillation can be achieved with active pectoral, dual-coil transvenous ICD lead systems. If clinical predictors of atrial defibrillation efficacy with these lead systems were identified, they could be used to predict which patients may require more complex lead systems for atrial defibrillation, such as a coronary sinus electrode. METHODS: This was a prospective study of 135 consecutive patients undergoing initial ICD implant for standard indications. The lead system evaluated was a transvenous defibrillation lead with coils in the superior vena cava (SVC) and right ventricular apex (RV), and a left pectoral pulse generator emulator (CAN). The shocking pathway was RV-->SVC+CAN. Atrial DFT was measured using a step-up protocol. Clinical and echocardiographic parameters were evaluated as predictors of atrial DFT and multiple linear regression was performed. RESULTS: Mean atrial DFT was 4.6 +/- 3.8 J. Atrial DFT was < or =3 J in 70 patients (52%) and < or = 10 J in 97% of patients. The highest atrial DFT was 20 J (one patient). Left atrial size (r = 0.21, P = .01) and left ventricular end-diastolic diameter (r = 0.19, P = .02) were independent predictors of atrial DFT. However, these two predictors accounted for only 6% of the variability in atrial DFT. CONCLUSIONS: Clinical parameters are of limited use in predicting atrial DFT with a dual-coil, active pectoral ICD lead system. Because the RV--> SVC + CAN shocking pathway provides reliable atrial and ventricular defibrillation, this configuration should be preferred for combined atrial and ventricular ICDs.  相似文献   

6.
An automatic cardioverter-defibrillator could be implanted using an endocardial defibrillation lead system (consisting of a tripolar defibrillation electrode catheter in conjunction with an epicostal patch electrode) in 9 of 10 patients with sustained ventricular tachycardia or ventricular fibrillation. Six lead system complications were observed during a follow-up period of 51 +/- 36 weeks. Three catheter electrode conductor fractures occurred and manifested as oversensing and subsequent delivery of inappropriate shocks (1 patient), inability to defibrillate during electrophysiologic testing 3 months after implant (1 patient) and sudden cardiac death (1 patient). Asymptomatic patch electrode conductor fractures were detected on a routine chest roentgenogram in 2 patients. Endocardial defibrillation threshold testing performed at the time of implantation resulted in malfunction of a previously implanted permanent pacemaker pulse generator in 1 patient. Catheter and patch electrode replacement procedures were performed in 3 consenting patients under local anesthesia. Endocardial defibrillation thresholds after lead replacement were comparable to those obtained at time of initial implant. Serial clinical, roentgenographic and electrophysiologic evaluation should be included in follow-up procedures for endocardial defibrillation lead systems. Monitoring for deleterious effects of endocardial defibrillation threshold testing on previously implanted pacemaker systems should be performed at the time of implant and during follow-up. Improved lead designs are necessary for long-term use of endocardial defibrillation electrodes, but replacement procedures are feasible without thoracotomy.  相似文献   

7.
OBJECTIVES: The purpose of this study was to compare transvenous atrial defibrillation thresholds with lead configurations consisting of an active left pectoral electrode and either single or dual transvenous coils. BACKGROUND: Low atrial defibrillation thresholds are achieved using complex lead systems including coils in the coronary sinus. However, the efficacy of more simple ventricular defibrillation leads with active pectoral pulse generators to defibrillate atrial fibrillation (AF) is unknown. METHODS: This study was a prospective, randomized assessment of shock configuration on atrial defibrillation thresholds in 32 patients. The lead system was a dual coil Endotak DSP lead with a left pectoral pulse generator emulator. Shocks were delivered either between the right ventricular coil and an active can in common with the proximal atrial coil (triad) or between the atrial coil and active can (transatrial). RESULTS: Delivered energy at defibrillation threshold was 7.1 +/- 6.0 J in the transatrial configuration and 4.0 +/- 4.2 J in the triad configuration (p < 0.005). Moreover, a low threshold (< or = 3 J) was observed in 69% of subjects in the triad configuration but only 47% in the transatrial configuration. Peak voltage and shock impedance were also lowered significantly in the triad configuration. Left atrial size was the only clinical predictor of the defibrillation threshold (r = 0.57, p < 0.002). CONCLUSIONS: These results indicate that low atrial defibrillation thresholds can be achieved using a single-pass transvenous ventricular defibrillation lead with a conventional ventricular defibrillation pathway. These data support the development of the combined atrial and ventricular defibrillator system.  相似文献   

