首页 | 本学科首页   官方微博 | 高级检索  
相似文献
 共查询到20条相似文献,搜索用时 62 毫秒
1.
Dogan N  Glasgow GP 《Medical physics》2003,30(12):3091-3096
This study investigates the surface dose and build-up region dosimetry for oblique IMRT beams. The dependence of surface and build-up region doses of 0 degrees (perpendicular incidence) and 75 degrees (oblique incidence) IMRT fields on field size was measured and compared with open field dosimetry. Measurements were performed using a parallel-plate chamber and KODAK EDR2 films in a polystyrene phantom for a 6 cm x 6 cm and a 12 cm x 12 cm, 6 MV photon beam at depths of 0 mm (surface) through dmax. Data were normalized to the dmax value of each field. Four intensity modulated delivery patterns were created and delivered using step-and-shoot IMRT: (1) six static 1 cm x 6 cm strips (IMRTstrip), (2) 12 static 1 cm x 12 cm strips (IMRTstrip), (3) intensity modulated beam patterns created by using the inverse planning optimization software (IMRTopt) for 6 cm x 6 cm, and (4) IMRTopt for 12 cm x 12 cm field sizes. The percent depth doses (PDDs) of 0 degrees, 6 cm x 6 cm IMRTstrip beam at the surface and 5 mm were lower by 8.8% and 1.6%, respectively, compared to the open field. The PDDs of 75 degrees, 6 cm x 6 cm IMRTstrip beam at the surface and 5 mm were lower by 6.7% and 2.4%, respectively, compared to the open field. This study showed that IMRT itself is not contributing to greater skin doses.  相似文献   

2.
This work is intended to investigate the application and accuracy of micro-MOSFET for superficial dose measurement under clinically used MV x-ray beams. Dose response of micro-MOSFET in the build-up region and on surface under MV x-ray beams were measured and compared to Monte Carlo calculations. First, percentage-depth-doses were measured with micro-MOSFET under 6 and 10 MV beams of normal incidence onto a flat solid water phantom. Micro-MOSFET data were compared with the measurements from a parallel plate ionization chamber and Monte Carlo dose calculation in the build-up region. Then, percentage-depth-doses were measured for oblique beams at 0 degrees-80 degrees onto the flat solid water phantom with micro-MOSFET placed at depths of 2 cm, 1 cm, and 2 mm below the surface. Measurements were compared to Monte Carlo calculations under these settings. Finally, measurements were performed with micro-MOSFET embedded in the first 1 mm layer of bolus placed on a flat phantom and a curved phantom of semi-cylindrical shape. Results were compared to superficial dose calculated from Monte Carlo for a 2 mm thin layer that extends from the surface to a depth of 2 mm. Results were (1) Comparison of measurements with MC calculation in the build-up region showed that micro-MOSFET has a water-equivalence thickness (WET) of 0.87 mm for 6 MV beam and 0.99 mm for 10 MV beam from the flat side, and a WET of 0.72 mm for 6 MV beam and 0.76 mm for 10 MV beam from the epoxy side. (2) For normal beam incidences, percentage depth dose agree within 3%-5% among micro-MOSFET measurements, parallel-plate ionization chamber measurements, and MC calculations. (3) For oblique incidence on the flat phantom with micro-MOSFET placed at depths of 2 cm, 1 cm, and 2 mm, measurements were consistent with MC calculations within a typical uncertainty of 3%-5%. (4) For oblique incidence on the flat phantom and a curved-surface phantom, measurements with micro-MOSFET placed at 1.0 mm agrees with the MC calculation within 6%, including uncertainties of micro-MOSFET measurements of 2%-3% (1 standard deviation), MOSFET angular dependence of 3.0%-3.5%, and 1%-2% systematical error due to phantom setup geometry asymmetry. Micro-MOSFET can be used for skin dose measurements in 6 and 10 MV beams with an estimated accuracy of +/- 6%.  相似文献   

