首页 | 本学科首页   官方微博 | 高级检索  
     

TomoTherapy HI-ART加速器剂量输出稳定性检测与评价
引用本文:管秋,杨波,王欣海,李文博,李楠,邱杰. TomoTherapy HI-ART加速器剂量输出稳定性检测与评价[J]. 协和医学杂志, 2013, 4(4): 404-407. DOI: 10.3969/j.issn.1674-9081.2013.04.012
作者姓名:管秋  杨波  王欣海  李文博  李楠  邱杰
作者单位:中国医学科学院 北京协和医学院 北京协和医院放射治疗科, 北京 100730
摘    要:  目的  对TomoTherapy HI-ART加速器剂量输出的稳定性进行检测和评价。  方法  利用A1SL电离室、矩形固体水模体与圆柱形固体水模体, 参照美国医学物理学会TG-148号报告, 对TomoTherapy HI-ART加速器质控规程中静态剂量输出、动态剂量输出、射线质稳定性进行检测, 并对结果进行分析。  结果  静态剂量输出结果偏差平均值为(-0.6±1.1)%, 有89.1%的数据在±2%的允许偏差范围之内, 数据分布无明显规律。动态剂量输出结果偏差平均值为(1.0±1.1)%, 有87.1%的数据在±2%的允许偏差范围之内。射线质较为稳定, 偏差平均值为(-0.01±1.26)%, 80%的结果在±1%的偏差范围之内, 96%的结果在±2%的允许偏差范围之内。  结论  TomoTherapy HI-ART加速器剂量输出稳定性略低于常规加速器, 剂量输出易发生波动, 因此每日对其输出量进行系统检测十分重要。

关 键 词:螺旋断层调强放疗   剂量输出   质量控制
收稿时间:2013-08-15

Quality Assurance Procedure for the Output Constancy of TomoTherapy HI-ART
GUAN Qiu,YANG Bo,WANG Xin-hai,LI Wen-bo,LI Nan,QIU Jie. Quality Assurance Procedure for the Output Constancy of TomoTherapy HI-ART[J]. Medical Journal of Peking Union Medical College Hospit, 2013, 4(4): 404-407. DOI: 10.3969/j.issn.1674-9081.2013.04.012
Authors:GUAN Qiu  YANG Bo  WANG Xin-hai  LI Wen-bo  LI Nan  QIU Jie
Affiliation:Department of Radiation Oncology, Peking Union Medical College Hospital, Chinese Academy of Medical Sciences & Peking Union Medical College, Beijing 100730, China
Abstract:Objective To test and assess the output constancy of TomoTherapy HI-ART. Methods Ac- cording to TG-148 of American Association of Physicists in Medicine (AAPM) report and by using A1SL ionization chamber, slab solid water phantom, and cheese phantom, we tested and analyzed the constancy of static output, ro- tational output, and energy of TomoTherapy HI-ART. Results The average output variation was ( - 0. 6 ± 1.1 ) % for the static condition, and 89. 1% of the collected output data were within the action level of ±2%. The aver- age variation of rotational output was ( 1.0 ± 1.1 ) % , and 87. 1% of the collected data were within ± 2%. An average energy variation was ( - 0. 01 ±1.26) %, and 80% data were within + 1% and 96% data within ± 2%, which showed good constancy and reproducibility. Conclusions The output constancy of TomoTherapy HI-ART is less stable than conventional Linac, so it is important to test static and rotational output everyday to ensure the accuracy.
Keywords:helical tomotherapy  output  quality assurance
本文献已被 维普 等数据库收录!
点击此处可从《协和医学杂志》浏览原始摘要信息
点击此处可从《协和医学杂志》下载免费的PDF全文
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号