首页 | 本学科首页   官方微博 | 高级检索  
相似文献
 共查询到20条相似文献,搜索用时 31 毫秒
1.
BACKGROUND: Most manufacturers of blood glucose monitoring equipment do not give advice regarding the use of their meters and strips onboard aircraft, and some airlines have blood glucose testing equipment in the aircraft cabin medical bag. Previous studies using older blood glucose meters (BGMs) have shown conflicting results on the performance of both glucose oxidase (GOX)- and glucose dehydrogenase (GDH)-based meters at high altitude. The aim of our study was to evaluate the performance of four new-generation BGMs at sea level and at a simulated altitude equivalent to that used in the cabin of commercial aircrafts. Methodology/Principal Findings: Blood glucose measurements obtained by two GDH and two GOX BGMs at sea level and simulated altitude of 8000 feet in a hypobaric chamber were compared with measurements obtained using a YSI 2300 blood glucose analyzer as a reference method. Spiked venous blood samples of three different glucose levels were used. The accuracy of each meter was determined by calculating percentage error of each meter compared with the YSI reference and was also assessed against standard International Organization for Standardization (ISO) criteria. Clinical accuracy was evaluated using the consensus error grid method. The percentage (standard deviation) error for GDH meters at sea level and altitude was 13.36% (8.83%; for meter 1) and 12.97% (8.03%; for meter 2) with p = .784, and for GOX meters was 5.88% (7.35%; for meter 3) and 7.38% (6.20%; for meter 4) with p = .187. There was variation in the number of time individual meters met the standard ISO criteria ranging from 72-100%. Results from all four meters at both sea level and simulated altitude fell within zones A and B of the consensus error grid, using YSI as the reference. CONCLUSIONS: Overall, at simulated altitude, no differences were observed between the performance of GDH and GOX meters. Overestimation of blood glucose concentration was seen among individual meters evaluated, but none of the results obtained would have resulted in dangerous failure to detect and treat blood glucose errors or in giving treatment that was actually contradictory to that required.  相似文献   

2.

Background

Concerns have been raised about the use of point-of-care (POC) glucose meters in the hospital setting.Accuracy has been questioned especially in critically ill patients. Although commonly used in intensive care units and operating rooms, POC meters were not approved by the Food and Drug Administration for such use.Data on POC glucose meter performance during anesthesia are lacking. We evaluated accuracy of a POC meter in the intraoperative setting.

Methods

We retrospectively reviewed 4,333 intraoperative records in which at least one intraoperative glucose was measured using electronic medical records at a large academic hospital. We evaluated the accuracy of a POC glucose meter (ACCU-CHEK® Inform, Roche Pharmaceuticals) based on the 176 simultaneous central laboratory (CL) blood glucose (BG) measurements that were found (i.e., documented collection times within 5 minutes). Point-of-care and central lab BG differences were analyzed by Bland-Altman and revised error grid analysis (rEGA).

Results

Mean POC BG was 163.4 ± 64.7 mg/dl [minimum (min) 48 mg/dl, maximum (max) 537 mg/dl] and mean CL BG was 162.6 ± 65.1 mg/dl (min 44 mg/dl, max 502 mg/dl). Mean absolute difference between POC and CL BG was24.3 mg/dl. Mean absolute relative difference was 16.5% with standard deviation 26.4%. Point-of-care measurements showed a bias of 0.8 relative to the corresponding CL value, with a precision of 39.0 mg/dl. Forty (23%) POC BG values fell outside the Clinical and Laboratory Standards Institute guideline and 3.4% POC measurements fell in zones C and D of the rEGA plot.

