首页 | 本学科首页   官方微博 | 高级检索  
检索        


Performance of a Computer System for Recording and Analysing Eye Gaze Position Using an Infrared Light Device
Authors:Z Ramdane-Cherif  A Naït-Ali  J F Motsch  M O Krebs
Institution:1. INSERM - E 0117, Service Hospitalo-Universitaire, H?pital Ste-Anne, Paris, France
2. Laboratoire Ingénierie des Signaux Neuraux Sensoriels, Université, Paris 12, France
3. Laboratoire d'Etude et de Recherche en Instrumentation, Signaux et Systèmes, Traitement du Signal et Instrumentation Médicale, Université Paris 12, France
Abstract:Our aim is to develop a computer system in order to study the visual scanning and eye fixation during exploration of image. The system uses an infrared device to detect the horizontal and vertical eye movements. In this paper, our interest is concentrated mainly on the calibration procedure. More precisely, on the correction of the nonlinearity in the oculomotor for individual subject's eyes. An adaptive calibration algorithm is developed. The algorithm uses a polynomial model and the optimal correction is obtained by using a mean square error criterion. The paper presents also a method to correct head movements. Experiments show that the proposed approach is fast and accurate for the study of visual perception and recognition processes.
Keywords:Eye movement tracking  visual attention  visual exploration  position estimation