3.
The effect of beam obliquity on the surface relative dose profiles for the tangential photon beams was studied. The 6 and 15 MV photon beams with 4 x 4 and 10 x 10 cm2 field sizes produced by a Varian 21 EX linear accelerator were used. Phase-space models of the photon beams were created using Monte Carlo simulations based on the EGSnrc code, and were verified using film measurements. The relative dose profiles in the phantom skin, at 2 mm depth from the surface of the half-phantom geometry, or HPG, were calculated for increasing gantry angles from 270 to 280 deg clockwise. Relative dose profiles of a full phantom enclosing the whole tangential beam (full phantom geometry, or FPG) were also calculated using Monte Carlo simulation as a control for comparison. The results showed that, although the relative dose profiles in the phantom skin did not change significantly with an oblique beam using a FPG, the surface relative depth dose was increased for the HPG. In the HPG, with 6 MV photon beams and field size = 10 x 10 cm2, when the beam angle, starting from 270 deg, was increased from 1 to 3 deg, the relative depth doses in the phantom skin were increased from 68% to 79% at 10 cm depth. This increase in dose was slightly larger than the dose from 15 MV photon beams with the same field size and beam angles, where the relative depth doses in phantom skin were increased from 81% to 87% at 10 cm depth. A parameter called the percent depth dose (PDD) ratio, defined as the relative depth dose from the HPG to the relative depth dose from the FPG at a given depth along the phantom skin, was used to evaluate the effect of the phantom-air interface. It is found that the PDD ratio increased significantly when the beam angle was changed from zero to 1-3 degrees. Moreover, the PDD ratio, for a given field size, experienced a greater increase for 6 MV than for 15 MV. For the same photon beam energy, the PDD ratio increased more with a 4 x 4 cm2 field compared to 10 x 10 cm2. The results in this study will be useful for physicists and dosimetrists to predict the surface relative dose variations when using clinical tangential-like photon beams in radiation therapy.  相似文献   

4.
The problem of obliquely incident beams in electron-beam treatment planning   总被引:1,自引:0,他引:1  
Oblique incidence of an electron beam can alter the central axis depth dose. The incident beam can be considered to be an integration of many pencil beams or slit beams. Depending on the depth in the phantom, neighboring pencil beams may have a greater or lesser contribution to the dose at a point on the central axis compared to the contribution under normal incidence. The effect has been studied experimentally and theoretically. For 6- and 9-MeV electron beans, oblique incidence is found to produce an increased dose at shallow depths and a decreased dose at normal treatment depths.  相似文献   

5.
The purpose of this study is to provide detailed characteristics of incident photon beams for different field sizes and beam energies. This information is critical to the future development of accurate treatment planning systems. It also enhances our knowledge of radiotherapy photon beams. The EGS4 Monte Carlo code, BEAM, has been used to simulate 6 and 18 MV photon beams from a Varian Clinac-2100EX accelerator. A simulated realistic beam is stored in a phase space data file, which contains details of each particle's complete history including where it has been and where it has interacted. The phase space files are analysed to obtain energy spectra, angular distribution, fluence profile and mean energy profiles at the phantom surface for particles separated according to their charge and history. The accuracy of a simulated beam is validated by the excellent agreement between the Monte Carlo calculated and measured dose distributions. Measured depth-dose curves are obtained from depth-ionization curves by accounting for newly introduced chamber fluence corrections and the stopping-power ratios for realistic beams. The study presents calculated depth-dose components from different particles as well as calculated surface dose and contribution from different particles to surface dose across the field. It is shown that the increase of surface dose with the increase of the field size is mainly due to the increase of incident contaminant charged particles. At 6 MV, the incident charged particles contribute 7% to 21% of maximum dose at the surface when the field size increases from 10 x 10 to 40 x 40 cm2. At 18 MV, their contributions are up to 11% and 29% of maximum dose at the surface for 10 x 10 cm2 and 40 x 40 cm2 fields respectively. However, the fluence of these incident charged particles is less than 1% of incident photon fluence in all cases.  相似文献   