Conclusions

The tested POC glucose meter performed poorly compared to a CL analyzer intraoperatively. Perioperative clinicians should be aware of limitations of specific POC glucose meters, and routine use of POC glucose meters as sole measurement devices in the intraoperative period should be carefully considered.  相似文献   

3.
A 6–week clinical evaluation of the One Touch II blood glucose meter was performed in a medical ward specializing in diabetic patients and compared to the existing Glucometer II system. Nursing staff trained in both methods performed 267 capillary blood glucose measurements using the two meters. A reference capillary blood glucose was performed simultaneously on 129 of these occasions using a Yellow Springs Instrument analyser. The One Touch II correlated better with the reference (r = 0.97) than the Glucometer II (r = 0.86) and was equally close or closer to the Yellow Springs Instrument on 82 % of occasions. A greater proportion of samples measured by the One Touch II were within 15% of the reference value (83 % vs 66 %). There were also 17 cases where the One Touch II was able to give a measurement which was outwith the analytical range of the Glucometer II. A questionnaire revealed the nursing staff unanimously preferred using the One Touch II to their present meter. It is concluded that in a ward setting the One Touch II meter showed improved accuracy, usefulness in its wider analytical range, and a greater user preference in comparison to the Glucometer II.  相似文献   

4.
《Primary Care Diabetes》2021,15(5):848-852
BackgroundPersonal glucose meters, primarily designed for self-control of glucose concentration in patients with diabetes, are frequently used in clinical practice as point-of-care equipment. The procedure of washing hands with water and soap before testing, as recommended by the manufacturer of personal glucose meters, in hospital wards is often difficult to fulfil and is replaced by disinfecting the place of blood sampling with isopropyl alcohol.AimThe purpose of the research was to evaluate the effect of different disinfecting procedures on glycemia measurement in capillary blood using personal glucose meters.Materials and methodsFour measurements of glycemia were taken in each of 50-volunteer group using Accu-Check Performa (Roche, Swiss) glucose meter using different procedures: washing hands with soap and drying them (1), disinfection with isopropyl alcohol and waiting for the disinfectant to evaporate (2) and the immediate puncture after disinfection. In the latter case two kind of single-use lancets were use – typical for adult (3) and smaller preferred by diabetics or children (4).ResultsPaired T-student test showed no statistically significant difference between measurements 1st and 2nd 95% CI (−4.282 to 0.322), p = 0.09. However, the comparison between 1st/3rd and 1st/4th showed statistically significant difference, 95% CI (−4.964 to −0.796), p = 0,0077 (p < 0,01) and 95% CI (−7.842 to −2.917), p = 0.00006 (p < 0.001).ConclusionsThe procedure of disinfection when fluid completely evaporates has no effect on glucose measurements. Sampling the wet finger influences results of glucose measurement but observed changes were clinically irrelevant.  相似文献   

5.

Background

Self-monitoring of blood glucose (SMBG) remains an important component of diabetes management, engendering a need for affordable blood glucose (BG) meters that are accurate, precise, and convenient. The CONTOUR® TS is a BG meter that endeavors to meet this need. It uses glucose dehydrogenase/flavin dinucleotide chemistry, automatic test strip calibration, and autocompensation for hematocrit along with the ease of use that has come to be expected of a modern meter. The objective of this clinical trial was to determine whether the CONTOUR TS system met these criteria.

Methods

The system was evaluated at a single clinical site with 106 subjects with type 1 or type 2 diabetes. Blood glucose values ranged from 60 to 333 mg/dl over all subjects. Both lay users and health care professionals (HCPs) tested the meters, with test strips from three different lots. Results were compared to a reference analyzer of verified precision and accuracy. Forty-nine of the subjects also participated in a home study of the meter. Lay users learned to use the system without assistance and were surveyed on its use at the end of the study.

Results

When used with capillary blood, both subjects and HCPs obtained results that exceeded the International Organization for Standardization 15197:2003 criteria, (i.e., ≥95% of values fell within 20% or 15 mg/dl of the laboratory value for BG levels greater than or less than 75 mg/dl, respectively). Specifically, lay users achieved 97.9% and HCPs 98.6%. When used with venous blood, 99.8% of measurements were within the criteria. All measurements for both capillary and venous blood fell into zones A or B of the Parkes error grid, deemed clinically accurate. Hematocrit was found to have no influence on BG measurements. A large majority of the subjects found the system easy to learn and to use.