8.
BACKGROUND: Atrial defibrillation can be achieved with a conventional dual-coil, active pectoral implantable cardioverter-defibrillator (ICD) lead system. Shocking vectors that incorporate an additional electrode in the CS have been used, but it is unclear if they improve atrial DFTs. OBJECTIVE: The objective of this prospective, randomized study was to determine if a coronary sinus (CS) electrode reduces atrial defibrillation thresholds (DFTs). METHODS: This was a prospective study of 36 patients undergoing initial ICD implant for standard indications. A defibrillation lead with superior vena cava (SVC) and right ventricular (RV) shocking coils was implanted in the RV. An active can emulator (Can) was placed in a pre-pectoral pocket. A lead with a 4 cm long shocking coil was placed in the CS. Atrial DFTs were determined in the following 3 shocking configurations in each patient, with the order of testing randomized: RV --> SVC + Can (Ventricular Triad), distal CS --> SVC + Can (Distal Atrial Triad), and proximal CS --> SVC + Can (Proximal Atrial Triad). RESULTS: The Proximal and Distal Atrial Triad configurations were both associated with significant reductions in peak current (p < 0.01), but this effect was offset by significant increases in shock impedance (p < 0.01), resulting in no net change in the peak voltage or DFT energy in comparison to the Ventricular Triad configuration (Ventricular Triad: 4.9 +/- 6.6 J, Proximal Atrial Triad: 3.3 +/- 4.1J, Distal Atrial Triad: 4.4 +/- 6.7 J, p > 0.2). CONCLUSION: Shocking vectors that incorporate a CS coil do not significantly improve atrial defibrillation efficacy. Since the Ventricular Triad shocking pathway provides reliable atrial and ventricular defibrillation, this configuration should be preferred for combined atrial and ventricular ICDs.  相似文献   

9.
We evaluated the early clinical performance of an implantable cardioverter/defibrillator with a nonepicardial lead system in patients with refractory ventricular tachycardia or ventricular fibrillation. Ten patients, mean age 67 years, mean left ventricular ejection fraction 35%, refractory to 5 +/- 2 antiarrhythmic drugs and with a history of prior cardiac surgery (7 patients), severe lung disease (2 patients), or renal failure (1 patient) underwent device and lead system implant. A tripolar electrode catheter with one sensing electrode and two defibrillating electrodes was placed in the right ventricular apex and a left thoracic submuscular patch electrode was used in an epicostal location. Defibrillation energy threshold was determined using dual- or triple-electrode configurations. Optimal patch electrode location was determined after temporary use of a cutaneous patch electrode prior to cardioverter/defibrillator implant. Electrophysiologic studies were performed before discharge and after 2 to 3 months to assess device function. Percutaneous insertion and placement of the electrode catheter was achieved in all patients. Defibrillation energy threshold testing was done using 1 to 4 (mean, 2.7) electrode configurations per patient and required 6 to 21 (mean, 13) ventricular fibrillation inductions and 8 to 56 (mean, 22) shocks per patient. In all patients, lowest reliable defibrillation energy threshold was obtained with a triple-electrode configuration (right ventricular common cathode with right atrial and thoracic patch as dual anodes) and bidirectional shocks (mean, 18 +/- 5 J). Optimal patch electrode position could be determined in 9 of 10 patients, and these 9 patients had cardioverter/defibrillator implant. Ventricular fibrillation termination with the first delivered shock at electrophysiologic study was documented in all patients. There was no perioperative mortality in device-implanted patients. Postoperative electrophysiologic studies before discharge (9 patients) and at 3 months (8 patients) continued to demonstrate successful defibrillation by the first device shock. During follow-up (range, 2 to 10 months; mean, 6 +/- 3 months), spontaneous device discharges occurred in 4 patients with inappropriate shocks due to electrode catheter fracture being documented in 1 patient. Antiarrhythmic drug therapy was withdrawn in 6 patients and reduced in 3 patients. We conclude, based on our preliminary experience, that an implantable cardioverter/defibrillator can be successfully used with a nonepicardial lead system for endocardial defibrillation in many patients. This lead system can be used with currently available pulse generators and should be considered at cardioverter/defibrillator implantation. It can be anticipated to reduce patient risk and hospital costs associated with this procedure.  相似文献   