6.
We have investigated the dependence of the measured optical density on the incident beam energy, field size and depth for a new type of film, Kodak extended dose range (Kodak EDR). Film measurements have been conducted over a range of field sizes (3 x 3 cm2 to 25 x 25 cm2) and depths (d(max) to 15 cm), for 6 MV and 15 MV photons within a solid water phantom, and the variation in sensitometric response (net optical density versus dose) has been reported. Kodak EDR film is found to have a linear response with dose, from 0 to 350 cGy, which is much higher than that typically seen for Kodak XV film (0-50 cGy). The variation in sensitometric response for Kodak EDR film as a function of field size and depth is observed to be similar to that of Kodak XV film; the optical density varied in the order of 2-3% for field sizes of 3 x 3 cm2 and 10 x 10 cm2 at depths of d(max), 5 cm and 15 cm in the phantom. Measurements for a 25 x 25 cm2 field size showed consistently higher optical densities at depths of d(max), 5 cm and 15 cm, relative to a 10 x 10 cm2 field size at 5 cm depth, with 4-5% differences noted at a depth of 15 cm. Fractional depth dose and profiles conducted with Kodak EDR film showed good agreement (2%/2 mm) with ion chamber measurements for all field sizes except for the 25 x 25 cm2 at depths greater than 15 cm, where differences in the order of 3-5% were observed. In addition, Kodak EDR film measurements were found to be consistent with those of Kodak XV film for all fractional depth doses and profiles. The results of this study indicate that Kodak EDR film may be a useful tool for relative dosimetry at higher dose ranges.  相似文献   

7.

Purpose:

To investigate the optimal sensitometric curves of extended dose range (EDR2) radiographic film in terms of depth, field size, dose range and processing conditions for dynamic intensity modulated radiation therapy (IMRT) dosimetry verification with 6 MV X-ray beams.

Materials and methods:

A Varian Clinac 23 EX linear accelerator with 6 MV X-ray beam was used to study the response of Kodak EDR2 film. Measurements were performed at depths of 5, 10 and 15 cm in MedTec virtual water phantom and with field sizes of 2x2, 3x3, 10x10 and 15x15 cm2. Doses ranging from 20 to 450 cGy were used. The film was developed with the Kodak RP X-OMAT Model M6B automatic film processor. Film response was measured with the Vidar model VXR-16 scanner. Sensitometric curves were applied to the dose profiles measured with film at 5 cm in the virtual water phantom with field sizes of 2x2 and 10x10 cm2 and compared with ion chamber data. Scanditronix/Wellhofer OmniProTM IMRT software was used for the evaluation of the IMRT plan calculated by Eclipse treatment planning.

Results:

Investigation of the reproducibility and accuracy of the film responses, which depend mainly on the film processor, was carried out by irradiating one film nine times with doses of 20 to 450 cGy. A maximum standard deviation of 4.9% was found which decreased to 1.9% for doses between 20 and 200 cGy. The sensitometric curves for various field sizes at fixed depth showed a maximum difference of 4.2% between 2x2 and 15x15 cm2 at 5 cm depth with a dose of 450 cGy. The shallow depth tended to show a greater effect of field size responses than the deeper depths. The sensitometric curves for various depths at fixed field size showed slightly different film responses; the difference due to depth was within 1.8% for all field sizes studied. Both field size and depth effect were reduced when the doses were lower than 450 cGy. The difference was within 2.5% in the dose range from 20 to 300 cGy for all field sizes and depths studied. Dose profiles measured with EDR2 film were consistent with those measured with an ion chamber. The optimal sensitometric curve was acquired by irradiating film at a depth of 5 cm with doses ranging from 20 to 450 cGy with a 3×3 cm2 multileaf collimator. The optimal sensitometric curve allowed accurate determination of the absolute dose distribution. In almost 200 cases of dynamic IMRT plan verification with EDR2 film, the difference between measured and calculated dose was generally less than 3% and with 3 mm distance to agreement when using gamma value verification.