Conclusions

The CONTOUR TS BG meter system gave accurate and reproducible results with both capillary and venous blood; subjects learned to use the meter system by following the user guide and quick reference guide.  相似文献   

6.

Background

Monitoring blood glucose levels is an integral part of routine diabetes management. To minimize the risk of transmission of bloodborne pathogens during monitoring, the Centers for Disease Control and Prevention (CDC) recommends that glucose meters be disinfected after each use whenever they are used to test multiple patients. The objective of this study is to assess the compatibility of some common disinfectants with certain blood glucose meter systems.

Methods

We tested six disinfectants for adverse impact on meter performance or the exterior meter surfaces.The disinfectants tested were 0.525% sodium hypochlorite, 20% 2-propanol and 10% ethanol, 17.2% isopropanol, 55% isopropanol, 70% isopropanol, and hydrogen peroxide. To assess meter performance, we tested OneTouch® Ultra® blood glucose monitoring systems with control solution before and after application of either water or disinfectant. To assess the effect on exterior meter surfaces, we performed a soaking test to simulate long-term exposure to disinfectant.

Results

Paired t-test results showed that the control solution data associated with disinfectant and with water application were not significantly different for each meter type. However, most of the meter types were adversely affected by hydrogen peroxide and/or by the higher concentrations of alcohol-based disinfectants.

Conclusions

Although none of the six disinfectants affected meter performance, hydrogen peroxide and isopropanol >20% adversely affected the exterior surfaces of the tested meters. When complying with CDC instructions for meter disinfection, users should use caution and choose disinfectants that have been validated by the meter manufacturer.  相似文献   

7.
A comparison of blood glucose meters in Australia   总被引:1,自引:0,他引:1  
OBJECTIVE: To assess the accuracy and precision of the five currently available blood glucose meters in Australia. DESIGN AND SETTING: Control solutions from manufacturers were used to determine the precision for each meter. Glucose levels in capillary blood samples from 49 patients attending a diabetes clinic were measured with each meter and with a laboratory reference method. OUTCOME MEASURES: The coefficient of variation was calculated to determine precision. Bias, Error Grid analysis, and Bland-Altman plots were used to determine accuracy. RESULTS: The CVs of most meters were acceptable at <5%. Bias ranged from 4.0 to 15.5% with only 1 meter satisfying the American Diabetes Association recommendation of <5% bias. Error Grid analysis showed that 94-100% of readings were clinically accurate, and that none of the differences from the reference method would lead to clinical errors. Bland-Altman plots showed that for two meters the magnitude of the difference between the meter and the reference method increased with increasing glucose values, but did not change significantly with glucose level for the other 3 meters. CONCLUSIONS: Currently available blood glucose meters show acceptable precision, and any errors (with respect to a laboratory method) are highly unlikely to lead to clinical errors. However, only the CareSens meter achieved a bias of less than 5%.  相似文献   

8.
AIMS: To examine patients' perspectives on ease of use and pain with the MediSense alternate site blood glucose testing device (Soft-Sense) compared with their current glucose testing method, and to evaluate the analytical performance of the MediSense device with the laboratory reference method. METHODS: Study participants were shown how to use the Soft-Sense glucose device and asked to perform two tests on their forearm. A capillary sample was collected from their finger and tested on the external port of the Soft-Sense meter and a laboratory method (YSI Glucose Analyser). Finally, one drop of blood was also directly tested from the finger onto the external port. Patients completed a questionnaire comparing ease of use and associated pain of their current testing method with the Soft-Sense meter. RESULTS: Patients preferred the Soft-Sense device to their own for ease of use and for less pain (93% found it easier to use and 96% less painful; P < 0.001). Glucose results correlated closely with the laboratory method (mean absolute percentage bias for the forearm 11.0%, finger 6.0%, and collected capillary sample 5.7%). Error grid analysis showed that all Soft-Sense results were clinically acceptable. CONCLUSIONS: Patients prefer the Soft-Sense alternate site testing device to their existing measuring method. The device accurately measures whole blood glucose.  相似文献   