10.
INTRODUCTION: Lead systems that include an active pectoral pulse generator are now standard for initial defibrillator implantations. However, the optimal transvenous lead system and coil location for such active can configurations are unknown. The purpose of this study was to evaluate the benefit and optimal position of a superior vena cava (SVC) coil on defibrillation thresholds with an active left pectoral pulse generator and right ventricular coil. METHODS AND RESULTS: This prospective, randomized study was performed on 27 patients. Each subject was evaluated with three lead configurations, with the order of testing randomized. Biphasic shocks were delivered between the right ventricular coil and an active can alone (unipolar), or an active can in common with the proximal coil positioned either at the right atrial/SVC junction (low SVC) or in the left subclavian vein (high SVC). Stored energies at defibrillation threshold were higher for the single-coil, unipolar configuration (11.2 +/- 6.6 J) than for the high (8.9 +/- 4.2 J) or low (8.5 +/- 4.2 J) SVC configurations (P < 0.01). Moreover, 96% of subjects had low (< or = 15 J) thresholds with the SVC coil in either position compared with 81% for the single-coil configuration. Shock impedance (P < 0.001) was increased with the unipolar configuration, whereas peak current was reduced (P < 0.001). CONCLUSION: The addition of a proximal transvenous coil to an active can unipolar lead configuration reduces defibrillation energy requirements. The position of this coil has no significant effect on defibrillation thresholds.  相似文献   

11.
INTRODUCTION: Despite major technological advances, structural problems in implantable cardioverter defibrillator (ICD) endocardial sensing leads remain a significant problem. There are two types of ICD sensing leads: (1) dedicated bipolar leads and (2) integrated lead systems that include defibrillation coils. The long-term performance of these two lead systems has not been directly compared. METHODS AND RESULTS: We prospectively examined the incidence of lead failure manifested by inappropriate arrhythmia detection in 247 consecutive patients undergoing abdominal ICD implant at a single center between 1991 and 1995. A total of 107 patients received BT-10 (dedicated bipolar) leads and 140 patients received Endotak (integrated bipolar) leads. Over a mean follow-up of 860 +/- 442 days, there were 19 (17.8%) lead failures with the BT-10 lead (261 to 1,505 days postimplant) compared with only 6 (4.3%; P < 0.01) with the Endotak lead (410 to 1,211 days postimplant). Lead failure was due to an insulation defect in all cases, with the problem occurring in the proximal lead (within the pulse generator pocket) in all but one case. Lead survival was significantly better with the Endotak lead (P = 0.015, risk ratio = 3.0, 95% confidence intervals 1.2 to 7.6). CONCLUSION: Late lead failure due to insulation defects in BT-10 sensing leads (causing inappropriate ICD activation) is a relatively common and progressive phenomenon, with difficulties becoming apparent as long as 4 years after implant. This problem is a likely cause of inappropriate shocks in patients with BT-10 leads. Implantation of a new sensing lead should be considered at the time of elective pulse generator replacement, even in the absence of demonstrable oversensing.  相似文献   

12.
BACKGROUND. Several clinical studies have demonstrated a general superiority of biphasic waveform defibrillation compared with monophasic waveform defibrillation using epicardial lead systems. To test the breadth of utility of biphasic waveforms in humans, a prospective, randomized evaluation of defibrillation efficacy of monophasic and single capacitor biphasic waveform pulses was performed for two distinct nonthoracotomy lead systems as well as for an epicardial electrode system in 51 cardiac arrest survivors undergoing automatic defibrillator implantation. METHODS AND RESULTS. The configurations tested consisted of a right ventricular-left ventricular (RV-LV) epicardial patch-patch system, an RV catheter-chest patch (CP) nonthoracotomy system, and a coronary sinus (CS) catheter-RV catheter nonthoracotomy system. For each configuration, the defibrillation current and voltage waveforms were recorded via a digital oscilloscope to measure defibrillation threshold voltage, current, resistance, and stored energy. Biphasic waveform defibrillation proved more efficient than monophasic waveform defibrillation for the epicardial RV-LV system (4.8 +/- 4.1 versus 6.7 +/- 4.9 J, p = 0.047) and the nonthoracotomy RV-CP system (23.4 +/- 11.1 versus 34.3 +/- 10.4 J, p = 0.0042). Biphasic waveform defibrillation thresholds were not significantly lower than monophasic waveform defibrillation thresholds for the CS-RV nonthoracotomy system (15.6 +/- 7.2 versus 20.0 +/- 11.5 J, p = 0.11). Biphasic waveform defibrillation proved more efficacious than monophasic waveform defibrillation in 13 of 20 patients (65%) with RV-LV epicardial patches, 10 of 15 patients (67%) with an RV-CP nonthoracotomy system, and nine of 16 patients (56%) with an RV-CS nonthoracotomy system. CONCLUSIONS. Biphasic pulsing was useful with nonthoracotomy lead systems as well as with epicardial lead systems. However, the degree of biphasic waveform defibrillation superiority appeared to be electrode system dependent. Furthermore, for a few individuals, biphasic waveform defibrillation proved less efficient than monophasic waveform defibrillation, regardless of the lead system used.  相似文献   

