操作者因素对Lenstar测量眼轴长度可重复性的影响
引用本文:彭显耀,孙瑞竹,吕帆. 操作者因素对Lenstar测量眼轴长度可重复性的影响[J]. 温州医科大学学报, 2016, 46(9): 649-651,655
作者姓名:彭显耀  孙瑞竹  吕帆
作者单位:温州医科大学附属眼视光医院临床与流行病学研究中心,浙江温州325027
摘    要:
目的:研究操作者因素对Lenstar测量眼轴长度(AL)可重复性的影响。方法:共纳入60名受试者(60只左眼),每位受试者进行2次检查,由操作者I和操作者II完成,每次检查进行至少3次连续测量。计算出所有受试者2次检查的标准差(SD)和变异系数(CV),用来评价操作者I和操作者II测量AL的可重复性;计算出操作者I和操作者II间的AL差值及其95%置信区间(95%CI)来评价2位操作者间的可重复性。结果:操作者I和操作者II测量的AL值为(24.05±0.02)mm和(24.05±0.02)mm,CV为0.07%和0.06%。操作者I和操作者II间AL差值为(0.00±0.03)mm。操作者I和操作者II间AL差值的95%CI为0.00±0.06。结论:不同操作者带来的差异并不影响Lenstar测量AL的可重复性。

关 键 词:眼轴长度  操作者  可重复性
  
收稿时间:2016-02-01

The influence of operators for the repeatability and reproducibility of axial length measured by Lenstar
PENG Xianyao,SUN Ruizhu,LYU Fan.. The influence of operators for the repeatability and reproducibility of axial length measured by Lenstar[J]. JOURNAL OF WENZHOU MEDICAL UNIVERSITY, 2016, 46(9): 649-651,655
Authors:PENG Xianyao  SUN Ruizhu  LYU Fan.
Affiliation:The Clinical & Epidemiological Eye Research Center, the Eye Hospital of Wenzhou Medical University, Wenzhou, 325027;
Abstract:
Objective: To explore the influence of different operators to the repeatability and reproducibility of Lenstar in axial length measurements. Methods: Sixty left eyes of 60 participants were enrolled. Each participant was performed two measurements which was finished by operator I and operator II, each measurement including at least three continue measure. The SD (standar deviation) and CV (coefficient of variation) of all the participants was calculated to evaluate the reproducibility of operator I and operator II. The difference between operator I and operator II in AL measurement and it’s 95%CI (confidence interval) was calculated to evaluate the reproducibility between operator I and operator II. Results: The Mean±SD of operator I and operator II were (24.05±0.02) mm and (24.05±0.02) mm. The CV of operator I and operator II were 0.07% and 0.06%. The Mean±SD of the difference of AL between operator I and operator II was (0.00±0.03) mm. The 95%CI of the difference of AL between operator I and operator II was 0.00±0.06. Conclusion: The factor of operators don’t affect the repeatability and reproducibility of Lenstar.
Keywords:axial length  operator  reproducibility  