Validation of 5 routine assays for serum free testosterone with a candidate reference measurement procedure based on ultrafiltration and isotope dilution-gas chromatography-mass spectrometry |
| |
Authors: | Van Uytfanghe K Stöckl D Kaufman J M Fiers T De Leenheer A Thienpont L M |
| |
Institution: | Laboratory for Analytical Chemistry, Faculty of Pharmaceutical Sciences, Ghent University, Harelbekestraat 72, B-9000 Gent, Belgium. |
| |
Abstract: | BACKGROUND/METHOD: The analytical validity of free testosterone (FTe) analog immunoassays is subject to much controversy. We revisited the validation of 4 analog assays and 1 FTe calculation procedure with a metrologically traceable reference measurement procedure (RMP) based on ultrafiltration and isotope dilution-mass spectrometry for direct measurement of Te in the ultrafiltrate. To this end, we performed split-sample measurements of 40 male sera. RESULTS: Deming regression showed that 3 of the immunoassays had moderate to good correlation (0.8474 < or = r < or = 0.9241) with the RMP; however, the slope was markedly below 1. The FTe calculation procedure was in good agreement with this result. The Sy/x values for all assays were higher than the combined imprecision values, which indicate their susceptibility to matrix-related effects. CONCLUSIONS: The study demonstrated substantial differences in analytical quality of FTe assays; however, the results suggested that after extending the validation with a larger variety of samples, recalibration of some analog assays might be worth further investigation. |
| |
Keywords: | Serum free testosterone Reference measurement procedure Ultrafiltration Isotope dilution-gas chromatography-mass spectrometry Method comparison |