Comparison of Preoperative and Postoperative Ocular Biometry in Eyes with Phakic Intraocular Lens Implantations |
| |
Authors: | Joo Youn Shin Jae Bum Lee Kyoung Yul Seo Eung Kweon Kim Tae-im Kim |
| |
Affiliation: | 1.The Institute of Vision Research, Department of Ophthalmology, Yonsei University College of Medicine, Seoul, Korea.;2.Yonsei Plus Eye Center, Seongnam, Korea. |
| |
Abstract: |
PurposeTo compare preoperative and postoperative ocular biometry in patients with iris-fixated phakic intraocular lens (pIOLs): Artisan and Artiflex.Materials and MethodsThis study included 40 eyes with Artisan and 36 eyes with Artiflex pIOL implants. Anterior chamber depth (ACD) and axial length (AL) were measured by applanation ultrasonography (A-scan) and partial coherence interferometry (IOLMaster) preoperatively and 3 months after pIOL implantation.ResultsACD measurements after Artisan or Artiflex pIOL implantation were smaller than preoperative measurements. Specifically, the difference after Artisan pIOL implantation was -1.07±0.17 mm by A-scan and -0.08±0.08 mm by IOLMaster. The difference after Artiflex pIOL implantation was -1.31±0.15 mm by A-scan and -0.05±0.07 mm by IOLMaster. After Artisan pIOL implantation, differences in AL measurements by A-scan were insignificant (difference: -0.03±0.15 mm), whereas postoperative AL measurements by IOLMaster were significantly longer than preoperative measurements (difference: 0.12±0.07 mm). After Artiflex pIOL implantation, AL measurements by both A-scan and IOLMaster were significantly longer than preoperative measurements (difference: 0.09±0.16 mm by A-scan and 0.07±0.10 mm by IOLMaster). In the Artiflex group, differences in AL measurements by A-scan correlated with the central thickness of the Artiflex pIOL.ConclusionACD and AL measurements were influenced by iris-fixated phakic IOL implantation. |
| |
Keywords: | Biometry phakic intraocular lenses anterior chamber depth axial length |
|
|