Conclusion:

EDR2 film can be used for accurate verification of composite isodose distributions of dynamic IMRT when the optimal sensitometric curve has been established.  相似文献   

8.
The purpose of this work is to measure the dose outside the treatment field for conformal CyberKnife treatments, to compare the results to those obtained for similar treatments delivered with gamma knife or intensity-modulated radiation therapy (IMRT), and to investigate the sources of peripheral dose in CyberKnife radiosurgery. CyberKnife treatment plans were developed for two hypothetical lesions in an anthropomorphic phantom, one in the thorax and another in the brain, and measurements were made with LiF thermoluminescent dosimeters (TLD-100 capsules) placed within the phantom at various depths and distances from the irradiated volume. For the brain lesion, gamma knife and 6-MV IMRT treatment plans were also developed, and peripheral doses were measured at the same locations as for the CyberKnife plan. The relative contribution to the CyberKnife peripheral dose from inferior- or superior-oblique beams entering or exiting through the body, internally scattered radiation, and leakage radiation was assessed through additional experiments using the single-isocenter option of the CyberKnife treatment-planning program with different size collimators. CyberKnife peripheral doses (in cGy) ranged from 0.16 to 0.041% (+/- 0.003%) of the delivered number of monitor units (MU) at distances between 18 and 71 cm from the field edge. These values are two to five times larger than those measured for the comparable gamma knife brain treatment, and up to a factor of four times larger those measured in the IMRT experiment. Our results indicate that the CyberKnife peripheral dose is due largely to leakage radiation, however at distances less than 40 cm from the field edge, entrance, or exit dose from inferior- or superior-oblique beams can also contribute significantly. For distances larger than 40 cm from the field edge, the CyberKnife peripheral dose is directly related to the number of MU delivered, since leakage radiation is the dominant component.  相似文献   

9.
Sharpe MB  Miller BM  Yan D  Wong JW 《Medical physics》2000,27(12):2719-2725
Two linear accelerators have been commissioned for delivering IMRT treatments using a step-and-shoot approach. To assess beam startup stability for 6 and 18 MV x-ray beams, dose delivered per monitor unit (MU), beam flatness, and beam symmetry were measured as a function of the total number of MU delivered at a clinical dose rate of 400 MU per minute. Relative to a 100 MU exposure, the dose delivered per MU by both linear accelerators was found to be within +/-2% for exposures larger than 4 MU. Beam flatness and symmetry also met accepted quality assurance standards for a minimum exposure of 4 MU. We have found that the performance of the two machines under study is well suited to the delivery of step-and-shoot IMRT. A system of dose calculation has also been commissioned for applying head scatter corrections to fields as small as 1x1 cm2. The accuracy and precision of the relative output calculations in water was validated for small fields and fields offset from the axis of collimator rotation. For both 6 and 18 MV x-ray beams, the dose per MU calculated in a water phantom agrees with measured data to within 1% on average, with a maximum deviation of 2.5%. The largest output factor discrepancies were seen when the actual radiation field size deviated from the set field size. The measured output in water can vary by as much 16% for 1x1 cm2 fields, when the measured field size deviates from the set field size by 2 mm. For a 1 mm deviation, this discrepancy was reduced to 8%. Steps should be taken to ensure collimator precision is tightly controlled when using such small fields. If this is not possible, very small fields should not contribute to a significant portion of the treatment, or uncertainties in the collimator position may effect the accuracy of the dose delivered.  相似文献   

