1.
The purpose of this investigation is to study the feasibility of using an alternative method to commission stereotactic radiosurgery beams shaped by micro multi-leaf collimators by using Monte Carlo simulations to obtain beam characteristics of small photon beams, such as incident beam particle fluence and energy distributions, scatter ratios, depth-dose curves and dose profiles where measurements are impossible or difficult. Ionization chambers and diode detectors with different sensitive volumes were used in the measurements in a water phantom and the Monte Carlo codes BEAMnrc/DOSXYZnrc were used in the simulation. The Monte Carlo calculated data were benchmarked against measured data for photon beams with energies of 6 MV and 10 MV produced from a Varian Trilogy accelerator. The measured scatter ratios and cross-beam dose profiles for very small fields are shown to be not only dependent on the size of the sensitive volume of the detector used but also on the type of detectors. It is known that the response of some detectors changes at small field sizes. Excellent agreement was seen between scatter ratios measured with a small ion chamber and those calculated from Monte Carlo simulations. The values of scatter ratios, for field sizes from 6 x 6 mm2 to 98 x 98 mm2, range from 0.67 to 1.0 and from 0.59 to 1.0 for 6 and 10 MV, respectively. The Monte Carlo calculations predicted that the incident beam particle fluence is strongly affected by the X-Y-jaw openings, especially for small fields due to the finite size of the radiation source. Our measurement confirmed this prediction. This study demonstrates that Monte Carlo calculations not only provide accurate dose distributions for small fields where measurements are difficult but also provide additional beam characteristics that cannot be obtained from experimental methods. Detailed beam characteristics such as incident photon fluence distribution, energy spectra, including composition of primary and scattered photons, can be independently used in dose calculation models and to improve the accuracy of measurements with detectors with an energy-dependent response. Furthermore, when there are discrepancies between results measured with different detectors, the Monte Carlo calculated values can indicate the most correct result. The data set presented in this study can be used as a reference in commissioning stereotactic radiosurgery beams shaped by a BrainLAB m3 on a Varian 2100EX or 600C accelerator.  相似文献   

2.
Ding GX 《Medical physics》2002,29(11):2459-2463
This study investigates a possible cause of reported significant dose discrepancies between Monte Carlo calculations and measurements in the buildup region for high-energy photon beams in large fields. A proposed hypothesis was that the discrepancy was caused by a source of electrons in the accelerator head that was not fully accounted for in the treatment head simulation. In this investigation, a lead foil is added just below the accelerator head in order to study this hypothesis. The lead foil effectively removes charged particles generated inside the accelerator head. The charged particles generated by the lead foil can be accounted for fully because the simple geometry can be simulated accurately. An 18 MV photon beam from a Varian Clinac-2100EX is measured using a WELLHOFER WP700 beam scanner with an IC-10 ionization chamber (cavity radius=3 mm). The BEAM Monte Carlo code is used in the incident beam simulations. Both EGS4/DOSXYZ and EGSnrc/DOSRZnrc are used in the dose calculations in a water phantom. The Monte Carlo calculated depth-dose curve is scaled so that it has the same values at 10 cm depth as the measured curve. It is found that the discrepancies between Monte Carlo calculations and measurements remain significant in the buildup region even after applying necessary corrections to the measured data. The discrepancies have only been modestly decreased with the lead foil in place compared to the 40 x 40 cm2 open field. At a depth of 1 cm, discrepancies of about 5% are still observed in the buildup region for the field with the lead foil. Therefore a new explanation for the unresolved discrepancy remains to be found.  相似文献   

