Evaluation of noise equivalent count parameters as indicators of adult whole-body FDG-PET image quality |
| |
Authors: | Garry M. McDermott Fahmid U. Chowdhury Andrew F. Scarsbrook |
| |
Affiliation: | 1. Department of Medical Physics and Engineering, Level 1, Bexley Wing, St. James’ University Hospital, Leeds, LS9 7TF, UK 2. Department of Nuclear Medicine, Level 1, Bexley Wing, St. James’ University Hospital, Leeds, UK
|
| |
Abstract: |
Objectives The aim of this study was to assess variation of qualitative and quantitative PET/CT image quality parameters with acquisition time, injection activity and body mass for a representative group of adults undergoing whole-body PET/CT imaging. Methods PET scan data from sixty patients were reconstructed with a scan time of 1, 2 and 3 min/bed position. These images were visually scored and three quantitative parameters were calculated: noise equivalent counts per axial length (NECpatient), noise equivalent count density (NECdensity) and liver signal to noise ratio (liver SNR). The ability of the three quantitative parameters to discriminate qualitative image quality was assessed using ROC analysis. Results The quantitative parameters were shown to discriminate images of good/excellent quality from those of poorer image quality with a high degree of accuracy (ROC area >0.9); further, NECpatient had significantly higher discrimination than either NECdensity or liver SNR (ROC area = 0.97). Conclusions NECpatient, NECdensity and liver SNR all have high discrimination for qualitatively assessed PET image quality. NECpatient in particular is an effective objective indicator of patient image quality, which will help to assess and standardise scan protocols for purposes such as multi-centre research trials. |
| |
Keywords: | |
本文献已被 SpringerLink 等数据库收录! |
|