1岁以下婴儿ICare手持回弹式眼压计眼压正常值及RetCam检查对其影响的研究
引用本文:郑姣 穆歌 赵欢欢 高彩凤 张振瑜 徐昊立. 1岁以下婴儿ICare手持回弹式眼压计眼压正常值及RetCam检查对其影响的研究[J]. 眼科, 2022, 31(5): 345-348. DOI: 10.13281/j.cnki.issn.1004-4469.2022.05.005
作者姓名:郑姣 穆歌 赵欢欢 高彩凤 张振瑜 徐昊立
作者单位:广东省妇幼保健院眼科,广州 511400
基金项目:广东省医学科学技术研究基金项目(A2020421)
摘    要:目的 分析1岁以下婴儿正常眼压值及眼底广域数字成像系统(RetCam3)视网膜筛查对眼压的影响。设计前瞻性病例系列。研究对象2019~2020年于广东省妇幼保健院行筛查的≤1岁婴儿285例(570眼),其中男性176例。选择同期在我院门诊就诊且眼压正常的成年人102例(204眼)作为对照。方法使用ICare手持回弹式眼压计测量婴儿RetCam3筛查前后的眼压值,比较筛查前后眼压变化及其与成年人的异同。主要指标眼压。结果婴儿双眼平均眼压(13.04±3.04) mmHg,左、右眼眼压差异无统计学意义(t=-1.68,P=0.10)。成年人双眼平均眼压(16.82±2.36)mmHg,婴儿眼压较成年人低(t=-18.14,P<0.01)。不同性别婴儿的眼压比较无统计学意义(t=0.20,P=0.85)。行RetCam3眼底检查后眼压(13.93±3.20) mmHg较检查前眼压(12.88±3.12) mmHg升高(t=5.09,P<0.01),出生时胎龄<32周组眼底检查前后眼压的差值(1.64±3.76)mmHg高于≥32周组(0.75±3.22)mmHg(t=2.1...

关 键 词:婴儿  眼压  RetCam
收稿时间:2022-03-15

Study on the normal intraocular pressure in infants under 1 year old by ICare hand-held rebound tonometer and the effect of retinal examination with RetCam on it
Zheng Jiao,Mu Ge,Zhao Huanhuan,Gao Caifeng,Zhang Zhenyu,Xu Haoli. Study on the normal intraocular pressure in infants under 1 year old by ICare hand-held rebound tonometer and the effect of retinal examination with RetCam on it[J]. Ophthalmology in China, 2022, 31(5): 345-348. DOI: 10.13281/j.cnki.issn.1004-4469.2022.05.005
Authors:Zheng Jiao  Mu Ge  Zhao Huanhuan  Gao Caifeng  Zhang Zhenyu  Xu Haoli
Affiliation:Department of Ophthalmology, Guangdong Women and Children Hospital, Guangzhou 511400, China
Abstract:Objective To observ the normal intraocular pressure (IOP) of infants under 1 year old and to analyze the effect of retinal examination with wide field digital fundus imaging system (RetCam3) on their IOP. Design Prospective case series. Participants 285 cases (570 eyes) of ≤1 year old infants born at 24~40 weeks of pregnancy, wherein 176 were males, and 102 adult cases (204 eyes) with normal IOP, in Guangdong Women and Children Hospital between 2019 to 2020. Methods IOPs before and after RetCam3 retinal examinations in infacts as well as adults were measured by the ICare hand-held rebound tonometer. The differences of IOP between before and after RetCam3 retinal examinations and between infacts and adults were compared. Main Outcome Measures IOPs. Results The mean IOP of both eyes of infants was (13.04±3.04) mmHg, there was no significant difference between left and right IOP (t=-1.68,P=0.10). The both eyes mean IOP of adults was (16.82±2.36) mmHg. The IOP of infants was lower than that of adults (t=-18.14,P<0.01). There was no statistical difference of IOP between different genders for infants (t=0.20,P=0.85). The IOP after RetCam3 examination was (13.93±3.20) mmHg, which was higher than that before examination (12.88±3.12 mmHg) (t=5.09,P<0.01). The difference of IOP before and after fundus examination was more higher in <32 weeks group (1.64±3.76 mmHg) than ≥32 weeks group (0.75±3.22 mmHg) (t=2.12,P=0.04).  Conclusion The mean IOP of infants ≤1 year old is lower than adults. RetCam3 retinal examinationcan cause mild IOP increase in infants, and the effect is more obvious in the infants with smaller gestational age. (Ophthalmol CHN, 2022, 31: 345-348)
Keywords:infants  intraocular pressure  RetCam  