Internal Defibrillation with Smaller Capacitors: A Prospective Randomized Cross-Over Comparison of Defibrillation Efficacy Obtained with 90-μF and 125-μF Capacitors in Humans |
| |
Authors: | MICHAEL BLOCK,M.D.,,DIETER HAMMEL,M.D.,DIRK BÖ CKER,M.D.,,MARTIN BORGGREFE,M.D.,,THOMAS SEIFERT,M.D.,,CHRISTIAN FASTENRATH,M.D.,,HANS H. SCHELD,M.D.,GÜ NTER BREITHARDT,M.D. |
| |
Affiliation: | Department of Cardiology/Angiology, Hospital of the Westfälische Wilhelms-University of Münster, and;Department of Cardiovascular Surgery and Institute for Research in Arteriosclerosis, Münster, Germany |
| |
Abstract: | Defibrillation with Small Capacitance. Introduction: The size of current implantable cardioverter defibrillators (ICD) is still large in comparison to pacemakers and thus not convenient for pectoral implantation. One way to reduce ICD size is to defibrillate with smaller capacitors. A trade-off exists, however, since smaller capacitors may generate a lower maximum energy output. Methods and Results: In a prospective randomized cross-over study, the step-down defibrillation threshold (DFT) of an experimental 90-μF biphasic waveform was compared to a standard 125-μF biphasic waveform. The 90-μF capacitor delivered the same energy faster and with a higher peak voltage but provided only a maximum energy output of 20 instead of 34 J. DFTs were determined intraoperatively in 30 patients randomized to receive either an endocardial (n = 15) or an endocardial-subcutaneous array (n = 15) defibrillation lead system. Independent of the lead system used, energy requirements did not differ at DFT for the experimental and the standard waveforms (10.3 ± 4.1 and 9.5 ± 4.9 J, respectively), but peak voltages were higher for the experimental waveform than for the standard waveform (411 ± 80 and 325 ± 81 V, respectively). For the experimental waveform the DFT was 10 J or less using an endocardial lead-alone system in 10 (67%) of 15 patients and in 12 (80%) of 15 patients using an endocardial-subcutaneous array lead system. Conclusions : A shorter duration waveform delivered by smaller capacitors does not increase defibrillation energy requirements and might reduce device size. However, the smaller capacitance reduces the maximum energy output. If a 10-J safety margin between DFT and maximum energy output of the ICD is required, only a subgroup of patients will benefit from 90-μF ICDs with DFTs feasible using current defibrillation lead systems. |
| |
Keywords: | implantable cardioverter defibrillator ventricular fibrillation endocardial defibrillation defibrillation threshold capacitance |
|
|