眼前段光学相干断层扫描仪与超声生物显微镜测量中央前房深度的比较
引用本文:李媚,杨晔,蔡小于,曾阳发,曹丹,刘杏. 眼前段光学相干断层扫描仪与超声生物显微镜测量中央前房深度的比较[J]. 现代临床医学生物工程学杂志, 2009, 16(4): 15-19. DOI: 10.3760/cma.j.issn.1674-1927.2010.01.004
作者姓名:李媚  杨晔  蔡小于  曾阳发  曹丹  刘杏
作者单位:中山大学中山眼科中心眼科学国家重点实验室,广州,510060;
摘    要:
Objective To compare the difference and repeatability between anterior segment optical coherence tomography (AS-OCT) and ultrasound biomicroscope (UBM) in measuring central anterior chamber depth (ACD). Methods Twenty-nine patients (32 attacked eyes and 23 unaffected fellow eyes) with acute primary angle closure glaucoma (APACG) and 41 patients (74 eyes) with chronic primary angle closure glaucoma (CPACG) were included in this study. AS-OCT and UBM were used to measure the ACD of these APACG and CPACG subjects. Pair sample t test and correlation analysis were introduced to analyze the difference and correlation of the ACD values between the 2 instruments. ACD of one complete randomly selected eye was measured 10 times by the above 2 instruments, respectively, and the repeatability of 2 instruments was analyzed. Results There was no significant difference in ACD values of attacked eyes of APACG between AS-OCT and UBM. In unaffected fellow eyes and CPACG eyes, the ACD values measured by AS-OCT were deeper than that measured by UBM[ (1.923±0.287) mm vs (1.860±0.243) mm, (2.017±0.054) mm vs (1.913±0.052) mm, both P=0.012]. Positive correlation between AS-OCT and UBM in measuring ACD was found (r=0.928, P<0.001). The coefficient of variation of ACD was 0.66% as measured by AS-OCT, and 0.82% as by UBM. Conclusions The ACD measured by AS-OCT can be deeper than that by UBM. Excellent correlation is found between AS-OCT and UBM measurements. AS-OCT appears to be more repeatable in measuring ACD than UBM.

关 键 词:眼前段光学相干断层扫描仪   超声生物显微镜   前房深度   青光眼   诊断技术,眼科   

Measurement of central anterior chamber depth with anterior segment optical coherence tomography versus ultrasound biomicroscope
LI Mei,YANG Ye,CAI Xiao-yu,ZENG Yang-fa,CAO Dan,LIU Xing. Measurement of central anterior chamber depth with anterior segment optical coherence tomography versus ultrasound biomicroscope[J]. Journal of Modern Clinical Medical Bioengineering, 2009, 16(4): 15-19. DOI: 10.3760/cma.j.issn.1674-1927.2010.01.004
Authors:LI Mei  YANG Ye  CAI Xiao-yu  ZENG Yang-fa  CAO Dan  LIU Xing
Abstract:
Keywords:Anterior segment optical coherence tomographyUltrasound biomicroscopeAnterior chamber depthGlaucomaDiagnostic techniques  ophthalmological