Semicontinuous intra-abdominal pressure measurement using an intragastric Compliance catheter
Authors:Jan J. De Waele  F. Berrevoet  K. Reyntjens  P. Pletinckx  I. De laet  E. Hoste
Affiliation:(1) Intensive Care Unit, Ghent University Hospital, De Pintelaan 185, 9000 Gent, Belgium;(2) Department of Anesthesiology, Ghent University Hospital, De Pintelaan 185, 9000 Ghent, Belgium;(3) Department of Surgery, Ghent University Hospital, De Pintelaan 185, 9000 Ghent, Belgium
Abstract:
Objective To compare intra-abdominal pressure (IAP) measurements obtained from an intragastric Compliance catheter with the pressure measured directly in the abdominal cavity. Design and setting Prospective cohort study in an operating room of the Ghent University Hospital Patients Seven patients undergoing elective laparoscopic cholecystectomy. Interventions IAP was obtained from both an intragastric catheter and directly from the peritoneal cavity at 1-minute intervals in patients undergoing elective cholecystectomy and compared using Bland-Altman analysis. Measurements and results In 156 paired measurements obtained from 7 patients the mean difference between IAPgastric and IAPref was 0.12 ±  0.70 mmHg (95% CI 0.01–0.23). Conclusions IAP measured using an intragastric Compliance catheter reliably reflects the reference IAP in patients undergoing laparoscopic cholecystectomy. Electronic supplementary material The online version of this article (doi:) contains supplementary material, which is available to authorized users. J.J. De Waele is supported by a Clinical Doctoral Grant of the Fund for Scientific Research-Flanders (Belgium; F.W.O.-Vlaanderen).
Keywords:Intra-abdominal pressure  Intra-abdominal hypertension  Abdominal compartment syndrome  Surgery  Trauma  Critically ill patients
本文献已被 PubMed SpringerLink 等数据库收录!
正在获取相似文献,请稍候...
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号