高场强磁共振乳腺成像扫描技术的优化
引用本文:陈永刚,徐凯,赵洪兰.高场强磁共振乳腺成像扫描技术的优化[J].现代保健,2009(9):1-3.
作者姓名:陈永刚  徐凯  赵洪兰
作者单位:[1]安徽省淮北市人民医院,235000 [2]徐州医学院附属医院,235000
摘    要:目的评价不同扫描序列及参数在乳腺MR成像中的应用价值。方法对42名乳腺疾病患者采用FSE—IR和FSE—XLT1WI横轴位、脂肪抑制FSE—XL T2WI矢状位平扫和3D—FSPGR T1WI双乳横轴位动态增强扫描,对各扫描序列的应用价值进行评价。结果不同序列对病变的检出率分别为FSE—IR93.33%、FSE—XLT2WI93.33%、FSE—XLTIWI42.22%、3D—FSPGRT1WI动态增强为82.22%。结论3.0TMR乳腺成像,联合使用FSE—IR序列进行脂肪抑制FSE—XLT2WI矢状位平扫和3D—FSPGR序列动态增强,可在较短的扫描时间内获得优质的扫描图像.

关 键 词:乳腺  临床分析  治疗方法  磁共振成像

Optimization of Breast Magnetic Resonance Imaging technology at 3.0 Tesla System
Institution:ZHAO Hong -lan( Huaibei People's Hospital ,Huaibei 235000, China )
Abstract:Objective To evaluate the application value of different sequences in breast magnetic resonance imaging, and compare the effect of different parameters on imaging quality. Methods Multiple sequences and parameters were performed with 8 -channel breast coil at 3.0T (GE). Forty -two patients were performed by FSE -IR and non fat suppressed FSE - XL T1WI in axial position, FSE - XL T2WI with fat suppression in sagittal position and 3D - FSPGR for dynamic contrast - enhancement in axial position. Then to evaluate the application value of above sequences by the number of displayed lesions and the imaging quality. Results The detectability of the different sequences for breast lesions was 93.33% by FSE - IR and FSE - XL T2WI, 42.22% by FSE - XL T1WI and 82.22% by 3D - FSPGR. Compared by imaging quality, there were no significant differences among the sequences ( P 〉 0.05 ). Conclusion The combination of FSE - IR, FSE - XL T2 WI and contrast - enhanced dynamic 3D - FSPGR sequences at 3.0T for breast MRI can obtain excellent imaging quality in short scanning time.
Keywords:Magnetic resonance imaging  Breast
本文献已被 维普 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号