3.
This study presents measured neutron dose using a neutron dosimeter in a water phantom and investigates a hypothesis that neutrons in a high-energy photon beam may be responsible for the reported significant dose discrepancies between Monte Carlo calculations and measurements at the build-up region in large fields. Borated polyethylene slabs were inserted between the accelerator head and the phantom in order to remove neutrons generated in the accelerator head. The thickness of the slab ranged from 2.5 cm to 10 cm. A lead slab of 3 mm thickness was also used in the study. The superheated drop neutron dosimeter was used to measure the depth-dose curve of neutrons in a high-energy photon beam and to verify the effectiveness of the slab to remove these neutrons. Total dose measurements were performed in water using a WELLHOFER WP700 beam scanner with an IC-10 ionization chamber. The Monte Carlo code BEAM was used to simulate an 18 MV photon beam from a Varian Clinac-2100EX accelerator. Both EGS4/DOSXYZ and EGSnrc/DOSRZnrc were used in the dose calculations. Measured neutron dose equivalents as a function of depth per unit total dose in water were presented for 10 x 10 and 40 x 40 cm2 fields. The measured results have shown that a 5-10 cm thick borated polyethylene slab can reduce the neutron dose by a factor of 2 when inserted between the accelerator head and the detector. In all cases the measured neutron dose equivalent was less than 0.5% of the photon dose. In order to study if the ion chamber was highly sensitive to the neutron dose, we have investigated the disagreement between the Monte Carlo calculated and measured central-axis depth-dose curves in the build-up region when different shielding materials were used. The result indicated that the IC-10 chamber was not highly sensitive to the neutron dose. Therefore, neutrons present in a high-energy photon beam were unlikely to be responsible for the reported discrepancies in the build-up region for large fields.  相似文献   

4.
In some linear accelerators, the charge collected by the monitor ion chamber is partly caused by backscattered particles from accelerator components downstream from the chamber. This influences the output of the accelerator and also has to be taken into account when output factors are derived from Monte Carlo simulations. In this work, the contribution of backscattered particles to the monitor ion chamber response of a Varian 2100C linac was determined for photon beams (6, 10 MV) and for electron beams (6, 12, 20 MeV). The experimental procedure consisted of charge integration from the target in a photon beam or from the monitor ion chamber in electron beams. The Monte Carlo code EGS4/BEAM was used to study the contribution of backscattered particles to the dose deposited in the monitor ion chamber. Both measurements and simulations showed a linear increase in backscatter fraction with decreasing field size for photon and electron beams. For 6 MV and 10 MV photon beams, a 2-3% increase in backscatter was obtained for a 0.5 x 0.5 cm2 field compared to a 40 x 40 cm2 field. The results for the 6 MV beam were slightly higher than for the 10 MV beam. For electron beams (6, 12, 20 MeV), an increase of similar magnitude was obtained from measurements and simulations for 6 MeV electrons. For higher energy electron beams a smaller increase in backscatter fraction was found. The problem is of less importance for electron beams since large variations of field size for a single electron energy usually do not occur.  相似文献   

5.
A simple analytical approach has been developed to model extrafocal radiation and monitor chamber backscatter for clinical photon beams. Model parameters for both the extrafocal source and monitor chamber backscatter are determined simultaneously using conventional measured data, i.e., in-air output factors for square and rectangular fields defined by the photon jaws. The model has been applied to 6 MV and 15 MV photon beams produced by a Varian Clinac 2300C/D accelerator. Contributions to the in-air output factor from the extrafocal radiation and monitor chamber backscatter, as predicted by the model, are in good agreement with the measurements. The model can be used to calculate the in-air output factors analytically, with an accuracy of 0.2% for symmetric or asymmetric rectangular fields defined by jaws when the calculation point is at the isocenter and 0.5% when the calculation point is at an extended SSD. For MLC-defined fields, with the jaws at the recommended positions, calculated in-air output factors agree with the measured data to within 0.3% at the isocenter and 0.7% at off-axis positions. The model has been incorporated into a Monte Carlo dose algorithm to calculate the absolute dose distributions in patients or phantoms. For three MLC-defined irregular fields (triangle shape, C-shape, and L-shape), the calculations agree with the measurements to about 1% even for points at off-axis positions. The model will be particularly useful for IMRT dose calculations because it accurately predicts beam output and penumbra dose.  相似文献   

6.
A method for commissioning an EGSnrc Monte Carlo simulation of medical linac photon beams through wide-field lateral profiles at moderate depth in a water phantom is presented. Although depth-dose profiles are commonly used for nominal energy determination, our study shows that they are quite insensitive to energy changes below 0.3 MeV (0.6 MeV) for a 6 MV (15 MV) photon beam. Also, the depth-dose profile dependence on beam radius adds an additional uncertainty in their use for tuning nominal energy. Simulated 40 cm x 40 cm lateral profiles at 5 cm depth in a water phantom show greater sensitivity to both nominal energy and radius. Beam parameters could be determined by comparing only these curves with measured data.  相似文献   

