Multiple injuries: Coma and fractures of the extremities |
| |
Authors: | Sanker P R A Frowein K -E Richard |
| |
Institution: | (1) Department of Neurosurgery and Neuroradiology, Evangelical Hospital of Oldenburg, Oldenburg, West Germany;(2) Nordwest-Krankenhaus Sanderbusch, Neurochirurgische Abteilung, D-2945 Sande 1, West-Germany |
| |
Abstract: | Direct coronal CT scanning can be efficiently performed by a new co-table equipment, which requires a comfortable prone positioning of the patient, whereby the patient must only extend his head slightly. The co-tableunit is linked to the original CT table board directly within a few minutes, coming through the opening of the gantry from the rear. This practical co-table method serves to reduce motion artifacts and to improve image quality significantly. By this method coronal follow-up CT scans are exactly reproducable. |
| |
Keywords: | Coronal CT scanning co-table image quality motion artifacts prone position |