1.5TMR乳腺扩散加权成像b值的优化
引用本文:张静,安宁豫,程流泉,蔡幼铨,张爱莲,李席如,郭勇.1.5TMR乳腺扩散加权成像b值的优化[J].中华放射学杂志,2011,45(10).
作者姓名:张静  安宁豫  程流泉  蔡幼铨  张爱莲  李席如  郭勇
作者单位:1. 海军总医院放射科 100048
2. 解放军总医院放射科, 北京,100853
摘    要:目的 通过分析水模、正常乳腺腺体、乳腺良性及恶性病变的ADC值及图像信噪比(SNR)随b值的变化规律,探讨1.5 TMR乳腺DWI合理的b值取值范围.方法 对32例经病理证实的乳腺病变(恶性18例,良性14例)及对侧正常腺体进行乳腺MR检查,采用EPI-DWI序列;b值分别采用0、50、100、200、400、600、800、1000、1200、1400、1600、1800、2000、2200、2400、2600 s/mm2.测量不同b值下水模、正常乳腺腺体、乳腺良性及恶性病变的平均ADC值和图像SNR,采用Pearson相关分析法分析不同b值时的变化规律.结果 DWI的SNR均随b值的增加逐渐下降,二者呈负相关(r=-0.802,P<0.01),乳腺良、恶性病变的ADC值均随着b值的增加而下降(r=-0.923和-0.855,P<0.01);当b值取800~1000 s/mm2时,恶性病变与良性病变和正常腺体之间的ADC值差异最大(0.7×10-3mm2/s);当b值>1400 s/mm2,差异逐渐减小.结论 取b值800~ 1000 s/mm2时,既能取得良好的图像质量,又能有效地鉴别乳腺良、恶性病变,是1.5 TMR乳腺DWI最合理的b值取值范围.

关 键 词:乳腺肿瘤  磁共振成像  扩散

Optimization of b-value in breast diffusion-weighted 1.5 T MR imaging ZHANG Jing
ZHANG Jing,AN Ning-yu,CHENG Liu-quan,CAI You-quan,ZHANG Ai-lian,LI Xi-ru,GUO Yong.Optimization of b-value in breast diffusion-weighted 1.5 T MR imaging ZHANG Jing[J].Chinese Journal of Radiology,2011,45(10).
Authors:ZHANG Jing  AN Ning-yu  CHENG Liu-quan  CAI You-quan  ZHANG Ai-lian  LI Xi-ru  GUO Yong
Abstract:Objective To optimize the b-value of breast diffusion-weighted MRI (DW-MRI) at 1.5T by applying a range of b values and comparing the apparent diffusion coefficient (ADC) and signal-to-noise ratio (SNR) on a phantom,disease-free breast tissues,and benign and malignant lesions.Methods A phantom and 32 women with pathologically confirmed malignant ( 18 ) and benign ( 14 ) lesions were examined using EPI-DWI with different b values on a 1.5 T MR scanner.The b-value of EPI-DWI was 0,50,100,200,400,600,800,1000,1200,1400,1600,1800,2000,2200,2400,and 2600 s/mm2,respectively.The SNR and ADC values of the phantom,disease-free breast tissues,and benign and malignant lesions were measured.The correlation between the b-value and ADC or SNR of each image was analyzed.Results The SNR of DWI decreased as the b-value increased,showing an inverse correlation (r =-0.802,P <0.01 ).The ADC values of benign and malignant lesion decreased as the b-value increased (r =-0.923 and -0.855,P <0.01 ).The maximum difference in ADC between malignant and benign lesions was observed when the b-value is between 800 and 1000 s/mm2 and diminished when the b-value was greater than 1400 s/mm2.Conclusion For good image quality and valid differentiation between malignant and benign lesions,the optimized b-value of DWI at 1.5 T is between 800 s/mm2 and 1000 s/mm2.
Keywords:Breast neoplasm  Diffusion magnetic resonance imaging