10.
Cheng CW  Das IJ 《Medical physics》2002,29(2):226-230
In the step-and-shoot delivery of an IMRT plan with a Siemens Primus accelerator, radiation is turned off by desynchronizing the injector while the field parameters are being changed. When the machine is ready again a trigger pulse is sent to the injector to start the beam instantaneously. The objective of this study is to investigate the beam characteristics of the machine operating in the IMRT mode and to study the effect of the Initial Pulse Forming Network (IPEN) on the dark current. The central axis (CAX) output for a 10 x 10 cm2 field over the range 1-100 MU was measured with an ion chamber in a polystyrene phantom for both 6 and 15 MV x rays. Beam profiles were also measured over the range of 2-40 MU with the machine operating in the IMRT mode and compared with those in the normal mode. By adjusting the IPFN value, dark current radiation (DCR) was measured using ion chamber measurements. For both the normal and IMRT modes, dose versus MU is nonlinear in the range 1-5 MUs. Above 5 MU, dose varies linearly with MU for both 6 and 15 MV x rays. For stability of dose profiles, the 2 MU-IM group exhibit 20% variation from one subfield to another. The variation is about 5% for the 8 MU-IM group and <5% for 10 MU and higher. The results are similar in the normal treatment mode. With the IPFN at >80% of the PFN value, a spurious radiation associated with dark current at approximately 0.7% of the dose at isocenter for a 10 x 10 cm2 field is detected during the "PAUSE" state of the accelerator for 15 MV x rays. When the IPFN is lowered to <80% of the PFN value, no DCR is detected. For 6 MV x rays, no measurable DCR was detected regardless of the IPFN setting.  相似文献   

11.
With the increased use of cone beam CT (CBCT) for daily patient setup, the accumulated dose from CBCT may be significantly higher than that from simulation CT or portal imaging. The objective of this work is to measure the dose from daily pelvic scans with fixed technical settings and collimations. CBCT scans were acquired in half-fan mode using a half bowtie and x-rays were delivered in pulsed-fluoro mode. The skin doses for seven prostate patients were measured on an IRB-approved protocol. TLD capsules were placed on the patient's skin at the central axis of three beams: AP, left lateral (Lt Lat) and right lateral (Rt Lat). To avoid the ring artefacts centred in the prostate, the treatment couch was dropped 3 cm from the patient's tattoo (central axis). The measured AP skin doses ranged 3-6 cGy for 20-33 cm separation. The larger the patient size the less the AP skin dose. Lateral doses did not change much with patient size. The Lt Lat dose was approximately 4.0 cGy, which was approximately 40% higher than the Rt Lat dose of approximately 2.6 cGy. To verify this dose asymmetry, surface doses on an IMRT QA phantom (oval shaped, 30 cm x 20 cm) were measured at the same three sites using TLD capsules with 3 cm table-drop. The dose asymmetry was due to: (1) kV source rotation which always starts from the patient's Lt Lat and ends at Lt Lat. Gantry rotation gets much slower near the end of rotation but dose rate stays constant and (2) 370 degrees scan rotation (10 degrees scan overlap on the Lt Lat side). In vivo doses were measured inside a Rando pelvic heterogeneous phantom using TLDs. The left hip (femoral head and neck) received the highest doses of approximately 10-11 cGy while the right hip received approximately 6-7 cGy. The surface and in vivo doses were also measured for phantoms at the central-axis setup. The difference was less than approximately 12% to the table-drop setup.  相似文献   

12.
Computer algorithms for rotational therapy beams, in most cases, perform dose calculations by summing stored fixed beam data at finite angular steps. Such an algorithm, based on the Bentley beam model, was evaluated by comparing calculations with measured data for an 18-MV x-ray beam. Measurements were made in a specially constructed cylindrical water phantom of 15-cm radius using a 0.1-cm3 ionization chamber for an arc of 180 degrees and for a field size of 7.2 X 7.2 cm2 at 100-cm source-axis distance. This study revealed that the Bentley beam model, with fixed beams summed every 10 degrees, predicts the dose in the treatment volume, centered about the isocenter, with an accuracy of approximately 2%. However, dose at depths between the phantom surface and the treatment volume could be underestimated by as much as 10% (3% of isocenter). This was shown to be partially due to the truncated tails of the off-axis profiles in the Bentley model, which extend only 8 mm outside the edge of the radiation field, and the large angular increment of integration (10 degrees). Using beam profiles extending to 4 cm outside the edge of the radiation field and angular steps of 5 degrees or less for summation of fixed beams reduced errors to less than 5%. Therefore, extended beam profiles and smaller angular steps for summing fixed beams are recommended for photon rotation calculation when increased accuracy is required.  相似文献   