7.
The purpose of this study was to investigate beam output factors (OFs) for conformal radiation therapy and to compare the OFs measured with different detectors with those simulated with Monte Carlo methods. Four different detectors (diode, diamond, pinpoint and ionization chamber) were used to measure photon beam OFs in a water phantom at a depth of 10 cm with a source-surface distance (SSD) of 100 cm. Square fields with widths ranging from 1 cm to 15 cm were observed; the OF for the different field sizes was normalized to that measured at a 5 cm x 5 cm field size at a depth of 10 cm. The BEAM/EGS4 program was used to simulate the exact geometry of a 6 MV photon beam generated by the linear accelerator, and the DOSXYZ-code was implemented to calculate the OFs for all field sizes. Two resolutions (0.1 cm and 0.5 cm voxel size) were chosen here. In addition, to model the detector four kinds of material, water, air, graphite or silicon, were placed in the corresponding voxels. Profiles and depth dose distributions resulting from the simulation show good agreement with the measurements. Deviations of less than 2% can be observed. The OF measured with different detectors in water vary by more than 35% for 1 cm x 1 cm fields. This result can also be found for the simulated OF with different voxel sizes and materials. For field sizes of at least 2 cm x 2 cm the deviations between all measurements and simulations are below 3%. This demonstrates that very small fields have a bad effect on dosimetric accuracy and precision. Finally, Monte Carlo methods can be significant in determining the OF for small fields.  相似文献   

8.
Stereotactic radiosurgery with several static conformal beams shaped by a micro multileaf collimator (microMLC) is used to treat small irregularly shaped brain lesions. Our goal is to perform Monte Carlo calculations of dose distributions for certain treatment plans as a verification tool. A dedicated microMLC component module for the BEAMnrc code was developed as part of this project and was incorporated in a model of the Varian CL2300 linear accelerator 6 MV photon beam. As an initial validation of the code, the leaf geometry was visualized by tracing particles through the component module and recording their position each time a leaf boundary was crossed. The leaf dimensions were measured and the leaf material density and interleaf air gap were chosen to match the simulated leaf leakage profiles with film measurements in a solid water phantom. A comparison between Monte Carlo calculations and measurements (diode, radiographic film) was performed for square and irregularly shaped fields incident on flat and homogeneous water phantoms. Results show that Monte Carlo calculations agree with measured dose distributions to within 2% and/or 1 mm except for field size smaller than 1.2 cm diameter where agreement is within 5% due to uncertainties in measured output factors.  相似文献   

9.
The purpose of this study is to provide detailed characteristics of incident photon beams for different field sizes and beam energies. This information is critical to the future development of accurate treatment planning systems. It also enhances our knowledge of radiotherapy photon beams. The EGS4 Monte Carlo code, BEAM, has been used to simulate 6 and 18 MV photon beams from a Varian Clinac-2100EX accelerator. A simulated realistic beam is stored in a phase space data file, which contains details of each particle's complete history including where it has been and where it has interacted. The phase space files are analysed to obtain energy spectra, angular distribution, fluence profile and mean energy profiles at the phantom surface for particles separated according to their charge and history. The accuracy of a simulated beam is validated by the excellent agreement between the Monte Carlo calculated and measured dose distributions. Measured depth-dose curves are obtained from depth-ionization curves by accounting for newly introduced chamber fluence corrections and the stopping-power ratios for realistic beams. The study presents calculated depth-dose components from different particles as well as calculated surface dose and contribution from different particles to surface dose across the field. It is shown that the increase of surface dose with the increase of the field size is mainly due to the increase of incident contaminant charged particles. At 6 MV, the incident charged particles contribute 7% to 21% of maximum dose at the surface when the field size increases from 10 x 10 to 40 x 40 cm2. At 18 MV, their contributions are up to 11% and 29% of maximum dose at the surface for 10 x 10 cm2 and 40 x 40 cm2 fields respectively. However, the fluence of these incident charged particles is less than 1% of incident photon fluence in all cases.  相似文献   

