首页 | 本学科首页   官方微博 | 高级检索  
检索        


A thermoluminescent dosimetry postal dose inter-comparison of radiation therapy centres in Malaysia
Authors:Email author" target="_blank">P?RassiahEmail author  K?H?Ng  L?A?DeWerd  K?Kunugi
Institution:Clinical Oncology Unit, University of Malaya Medical Centre, Kuala Lumpur, Malaysia. rassiah@uthscsa.edu
Abstract:A thermoluminescent dosimetry (TLD) postal dose inter-comparison was carried out amongst radiotherapy centres in Malaysia. The aim of this TLD inter-comparison was to compare the uniformity involved in the measurement of absorbed dose among the participating centres. A set of 5 TLD chips placed within acrylic trays were mailed to all participating centres for irradiation to an absorbed dose to water of 2 Gy. Measurements were made for 6 MV and 60Co photon beams. Results show an agreement of +/- 5% for all but three radiotherapy centres. The ratios of the TLD readings to that of the reference centre are comparable with other national/regional dose inter-comparisons. The importance of a proper ongoing quality assurance program is essential in maintaining the consistency and uniformity of doses delivered.
Keywords:
本文献已被 PubMed SpringerLink 等数据库收录!