ArcCHECK半导体探头特性及在容积调强弧形治疗剂量验证应用研究
引用本文:李成强,李光俊,冀传仙,肖江洪,郭昌,全红,柏森.ArcCHECK半导体探头特性及在容积调强弧形治疗剂量验证应用研究[J].中华放射肿瘤学杂志,2013,22(3):253-257.
作者姓名:李成强  李光俊  冀传仙  肖江洪  郭昌  全红  柏森
作者单位:430072 武汉大学物理科学与技术学院(李成强、郭昌、全红);610064 成都,四川大学华西医院肿瘤放射物理技术中心(李光俊、冀传仙、肖江洪、柏森)
基金项目::国家自然科学基金项目(81101697)
摘    要:目的 研究ArcCHECK半导体探头特性及在容积调强弧形治疗(VMAT)剂量验证应用。
方法 用PTWRW3型固体水模体对ArcCHECK探头测量固有敏感性、稳定性、剂量响应、剂量率响应、每脉冲剂量响应、射野大小依赖性、角度依赖性,并与PTW31010型0.125 cm3指形电离室测量值或VMAT计划系统计算值比较。随机选取211例已验证过的VMAT计划,分析计划与测量剂量分布的γ通过率差异,两两比较采用配对t检验。
结果
除角度依赖性外ArcCHECK探头其余特性均符合临床验证要求,当射束从探头底部入射时探头响应最小(180°时约为-3.9%),射束从两侧入射时响应最大(255°时约为 7.7%)。113例鼻咽癌、48例宫颈癌和50例直肠癌VMAT计划3 mm3%的γ平均通过率分别为93.5%、95.7%和97.5%,两两比较t=-12.69~-4.88,P均<0.01。
结论ArcCHECK半导体探头进行VMAT剂量验证前需精心校正,计划复杂度是影响VMAT计划验证通过率主要因素。

关 键 词:ArcCHECK半导体探头    容积调强弧形治疗    剂量验证  
收稿时间:2012-12-13

The characteristics and clinical application of the ArcCHECK diode array for volumetric-modulated arc therapy verification
LI Cheng-qiang,LI Guang-jun,JI Chuan-xian,XIAO Jiang-hong,GUO Chang,QUAN Hong,BAI Sen.School of Physics and Technology,Wuhan University,Wuhan,ChinaCorresponding author:BAI Sen,Radiation Physics Center,Cancer Center,West China Hospital,Sichuan University,Chengdu,China,Email:Tonybaisen@yahoo.com.cn.The characteristics and clinical application of the ArcCHECK diode array for volumetric-modulated arc therapy verification[J].Chinese Journal of Radiation Oncology,2013,22(3):253-257.
Authors:LI Cheng-qiang  LI Guang-jun  JI Chuan-xian  XIAO Jiang-hong  GUO Chang  QUAN Hong  BAI SenSchool of Physics and Technology  Wuhan University  Wuhan  ChinaCorresponding author:BAI Sen  Radiation Physics Center  Cancer Center  West China Hospital  Sichuan University  Chengdu  China  Email:Tonybaisen@yahoocomcn
Institution:TonybaBAI Sen.School of Physics and Technology, Wuhan University,Wuhan 430072,ChinaCorresponding author:BAI Sen, Radiation Physics Center, Cancer Center, West China Hospital, Sichuan University, Chengdu 610064,
Abstract:Objective To evaluate the characteristics and clinical suitability of the ArcCHECK diode array for volumetric-modulated arc therapy (VMAT) verification. Methods The intrinsic sensitivity, short and long term reproducibility, dose and dose rate dependence, dose per pulsed dependence, field sizes dependence and directional response of the diodes were measured. The results of the diodes were compared with the measurement results of an ionization chamber and calculated results of treatment planning system. Gamma index was used to analyze the dose difference between the calculation and measurement for random selected 211 verified VMAT plans. Results The ArcCHECK performed well for all tests except directional dependence, which varies from a minimum of -3.9%(seen only when the beam was incident on the diode at 180°) to a maximum of 7.7%(approximately at 255°). Average gamma analysis passing rates with 3 mm/3% for 113 nasopharyngeal cancer, 48 cervical cancer and 50 rectal cancer patients VMAT plans were 93.5%,95.7% and 97.5%, respectively,statistical significance has been found between either two of the groups (t=-12.69—-4.88,all P<0.01). Conclusions With proposed calibration method, the ArcCHECK diode array is very suitable for VMAT pretreatment verification,the complexity of VMAT plan is the main factor affecting the passing rate.
Keywords:ArcCHECK semiconductor probe  Volumetric-modulated arc therapy  Dose verification