10.
Clinical implementation of a Monte Carlo treatment planning system.   总被引:4,自引:0,他引:4  
The purpose of this study was to implement the Monte Carlo method for clinical radiotherapy dose calculations. We used the EGS4/BEAM code to obtain the phase-space data for 6-20 MeV electron beams and 4, 6, and 15 MV photon beams for Varian Clinac 1800, 2100C, and 2300CD accelerators. A multiple-source model was used to reconstruct the phase-space data for both electron and photon beams, which retained the accuracy of the Monte Carlo beam data. The multiple-source model reduced the phase-space data storage requirement by a factor of 1000 and the accelerator simulation time by a factor of 10 or more. Agreement within 2% was achieved between the Monte Carlo calculations and measurements of the dose distributions in homogeneous and heterogeneous phantoms for various field sizes, source-surface distances, and beam modulations. The Monte Carlo calculated electron output factors were within 2% of the measured values for various treatment fields while the heterogeneity correction factors for various lung and bone phantoms were within 1% for photon beams and within 2% for electron beams. The EGS4/DOSXYZ Monte Carlo code was used for phantom and patient dose calculations. The results were compared to the dose distributions produced by a conventional treatment planning system and an intensity-modulated radiotherapy inverse-planning system. Significant differences (>5% in dose and >5 mm shift in isodose lines) were found between Monte Carlo calculations and the analytical calculations implemented in the commercial systems. Treatment sites showing the largest dose differences were for head and neck, lung, and breast cases.  相似文献   

11.
This work is intended to investigate the application and accuracy of micro-MOSFET for superficial dose measurement under clinically used MV x-ray beams. Dose response of micro-MOSFET in the build-up region and on surface under MV x-ray beams were measured and compared to Monte Carlo calculations. First, percentage-depth-doses were measured with micro-MOSFET under 6 and 10 MV beams of normal incidence onto a flat solid water phantom. Micro-MOSFET data were compared with the measurements from a parallel plate ionization chamber and Monte Carlo dose calculation in the build-up region. Then, percentage-depth-doses were measured for oblique beams at 0 degrees-80 degrees onto the flat solid water phantom with micro-MOSFET placed at depths of 2 cm, 1 cm, and 2 mm below the surface. Measurements were compared to Monte Carlo calculations under these settings. Finally, measurements were performed with micro-MOSFET embedded in the first 1 mm layer of bolus placed on a flat phantom and a curved phantom of semi-cylindrical shape. Results were compared to superficial dose calculated from Monte Carlo for a 2 mm thin layer that extends from the surface to a depth of 2 mm. Results were (1) Comparison of measurements with MC calculation in the build-up region showed that micro-MOSFET has a water-equivalence thickness (WET) of 0.87 mm for 6 MV beam and 0.99 mm for 10 MV beam from the flat side, and a WET of 0.72 mm for 6 MV beam and 0.76 mm for 10 MV beam from the epoxy side. (2) For normal beam incidences, percentage depth dose agree within 3%-5% among micro-MOSFET measurements, parallel-plate ionization chamber measurements, and MC calculations. (3) For oblique incidence on the flat phantom with micro-MOSFET placed at depths of 2 cm, 1 cm, and 2 mm, measurements were consistent with MC calculations within a typical uncertainty of 3%-5%. (4) For oblique incidence on the flat phantom and a curved-surface phantom, measurements with micro-MOSFET placed at 1.0 mm agrees with the MC calculation within 6%, including uncertainties of micro-MOSFET measurements of 2%-3% (1 standard deviation), MOSFET angular dependence of 3.0%-3.5%, and 1%-2% systematical error due to phantom setup geometry asymmetry. Micro-MOSFET can be used for skin dose measurements in 6 and 10 MV beams with an estimated accuracy of +/- 6%.  相似文献   

12.
This study presents a theoretical approach to the dosimetry for small and non-equilibrium radiation fields. We applied the newly developed VMCBC algorithm to the dosimetry for megavoltage photon beams using Monte Carlo techniques. The approach assumes that a Monte Carlo simulated beam can be calibrated per incident particle at the target in an x-ray tube or in an accelerator head. Since the geometry of the accelerator head and beam defining systems can be modeled in detail, the output of a radiation beam can be accurately related to the number of incident particles through particle transport calculations. The proposed methodology is benchmarked and validated using existing radiosurgery beam commissioning data, which were experimentally measured for narrow beams defined by conical collimators with diameters ranging from 7.5 mm to 30 mm. The Monte Carlo predicted beam outputs agree with the measurement values within the uncertainty of the experiments. The Monte Carlo approach developed and introduced in this study allows the user to perform absolute radiation dosimetry in addition to relative dose distributions at locations where charged-particle equilibrium (CPE) does not exist, such as radiation dose from a narrow stereotactic radiosurgery beam, and where experimental measurements are difficult. The BEAMnrc/DOSXYZnrc code was employed in the Monte Carlo simulations.  相似文献   

