首页 | 本学科首页   官方微博 | 高级检索  
     


A new capnograph based on an electro acoustic sensor
Authors:M. Folke  B. Hök
Affiliation:1.Department of Computer Science and Electronics,M?lardalen University,V?ster?s,Sweden;2.H?k Instrument AB,V?ster?s,Sweden;3.IDE,M?lardalen University,V?ster?s,Sweden
Abstract:End tidal carbon dioxide measurements with an electro acoustic capnograph prototype have been demonstrated. The aim of this study was to verify that it is possible to obtain an adequate capnogram using the prototype and to investigate the influence of ambient temperature and humidity variations. By simultaneous measurements with a reference capnograph, on subjects performing exercise, hypo- and hyperventilation, PETCO2 readings from the reference were compared with the output signal from the prototype. The capnogram from the prototype correlated well with the reference in terms of breath time. The first parts of the expiration and inspiration phases were steeper for the reference than the prototype. The output signal from the prototype correlated well with the reference PETCO2 readings with a correlation coefficient of 0.93 at varied temperature and relative humidity.
Keywords:End tidal  Carbon dioxide  Electro acoustic sensor  Capnometry  Respiratory monitoring
本文献已被 PubMed SpringerLink 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号