FEM assisted evaluation of PMMA and Ti6Al4V as materials for cranioplasty resulting mechanical behaviour and the neurocranial protection
Authors:Tsouknidas Alexander  Maropoulos Stergios  Savvakis Savvas  Michailidis Nikolaos
Affiliation:Laboratory for Machine Tools and Manufacturing Engineering, School of Polytechnics, Aristoteles University of Thessaloniki, Thessaloniki, Greece. alextso@auth.gr
Abstract:
Recent advances in Computer Aided Design and Manufacturing techniques (CAD/CAM) have facilitated the rapid and precise construction of customized implants used for craniofacial reconstruction. Data of the patients' trauma, acquired through Computer Topographies (CT), provide sufficient information with regard to the defect contour profile, thus allowing a thorough preoperative evaluation whilst ensuring excellent implant precision. During the selection, however, of a suitable implant material for the specific trauma, the mechanical aspects of the implant have to be considered. This investigation aims to assess the mechanical strength, the shock resistance and the critical deflection of cranial implants manufactured with two commonly used materials, Polymethylmethacrylate (PMMA) and Ti6Al4V. Even though the strength properties of Ti-alloys are far superior to those of PMMA, there are several aspects that may act in advantage of PMMA, e.g., it is known that discontinuities in the elastic modulus of adjoined parts (bone-implant) lead to bone resorption thus loosening the fixation of the implant over time.The implant design and fixation was the same in both cases allowing a direct comparison of the implant behavior for various loads. Finite Element Methods (FEM) assisted procedures were employed, providing a valuable insight to the neurocranial protection granted by these implants.
Keywords:
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号