13.
The impact of the oblique electron beam on the lateral buildup ratio (LBR), used in the electron pencil beam model to predict the per cent depth dose (PDD) and dose per monitor unit (MU) for an irregular electron field, was examined using Monte Carlo simulation. The EGSnrc-based Monte Carlo code was used to model electron beams produced by a Varian 21 EX linear accelerator for different beam energies, angles of obliquity and field sizes. The Monte Carlo phase space model was verified by measurements using electron diode and radiographic film. For PDDs of oblique electron beams, it is found that the depth of maximum dose (d(m)) shifts towards the surface as the beam obliquity increases. Moreover, for increasing the beam angle of obliquity, the depth doses just beyond d(m) decrease with depth. The depth doses then increase eventually in a deeper depth close to the practical range. The LBRs and pencil beam radial spread function, calculated using PDDs with different field sizes, are found varying with electron beam energies, angles of obliquity and cutout diameters. It is found that LBR increases along the normalized depth when the beam angle of obliquity increases. This results in a decrease of the radial spread function with an increase of beam obliquity. When the size of the electron field increases, the variation of LBR with beam angle of obliquity decreases. It should be noted that when calculating dose per MU for an oblique electron beam with an irregular field misunderstanding and neglecting the effect of beam obliquity would lead to a significant deviation. A database of LBRs for oblique electron beams can be created using Monte Carlo simulation conveniently and is recommended when an oblique beam is used in electron radiotherapy.  相似文献   

14.
Sufficient skin dose needs to be delivered by a radiotherapy chest wall treatment regimen to ensure the probability of a near surface tumor recurrence is minimized. To simulate a chest wall treatment a hemicylindrical solid water phantom of 7.5 cm radius was irradiated with 6 MV x-rays using 20x20 cm2 and 10x20 cm2 fields at 100 cm source surface distance (SSD) to the base of the phantom. A surface dose profile was obtained from 0 to 180 degrees, in 10 degrees increments around the circumference of the phantom. Dosimetry results obtained from radiochromic film (effective depth of 0.17 mm) were used in the investigation, the superficial doses were found to be 28% (of Dmax) at the 0 degrees beam entry position and 58% at the 90 degrees oblique beam position. Superficial dose results were also obtained using extra thin thermoluminescent dosimeters (TLD) (effective depth 0.14 mm) of 30% at 0 degrees, 57% at 90 degrees, and a metal oxide semiconductor field effect transistor (MOSFET) detector (effective depth 0.5 mm) of 43% at 0 degrees, 62% at 90 degrees. Because the differences in measured superficial doses were significant and beyond those related to experimental error, these differences are assumed to be mostly attributable to the effective depth of measurement of each detector. We numerically simulated a bolus on/bolus off technique and found we could increase the coverage to the skin. Using an alternate "bolus on," "bolus off" regimen, the skin would receive 36.8 Gy at 0 degrees incidence and 46.4 Gy at 90 degrees incidence for a prescribed midpoint dose of 50 Gy. From this work it is evident that, as the circumference of the phantom is traversed the SSD increases and hence there is an inverse square fluence fall-off, this is more than offset by the increase in skin dose due to surface curvature to a plateau at about 90 degrees. Beyond this angle it is assumed that beam attenuation through the phantom and inverse square fall-off is causing the surface dose to reduce.  相似文献   

15.
Estimation of surface dose is very important for patients undergoing radiation therapy. In this work we investigate the dose at the surface of a water phantom and at a depth of 0.007 cm, the practical reference depth for skin as recommended by ICRP and ICRU, with ultra-thin TLDs and Monte Carlo calculations. The calculations and measurements were carried out for fields ranging from 5 x 5 cm2 to 20 x 20 cm2 for 6 MV, 10 MV and 18 MV photon beams. The variation of the surface dose with angle of incidence and field size was investigated. Also, the exit dose was computed and measured for the same fields and angles of incidence. The dose at the ICRU reference depth was computed. Good agreement (+/-5%) was achieved between measurements and calculations. The surface dose at the entrance increased with the angle of incidence and/or the field size. The exit dose decreased with the angle of incidence but it increased with field size. The dose at the surface of the patient is mostly dependent on the beam energy, modality and beam obliquity rather than the field size and field separation. By correlating TLD measurements with Monte Carlo calculations, we were able to predict the dose at the skin surface with good accuracy. Knowing the dose received at the surface of the patient can lead to prediction of skin reactions helping with the design of new treatment techniques and alternative dose fractionation schemes.  相似文献   

