首页 | 本学科首页   官方微博 | 高级检索  
     


Evaluation of an automated immunodiagnostic assay, VIDAS Rotavirus, for detection of rotavirus in fecal specimens.
Authors:P H Dennehy   T E Schutzbank     G M Thorne
Abstract:Four hundred sixty-four fecal specimens were tested for rotavirus by three immunoassays, VIDAS Rotavirus (bioMérieux Vitek, Hazelwood, Mo.), Rotaclone (Cambridge Biotech Corporation, Worcester, Mass.), and Pathfinder Rotavirus (Sansfi Diagnostics Pasteur, Chaska, Minn.). Twenty-seven discordant specimens were evaluated by electron microscopy. The sensitivities, specificities, positive predictive values, and negative predictive values were 98, 99.3, 98.7, and 99%, respectively, for VIDAS Rotavirus; 100, 99, 98.1, and 100%, respectively, for Rotaclone; and 100, 92.4, 87.3, and 100%, respectively for Pathfinder Rotavirus.
Keywords:
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号