13.
Shih R  Li XA  Chu JC 《Medical physics》2001,28(4):612-619
The purpose of this study is to analyze the characteristics of dynamic wedges (DW) and to compare DW to physical wedges (PW) in terms of their differences in affecting beam spectra, energy fluence, angular distribution, contaminated electrons, and dose distributions. The EGS4/BEAM Monte Carlo codes were used to simulate the exact geometry of a 6 MV beam and to calculate 3-D dose distributions in phantom. The DW was simulated in accordance with the segmented treatment tables (STT). The percentage depth dose curves and beam profiles for PW, DW, and open fields were measured and used to verify the Monte Carlo simulations. The Monte Carlo results were found to agree within 2% with the measurements performed using film and ionizing chambers in a water phantom. The present EGS4 calculation reveals that the effects of a DW on beam spectral and angular distributions, as well as electron contamination, are much less significant than those for a PW. For the 6 MV photon beam, a 45 degrees PW can result in a 30% increase in mean photon energy due to the effect of beam hardening. It can also introduce a 5% dose reduction in the build-up region due to the reduction of contaminated electrons by the PW. Neither this mean-energy increase nor such dose reduction is found for a DW. Compared to a DW, a PW alters the photon-beam spectrum significantly. The dosimetric differences between a DW and a PW are significant and clearly affect the clinical use of these beams. The data presented may be useful for DW commissioning.  相似文献   

14.
The effect of beam obliquity on the surface relative dose profiles for the tangential photon beams was studied. The 6 and 15 MV photon beams with 4 x 4 and 10 x 10 cm2 field sizes produced by a Varian 21 EX linear accelerator were used. Phase-space models of the photon beams were created using Monte Carlo simulations based on the EGSnrc code, and were verified using film measurements. The relative dose profiles in the phantom skin, at 2 mm depth from the surface of the half-phantom geometry, or HPG, were calculated for increasing gantry angles from 270 to 280 deg clockwise. Relative dose profiles of a full phantom enclosing the whole tangential beam (full phantom geometry, or FPG) were also calculated using Monte Carlo simulation as a control for comparison. The results showed that, although the relative dose profiles in the phantom skin did not change significantly with an oblique beam using a FPG, the surface relative depth dose was increased for the HPG. In the HPG, with 6 MV photon beams and field size = 10 x 10 cm2, when the beam angle, starting from 270 deg, was increased from 1 to 3 deg, the relative depth doses in the phantom skin were increased from 68% to 79% at 10 cm depth. This increase in dose was slightly larger than the dose from 15 MV photon beams with the same field size and beam angles, where the relative depth doses in phantom skin were increased from 81% to 87% at 10 cm depth. A parameter called the percent depth dose (PDD) ratio, defined as the relative depth dose from the HPG to the relative depth dose from the FPG at a given depth along the phantom skin, was used to evaluate the effect of the phantom-air interface. It is found that the PDD ratio increased significantly when the beam angle was changed from zero to 1-3 degrees. Moreover, the PDD ratio, for a given field size, experienced a greater increase for 6 MV than for 15 MV. For the same photon beam energy, the PDD ratio increased more with a 4 x 4 cm2 field compared to 10 x 10 cm2. The results in this study will be useful for physicists and dosimetrists to predict the surface relative dose variations when using clinical tangential-like photon beams in radiation therapy.  相似文献   