16.
Film dosimetry provides a convenient tool to determine dose distributions, especially for verification of IMRT plans. However, the film response to radiation shows a significant dependence on depth, energy and field size that compromise the accuracy of measurements. Kodak's XV2 film has a low saturation dose (approximately 100 cGy) and, consequently, a relatively short region of linear dose-response. The recently introduced Kodak extended range EDR2 film was reported to have a linear dose-response region extending to 500 cGy. This increased dose range may be particularly useful in the verification of IMRT plans. In this work, the dependence of Kodak EDR2 film's response on the depth, field size and energy was evaluated and compared with Kodak XV2 film. Co-60, 6 MV, 10 MV and 18 MV beams were used. Field sizes were 2 x 2, 6 x 6, 10 x 10, 14 x 14, 18 x 18 and 24 x 24 cm2. Doses for XV2 and EDR2 films were 80 cGy and 300 cGy, respectively. Optical density was converted to dose using depth-corrected sensitometric (Hurter and Driffield, or H&D) curves. For each field size, XV2 and EDR2 depth-dose curves were compared with ion chamber depth-dose curves. Both films demonstrated similar (within 1%) field size dependence. The deviation from the ion chamber for both films was small forthe fields ranging from 2 x 2 to 10 x 10 cm2: < or =2% for 6, 10 and 18 MV beams. No deviation was observed for the Co-60 beam. As the field size increased to 24 x 24 cm2, the deviation became significant for both films: approximately 7.5% for Co-60, approximately 5% for 6 MV and 10 MV, and approximately 6% for 18 MV. During the verification of IMRT plans, EDR2 film showed a better agreement with the calculated dose distributions than the XV2 film.  相似文献   

17.
The peripheral dose outside the applicators in electron beams was studied using a Varian 21 EX linear accelerator. To measure the peripheral dose profiles and point doses for the applicator, a solid water phantom was used with calibrated Kodak TL films. Peak dose spot was observed in the 4 MeV beam outside the applicator. The peripheral dose peak was very small in the 6 MeV beam and was ignorable at higher energies. Using the 10 x 10 cm(2) cutout and applicator, the dose peak for the 4 MeV beam was about 12 cm away from the field central beam axis (CAX) and the peripheral dose profiles did not change with depths measured at 0.2, 0.5 and 1 cm. The peripheral doses and profiles were further measured by varying the angle of obliquity, cutout and applicator size for the 4 MeV beam. The local peak dose was increased with about 3% per degree angle of obliquity, and was about 1% of the prescribed dose (angle of obliquity equals zero) at 1 cm depth in the phantom using the 10 x 10 cm(2) cutout and applicator. The peak dose position was also shifted 7 mm towards the CAX when the angle of obliquity was increased from 0 to 15 degrees.  相似文献   