9.
The aim of the study was to evaluate the precision and accuracy of the ExacTech home blood glucose meter when used with either capillary or venous blood and to compare this with a reference whole blood glucose assay. Non-fasting glucose measurements were used since a validation study showed no capillary-venous differences between fasting and post-prandial states. In a cross-sectional study, blood was taken from 182 patients and measured in duplicate on three batches of strips. Altogether we analysed 1089 readings. The regression of the data from capillary blood samples (meter vs reference method) had a correlation coefficient, of 0.93, and a mean bias of 0.2 mmol l-1. The corrected 90% confidence interval was +/- 1.5 mmol l-1 overall, and +/- 0.9 mmol l-1 for readings under 7.0 mmol l-1. Regression of the data from venous blood samples (meter vs reference method) had a correlation coefficient of 0.93 and a slope of x 1.1. The corrected 90% confidence interval was +/- 1.7 mmol l-1. Thus venous blood may be used even though the meter is calibrated for capillary samples but the value must be corrected by dividing by 1.1. Error-grid analysis showed that day-to-day clinical decisions could be made on the basis of ExacTech readings, although a diagnosis of borderline diabetes may not be possible.  相似文献   

10.

Background

The goal of diabetes treatment is maintaining near normoglycemia based on self-monitoring of blood glucose (SMBG). In this study, an evaluation of the analytical performance of the coulometry-based Optium Omega™ glucose meter designed for SMBG has been carried out.

Methods

The assessment of precision and between-lot variability was based on glucose measurements in ethylene-diaminetetraacetic acid venous blood samples. Glucose concentrations measured in 289 fresh capillary blood samples using the Omega glucose meter and the Biosen C_line analyzer were compared.

Results

Within-run imprecision coefficient of variation for the lower and higher glucose concentrations amounted to 5.09 and 2.1%, respectively. The relative lot-dependent differences found for the lower and higher glucose concentrations were equal to 6.8 and 2.6%, respectively. The glucose meter error calculated for various concentration ranges amounted from 2.22 to 4.48%. The glucose meter error met the accuracy criteria recommended by the International Organization for Standardization and the American Diabetes Association. The Passing-Bablok agreement test and error grid analysis with 96% of results in zone A indicated good concordance of results, including glucose concentrations below 100 mg/dl.

Conclusions

The evaluated Optium Omega glucose meter fits the analytical requirements for its use in blood glucose monitoring in diabetes patients.  相似文献   

11.
The accuracy of capillary blood glucose monitoring has been well demonstrated when applied by patients in the outpatient setting or by specially trained nurses on inpatient diabetes units. In order to determine the applicability of this technique in the more general hospital setting, a program was initiated for instructing general staff nurses in the use of Chemstrips bG strips and the Accu-Chek bG meter. As a pilot study, nurses on four general medical and surgical hospital floors performed capillary glucose determinations within 15 minutes of the drawing of venous blood samples for determination of plasma glucose in the hospital's Chemistry Laboratory. Two hundred ten paired measurements were made by 31 nurses. Linear regression analysis yielded a Pearson correlation coefficient of 0.96 between bedside and laboratory glucose measurements. The mean percent deviation between the two values was 7.9 percent. Acceptance by both nurses and patients was high. Properly supervised capillary glucose monitoring can provide a valuable adjunct to the care of hospitalized patients with diabetes.  相似文献   

