首页 | 本学科首页   官方微博 | 高级检索  
相似文献
 共查询到20条相似文献,搜索用时 31 毫秒
1.
S Kim  C Liu  C Chen  J R Palta 《Medical physics》1999,26(6):949-955
A simple algorithm was developed for calculation of the in-air output at various source-to-detector distances (SDDs) on the central axis for wedged fields. In the algorithm we dealt independently with two effective sources, one for head scatter and the other for wedge scatter. Varian 2100C with 18 and 8 MV photon beams was used to examine this algorithm. The effective source position for head scatter for wedged fields was assumed to be the same as that for open fields, and the effective source position for wedge scatter was assumed to be a certain distance upstream from the physical location of the wedge. The shift of the effective source for wedge scatter, w, was found to be independent of field size. Moreover, we observed no systematic dependency of w on wedge angle or beam energy. One value, w = 5.5 cm, provided less than 1% difference in in-air outputs through the whole experimental range, i.e., 6 x 6 to 20 x 20 cm2 field size (15 x 20 cm2 for 60 degrees wedge), 15 degrees-60 degrees wedge angle, 80-130 cm SDD, and both 18 and 8 MV photon beams. This algorithm can handle the case in which use of a tertiary collimator with an external wedge makes the field size for the determination of wedge scatter different from that for head scatter. In this case, without the two-effective-source method, the maximum of 4.7% and 2.6% difference can be given by the inverse square method and one-effective-source method in a 45 degrees wedged field with 18 MV. Differences can be larger for thicker wedges. Enhanced dynamic wedge (EDW) fields were also examined. It was found that no second effective source is required for EDW fields.  相似文献   

2.
The behavior of the effective source position and the correction factor associated with the collimator opening (head-scatter factor) were investigated for the 6- and 25-MV x-ray beams of a linear accelerator. The primary photon fluence was measured in air for square field sizes from 5 x 5 cm to 40 x 40 cm at distances from the nominal source of 80 to 140 cm, for open and wedged fields (wedge angle 60 deg). An inverse-square analysis shows that, for open fields, the effective source position of the accelerator is about the same (approximately 1 cm downstream) at 6 and 25 MV, for all field sizes. For the wedged fields, the effective source position depends on field size and ranges from about 2 to 4 cm. The head-scatter correction factors for given collimator settings were found to be essentially independent of distance at both energies.  相似文献   

3.
Mini-phantoms are an important tool for measurement of basic head scatter parameters in high-energy photon beams, and recently they have also been used for beam quality specification. Therefore the feasibility and reliability of basic beam parameter acquisition using only a mini-phantom is checked in 6, 18 and 25 MV photon beams. These parameters include head scatter correction factors, phantom scatter correction factors, total scatter correction factors, wedge factors, off-axis ratios, as well as beam attenuation coefficients and beam hardening coefficients. In order to specify beam quality variations and beam quality modifications by a wedge, two different methods are compared: the first method uses a constant source to chamber distance of 1 m, the second method refers to narrow beam geometry. Mu values derived with two different beam quality specification methods show a systematic deviation. However, relative variations of the attenuation coefficient within the beam and the associated beam quality modifications observed with the two methods show good agreement in open and wedged beams. Phantom scatter correction factors are calculated from measured head scatter correction factors and total scatter correction factors as well as from attenuation coefficients. Measured and calculated phantom scatter correction factors agree within 1% with the values given in literature. For 18 and 25 MV photon beam, wedge factors measured in water or in the mini-phantom agree within 0.5%, but maximum deviations of approximately 1.5% are observed at 6 MV for the largest field sizes. It is demonstrated that the determination of several beam data related to full scatter conditions does not necessarily require the availability of a full scatter phantom. The mini-phantom is a reliable but very cheap and simple tool. It offers versatile possibilities to measure, check and verify basic beam parameters in high-energy photon beams.  相似文献   

