Precil XC-40型全自动动态血沉测试仪的应用评价
引用本文:彭慧文,孙艳虹,张文迎,谭坪海,陆庆红,钟威,郑丹. Precil XC-40型全自动动态血沉测试仪的应用评价[J]. 检验医学与临床, 2013, 10(9): 1080-1081,1083
作者姓名:彭慧文  孙艳虹  张文迎  谭坪海  陆庆红  钟威  郑丹
作者单位:彭慧文 (中山大学附属一院黄埔院区检验科,广州,510700); 孙艳虹 (中山大学附属一院黄埔院区检验科,广州,510700);张文迎 (中山大学附属一院黄埔院区检验科,广州,510700);谭坪海 (中山大学附属一院黄埔院区检验科,广州,510700);陆庆红 (中山大学附属一院黄埔院区检验科,广州,510700);钟威 (中山大学附属一院黄埔院区检验科,广州,510700);郑丹 (中山大学附属一院黄埔院区检验科,广州,510700);
摘    要:
目的比较Precil XC-40型全自动动态血沉测试仪测定结果与传统魏氏法测定结果的相关性和线性。方法分别用自动血沉仪法和传统魏氏法测定238例临床患者1h红细胞沉降率(ESR)。结果经统计学处理,238例ESR在两种方法的测定结果经检验,差异无统计学意义(P>0.05)。在正常参考值范围内,二者测定结果一致。但当ESR>15mm/h后,Precil XC-40型全自动动态血沉测试仪测定中位数高于魏氏法,而且随着测定值增加,二者之间差异越明显。结论全自动动态血沉测试仪对正常标本测定结果与魏氏法接近,而在贫血、脂血标本以及ESR明显升高的标本则应以魏氏法为金标准。

关 键 词:红细胞沉降率  全自动动态血沉测试仪  魏氏法

Evaluation of Precil XC-40 automatic dynamic sedimentation analyzer
PENG Hui-wen,SUN Yan-hong,ZHANG Wen-ying,TAN Ping-hai,LU Qing-hong,ZHONG Wei,ZHENG Dan. Evaluation of Precil XC-40 automatic dynamic sedimentation analyzer[J]. Laboratory Medicine and Clinic, 2013, 10(9): 1080-1081,1083
Authors:PENG Hui-wen  SUN Yan-hong  ZHANG Wen-ying  TAN Ping-hai  LU Qing-hong  ZHONG Wei  ZHENG Dan
Affiliation:(Clinical Laboratory, Huangpu Department Of First Affiliated Hospital of Sun Yat-sen University ,Guangzhou,Guangdong 510700,China)
Abstract:
Objective To understand the correlation and linear between the detection results of Precil XC-40 automatic dynamic sedimentation analyzer and traditional Westergren method. Methods 238 cases of lh erythrocyte sedimentation rate (ESR) were detected by Precil XC-40 automatic monitor analyzer and traditional Westergren method. Results In normal reference range, there was no significant difference of the detection results of ESR be- tween the two methods (P〉0.05). But when ESR larger than 15 ram/h, the median of XC-40 method was higher than that of Westergren method, and with the increasing of detected value, the differences became more and more obvious. Conclusion Automatic dynamic ESR test method might be identical with Westergren method for the detec- tion of samples in normal reference range, but for the detection of anemia and lipid turbidity samples and samples with increased ESR level, Westergren method should be taken as gold-standard.
Keywords:erythrocyte sedimentation rate  auto dynamic blood sediment apparatus  Westergren method