13.
Neuzner J 《Herz》2005,30(7):601-606
The automatic detection and termination of ventricular fibrillation is still the key function of implantable cardioverter defibrillator (ICD) therapy. The progress in generator and lead technology has overcome limitations in defibrillation efficacy in early transvenous defibrillator devices. Current, active pectoral biphasic devices provide a high defibrillation efficacy. More than 90% of all patients will meet accepted implantation criteria without any intraoperative system modifications. Is this enough to abandon the intraoperative assessment of defibrillation efficacy? Arguments for abandoning intraoperative device testing include: reduction of perioperative complications, time and cost saving, no worse prognosis for defibrillator patients with borderline defibrillation efficacy, DFT testing might be a barrier to an easy access to ICD implantation. Abandoning intraoperative assessment of defibrillation efficacy may result in inadequate defibrillation safety in up to 9% of all patients. The noninferior outcome of patients with nonadequate defibrillation efficacy is not already proven. Intraoperative device testing could be limited to a small number of VF inductions, the safety of these protocols is well established. A significant time and cost reduction is not really existing. The abandoning of defibrillation testing will not lead to an increase in ICD implant capacity. The intraoperative assessment of defibrillation efficacy should be an important part of ICD implantation.  相似文献   

14.
Serial DFT Measures in Man. Introduction; The defibrillation threshold (DFT) may change throughout the first year following implantation of a cardioverter defibrillator, but it remains uncertain if changes are a consequence of changes in clinical condition or are related to fundamental alterations at the electrode-tissue interface. The purpose of this study was to evaluate the extent and time course of DFT changes over the first year following implantable cardioverter defibrillator (ICD) surgery when extraneous clinical and device variables potentially affecting the DFT were excluded. Methods and Results.: We prospectively enrolled 61 patients undergoing epicardial or non-thoracotomy/transvenous ICD therapy into a series of follow-up studies where the DFT was measured at implant and at 1, 6,12, and 52 weeks following implantation in a uniform manner. Stored energy DFT was measured and recorded for all patients. Patient exclusion criteria were: (1) inability to complete all five measures of the DFT; (2) institution of Class I or Class III antiarrhythmic drugs at any time during the study; (3) lead system changes (relocation or new leads) or programming changes in pulse width or current pathway; or (4) development of a significant change in their clinical status, such as decompensated congestive heart failure or acute ischemia. Only 20 of the 61 patients satisfied the criteria required to complete the study. Two of the excluded patients developed high DFTs, which required reprogramming of the current pathway. Eight patients had an epicardial lead system, and 12 had a nonthoracotomy lead system. The rise in DFT over the first 12 weeks was significant for the eight epicardial lead system patients (P = 0.05) and for the 12 nonthoracotomy lead system patients (P = 0.004). The peak rise in DFT occurred at 1 week for the patients with an epicardial lead system (3.4 ± 1.8 J to 7.9 ± 3.8 J) and at 12 weeks for the patients with a transvenous lead system (10.3 ± 5.3 J to 16.1 ± 7.4 J). Conclusions: This study confirms a transient significant rise in the DFT in the first 12 weeks following ICD surgery that partially returns to the implant value over the remainder of the year. Because specific clinical and technical variables were excluded from this study, the observations made in this patient population suggest that the rise in DFT may be a consequence of changes at the electrode-tissue interface.  相似文献   

