SRS MapCHECK半导体矩阵用于射波刀脊柱计划剂量验证的应用研究
引用本文:祝小惠,梁志文,曹婷,韩军,王皓宁,龙浩,杨晶.SRS MapCHECK半导体矩阵用于射波刀脊柱计划剂量验证的应用研究[J].中国医学物理学杂志,2022,0(4):415-420.
作者姓名:祝小惠  梁志文  曹婷  韩军  王皓宁  龙浩  杨晶
作者单位:1.华中科技大学同济医学院附属协和医院肿瘤中心, 湖北 武汉 430022; 2.中南民族大学电子信息工程学院, 湖北 武汉 430074
摘    要:目的:评估SRS MapCHECK半导体矩阵探测器用于射波刀脊柱计划剂量验证工作的适用性。方法:将SRS MapCHECK探测器安装于专用模体StereoPHAN开展研究工作,测试了射波刀固定和Iris可变准直器的单野计划,以及脊柱临床计划的质量保证(QA)计划,采用SNC Patient软件对比分析实测与计划剂量分布之间的差异,分别计算在2 mm/5%、2 mm/3%和2 mm/2%标准下的γ通过率。结果:在绝对剂量分析模式2 mm/5%,2 mm/3%和2 mm/2%标准下,固定和Iris可变准直器单野计划的γ通过率均为100.0%,脊柱QA计划的平均γ通过率分别为(99.3±1.2)%、(96.5±2.7)%和(93.5±5.3)%。结论:SRS MapCHECK半导体矩阵探测器在2 mm/5%和2 mm/3%的γ分析标准下适合用于开展射波刀脊柱临床计划的剂量验证工作。

关 键 词:SRS  MapCHECK  半导体矩阵  射波刀  脊柱临床计划  剂量验证  γ通过率

Application of SRS MapCHECK semiconductor detector array in dose verification for spine CyberKnife plan
ZHU Xiaohui,LIANG Zhiwen,CAO Ting,HAN Jun,WANG Haoning,LONG Hao,YANG Jing.Application of SRS MapCHECK semiconductor detector array in dose verification for spine CyberKnife plan[J].Chinese Journal of Medical Physics,2022,0(4):415-420.
Authors:ZHU Xiaohui  LIANG Zhiwen  CAO Ting  HAN Jun  WANG Haoning  LONG Hao  YANG Jing
Institution:1. Cancer Center, Union Hospital, Tongji Medical College, Huazhong University of Science and Technology, Wuhan 430022, China 2. School of Electronic and Information Engineering, South-Central Minzu University, Wuhan 430074, China
Abstract:Abstract: Objective To evaluate the applicability of SRS MapCHECK semiconductor detector in dose verification for spine CyberKnife plan. Methods SRS MapCHECK detector was installed in the specific StereoPHAN phantom to conduct research. The single-field plans delivered by CyberKnife fixed and Iris variable collimators, as well as the quality assurance (QA) plans for spine clinical plans were tested. SNC Patient software was used to compare and analyze the differences between the measured and planned dose profiles, and the gamma passing rates for the criteria of 2 mm/5%, 2 mm/3% and 2 mm/2% were calculated. Results In absolute dose analysis mode and under the criteria of 2 mm/5%, 2 mm/3% and 2 mm/2%, all the gamma passing rates of the single-field plans delivered by fixed or Iris variable collimators reached 100.0%, and the average gamma passing rates of the spine QA plans were 99.3%±1.2%, 96.5%±2.7% and 93.5%±5.3%, respectively. Conclusion SRS MapCHECK semiconductor detector is suitable for the quick dose verification for spine CyberKnife plans under the gamma criteria of 2 mm/5% and 2 mm/3%.
Keywords:Keywords: SRS MapCHECK semiconductor array CyberKnife spine clinical plan dose verification gamma passing rate