4.
Compared with a set of physical photon wedges, a non physical wedge (virtual or dynamic wedge), realized by a moving collimator jaw, offers an alternative that allows creation of a wedged field with any arbitrary wedge angle instead of the traditional four physical wedges (15 degrees, 30 degrees, 45 degrees and 60 degrees). It is commonly assumed that non-physical wedges do not alter the photon spectrum compared with physical wedges that introduce beam hardening and loss of dose uniformity in the unwedged direction. In this study, we investigated the influence of a virtual wedge on the photon spectra of a 6-10 MV Siemens MD2 accelerator with the Monte Carlo code EGS4/BEAM. Good agreement was obtained between calculated and measured lateral dose profiles at the depth of maximum dose and at 10 cm depth for 20 x 20 cm2 fields for 6 and 10 MV photon beams. By comparing Monte Carlo models of a physical wedge and the virtual wedge that was studied in this work, it is confirmed that the latter has an insignificant effect on the beam quality, whereas the former can introduce significant beam hardening.  相似文献   

5.
A method is presented for calculation of a two-dimensional function, T(wedge)(x,y), describing the transmission of a wedged photon beam through a patient. This in an extension of the method that we have published for open (nonwedged) fields [Med. Phys. 25, 830-840 (1998)]. Transmission functions for open fields are being used in our clinic for prediction of portal dose images (PDI, i.e., a dose distribution behind the patient in a plane normal to the beam axis), which are compared with PDIs measured with an electronic portal imaging device (EPID). The calculations are based on the planning CT scan of the patient and on the irradiation geometry as determined in the treatment planning process. Input data for the developed algorithm for wedged beams are derived from (the already available) measured input data set for transmission prediction in open beams, which is extended with only a limited set of measurements in the wedged beam. The method has been tested for a PDI plane at 160 cm from the focus, in agreement with the applied focus-to-detector distance of our fluoroscopic EPIDs. For low and high energy photon beams (6 and 23 MV) good agreement (approximately 1%) has been found between calculated and measured transmissions for a slab and a thorax phantom.  相似文献   

6.
A M Kalend  A Wu  V Yoder  A Maitz 《Medical physics》1990,17(4):701-704
It is known experimentally that a wedge transmission factor depends upon the field size and depth of measurement in particular. Dependence of the transmission upon depth has been attributed to a hardening of the incident beam through the filter, which preferentially absorbs the low-energy photon of the bremsstrahlung component of that beam. We have attempted to separate this hardening effect from that of increased phantom scatter due to dose gradient induced by the wedge filter. Using an experimental wedge machined from cerrobend, the filter transmission at depth is measured and redefined relative to an "equally hardened" beam, obtained by filtering through a flat slab of equal thickness at the center of the wedge. Results of the Co-60, 4-, and 8-MV wedged beams indicate that nearly half of the increase in the transmission at depth is due to the effect of dose-gradient scatter in polystyrene phantom. Based on a simple relationship between primary and scattering radiation, an algebraic presentation is indeed in support of the dose gradient resulting in apparent increase in the wedge factors, at depth.  相似文献   

7.
The Philips SL25 accelerator is a multimodality machine offering asymmetric collimator jaws and a new type of beam bending and transport system. It produces photon beams, nominally at 6 and 25 MV, and a scattered electron beam with nine selectable energies between 4 and 22 MeV. Dosimetric characteristics for the 6- and 25-MV photon beams are presented with respect to field flatness, surface and depth dose characteristics, isodose distribution, field size factors for both open and wedged fields, and narrow beam transmission data in different materials.  相似文献   

8.
An ESTRO booklet and a report of the Netherlands Commission on Radiation Dosimetry have been published recently describing empirical methods for monitor unit (MU) calculations in symmetrical high-energy photon beams. Both documents support the same basic ideas; firstly the separation of head scatter and volume scatter components and secondly the determination of head scatter quantities in a mini-phantom. Based on these ideas the methods previously described for MU calculations in symmetrical beams are extended to asymmetrical open and wedged beams in isocentric treatment conditions. All required dosimetric parameters (normalized head scatter factors, phantom scatter correction factors, wedge factors, off-axis ratios, quality index, and depth dose parameters) are determined as a function of beam axis position in order to study their off-axis dependence. Measurements are performed for 6 MV and 18 MV photon beams provided by two different dual-energy linear accelerators, a GE Saturne 42 and a Varian 2100 CD linac.  相似文献   