15.
AIM: To review techniques of implantable cardioverter-defibrillators (ICD) in patients after Mustard surgery for arterial transposition. METHODS AND RESULTS: Retrospective analysis of all Mustard patients receiving ICDs at our institution. Five patients (median age 24 years, range 19-35, 3 male) with systemic right ventricular dysfunction (sRV) dysfunction and New York Heart Association (NYHA) II and III, received ICDs. Implantation was performed transvenously in three patients, epicardial patches and subcutaneous arrays at surgery in two patients. Two patients required lead extraction and baffle stent angioplasty before ICD implantation. Defibrillation vectors incorporating the anterior sRV mass [i.e., sub-pulmonary left ventricle (pLV) to generator can, and between epicardial defibrillator patches], consistently achieved a minimum 10 joule(J) safety margin during defibrillation threshold (DFT) testing. Subcutaneous arrays and endocardial vectors that included a superior vena cava (SVC) electrode were less effective. One patient developed pulmonary oedema post-procedure. At a median 20 months, all patients were alive and in NYHA class II. Follow-up over 24 months documented multiple non-sustained ventricular tachycardia (VT) in the group and one patient had recurrent VT with aborted device therapy. CONCLUSION: Defibrillator implantation in Mustard patients is challenging. Sub-optimal defibrillation should be anticipated and can be overcome using vectors which integrate the RV mass and high-energy devices. A staged procedure involving pre-implant interventions or separate DFT tests, where indicated, may be better tolerated by patients.  相似文献   

16.
AIMS: In patients with implantable cardioverter-defibrillators (ICD), the goals of lowering the defibrillation threshold (DFT) can be achieved by means of higher defibrillation safety margins, more rapid charging of capacitors, improved battery longevity, implying smaller devices. Whether an increase in the electrically active surface of ICD leads by fractal coating results in decreased DFTs is unknown. METHODS AND RESULTS: In this prospective randomized cross-over study the defibrillation efficacy of a novel right ventricular endocardial defibrillation electrode fractally coated with iridium was compared with an uncoated but otherwise identical electrode in 30 patients undergoing ICD implantation. In each patient, DFT testing was performed twice according to a binary search protocol introducing the two different electrodes in a random order. The mean DFT was 8.4 +/- 4.1 J with the fractally coated lead and 9.6 +/- 3.6 J using the uncoated lead. The improvement of 1.2 J was statistically not significant (P = 0.11). No differences were observed between the patients with an improved DFT (n =12) and those with an unchanged or worsened DFT (n = 18) concerning age, underlying cardiac disease, NYHA class, or left ventricular ejection fraction, respectively. CONCLUSION: Increasing the electrical surface of defibrillation leads by fractal coating does not lead to a substantial clinically relevant reduction in defibrillation thresholds. Defibrillation impedance is not influenced by the increased electrical surface of the defibrillation lead.  相似文献   

17.
INTRODUCTION: Placement of implantable cardioverter defibrillators (ICDs) has been simplified by using the shell of a pectorally implanted ICD as a defibrillation electrode in combination with an endocardial right ventricular defibrillation lead. However, a sufficiently low defibrillation threshold (DFT) cannot be obtained in a few patients. Therefore, alternative approaches were systematically tested in the Active Can versus Array Trial (ACAT). METHODS AND RESULTS: In the first of two prospective randomized studies, the DFT of a subcutaneous left dorsolateral array anode introduced from a pectoral incision was compared to that of a standard active can anode in 68 patients. Intraoperatively, the DFT was determined twice in each patient using either the active can or, in patients with a subcutaneous array lead, once with all three fingers and once omitting the middle finger of the array. The second prospective randomized study included 40 patients. DFT also was determined twice in each patient using an active can in a left pectoral position as anode alone and combined with a left dorsolateral array electrode with two fingers. In ACAT I, stored energy at DFT decreased from 13.1+/-7.7 J (active can) to 9.6+/-6.1 J (three-finger array lead) (P = 0.04), impedance decreased from 53+/-8 ohms to 40+/-6 ohms (P < 0.0001). Omitting the middle finger of the array lead, stored energy at DFT increased by 0.9 J (P = 0.47) and impedance by 2 ohm (P < 0.0001). In ACAT II, DFT and impedance using an active can device were significantly lower when a two-finger array lead was added that decreased stored energy at DFT from 10.1+/-5.2 J to 6.9+/-3.9 J (P = 0.001) and impedance from 56+/-5 1 to 42+/-5 l (P < 0.0001). CONCLUSION: In combination with a right ventricular defibrillation electrode, a left pectoral subcutaneous array lead improves defibrillation efficacy if used instead of, or in addition to, a left pectoral active can ICD device. Implantation of the array lead can be simplified by using two instead of three fingers, without a significant loss of defibrillation efficacy.  相似文献   