18.
An add-on multileaf collimator for electrons (eMLC) has been developed that provides computer-controlled beam collimation and isocentric dose delivery. The design parameters result from the design study by Gauer et al (2006 Phys. Med. Biol. 51 5987-6003) and were configured such that a compact and light-weight eMLC with motorized leaves can be industrially manufactured and stably mounted on a conventional linear accelerator. In the present study, the efficiency of an initial computer-controlled prototype was examined according to the design goals and the performance of energy- and intensity-modulated treatment techniques. This study concentrates on the attachment and gantry stability as well as the dosimetric characteristics of central-axis and off-axis dose, field size dependence, collimator scatter, field abutment, radiation leakage and the setting of the accelerator jaws. To provide isocentric irradiation, the eMLC can be placed either 16 or 28 cm above the isocentre through interchangeable holders. The mechanical implementation of this feature results in a maximum field displacement of less than 0.6 mm at 90 degrees and 270 degrees gantry angles. Compared to a 10 x 10 cm applicator at 6-14 MeV, the beam penumbra of the eMLC at a 16 cm collimator-to-isocentre distance is 0.8-0.4 cm greater and the depth-dose curves show a larger build-up effect. Due to the loss in energy dependence of the therapeutic range and the much lower dose output at small beam sizes, a minimum beam size of 3 x 3 cm is necessary to avoid suboptimal dose delivery. Dose output and beam symmetry are not affected by collimator scatter when the central axis is blocked. As a consequence of the broader beam penumbra, uniform dose distributions were measured in the junction region of adjacent beams at perpendicular and oblique beam incidence. However, adjacent beams with a high difference in a beam energy of 6 to 14 MeV generate cold and hot spots of approximately 15% in the abutting region. In order to improve uniformity, the energy of adjacent beams must be limited to 6 to 10 MeV and 10 to 14 MeV respectively. At the maximum available beam energy of 14 MeV, radiation leakage results mainly from the intraleaf leakage of approximately 2.5% relative dose which could be effectively eliminated at off-axis distances remote from the field edge by adjusting the jaw field size to the respective opening of the eMLC. Additionally, the interleaf and leaf-end leakage could be reduced by using a tongue-and-groove leaf shape and adjoining the leaf-ends off-axis respectively.  相似文献   

19.
The higher sensitivity to low-energy scattered photons of radiographic film compared to water can lead to significant dosimetric error when the beam quality varies significantly within a field. Correcting for this artifact will provide greater accuracy for intensity modulated radiation therapy (IMRT) verification dosimetry. A procedure is developed for correction of the film energy-dependent response by creating a pencil beam kernel within our treatment planning system to model the film response specifically. Film kernels are obtained from EGSnrc Monte Carlo simulations of the dose distribution from a 1 mm diameter narrow beam in a model of the film placed at six depths from 1.5 to 40 cm in polystyrene and solid water phantoms. Kernels for different area phantoms (50 x 50 cm2 and 25 x 25 cm2 polystyrene and 30 x 30 cm2 solid water) are produced. The Monte Carlo calculated kernel is experimentally verified with film, ion chamber and thermoluminescent dosimetry (TLD) measurements in polystyrene irradiated by a narrow beam. The kernel is then used in convolution calculations to, predict the film response in open and IMRT fields. A 6 MV photon beam and Kodak XV2 film in a polystyrene phantom are selected to test the method as they are often used in practice and can result in large energy-dependent artifacts. The difference in dose distributions calculated with the film kernel and the water kernel is subtracted from film measurements to obtain a practically film artifact free IMRT dose distribution for the Kodak XV2 film. For the points with dose exceeding 5 cGy (11% of the peak dose) in a large modulated field and a film measurement inside a large polystyrene phantom at depth of 10 cm, the correction reduces the fraction of pixels for which the film dose deviates from dose to water by more than 5% of the mean film dose from 44% to 6%.  相似文献   

20.
The use of rigid carbon fiber couch inserts in radiotherapy treatment couches is a well-established method of reducing patient set-up errors associated with couch sag. Several published studies have described such inserts as radiotranslucent with negligible attenuation of the radiation field. Most of these studies were conducted with the radiation field normally incident on the couch and there appears to be no evidence in the literature of the effect of the gantry angle on the extent of beam attenuation by the carbon fiber insert alone during external beam radiotherapy. In this study we examined the magnitude of this effect over a range of posterior oblique gantry angles using a cylindrical solid water phantom containing an ionization chamber placed isocentrically. It was found that a 6 MV photon beam, field size 10 x 5 cm, was attenuated significantly as the gantry angle approached the plane of the couch, from 2% at normal incidence and reaching 9% attenuation at angle of incidence 70 degrees. This could have serious implications regarding dose to the treatment volume for treatments requiring posterior oblique angles of incidence with a possible correction factor necessary in monitor unit calculations.  相似文献   

设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号