12.
Patients managed in the intensive care units (ICUs) and general wards of the hospital experience a high incidence of hyperglycemia, hypoglycemia, and glycemic variability, despite significant hospital resources devoted to glucose control. Optimized glucose meters and monitoring systems are required to improve the safety and efficacy of insulin delivery and glucose control in the hospital. Safe insulin dosing requires timely and accurate glucose measurements, especially during dynamic changes in nutrition, insulin sensitivity, and physiological stress. In the current issue of Journal of Diabetes Science and Technology, Mitsios and coauthors describe the analytical accuracy of the new Accu-Check® Inform II blood glucose (BG) monitoring system commercialized by F. Hofmann-La Roche Ltd. The point-of-care glucose meter achieved the desired degree of accuracy and precision, as defined by Clinical and Laboratory Standards Institute POCT12-A3 guidelines when evaluated using venous blood from 600 critically ill patients from multiple ICUs at two medical centers. Venous whole blood samples were used to obtain glucose meter results in duplicate. The remaining blood sample was centrifuged to obtain plasma for central hospital laboratory testing using the hexokinase method within 5 min of meter testing. A total of 98.8% of the 1200 Accu-Check Inform II meter’s glucose values were within ±12.5% (±12 mg/dl) of the mean laboratory glucose value, and 99.8% were within ±20% (±20 mg/dl), thus meeting the Clinical and Laboratory Standards Institute criteria. Future studies are required to evaluate the clinical performance of the new BG monitoring system in the intended-use patient populations and critical care environments, using arterial, peripheral venous, central venous, and capillary blood samples.  相似文献   

13.

Background

Blood glucose (BG) meters used for assisted monitoring of blood glucose (AMBG) require different attributes compared with meters designed for home use. These include safety considerations (i.e., minimized risk of blood-borne pathogen transmission), capability for testing multiple blood sample types, and enhanced performance specifications. The OneTouch® Verio™Pro+ BG meter is designed to incorporate all of these attributes.

Methods

Meter accuracy was assessed in clinical studies with arterial, venous, and capillary blood samples with a hematocrit range of 22.9–59.8%. The effect of interferents, including anticoagulants, on accuracy was evaluated. The meter disinfection protocol was validated, and instructions for use and user acceptance of the system were assessed.

Results

A total of 97% (549/566) of BG measures from all blood sample types and 95.5% (191/200) of arterial blood samples were within ±12 mg/dl or 12.5% of reference measurements. The system was unaffected by 4 anticoagulants and 57 of 59 endogenous and exogenous compounds; it was affected by 2 compounds: pralidoxime iodide and xylose. Bleach wipes were sufficient to disinfect the meter. Users felt that the meter''s quality control (QC) prompts would help them to comply with regulatory requirements.

Conclusions

The meter provided accurate measurements of different blood samples over a wide hematocrit range and was not affected by 57 physiologic and therapeutic compounds. The QC prompts and specific infection-mitigating design further aid to make this meter system practical for AMBG in care facilities.  相似文献   

14.

Background

Patients and physicians expect accurate whole blood glucose monitoring even when patients are anemic, are undergoing peritoneal dialysis, or have slightly elevated ascorbate levels. The objective of this study was to estimate analytical error in two consumer and two hospital glucose meters contributed by variations in hematocrit, maltose, ascorbate, and imprecision.

Method

The influence of hematocrit (20–60%), maltose, and ascorbate were tested alone and in combination with each glucose meter and with a reference plasma glucose method at three concentrations of glucose. Precision was determined by consecutive analysis (n = 20) at three levels of glucose. Multivariate regression analysis was used to estimate the bias associated with the interferences, alone and in combination. Total analytical error was estimated as |% bias| + 1.96 (% imprecision).

Results

Three meters demonstrated hematocrit bias that was dependent upon glucose concentration. Maltose had profound concentration-dependent positive bias on the consumer meters, and the extent of maltose bias was dependent on hematocrit. Ascorbate produced small but statistically significant biases on three meters. Coincident low hematocrit, presence of maltose, and presence of ascorbate increased the observed bias and was summarized by estimation of total analytical error. Among the four glucose meter devices assessed, estimates of total analytical error in glucose measurement ranged from 6 to 68% under the conditions tested.