18.
AIMS: To compare the atrial defibrillation threshold (DFT) for two electrode configurations in patients with drug refractory persistent atrial fibrillation (AF). METHODS AND RESULTS: 11 patients, 73% male, mean age 60.9 (range 38 to 83), underwent implantation of a Medtronic Jewel AF dual chamber defibrillator (model 7250). A step-up atrial DFT was performed in a randomized sequence for two electrode configurations: (1) Right atrial to distal coronary sinus electrode (RA > CS) and (2) defibrillator can to right ventricular and right atrial electrodes (CAN > RV + RA). The RA > CS configuration restored SR in 10 patients (91%). The CAN > RA + RV configuration restored SR in four patients (36%). The mean atrial DFT was significantly lower for the RA > CS than CAN > RA + RV configuration (10 +/- 7 Joules vs 25 +/- 6 Joules), P < 0.01. At 3 months post implantation, AF was reinduced and the protocol was repeated for the optimal electrode configuration. There was no significant difference in the atrial DFT compared with that at implant. CONCLUSION: The right atrium to coronary sinus electrode configuration significantly reduces the atrial DFT. The atrial DFT also remains stable at 3 months post-implantation. Patients with persistent AF undergoing insertion of an atrial defibrillator should have a coronary sinus electrode implanted.  相似文献   

19.
Introduction: Sudden death is prevalent in heart failure patients. We tested an implantable ventricular support device consisting of a wireform harness with one or two pairs of integrated defibrillation electrode coils. Methods and Results: The device was implanted into six pigs (36–44 kg) through a subxiphoid incision. Peak voltage (V) defibrillation thresholds (DFT) were determined for five test configurations compared with a control transvenous lead (RV to CanPect). Defibrillator can location (abdominal or pectoral) and common coil separation on the implant (0° or 60°) were studied.. The DFT for RV60 to LV60 + CanPect was significantly less than control (348 ± 57 vs 473 ± 27 V, P < 0.05). The DFTs for other vectors were similar to control except for RV0 to LV0 + CanAbd (608 ± 159 V). The device was implanted into 12 adult dogs for 42, 90, or 180 days with DFT and pathological examination performed at the terminal study. Cardiac pressures were determined at baseline, after implantation, and at the terminal study. The DFT was also determined in a separate group of four dogs at 42 days following implantation of the support device with one pair of defibrillation electrodes. The DFTs at implant and explant in dogs with one pair (8 ± 1.5 Joules [J] and 6 ± 1.9 J) or two pairs (8 ± 3.4 J and 7 ± 1.9 J) of defibrillation electrodes were not significantly different from each other but significantly less than control measured at the terminal study (18 ± 3.4 J). Left‐sided pressures were significantly decreased at explant but within expected normal ranges. Right‐sided pressures were not different except for RV systolic. Histopathology indicated mild to moderate epicardial inflammation and fibrosis, consistent with a foreign body healing response. Conclusions: This defibrillation‐enabled ventricular support system maintained mechanical functionality for up to 6 months while inducing typical chronic healing responses. The DFT was equal to or lower than a standard transvenous vector.  相似文献   

20.
Effects of chronic amiodarone therapy on defibrillation threshold.   总被引:5,自引:0,他引:5  
In a prospective and parallel, randomized study, the long-term stability of epicardial defibrillation threshold was evaluated in 22 patients, using a patch-patch lead configuration at the time of implantation and generator replacement. The concomitant antiarrhythmic drug treatment consisted of either mexiletine (720 mg/day) or amiodarone (400 mg/day) and was administered to patients in a randomized and parallel manner. During a mean follow-up of 24 +/- 6 months, the defibrillation threshold increased significantly from 14.3 +/- 2.8 to 17.9 +/- 5.3 J (p < 0.05) for the entire patient group. The increase in the chronic defibrillation threshold was due to a marked increase in defibrillation energy needs in the subgroup of patients receiving amiodarone. Whereas no significant change in the defibrillation threshold was documented in the subgroup of patients receiving mexiletine, the mean defibrillation threshold increased from 14.1 +/- 3.0 to 20.9 +/- 5.4 J (p < 0.001) in those receiving amiodarone. In all patients with increased defibrillation thresholds, reevaluation showed a reduction in the defibrillation threshold after discontinuation of antiarrhythmic drug therapy. The only variable associated with an increase in the chronic defibrillation threshold was amiodarone treatment. These findings suggest that the defibrillation threshold should be measured at each generator replacement and in case of a change in antiarrhythmic drug treatment. In particular, if amiodarone treatment is initiated, it is recommended that the defibrillation threshold should be reevaluated to ensure an adequate margin of safety.  相似文献   

设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号