Pentacam三维眼前节分析诊断系统 与Keratron Scout角膜像差仪测量 Kappa角的比较
引用本文:师蓉,赵炜,董泽红,孙丽娟,陈雪婷,董丽蓉,冯洁,王雨生.Pentacam三维眼前节分析诊断系统 与Keratron Scout角膜像差仪测量 Kappa角的比较[J].中华眼视光学与视觉科学杂志,2019,21(11):826-830.
作者姓名:师蓉  赵炜  董泽红  孙丽娟  陈雪婷  董丽蓉  冯洁  王雨生
作者单位:Rong Shi, Wei Zhao, Zehong Dong, Lijuan Sun, Xueting Chen, Lirong Dong, Jie Feng, Yusheng Wang
摘    要:目的:比较Pentacam三维眼前节分析诊断系统与Keratron Scout角膜像差仪测量屈光手术患者Kappa 角大小的差异。方法:系列病例研究。选取空军军医大学西京医院眼科屈光矫正中心术前筛查患 者98例(196眼)。在同一环境下进行Pentacam与Keratron Scout检查,记录2种检查方式的瞳孔直径、 Kappa角大小(采用直角坐标系表示),从而获得Kappa角在角膜的x轴和y轴偏移量,比较2种检查方 法测量的Kappa角大小的差异及2种检查方法的相关性。采用配对t检验以及Pearson相关分析进行 数据处理。结果:Pentacam与Keratron Scout测量近视患者的瞳孔直径分别为右眼(3.17±0.44)mm、 (3.19±0.53)mm,左眼(3.25±0.65)mm、(3.33±0.48)mm,二者差异均无统计学意义。Pentacam与 Keratron Scout测得双眼的Kappa角x轴以及y轴偏移量的差异均无统计学意义,2种方法测得的Kappa 角偏移量具有显著相关性(P<0.05),右眼x轴相关系数为0.966,y轴为0.969,左眼x轴为0.956,y轴 为0.913。结论:在瞳孔直径一致的条件下,Pentacam三维眼前节分析诊断系统与Keratron Scout角膜 像差仪测量近视以及近视散光患者Kappa角无差异,在手术设计中均可作为Kappa角调整的参考。

关 键 词:Pentacam  Keratron  Scout  瞳孔直径  kappa角  
收稿时间:2019-02-18

Comparison of the Angle Kappa Measured by Pentacam and Keratron Scout
Rong Shi,Wei Zhao,Zehong Dong,Lijuan Sun,Xueting Chen,Lirong Dong,Jie Feng,Yusheng Wang.Comparison of the Angle Kappa Measured by Pentacam and Keratron Scout[J].Chinese Journal of Optometry Ophthalmology and Visual Science,2019,21(11):826-830.
Authors:Rong Shi  Wei Zhao  Zehong Dong  Lijuan Sun  Xueting Chen  Lirong Dong  Jie Feng  Yusheng Wang
Institution:Department of Ophthalmology, Xijing Hospital, Institute of Ophthalmology of Chinese PLA, Air Force  Medical University, Xi'an 710032, China
Abstract:Objective: To investigate and analyze the difference in angle Kappa measured by Pentacam and Keratron Scout in refractive surgery patients. Methods: One hundred ninety-six eyes of 98 patients screened for refractive surgery were enrolled in this case series study. Pupil diameters and angle Kappa expressed in rectangular coordinates were measured by Pentacam and Keratron Scout. Data differences and correlation between the two examinations were compared and analyzed by a paired t test and Pearson correlation analysis. Results: Pupil diameters measured by Pentacam and Keratron Scout were 3.17±0.44 mm, 3.19±0.53 mm (right eye), and 3.25±0.65 mm, 3.33±0.48 mm (left eye), respectively. The differencesbetween the two instruments were not significant. The horizontal and vertical offset measured by the two instruments were not statistically different. There was a significant correlation in the offset of the angle Kappa measured by the two instruments (P<0.05). The correlation coefficients of the right eye horizontal offset, right eye vertical offset, left eye horizontal offset, left eye vertical offset were 0.966, 0.969, 0.956 and 0.913, respectively. Conclusion: There is no statistically significant difference between angle Kappa measured by Pentacam and Keratron Scout when pupil diameters are consistent. Both instruments can be used as a reference for the adjustment in angle Kappa in refractive surgery.
Keywords:Pentacam  Keratron Scout  pupil diameter  angle kappa  