9.
Monitor unit calculations for wedged asymmetric photon beams   总被引:1,自引:0,他引:1  
Algorithms for calculating monitor units (MUs) in wedged asymmetric high-energy photon beams as implemented in treatment planning systems have their limitations. Therefore an independent method for MU calculation is necessary. The aim of this study was to develop an empirical method to determine MUs for points at the centre of wedged fields, asymmetric in two directions. The method is based on the determination of an off-axis factor (OAF) that corrects for the difference in dose between wedged asymmetric and wedged symmetric beams with the same field size. Measurements were performed in a water phantom irradiated with 6 and 18 MV photon beams produced by Elekta accelerators, which are fitted with an internal motorized wedge that has a complex shape. The OAF perpendicular to the wedge direction changed significantly with depth for the 18 MV beam. Dose values measured for a set of 18 test cases were compared with those calculated with our method. The maximum difference found was 6.5% and in 15 cases this figure was smaller than 2.0%. The analytical method of Khan and the empirical method of Georg were also tested and showed errors up to 12.8%. It can be concluded that our simple formalism is able to calculate MUs in wedged asymmetric fields with an acceptable accuracy in most clinical situations.  相似文献   

10.
Dosimetric properties of Virtual Wedge (VW) and physical wedge (PW) in 6 and 23 MV photon beams from a Siemens Primus linear accelerator, including wedge factors, depth doses, dose profiles, peripheral doses and surface doses, are compared. While there is a great difference in absolute values of wedge factors, VW factors (VWFs) and PW factors (PWFs) have a similar trend as a function of field size. PWFs have a stronger depth dependence than VWF due to beam hardening in PW fields. VW dose profiles in the wedge direction, in general, match very well with PW, except in the toe area of large wedge angles with large field sizes. Dose profiles in the nonwedge direction show a significant reduction in PW fields due to off-axis beam softening and oblique filtration. PW fields have significantly higher peripheral doses than open and VW fields. VW fields have similar surface doses as the open fields while PW fields have lower surface doses. Surface doses for both VW and PW increase with field size and slightly with wedge angle. For VW fields with wedge angles 45 degrees and less, the initial gap up to 3 cm is dosimetrically acceptable when compared to dose profiles of PW. VW fields in general use less monitor units than PW fields.  相似文献   

11.
Due to the significant increase in the number of monitor units used to deliver a dynamic IMRT treatment, the total MLC leakage (transmission plus scatter) can exceed 10% of the maximum in-field dose. To avoid dosimetric errors, this leakage must be accurately accounted for in the dose calculation and conversion of optimized intensity patterns to MLC trajectories used for treatment delivery. In this study, we characterized the leaf end transmission and leakage radiation for Varian 80- and 120-leaf MLCs using Monte Carlo simulations. The complex geometry of the MLC, including the rounded leaf end, leaf edges (tongue-and-groove and offset notch), mounting slots, and holes was modeled using MCNP4b. Studies were undertaken to determine the leakage as a function of field size, components of the leakage, electron contamination, beam hardening and leaf tip effects. The leakage radiation with the MLC configured to fully block the field was determined. Dose for 6 and 18 MV beams was calculated at 5 cm depth in a water phantom located at 95 cm SSD, and normalized to the dose for an open field. Dose components were scored separately for radiation transmitted through and scattered from the MLC. For the 80-leaf MLC at 6 MV, the average leakage dose is 1.6%, 1.7%, 1.8%, and 1.9% for 5 x 5, 10 x 10, 15 x 15, and 20 x 20cm2 fields, respectively. For the 120-leaf MLC at 6 MV, the average leakage dose is 1.6%, 1.6%, 1.7%, and 1.9% for the same field sizes. Measured leakage values for the 120-leaf MLC agreed with calculated values to within 0.1% of the open field dose. The increased leakage with field size is attributed to MLC scattered radiation. The fractional electron contamination for a blocked MLC field is greater than that for an open field. The MLC attenuation significantly affects the photon spectrum, resulting in an increase in percent depth dose at 6 MV, however, little effect is observed at 18 MV. Both phantom scatter and the finite source size contribute to the leaf tip profile observed in phantom. The results of this paper can be applied to fluence-to-trajectory and trajectory-to-fluence calculations for IMRT.  相似文献   

