首页 | 本学科首页   官方微博 | 高级检索  
检索        

DA3200-ABI7500全自动核酸提取平台高敏 HCV…RNA定量检测性能验证
引用本文:冯飞雪,王大望,肇玉博,王凯,周嘉迪,王战争,王丽,马艳侠,史小武.DA3200-ABI7500全自动核酸提取平台高敏 HCV…RNA定量检测性能验证[J].现代检验医学杂志,2021,0(1):108-111.
作者姓名:冯飞雪  王大望  肇玉博  王凯  周嘉迪  王战争  王丽  马艳侠  史小武
作者单位:(1.陕西中医药大学附属第一医院检验科,陕西咸阳 712000;2.陕西中医药大学医学技术学院,陕西咸阳 712000;3.咸阳市中心血站检验科,陕西咸阳 712000)
摘    要:目的 验证 DA3200-ABI7500全自动核酸提取高敏检测 HCV RNA方法性能。方法 参考美国临床实验室标准化协会批准指南( CSLI-EP),采用 DA3200全自动核酸提取平台及荧光定量 PCR检测 HCV RNA,评价检测方法的核酸提取纯度、精密度、正确度、线性范围、分析灵敏度、特异度和抗污染能力等性能指标。结果 核酸提取后 A260nm/ A280nm比值为 2.02。高、低两个浓度样本的批间精密度和批内精密度 CV值均≤ 5%。正确度与抗干扰能力方面,标本实测值与靶值之间的偏离度≤ ±0.4lg。在 13~ 1.3 E+06 IU/ml范围内,检测结果与靶值有良好的线性关系(Y=0.984X-0.186, r2=0.998)。最低检出限为 20 IU/ml的检出率为 100%,且系统具有较强的特异度和抗污染能力。结论 基于 DA3200-ABI7500全自动核酸提取高敏检测 HCV RNA各项分析性能指标均较好,高敏检测对治疗过程的监测及治疗终点判定具有重要意义。

关 键 词:丙型肝炎病毒  HCV  RNA  高敏检测  性能验证

Performance Verification of DA3200-ABI7500 Automatic Nucleic Acid Extraction Platform for High Sensitive HCV RNA Quantitative Detection
FENG Fei-xue,WANG Da-wang,ZHAO Yu-bo,WANG Kai,ZHOU Jia-di,WANG Zhan-zheng,WANG Li,MA Yan-xia,SHI Xiao-wu.Performance Verification of DA3200-ABI7500 Automatic Nucleic Acid Extraction Platform for High Sensitive HCV RNA Quantitative Detection[J].Journal of Modern Laboratory Medicine,2021,0(1):108-111.
Authors:FENG Fei-xue  WANG Da-wang  ZHAO Yu-bo  WANG Kai  ZHOU Jia-di  WANG Zhan-zheng  WANG Li  MA Yan-xia  SHI Xiao-wu
Institution:(1.Department Clinical Laboratory, the First Hospital of Shaanxi University of Chinese Medicine, Shaanxi Xianyang 712000,China; 2. Academy of Medical Technology of Shaanxi University of Chinese Medicine, Shaanxi Xianyang 712000,China; 3. Department Clinicol Laboratory,Xianyang Central Blood Station, Shaanxi Xianyang 712000,China)
Abstract:Objective To evaluate the DA3200-ABI7500 automatic nucleic acid testing platform for highly sensitive HCV RNA quantitative detection.Methods Based on American standard CLSI-EP,the DA3200 automatic nucleic acid extraction platform and PCR analysis system was applied to high sensitivity detect HCV RNA,the precision,accuracy,linear range,detection ability and anti-pollution ability of this method was analyzed.Results In terms of nucleic acid extraction,the ratio of A260nm/A280nm was 2.02,and the intra-assay and inter-assay precision CV values of high and low concentration samples were both≤5%.In accuracy and anti-interference ability,the deviation between the measured value of the specimen and the target value was≤±0.4lg,and there was a good linear relationship between the test results and the target value in the range of 13~1.3 E+06 IU/ml(Y=0.984X-0.186,r^2=0.998).The detection rate of 20 IU/ml was 100%.Conclusion Based on the DA3200-ABI7500 automatic nucleic acid extraction platform,the high sensitivity HCV-RNA quantitative detection performance indicators were all good,and its high-sensitivity detection is of great significance for the monitoring of the treatment process and the determination of treatment endpoints.
Keywords:hepatitis C virus  HCV RNA  high sensitive detection  performance verificatio
本文献已被 维普 等数据库收录!
点击此处可从《现代检验医学杂志》浏览原始摘要信息
点击此处可从《现代检验医学杂志》下载免费的PDF全文
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号