不同形状超声乳化针头在扭动超声模式的应用
引用本文:郭作锋,史庆成,周衍文.不同形状超声乳化针头在扭动超声模式的应用[J].国际眼科杂志,2015,15(8):1441-1443.
作者姓名:郭作锋  史庆成  周衍文
作者单位:沈阳爱尔眼视光医院白内障科, 中国辽宁省沈阳市,110003
摘    要:目的:通过与Infinity超声乳化针头对比,观察Centurion双重弯曲形状设计的超声乳化针头的临床效果。
  方法:前瞻性对照研究,将年龄相关性白内障患者132例132眼,随机分为两组,试验组66例66眼采用Centurion超声乳化针头,对照组66例66眼采用Infinity超声乳化针头,两组均用Centurion超声乳化仪在同一手柄下,利用相同参数下完成常规的超声乳化联合人工晶状体植入术。记录两组术中的累积消耗能量( cumulative dissipated energy,CDE),观察术后1wk,1mo 的最佳矫正视力( best corrected visual acuity, BCVA )、角膜内皮密度( corneal endothelial cell density,CD)、中央角膜厚度( central corneal thickness,CCT)。采用独立样本t检验和χ2检验进行统计分析。
  结果:两组患者的CDE差异有统计学意义( P<0.05)。术后1wk时BCVA( LogMAR)试验组优于对照组,差异有统计学意义( P<0.05)。术后1 wk试验组CCT小于对照组,差异有统计学意义(P<0.01),但术后1mo两组无统计学差异(P>0.05)。
  结论:Centurion超声乳化针头能有效减少术中超声能量的释放,减少术后角膜内皮的损伤,对术后早期视力恢复更具有优势。

关 键 词:超声乳化白内障吸除术  白内障  累计消耗能量  超声乳化针头
收稿时间:2015/2/28 0:00:00
修稿时间:2015/7/15 0:00:00

Application of different shape tips in torsional microcoaxial phacoemulsification
Zuo-Feng Guo,Qing-Cheng Shi and Yan-Wen Zhou.Application of different shape tips in torsional microcoaxial phacoemulsification[J].International Journal of Ophthalmology,2015,15(8):1441-1443.
Authors:Zuo-Feng Guo  Qing-Cheng Shi and Yan-Wen Zhou
Institution:Department of Cataract,Shenyang Aier Eye Hospital, Shenyang 110003, Liaoning Province, China;Department of Cataract,Shenyang Aier Eye Hospital, Shenyang 110003, Liaoning Province, China;Department of Cataract,Shenyang Aier Eye Hospital, Shenyang 110003, Liaoning Province, China
Abstract:AIM: To compare the clinical outcomes of torsional microcoaxial phacoemulsification by using different shape tips, including Centurion ultrasound tips and Infinity tips.

METHODS:A randomized prospective study was conducted. 132 patients(132 eyes)with age-related cataract, were randomly divided into study group(Centurion ultrasound tips)and control group(Infinity ultrasound tips), 66 cases(66 eyes)in each group. Under the same parameter of the Centurion phacoemulsification instrument, all the patients received normal microcoaxial phacoemulsification and intraocular lens implantation. The cumulative dissipated energy(CDE)was recorded during the operation. The best corrected visual acuity(BCVA), corneal endothelial cell density(CD)and central corneal thickness(CCT)were compared at 1wk and 1mo after surgery. Statistical analysis of the data was performed using Student's t test and Chi-square test.

RESULTS: There was significant difference on CDE(P<0.05)between these two groups. On 1wk after surgery, the study group showed better BCVA(LogMAR)as compared that of control group, the difference was statistically significant(P<0.05). The CCT of study group was smaller than the control group on 1wk after surgery(P<0.01), but there was no significant difference on 1mo(P>0.05).

CONCLUSION: Centurion ultrasound tips can significantly reduce the cumulative dissipated energy and cause less damage to corneal endothelial cells. These findings suggest that centurion vision system phacoemulsification surgery could get earlier visual rehabilitation postoperative.

Keywords:phacoemulsification  cataract  cumulative dissipated energy  ultrasound tips