首页 | 本学科首页   官方微博 | 高级检索  
检索        

Sirius和Lenstar及A超测量中央角膜厚度的对比研究
引用本文:黄文志,王露,高宗银,杨为中.Sirius和Lenstar及A超测量中央角膜厚度的对比研究[J].广州医药,2014,45(6):14-17.
作者姓名:黄文志  王露  高宗银  杨为中
作者单位:广州市第一人民医院眼科,510180
基金项目:广州市医药卫生科技项目
摘    要:目的 比较Sirius、Lenstar及A超测量中央角膜厚度(CCT)的差异,并对测量结果进行一致性评价.方法 对52例(104眼)近视眼患者,依次采用Sirius、Lenstar及A超测量CCT.配对t检验和Pearson相关分析评价CCT测量值两两间的差异及相关关系,Bland-Altman分析评价一致性.结果 Sirius与A超、Lenstar与A超的CCT测量值之间差异均有统计学意义,而Sirius与Lenstar的CCT测量值比较差异无统计学意义.三种仪器测量CCT具有显著的正相关性.CCT测量值95%一致性界限分别为Sirius与A超(-16.7,8.0)μm、Lenstar与A超(-19.7,8.5)μm、Sirius与Lenstar(-12.8,15.3)μm.结论 Sirius、Lenstar与A超测量CCT之间具有较高的一致性,可作为角膜中央厚度测量的常规工具.Sirius和Lenstar测量值与A超相比,存在差异,前二者低估了中央角膜厚度.

关 键 词:A超  中央角膜厚度

Comparative study of central corneal thickness measurements with Sirius,Lenstar and Ultrasound pachymetry
Institution:Huang Wenzhi, Wang Lu, Gao Zongyin, et al. (Department of Ophthalmology, Guangzhou First People's Hospital. Guangzhou 510180, China)
Abstract:Objective To assess the differences and agreements among measurements of central corneal thickness (CCT) using Sirius, Lenstar and ultrasound pachymetry (USP) . Methods The CCT of 104 eyes from 52 patients with myopia were measured with Sirius, Lenstar and ultrasound pachymetry. Data were compared by using paired t-tests and the Pearson correlations. Bland-Altman analysis of all pairs were determined. Results The difference between Sirius and USP, Lenstar and USP were statistically significant (P 〈 0. 001 ), but the CCT measured by Sirius and Lenstar had no statistically significant difference (P 〉 0.05 ) . There was high correlation between the three instruments. The 95% limits of agreement for central cor- neal thickness measurements were - 16. 7μm to + 8. 0μm between Sirius and USP, - 19.7μm to + 8. 5μm between Lenstar and USP, and - 12. 8μm to + 15.3μm between Sirius and Lenstar. Conclusion Central corneal thickness obtained with the USP were similar to those obtained with both the Sirius and Lenstar. They all can be used as a routine tool of central corneal thickness measurement. But compared with USP, Sirius and Lenstar may underestimate the central corneal thickness.
Keywords:Sirius  Lenstar  Ultrasound pachymetry  Central corneal thickness
本文献已被 维普 万方数据 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号