12.
X A Li 《Medical physics》1999,26(6):962-966
The peak scatter factor (PSF) for a photon beam is defined as the ratio of the total dose and the primary dose at the depth of dose maximum. The values of the PSF for photon beams ranging from 60Co to 24 MV are calculated using the EGS4 Monte Carlo technique, to avoid measurement difficulties. The calculation shows that the effect of SSD on PSF for high energy photon beams is not significant for small fields, but can be as high as 1% for large fields. For the 60Co beam, the calculation agrees with the data tabulated on BJR Supplement 25 to within 0.8%. The BJR value (1.054) of 10 x 10 cm2 for 60Co is 0.6% lower than the present value due to the underestimation of scatters from the source capsule and collimators. For a given field size, PSF is varied by up to 2% when beam quality changes from 60Co to 24 MV. For normalized PSF, the values of BJR Supplement 25 (which are assumed to be the same for beams ranging from 60Co to 50 MV) agree well with the present calculation for small field sizes, but are higher than our data by up to 2% for large field sizes. The presently calculated PSFs are related to field size(s) by an empirical expression, PSF = 1 + ms/(s + n), where m and n are the fitting parameters. This equation describes the PSFs within 0.4% (0.15% on average).  相似文献   

13.
The depth dependence (up to 25 cm) of the in-phantom wedge transmission factor (WTF) has been determined for three medical linear accelerator x-ray beams with energies of 4, 6, and 10 MV containing 15 degrees-60 degrees (nominal) brass wedges. All measurements were made with a cylindrical ionization chamber in water, for a field size of 10 X 10 cm2 with a source-skin distance of 80 or 100 cm. We conclude that, for the accelerators studied, the WTF factor at depth is less than 2% different from that determined at dmax (for the nominal wedge angles and photon energies studied) unless the depth of interest is greater than 10 cm. Up to the maximum depth studied (25 cm) the relative wedge factor--that is, wedge factor at depth compared to that determined at dmax--was about equal to or less than 1.02 for the 15 degrees and 30 degrees wedges and any of the photon beam energies studied. For the seldom utilized combination of a nominal wedge angle in excess of 45 degrees with a depth greater than 10 cm, the WTF at depth can differ from the WTF determined at dmax, by up to 5%. Since the wedge transmission factor is reflective of relative percent dose data, our results also indicate that it is in error to use open field percent depth doses for certain combinations of wedge angle, photon energy, and depth.  相似文献   

14.
The separation of total absorbed dose into primary and scatter components is a commonly used technique in photon dose calculations. The primary dose component can be characterised by a measured narrow beam attenuation coefficient and a single normalisation value which establishes the relative proportion of the primary to the total dose at some reference depth and field size. The determination of this normalisation value from measured data requires an extrapolation of measured values for finite field sizes to obtain a zero field size value. We have used Monte Carlo simulations to score primary and scatter dose for photon beams of 4, 6, 10, 15 and 24 MV and report values of the scatter to primary ratio at the depth of dose maximum for the circular equivalent of a 10 cm x 10 cm field. These values have an uncertainty of less than 1% and can be used in lieu of extrapolation of measured data to establish the relative magnitude of the primary dose for a wide range of photon beam energies.  相似文献   

15.
Enhanced dynamic wedge factors (EDWF) are characterized by a strong field size dependence. In contrast to physical wedge factors, the EDWF decrease as the field size is increased: for 6 MV 60 degrees wedge, the EDWF decreases by 50% when the field size is increased from 4 x 4 cm2 to 20 x 20 cm2. A method that eliminates the field size dependence of EDWF was developed and investigated in this work. In this method, the wedged field shape is determined by a multileaf collimator. The initial position of the moving Y jaw is determined by the field size and the stationary Y jaw is kept fixed at 10 cm for field sizes < or = 20 cm in the wedged direction. For all other fields, the stationary Y jaw setting is determined by the field size. The modified method results in EDWF that are independent of field size, with no change in the wedge dose distribution when compared with the conventional use of EDW.  相似文献   