Conclusions

The susceptibility of glucose meters to clinically significant analytical biases is highly device-dependent, and low hematocrit exacerbated the observed analytical error.  相似文献   

15.

Background

This study evaluated the performance of the DIDGET® blood glucose monitoring system (BGMS) in the hands of its intended users: children, teens, and young adults with diabetes.

Methods

Finger stick capillary blood samples were tested in duplicate by subjects (with parent/guardian assistance, if needed) and health care professionals using the DIDGET BGMS, and results were compared with those obtained using a Yellow Springs Instruments (YSI) glucose analyzer. Modified venous blood samples (i.e., glycolyzed or spiked with glucose) were used to analyze meter performance under extreme glucose concentrations. Accuracy was assessed using International Organization for Standardization (ISO) 15197:2003 guidelines (i.e., 95% of meter results within ±15 mg/dl or ±20% of reference values).

Results

A total of 123 subjects aged 4 to 24 years with type 1 or type 2 diabetes were enrolled. The DIDGET meter achieved accuracy according to ISO 15197:2003 criteria: >97% of meter results were within ±15 mg/dl or ±20% of reference values. Regression analyses showed a high degree of correlation between meter and YSI results: coefficient of determination (R2) = 98.2% for all samples combined and 97.2% for capillary samples only. Clinical accuracy for combined samples was demonstrated by Parkes consensus error grid analyses; 100% of meter results were in zone A (98.5%) or zone B (1.5%). There was no difference in performance or accuracy across age subsets. Hematocrit values did not affect meter blood glucose results.

Conclusion

The DIDGET BGMS provided accurate test results across all age ranges in children, teens, and young adults with diabetes.  相似文献   

16.
目的 比较血糖仪与全自动生化仪检测末梢血和静脉血的一致性. 方法 选取2012年11~12月于我科门诊就诊的糖尿病和非糖尿病患者197例,使用3款血糖仪和日立7180全自动生化分析仪分别检测末梢血和静脉血血糖,并行线性回归分析和配对t检验. 结果 当红细胞压积(Hct)在35.1%~51.6%时,3款血糖仪与日立7180相关性顺序由高到低为罗氏Accuchek(末梢血:R2=0.968,静脉血:R2=0.966)、爱科莱GT-1920、强生One Touch.爱科莱GT-1920检测末梢血与日立7180的偏倚较小,检测静脉血的偏倚较大.罗氏Accuchek检测静脉血的均值低于日立7180.强生One Touch的偏倚约为10%. 结论3 款血糖仪与全自动生化分析仪相关性均较好,Hct可对两者的相关性造成影响.  相似文献   

17.
Glucose testing in the hospital with point-of-care devices presents multiple opportunities for error. Any device can fail under the right conditions. For glucose monitoring in the hospital, with thousands of operators, hundreds of devices, and dozens of locations involved, there is ample opportunity for errors that can impact the quality of test results. Errors can occur in any phase of the testing process: preanalytic, analytic, or postanalytic. Common sources of meter error include patient or methodology interferences, operator mistakes, environmental exposure, and device malfunction. Early models of glucose meters had few internal checks or capability to warn the operator of meter problems. The latest generation of glucose monitors has a number of internal checks and controls engineered into the testing process to prevent serious errors or warn the operator by suppressing test results. Some of these control processes are built into the software and data management system of the meters, others require the hospital to do something, such as regularly clean the meter or analyze control samples of known glucose concentration, to verify meter performance. Hospitals need to be aware of the potential for errors by understanding weaknesses in the testing process that could lead to erroneous results and take steps to prevent errors from occurring or to minimize the harm to patients when errors do occur. The reliability of a glucose result will depend on the balance of internal control features available from manufacturers in conjunction with the liquid control analysis and other control processes (operator training, device validation, and maintenance) utilized by the hospitals.  相似文献   

