VITEK-MS 质谱仪直接检测BacT/ALERT 3D仪阳性血培养瓶微生物的实验研究
引用本文:李聪,丁守甫,黄颖,王中新,沈继录,徐元宏. VITEK-MS 质谱仪直接检测BacT/ALERT 3D仪阳性血培养瓶微生物的实验研究[J]. 中国现代医学杂志, 2017, 27(23): 13-17
作者姓名:李聪  丁守甫  黄颖  王中新  沈继录  徐元宏
作者单位:安徽医科大学第一附属医院检验科,安徽合肥230022
基金项目:国家自然科学基金(No:81171606)
摘    要:
目的评估基质辅助激光解析电离飞行时间质谱(VITEK-MS)仪直接检测BacT/ALERT 3D仪阳性培养瓶微生物方法的可行性与准确性。方法选取2016 年9 月-2016 年11 月安徽医科大学第一附属医院微生物室BacT/ALERT 3D全自动培养仪当天报告阳性的血培养瓶,差速离心法处理后直接用VITEK-MS 进行检测,并与常规鉴定法进行比较。结果150 例阳性培养瓶中,除13 例培养瓶中无细菌生长外,VITEK-MS 直接鉴定法正确检测出97 例(70.8%)菌株,错误检测出10 例(7.3%)菌株,30 例(21.9%)鉴定无结果。其中,革兰阴性杆菌正确鉴定出61 例(88.4%)菌株,鉴定无结果8 例(11.6%);革兰阳性球菌正确鉴定出36 例(52.9%)菌株,鉴定错误10 例(14.7%),鉴定无结果22 例(32.4%)。结论利用VITEK-MS 直接鉴定阳性血培养瓶的方法准确性高、特异性强,革兰阴性杆菌鉴定准确率远高于革兰阳性球菌,且鉴定时间短(只需2 h)。该方法可以作为快速诊断阳性血培养瓶的新方法。

关 键 词:基质辅助激光解析电离飞行时间质谱;血培养;直接鉴定法
收稿时间:2017-02-13

VITEK-MS direct identification of bacteria from positive BacT/ALERT 3D blood culture bottles
Cong Li,Shou-fu Ding,Ying Huang,Zhong-xin Wang,Ji-lu Shen,Yuan-hong Xu. VITEK-MS direct identification of bacteria from positive BacT/ALERT 3D blood culture bottles[J]. China Journal of Modern Medicine, 2017, 27(23): 13-17
Authors:Cong Li  Shou-fu Ding  Ying Huang  Zhong-xin Wang  Ji-lu Shen  Yuan-hong Xu
Affiliation:Clinical Laboratory, the First Affiliated Hospital of Anhui Medical University,Hefei, Anhui 230022, China
Abstract:
Objective To evaluate the performance of matrix-assisted laser desorption/ionization time-offlightmass spectrometry system (VITEK-MS) for direct detection of bacteria in the positive bottles by BacT/ALERT 3D blood culture system and its identification rate. Methods All positive blood culture botlles reported by BacT/ALERT 3D blood culture system that had been obtained from the patients in the First Affiliated Hospital of Anhui Medical University from September to November 2016 were selected. The bacteria were directly detected by VITEK -MS after differential centrifugation, and compared with the results by traditional biochemical identification. Results Among 150 positive blood culture bottles, there was no bacteria growth in 13 bottles; in the rest 137 bottles, VITEK-MS correctly identified 97 (70.8%), wrongly detected 10(7.3%), and could not identify 30 (21.9%). For Gram-negative bacilli, the correct identification rate was 88.4% (61 cases) and 8 cases (11.6%) were not identified. In Gram-positive cocci, the correct and wrong identification rate were 52.9% (36 cases) and 14.7% (10 cases), and 22 cases (32.4%) could not be identified. Conclusions The method of using VITEK-MS to identify the positive blood culture bottles is accurate and specific. The accurate identification rate of Gram-negative bacilli is much higher than that of Gram-positive cocci. The time for identification is reduced to 2 h. Thus, this method can provide rapid and reliable identification of microorganisms in blood culture bottles.
Keywords:matrix-assisted laser desorption/ionization time-of-flight mass spectrometry   blood culture  direct identification