16.
Recent studies have indicated that radiotherapy treatments undertaken on a flattening filter-free (FFF) linear accelerator have a number of advantages over treatments undertaken on a conventional linear accelerator. In addition, 4 MV photon beams may give improved isodose coverage for some treatment volumes at air/tissue interfaces, compared to when utilizing the clinical standard of 6 MV photons. In order to investigate these benefits, FFF beams were established on an Elekta Beam Modulator linear accelerator for 4 MV photons. Commissioning beam data were obtained for open and wedged fields. The measured data were then imported into a treatment planning system and a beam model was commissioned. The beam model was optimized to improve dose calculations at shallow, clinically relevant depths. Following verification, the beam model was utilized in a treatment planning study, including volumetric modulated arc therapy, for a selection of lung, breast/chest wall and larynx patients. Increased dose rates of around 800 MU min(-1) were recorded for open fields (relative to 320 MU min(-1) for filtered open fields) and reduced head scatter was inferred from output factor measurements. Good agreement between planned and delivered dose was observed in verification of treatment plans. The planning study indicated that with a FFF beam, equivalent (and in some cases improved) isodose profiles could be achieved for small lung and larynx treatment volumes relative to 4 MV filtered treatments. Furthermore, FFF treatments with wedges could be replicated using open fields together with an 'effective wedge' technique and isocentre shift. Clinical feasibility of a FFF beam was therefore demonstrated, with beam modelling, treatment planning and verification being successfully accomplished.  相似文献   

17.
目的:探讨不同能量下,Varian21EX直线加速器中物理楔形因子和动态楔形因子受照射野大小和深度的影响。方法:在固体水膜体中利用0.6 cc电离室对6 MV和15 MV射线束下不同角度物理楔形板和动态楔形板分别测量加和不加楔形滤片时的剂量率来计算楔形因子。通过测量不同角度的物理楔形板和动态楔形板在固定照射野(10 cm×10 cm)的不同深度下的楔形因子来研究楔形因子随深度的变化规律。同时,对于楔形因子随射野大小的变化规律,还测量了不同角度的物理楔形板和动态楔形板在固定深度(d=10 cm)下的不同射野大小的楔形因子。为了更好地分析物理楔形因子与动态楔形因子的差异,引入了相对楔形因子NWF。结果:深度对于物理楔形板的楔形因子较为明显,深度增加时楔形因子增大,且随着楔形角的增大变化更明显。对于150、300、450、600的物理楔形板,当深度由最大深度增加到20 cm时对于6 MV能量楔形因子分别增加了1.86%、3.79%、4.99%、7.95%;对于15 MV能量1.29%、1.35%、1.49%、2.03%。而动态楔形因子随深度变化不明显,最大变化不到1%。射野大小对于物理楔形因子也有一定的影响,楔形因子随射野增加而增加,但是增加幅度不大;而对于动态楔形板,在6 MV和15 MV射线束下楔形因子受射野的增大都有明显的减小。对于100、150、200、250、300、450、600的动态楔形板,从参考射野(10 cm×10 cm)到最大射野,楔形因子分别减少了7.91%、11.04%、14.08%、16.96%、19.7%、28.03%、35.89%对于6 MV和5.72%、8.17%、10.41%、12.85%、15.08%、21.82%、30.59%对于15 MV能量。结论:对于物理楔形板,深度和射野大小都对物理楔形因子有影响,所以临床剂量计算时必须考虑深度和射野大小对物理楔形因子的影响并对它进行修正。对于动态楔形板,深度对动态楔形因子影响较小,在临床剂量计算时可以忽略;而射野大小对动态楔形因子影响比较明显,在临床剂量计算时只须考虑相对射野楔形因子。  相似文献   

