Clinical evaluation of the rapid nucleic acid amplification point-of-care test (Smart Gene SARS-CoV-2) in the analysis of nasopharyngeal and anterior nasal samples
Institution:1. Division of Infectious Diseases, Department of Medicine, Tsukuba Medical Center Hospital, 1-3-1 Amakubo Tsukuba, Ibaraki, 305-8558, Japan;2. Department of Infectious Diseases, University of Tsukuba Hospital, 2-1-1 Amakubo, Tsukuba, Ibaraki, 305-8576, Japan;3. Mizuho Medy Co., Ltd., 5-4 Fujinoki-machi, Tosu City, Saga, 841-0048, Japan;4. Akashi Internal Medicine Clinic, 3-1-63 Asahigaoka, Kashiwara, Osaka, 582-0026, Japan;5. Department of Clinical Laboratory, Tsukuba Medical Center Hospital, 1-3-1 Amakubo, Tsukuba, Ibaraki, 305-8558, Japan;6. Department of Respiratory Medicine, Tsukuba Medical Center Hospital, 1-3-1 Amakubo Tsukuba, Ibaraki, 305-8558, Japan;7. Department of Infectious Diseases, Faculty of Medicine, University of Tsukuba, 1-1-1 Tennodai, Tsukuba, Ibaraki, 305-8575, Japan
Abstract:IntroductionSmart Gene is a point-of-care (POC)-type automated molecular testing platform that can be performed with 1 min of hands-on-time. Smart Gene SARS-CoV-2 is a newly developed Smart Gene molecular assay for the detection of SARS-CoV-2. The analytical and clinical performance of Smart Gene SARS-CoV-2 has not been evaluated.MethodsNasopharyngeal and anterior nasal samples were prospectively collected from subjects referred to the local PCR center from March 25 to July 5, 2021. Two swabs were simultaneously obtained for the Smart Gene SARS-CoV-2 assay and the reference real-time RT-PCR assay, and the results of Smart Gene SARS-CoV-2 were compared to the reference real-time RT-PCR assay.ResultsAmong a total of 1150 samples, 68 of 791 nasopharyngeal samples and 51 of 359 anterior nasal samples were positive for SARS-CoV-2 in the reference real-time RT-PCR assay. In the testing of nasopharyngeal samples, Smart Gene SARS-CoV-2 showed the total, positive and negative concordance of 99.2% (95% confidence interval CI]: 98.4–99.7%), 94.1% (95% CI: 85.6–98.4%) and 99.7% (95% CI: 99.0–100%), respectively. For anterior nasal samples, Smart Gene SARS-CoV-2 showed the total, positive and negative concordance of 98.9% (95% CI: 97.2–99.7%), 98.0% (95% CI: 89.6–100%) and 99.0% (95% CI: 97.2–99.8%), respectively. In total, 5 samples were positive in the reference real-time RT-PCR assay and negative in the Smart Gene SARS-CoV-2 assay, whereas 5 samples were negative in the reference real-time RT-PCR assay and positive in the Smart Gene SARS-CoV-2 assay.ConclusionSmart Gene SARS-CoV-2 showed sufficient analytical performance for the detection of SARS-CoV-2 in nasopharyngeal and anterior nasal samples.
Keywords:Point-of-care test  Smart gene  Nucleic acid amplification test  QProbe  SARS-CoV-2  COVID-19
本文献已被 ScienceDirect 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号