A new capnograph based on an electro acoustic sensor |
| |
Authors: | M. Folke B. Hök |
| |
Affiliation: | 1.Department of Computer Science and Electronics,M?lardalen University,V?ster?s,Sweden;2.H?k Instrument AB,V?ster?s,Sweden;3.IDE,M?lardalen University,V?ster?s,Sweden |
| |
Abstract: | End tidal carbon dioxide measurements with an electro acoustic capnograph prototype have been demonstrated. The aim of this study was to verify that it is possible to obtain an adequate capnogram using the prototype and to investigate the influence of ambient temperature and humidity variations. By simultaneous measurements with a reference capnograph, on subjects performing exercise, hypo- and hyperventilation, PETCO2 readings from the reference were compared with the output signal from the prototype. The capnogram from the prototype correlated well with the reference in terms of breath time. The first parts of the expiration and inspiration phases were steeper for the reference than the prototype. The output signal from the prototype correlated well with the reference PETCO2 readings with a correlation coefficient of 0.93 at varied temperature and relative humidity. |
| |
Keywords: | End tidal Carbon dioxide Electro acoustic sensor Capnometry Respiratory monitoring |
本文献已被 PubMed SpringerLink 等数据库收录! |
|