Evaluation of myocardial function in patients with end-stage heart failure during support with the Jarvik 2000 left ventricular assist device. |
| |
Authors: | Markus Ferrari Kamuran A Kadipasaoglu Mihai Croitoru Jeff Conger Tim Myers Igor Gregoric Branislav Radovancevic George V Letsou O H Frazier |
| |
Affiliation: | Texas Heart Institute, Houston, TX 77225, USA. |
| |
Abstract: | In 2 patients with the Jarvik 2000 left ventricular assist device (LVAD), we assessed left ventricular systolic function through pressure-volume loops and E(max) at the beginning and end of the support period to potentially predict the possibility of pump removal without transplantation. Immediately before LVAD implantation and explantation, pressure and volume measurements were made with catheters and echocardiography, respectively, the E(max) being calculated from the slope of the pressure-volume loops, and the left ventricular ejection fraction (LVEF) being estimated by echocardiography. Transplantation was performed after 14 and 62 days, respectively, during which the LVEF increased by 75% (from 12% to 21%) in Patient 1 and remained unchanged (from 16% to 18%) in Patient 2, whereas the E(max) increased from 0.63 and 0.42 mm Hg/ml, respectively, to 1.31 and 1.07 mm Hg/ml, reflecting a 107% and 155% improvement. In these 2 cases, the E(max) was a more reliable indicator of intrinsic myocardial contractility than was the LVEF. |
| |