Influence of iterative reconstruction and dose levels on metallic artifact reduction: A phantom study within four CT systems
Authors:J Greffier  A Larbi  J Frandon  PA Daviau  JP Beregi  F Pereira
Institution:Department of Radiology, CHU Nîmes, Université Montpellier, Medical Imaging Group Nîmes, EA 2415, 30029 Nîmes cedex, France
Abstract:

Purpose

To compare metallic artifact reduction (MAR) algorithms proposed by four vendors according to the delivered dose and iterative level using a phantom study.

Methods

Four CT systems (Revolution GSI®, Ingenuity Elite®, Somatom Edge®, and Aquilion Prime®) equipped with MAR algorithms (Smart MAR®, O-MAR®, iMAR®, and SEMAR®) were compared. Acquisitions were performed with CIRS model 062M Phantom containing a titanium rod core insert using 120 kV and two dose levels (3 and 7 mGy). Images were reconstructed with and without MAR algorithms using standard “soft tissue” kernel for filtered back projection (FBP) and intermediary iterative level. Artifact propagation was assessed by counting the number of pixels containing a HU outside a defined threshold interval (> 100HU and < ?80 HU). Artifact correction was evaluated by computing the differences between images with and without MAR.

Results

Accuracy of NCT values increased significantly using MAR algorithms, IR, and high dose levels (P < 0.001). Image noise reduced ?31 ± 15 (SD) % (range: ?50%;?14%) with Smart-MAR®, ?28 ± 3 (SD) % (range ?31%; ?25%) with O-MAR®, ?32 ± 7 (SD) % (range: ?40%; ?24%) with iMARCN®, ?52 ± 8 (SD) % (range: ?60%; ?42%) with iMARTH® and ?29 ± 6 (SD) % (range: ?37%; ?23%) with SEMAR®. The number of pixels outside the threshold interval was also reduced using MAR algorithms. Each MAR algorithm corrected in distinct patterns, with satisfactory artifact correction for all MAR algorithms.

Conclusion

This study demonstrates that artifact correction using MAR algorithms differs according to the main manufacturers, although corrections are satisfactory for all systems. Corrections also improved by using IR and increasing the dose level.
Keywords:Metallic artifact reduction (MAR)  Computed tomography (CT)  Image quality  Iterative reconstruction  CT  Computed tomography  FBP  Filtered back projection  Iterative metal artifact reduction  IR  Iterative reconstruction  MAR  Metallic Artifact reduction  Metal Artifact reduction for orthopedic implants  ROI  Region of Interest  Single energy metal artifact reduction  Smart metallic artifact reduction