18.
The effect of beam obliquity on the surface relative dose profiles for the tangential photon beams was studied. The 6 and 15 MV photon beams with 4 x 4 and 10 x 10 cm2 field sizes produced by a Varian 21 EX linear accelerator were used. Phase-space models of the photon beams were created using Monte Carlo simulations based on the EGSnrc code, and were verified using film measurements. The relative dose profiles in the phantom skin, at 2 mm depth from the surface of the half-phantom geometry, or HPG, were calculated for increasing gantry angles from 270 to 280 deg clockwise. Relative dose profiles of a full phantom enclosing the whole tangential beam (full phantom geometry, or FPG) were also calculated using Monte Carlo simulation as a control for comparison. The results showed that, although the relative dose profiles in the phantom skin did not change significantly with an oblique beam using a FPG, the surface relative depth dose was increased for the HPG. In the HPG, with 6 MV photon beams and field size = 10 x 10 cm2, when the beam angle, starting from 270 deg, was increased from 1 to 3 deg, the relative depth doses in the phantom skin were increased from 68% to 79% at 10 cm depth. This increase in dose was slightly larger than the dose from 15 MV photon beams with the same field size and beam angles, where the relative depth doses in phantom skin were increased from 81% to 87% at 10 cm depth. A parameter called the percent depth dose (PDD) ratio, defined as the relative depth dose from the HPG to the relative depth dose from the FPG at a given depth along the phantom skin, was used to evaluate the effect of the phantom-air interface. It is found that the PDD ratio increased significantly when the beam angle was changed from zero to 1-3 degrees. Moreover, the PDD ratio, for a given field size, experienced a greater increase for 6 MV than for 15 MV. For the same photon beam energy, the PDD ratio increased more with a 4 x 4 cm2 field compared to 10 x 10 cm2. The results in this study will be useful for physicists and dosimetrists to predict the surface relative dose variations when using clinical tangential-like photon beams in radiation therapy.  相似文献   

19.
In a beam accessory configuration for a linear accelerator using a prototype multileaf collimator, newly designed wedges were mounted beyond the blocking tray. The isodose curves, depth of maximum dose, surface dose, and wedge transmission factors were measured for the wedges designed for this unique configuration. The same set of wedges was used for both 6- and 18-MV x rays. The shape of the wedged isodose curves was essentially unchanged from those produced by the conventional wedges located above the blocking tray. The isodose curves exhibited the desired wedge angles over the range of field sizes from 5 x 5 to 15 x 40 cm. In the 10 x 10-cm field, the average difference between the observed wedge angle and the desired wedge angle was 3.8 degrees. The surface doses ranged from 18% to 35% for the wedged 10 x 10-cm fields as compared with about 15% for the same open field. Dosimetrically the wedges were acceptable for clinical use.  相似文献   

20.
Compensators can be used as beam intensity modulation devices for intensity-modulated radiation therapy applications. In contrast with multileaf collimators, compensators introduce scatter and beam hardening into the therapeutic x-ray beam. The degree of scatter and beam filtering depends on the compensator material and beam energy. Pencil beam dose calculation models can be used to derive the shape of the compensator. In this study a novel way of incorporating the effect of compensator-induced scatter and beam filtration is presented. The study was conducted using 6, 8, and 15 MV polyenergetic pencil beams (PBs). The compensator materials that were studied included wax, brass, copper, and lead. The perturbation effects of the compensators on the PB dose profiles were built in the PB dose profiles and tested for regular fields containing a step compensator and benchmarked against DOSXYZnrc Monte Carlo calculated dose profiles. These effects include compensator beam filtration and Compton-scattered photons generated in the compensator materials that influence the resulting PB dose profiles. These data were obtained from DOSXYZnrc simulations. A Gaussian function was used to model off-axis scatter and an exponential function was used to model beam hardening at any radius, r. Dose profiles were calculated under a step compensator using the method that can model beam hardening and off-axis scatter, as well as a conventional method where the PB profiles are not adjusted, but a single effective attenuation coefficient is used instead to best match the dose profiles. Both sets of data were compared to the DOSXYZnrc data. Depth and profile dose data for 10 x 10 cm2 and 20 x 20 cm2 fields indicated that at 2 cm depth in water the method that takes compensator scatter into account agrees more closely with the DOSXYZnrc data compared to the data using only an effective attenuation coefficient. Further, it was found that the effective attenuation method can only replicate the DOSXYZnrc data at 10 cm depth where it was chosen to do so. At shallower depths the effective attenuation method overestimates the dose and beyond 10 cm depth it causes an underestimation in the dose. The scatter and beam hardening inclusion method does not exhibit such properties. The exclusion of scatter can lead to dose errors of up to 4 percent with a copper compensator at 5 cm depth for a 10 X 10 cm2 field under a thickness of 5 cm at 6 MV. For materials such as lead this discrepancy could be as high as 7 to 8 percent at 6 MV. For larger fields (20 X 20 cm2) the effect of in-phantom scatter reduces the differences between the dose profiles calculated with the mentioned methods.  相似文献   

设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号