15.
16.
This paper investigates a quality assurance (QA) phantom specially designed to verify the accuracy of dose distributions and monitor units (MU) calculated by clinical treatment planning optimization systems and by the Monte Carlo method for intensity-modulated radiotherapy (IMRT). The QA phantom is a PMMA cylinder of 30 cm diameter and 40 cm length with various bone and lung inserts. A procedure (and formalism) has been developed to measure the absolute dose to water in the PMMA phantom. Another cylindrical phantom of the same dimensions, but made of water, was used to confirm the results obtained with the PMMA phantom. The PMMA phantom was irradiated by 4, 6 and 15 MV photon beams and the dose was measured using an ionization chamber and compared to the results calculated by a commercial inverse planning system (CORVUS, NOMOS, Sewickley, PA) and by the Monte Carlo method. The results show that the dose distributions calculated by both CORVUS and Monte Carlo agreed to within 2% of dose maximum with measured results in the uniform PMMA phantom for both open and intensity-modulated fields. Similar agreement was obtained between Monte Carlo calculations and measured results with the bone and lung heterogeneity inside the PMMA phantom while the CORVUS results were 4% different. The QA phantom has been integrated as a routine QA procedure for the patient's IMRT dose verification at Stanford since 1999.  相似文献   

17.
Volume averaging and lack of electronic equilibrium complicate accurate dosimetry of small photon fields. In this paper the performance of the PinPoint ion chamber for characterizing small fields used in intensity-modulated radiotherapy (IMRT) was investigated and the results were compared with those obtained using the Markus ion chamber and a diamond detector. Sharp beam penumbras were measured for a 5 x 5 cm field defined using a cerrobend block mounted on the accelerator head. In addition, output factors were measured for a 6 MV photon beam and a variety of small rectangular fields collimated widthwise using the multileaf collimator (MLC) in combination with the back-up jaws. From this study, a reference field of 5 x 5 cm and a measuring depth of 5 cm are recommended. This is related to the over-response of the PinPoint chamber to low-energy Compton scattered photons, an effect that was investigated rigorously and turned out to limit the scope of this ionization chamber. However, taking into account some limitations, the PinPoint chamber is an excellent detector for output measurements in small fields down to 2 cm. In profile measurements the chamber causes a broadening of the measured penumbras but its spatial resolution is superior to that of the Markus chamber.  相似文献   

18.
Reference dosimetry of photon fields is a well-established subject and currently available protocols (such as the IAEA TRS-398 and AAPM TG-51) provide methods for converting the ionization chamber (IC) reading into dose to water, provided reference conditions of charged particle equilibrium (CPE) are fulfilled. But these protocols cannot deal with the build-up region, where the lack of CPE limits the applicability of the cavity theorems and so the chamber correction factors become depth dependent. By explicitly including the IC geometry in the Monte Carlo simulations, depth-dependent dose correction factors are calculated for a PTW 30001 0.6 cm(3) ion chamber in the build-up region of the 6 MV photon beam. The corrected percentage depth dose (PDD) agrees within 2% with that measured using the NACP 02 plane-parallel ion chamber in the build-up region at depths greater than 0.4 cm, where the Farmer chamber wall reaches the phantom surface.  相似文献   

19.
Pre-existing methods for photon beam spectral reconstruction are briefly reviewed. An alternative reconstruction method by scatter analysis for linear accelerators is introduced. The method consists in irradiating a small plastic phantom at standard 100 cm SSD and inferring primary beam energy spectral information based on the measurement with a standard Farmer chamber of scatter around the phantom at several specific scatter angles: a scatter curve is measured which is indicative of the primary spectrum at hand. A Monte Carlo code is used to simulate the scatter measurement set-up and predict the relative magnitude of scatter measurements for mono-energetic primary beams. Based on mono-energetic primary scatter data, measured scatter curves are analysed and the spectrum unfolded as the sum of mono-energetic individual energy bins using the Schiff bremsstrahlung model. The method is applied to an Elekta/SL18 6 MV photon beam. The reconstructed spectrum matches the Monte Carlo calculated spectrum for the same beam within 6.2% (average error when spectra are compared bin by bin). Depth dose values calculated for the reconstructed spectrum agree with physically measured depth dose data to within 1%. Scatter analysis is preliminarily shown to have potential as a practical spectral reconstruction method requiring few measurements under standard 100 cm SSD and feasible in any radiotherapy department using a phantom and a Farmer chamber.  相似文献   

20.
A Monte Carlo study on internal wedges using BEAM   总被引:1,自引:0,他引:1  
  相似文献   