18.
A widely used method in monitoring glycemic status of ICU patients is point-of-care (POC) monitoring devices. A possible limitation to this method is altered peripheral blood flow in patients in shock, which may result in over/underestimations of their true glycemic status. This study aims to determine the accuracy of blood glucose measurements with a POC meter compared to laboratory methods in critically ill patients in shock. POC blood glucose was measured with a glucose-1-dehydrogenase-based reflectometric meter. The reference method was venous plasma glucose measured by a clinical chemistry analyzer (glucose oxidase-based). Outcomes assessed were concordance to ISO 15197:2003 minimum accuracy criteria for glucose meters, bias in glucose measurements obtained by the 2 methods using Bland–Altman analysis, and clinical accuracy through modified error grid analysis. A total of 186 paired glucose measurements were obtained. ISO 2003 accuracy criteria were met in 95.7% and 79.8% of POC glucose values in the normotensive and hypotensive group, respectively. Mean bias for the normotensive group was –12.4 mg/dL, while mean bias in the hypotensive group was –34.9 mg/dL. POC glucose measurements within the target zone for clinical accuracy were 90.2% and 79.8% for the normotensive and hypotensive group, respectively. POC blood glucose measurements were significantly less accurate in the hypotensive subgroup of ICU patients compared to the normotensive group. We recommend a lower threshold in confirming POC blood glucose with a central laboratory method if clinically incompatible. In light of recently updated accuracy standards, we also recommend alternative methods of glucose monitoring for the ICU population as a whole regardless of blood pressure status.  相似文献   

19.
The article entitled “Hematocrit Compensation in Electrochemical Blood Glucose Monitoring Systems” by Teodorczyk and colleagues in this issue of Journal of Diabetes Science and Technology demonstrates that the OneTouch® Verio™ glucose meter meets current regulatory expectations for glucose meter performance and is relatively free from interference by hematocrit. The lack of influence of hematocrit on whole blood glucose results is a valuable attribute for hospital applications, where greater variation of hematocrit among patients is anticipated. The choice of reference method for evaluation of glucose meters is an important consideration, and it is not clear to what extent reference methods used to evaluate glucose meters are also free from hematocrit interferences.  相似文献   

20.
OBJECTIVES: To evaluate and characterize routine blood glucose monitoring practices in nursing homes and assisted living facilities (ALFs). DESIGN: Cross‐sectional, self administered survey and facility site visit. SETTING: Two hundred eighty‐nine licensed long‐term care facilities in Pinellas County, Florida. PARTICIPANTS: Stratified random sample of 48 long‐term care facilities (17% overall sample). MEASUREMENTS: Data on facility characteristics, infection control policies, staff practices, and equipment used for blood glucose monitoring. Differences between facilities in each stratum were compared and evaluated using the Pearson chi‐square or Fisher exact test. RESULTS: Fifteen nursing homes and 17 small and 16 large ALFs participated; 53 declined (48% participation rate). Bloodborne pathogen training (P=.02), hepatitis B vaccination (P=.003), and blood glucose monitoring (P<.001) policies were reported less often at ALFs. Staff glove use during blood glucose monitoring was lowest (50%) at small ALFs (P=.02). Reusable fingerstick devices intended for personal use were most often in use at ALFs (P<.001); four of 18 facilities (including 1 nursing home) were inappropriately using them for multiple residents. At 22 facilities (including all nursing homes), multiple residents shared blood glucose meters; only six (27%) reported cleaning them after each use. CONCLUSION: Despite existing recommendations, practices that facilitate bloodborne pathogen transmission during blood glucose monitoring were identified at nursing homes and ALFs. Infection control practices and polices were most often lacking at ALFs. Better training and oversight of blood glucose monitoring in long‐term care is needed to prevent transmission of bloodborne pathogens.  相似文献   

设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号